The VibraPEP™ Mucus Clearing Device is a Positive Expiratory Pressure, PEP Device. It was designed to exercise patient's lungs and to improve secretion clearance.
Device Story
VibraPEP is a mechanical Positive Expiratory Pressure (PEP) device used in hospitals, clinics, and home settings. It consists of a curved tube, mouthpiece with therapy selector, valve tube, and cap. During exhalation, the patient blows through the device; internal hose pressure increases until it buckles, causing the hose end to open and release pressure against the tube wall. This cycle repeats, creating oscillations in the air column to promote secretion clearance. The therapy selector allows adjustment of pressure and frequency. The device is operated by the patient. It provides a non-pharmacological method to assist in airway clearance, potentially benefiting patients with respiratory conditions requiring secretion management.
Clinical Evidence
Bench testing only. Comparative performance testing evaluated average pressure, pressure amplitude, frequency, and average flow rate at 10, 20, and 40 lpm. Additional testing included cleaning validation, age testing, and mechanical drop testing. Results demonstrate equivalence to the predicate device.
Technological Characteristics
Mechanical PEP device; curved tube design with internal valve tube. Materials identical to predicate (ISO 10993 compliant for permanent mucosal contact). Operates via passive fluid dynamics (hose buckling/oscillation). No energy source required. Manual adjustment via therapy selector. Non-electronic, non-software.
Indications for Use
Indicated for patients prescribed PEP therapy to exercise lungs and improve secretion clearance. Contraindications include hemodynamic instability, intracranial pressure > 20 mm Hg, acute sinusitis, recent facial/oral/skull surgery or trauma, epistaxis, esophageal surgery, active hemoptysis, untreated pneumothorax, nausea, and known/suspected tympanic membrane rupture or middle ear pathology.
Regulatory Classification
Identification
An incentive spirometer is a device that indicates a patient's breathing volume or flow and that provides an incentive to the patient to improve his or her ventilation.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an image of three stylized human profiles facing to the right, with flowing lines beneath them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 25, 2016
Medica Holdings, LLC c/o Paul Dryden Regulatory Consultant for Medica Holdings, LLC 5200 Meadows Road. Suite 150 Lake Oswego, Oregon 97035
Re: K153441
Trade/Device Name: VibraPEPTM Regulation Number: 21 CFR 868.5690 Regulation Name: Incentive Spirometer Regulatory Class: Class II Product Code: BWF Dated: February 22, 2016 Received: February 23, 2016
Dear Paul Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Erin I. Keith -S
Erin I. Keith M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control. and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known)
#### K153441
Device Name
#### VibraPEP™
Indications for Use (Describe)
The VibraPEP™ Mucus Clearing Device is a Positive Expiratory Pressure, PEP Device. It was designed to exercise patient's lungs and to improve secretion clearance.
Patient - Patients who have been prescribed PEP therapy
Environment - Hospital, clinics, physician offices, home setting
Type of Use (Select one or both, as applicable)
XX Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
PSC Publishing Services (301) 443-6740
{3}------------------------------------------------
| | 510(k) Summary<br>March 15, 2016<br>Page 1 of 5 |
|----------------------------|----------------------------------------------------------------------------------------------------------------------|
| Date Prepared | March 15, 2016 |
| Official Contact: | George Reed<br>Medica Holdings, LLC<br>5200 Meadows Road, Suite 150<br>Lake Oswego, OR 97035<br>Tel – 503-227-1900 |
| Proprietary or Trade Name: | VibraPEP™ |
| Common/Usual Name: | Spirometer, Therapeutic (Incentive) |
| Classification Name: | BWF - 21CFR868.5690, Class II |
| Predicate Device: | Pari – RC-Cornet – K983308 |
## Device Description:
The VibraPEP™ is a Positive End Expiratory (PEP) device in which the user exhales through the device and the device generates expiratory pressure (resistance) and the Tube Valve oscillates creates a range of frequencies within the air column to the patient which help to promote secretion clearance. The VibraPEP™ has been designed to be very similar to the predicate RC-Cornet device (K983308). The basic design, form, function and performance have been compared to the predicate and have been demonstrated to be functionally equivalent and with the same intended use as the predicate.
The device is comprised of several components:
- Curved tube
- Cap for end of Tube ●
- Mouthpiece with selector ●
- Valve tube
- Drying tool ●
## Principle of Operation
The VibraPEP™ is a curved tube in which a long valve is inserted. As the patient blows through the VibraPEP, the hose pressure increases and buckles at the bending of the tube. When the peak pressure is reached, the hose end opens and is catapulted against the wall releasing its pressure. This process is repeated, providing an oscillation effect during the entire exhalation phase. By rotating the therapy selector, pressure and flow can be adjusted to increase or decrease the pressure and frequency of PEP therapy.
## Indications for Use:
The VibraPEP™ Mucus Clearing Device is a Positive Expiratory Pressure. PEP Device. It was designed to exercise patient's lungs and to improve secretion clearance.
Patient - Patients who have been prescribed PEP therapy
Environment - Hospital, clinics, physician offices, home setting
{4}------------------------------------------------
# 510(k) Summary
March 15, 2016 Page 2 of 5
## Contraindications:
Although no absolute contraindications to the use of PEP Therapy have been reported, the following should be carefully evaluated before a decision is made to initiate therapy:
- Inability to tolerate increased work of breathing .
- . Hemodynamic instability
- Intracranial pressure (IPC) > 20 mm Hg ●
- Acute sinusitis
- Recent facial, oral or skull surgery or trauma
- Epistaxis
- Esophageal surgery
- Active hemoptysis
- . Untreated pneumothorax
- Nausea ●
- Known or suspected tympanic membrane rupture or other middle ear pathology .
## Substantial Equivalence
| Features | Predicate - RC-Cornet<br>K983308 | Proposed<br>VibraPEP™ |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use | The PARI RC Cornet Mucus Clearing Device<br>is a Positive Expiratory Pressure PEP Device<br>and it was designed to exercise patient's lungs<br>and to improve secretion clearance. The device<br>was designed to work with PARI LC<br>Nebulizers. | The VibraPEP™ Mucus Clearing Device<br>is a Positive Expiratory Pressure, PEP<br>Device. It was designed to exercise<br>patient's lungs and to improve secretion<br>clearance. |
| Environment of Use | Hospital, clinics, physician offices, home<br>setting | Hospital, clinics, physician offices, home<br>setting |
| Patient Population | General population | Patients who have been prescribed PEP<br>therapy |
| Contraindications | • Hemodynamic instability<br>• Intracranial pressure (IPC) > 20 mm Hg<br>• Acute sinusitis<br>• Recent facial, oral or skull surgery or<br>trauma<br>• Epistaxis<br>• Esophageal surgery<br>• Active hemoptysis<br>• Untreated pneumothorax<br>• Nausea<br>• Known or suspected tympanic membrane<br>rupture or other middle ear pathology | • Hemodynamic instability<br>• Intracranial pressure (IPC) > 20 mm Hg<br>• Acute sinusitis<br>• Recent facial, oral or skull surgery or<br>trauma<br>• Epistaxis<br>• Esophageal surgery<br>• Active hemoptysis<br>• Untreated pneumothorax<br>• Nausea<br>• Known or suspected tympanic membrane<br>rupture or other middle ear pathology |
| Principle of Operation | Tube valve that upon patient exhalation will<br>create PEP and oscillation<br><br>Adjustable to create different frequencies | Tube valve that upon patient exhalation<br>will create PEP and oscillation<br><br>Adjustable to create different frequencies |
{5}------------------------------------------------
## 510(k) Summary
March 15, 2016
| Page 3 of 5 | | | | | | |
|---------------------------------------------------------|-------------------------------------------------------------------------------------------------------|---------|---------|-----------------------------------------------------------------------------------------------------------------|---------|---------|
| Features | Predicate - RC-Cornet<br>K983308 | | | Proposed<br>VibraPEP™ | | |
| Use with a nebulizer | Pari LC nebulizer | | | No recommendation | | |
| Single patient,<br>multi-use | Yes | | | Yes | | |
| Cleaning method | Soap / water and Boiling water | | | Soap / water and Boiling water | | |
| Components that may<br>be used | Not specified<br>Tee adapter to be inserted between<br>nebulizer and mouthpiece<br>Pari LC nebulizers | | | No user supplied components | | |
| Components | Curved tube<br>Mouthpiece / selector<br>Tube Valve<br>Cap<br>Drying device | | | Curved tube<br>Mouthpiece / selector<br>Tube Valve<br>Cap<br>Drying device | | |
| Performance – Non-clinical | | | | | | |
| Materials<br>per ISO 10993 | Surface Contact<br>Mucosal membrane<br>Duration of Use – permanent (> 30 days) | | | Surface Contact<br>Mucosal membrane<br>Duration of Use – permanent (> 30 days)<br>Identical materials - K983308 | | |
| | Flow rate | | | Flow rate | | |
| | 10 lpm | 20 lpm | 40 lpm | 10 lpm | 20 lpm | 40 lpm |
| Ave. Pressure<br>(cmH2O)<br>across full range | 11 – 14 | 17 – 21 | 23 – 44 | 10 – 13 | 18 – 21 | 27 – 41 |
| Ave. Pressure<br>Amplitude (cmH2O)<br>across full range | 6 – 19 | 21 – 27 | 46 – 80 | 7 – 17 | 22 – 28 | 51 – 76 |
| Ave. Flow Rate (lpm)<br>across full range | 6 – 16 | 12 – 18 | 15 – 21 | 8 – 16 | 12 – 17 | 19 – 22 |
Note: The pressure, frequency and flow testing was performed on the bench at a constant flow rate for comparative purposes.
## Discussion of Substantial Equivalence
The VibraPEP™ oscillator PEP device is viewed as substantially equivalent to the predicate device because:
## Indications for Use -
The proposed indications for use are to exercise patient's lungs and to improve secretion clearance. Discussion - The indications for use are similar for the proposed device and the predicate - K983308 - RC-Cornet. The use with a nebulizer is not included.
## Patient Population -
The patient population has been clarified to be for patients who have been prescribed PEP therapy.
Discussion - The patient population is intended to be identical to the predicate device, however the predicate did not specify a patient population other than to imply general patient population. The proposed language is viewed as substantially equivalent with added clarification vs. the predicate - K983308 - RC-Cornet.
{6}------------------------------------------------
## Environment of Use -
The proposed environments of use are Hospital, clinics, physician offices, home setting. Discussion - The environment of use is identical for the proposed device and the predicate -K983308 - RC-Cornet.
## Technology -
The design as a curved tube with mouthpiece/selector, tube valve and cap is the identical principle of operation.
Discussion - The technology for generating oscillation and expiratory pressure is identical to the predicate - K983308 - RC-Cornet.
## Non-clinical Testing
## Materials -
The materials in patient contact and the gas pathway are identical to the predicate - RC-Cornet, K983308. We have evaluated the materials and their type of relations to the patient and determined that they are characterized as:
- Surface Contact .
- . Mucosal membrane
- Duration of Use permanent (> 30 days) .
Discussion - We have been supplied identification of the materials used by the predicate. We have selected the identical materials and certify that the manufacturing processes are identical for the proposed device and the predicate K983308 - RC-Cornet. We provided Material Certifications with the predicate which has the identical indications for use, patient population, environment of use, type of patient contact.
## Comparative Bench Testing
We performed a number of performance tests which included:
- . Comparative Performance at 10, 20, and 40 lpm
- o Average Pressure
- Average Pressure Amplitude o
- Frequency O
- Average Flow Rate o
- . Cleaning validation
- . Age testing
- Mechanical / Drop test ●
The comparative bench testing demonstrated that the VibraPEP™ is equivalent to the predicate K983308 - RC-Cornet. See Table 1 above.
The comparative testing demonstrates that the proposed device is substantially equivalent to the predicate device.
## Discussion of Differences and Substantial Equivalence Conclusion
As detailed above, the indications for use, patient population, environment of use, technology or principle of operation, and performance have been demonstrated to be substantially equivalent to the predicate.
{7}------------------------------------------------
# 510(k) Summary
March 15, 2016
## Page 5 of 5
There are no differences between the proposed VibraPEP™ and the predicate – K983308 – RC-Cornet and based upon the comparative performance testing we can conclude that there are no new safety or effectiveness concerns and thus can determine then to be substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.