The InspirationTM Ventilator System is indicated for use with a wide range of patients from infant through adult, requiring respiratory support for a wide range of clinical conditions in hospital-type facilities and intra-hospital transport.
Device Story
Inspiration Ventilator System provides continuous respiratory support via pressure-based and volume-based mandatory and spontaneous breaths. Device comprises metallic enclosure with user interface, display screen, and mobile stand. Operator programs ventilation parameters using function keys and Push & Turn control. Microprocessor processes inputs to control pneumatic system, delivering air/oxygen to patient. System operates on AC, internal, or external battery power, including integral compressor for transport. Modes include V-CMV, V-SIMV, P-CMV, P-SIMV, SPONT, and PSV. Communications interfaces (Nurse Call, RS-232, Ethernet) transmit data to workstations/PCs. Used in hospitals and intra-hospital transport by clinicians to manage patient ventilation, potentially improving respiratory outcomes.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing and comparison of product features and specifications to predicate devices.
Indicated for infant through adult patients requiring respiratory support for various clinical conditions in hospital settings and during intra-hospital transport.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the text 'KD2112'. The text appears to be handwritten in black ink on a white background. The characters are slightly slanted and have a casual, informal style.
NOV 2 2 2002
# Section 16. 510(k) Summary
# Section 16.a Date Summary Prepared
01 March 2002
# Section 16.b Company Information
Establishment:
6a Liosban Business Park Tuam Road Galway, Ireland
eVent Medical Ltd.,
Official Correspondent:
Mike Griffiths CEO eVent Medical Ltd., 800 Grand Avenue, Suite C-3 Carlsbad, CA 92008 760 497-3553 760-434-1334 (Fax)
# Section 16.c Name of Device
| Proprietary: | <i>Inspiration</i> ™ Ventilator System |
|-----------------|-----------------------------------------|
| Common/Usual: | Respiratory Ventilator |
| Classification: | Continuous Ventilator (§868.5895/73CBK) |
## Section 16.d Equivalent Devices
Substantial equivalence to the following legally marketed predicate devices with the same or similar indications for use has been demonstrated by a comparison of product features as described in the labeling and promotional literature for predicate devices and the Inspiration ~ Ventilator System, as well as testing to accepted industry standards. The predicate devices are as follows:
- 1. Servo Ventilator 300, Siemens-Elema AB, K960010
- 2. 760 Ventilator, Puritan Bennett Inc., K984379
## Section 16.e Device Description
The Inspiration™ Ventilator System is housed in a metallic enclosure, comprising clear anodized aluminum with painted finish and is mounted on a mobile stand. User controls and display screen are located on the front face of the Ventilator. Display and controls operation is described in the Ventilator Settings and Controls section of the User Manual (See Section 14 of this Submission). Ventilation parameters are programmed by the operator, using the function and navigation keys plus the Push & Turn control device on the Ventilator front panel, in conjunction with selectable display screens.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the word "CONFIDENTIAL" in all caps. The word is written in a bold, sans-serif font. The letters are slightly distressed, giving the word a textured appearance.
The programmed data from the User Interface is processed by the main microprocessor and stored in ventilator non-volatile memory. The microprocessor uses the stored data in the control of breath delivery.
Oxygen and air supply connections are located on the back of the Ventilator. The patient breathing circuit connection is located below the front face of the unit, together with a pneumatic outlet for driving an optional accessory jet nebulizer. The Ventilator stand serves to support the main Ventilator unit, breathing circuit tubing and accessory equipment, facilitating connection to the Ventilator.
The Inspirations Ventilator System incorporates three communications interfaces, namely, a Nurse Call interface, an RS-232 and an Ethernet interface. The Nurse Call interface comprises normally open and normally closed relay contacts (changeover switch) acting simultaneously with an alarm condition. Both the RS232 and Ethernet interfaces enable the same data transmission to receiving terminals such as nurse and physician workstations and PC equipment.
### Section 16.f Intended Use
The Inspiration -- Ventilator System is intended for use with a wide range of patients from infant through adult, requiring respiratory support for a wide range of clinical conditions in hospital, hospital-type facilities and intra-hospital transport.
The intended use, patient population and environment of use are the same or similar to the predicate devices, the Servo Ventilator 300, Siemens-Elema AB and the 760 Ventilator, Puritan Bennett Inc.
#### Section 16.g Technological Characteristics
The Inspiration Wentilator System provides continuous ventilation to patients requiring respiratory support by means of pressure-based and volume-based mandatory and spontaneous breaths. The system operates on AC, internal battery, or external battery power and includes an integral compressor. This enables the system to operate both in the hospital environment and also during intra-hospital transport.
Ventilation modes are selected using the display screens and associated controls. The following ventilation modes are available:
Volume-based synchronized mandatory ventilation; V-CMV.
Volume-based synchronized intermittent mandatory ventilation: V-SIMV
Pressure-based synchronized mandatory ventilation; P-CMV
Pressure-based synchronized intermittent mandatory ventilation; P-SIMV
Spontaneous ventilation: SPONT
Pressure support: PSV
The User Manual, Section 14 of this submission, describes the procedures for selecting individual modes, plus performance specifications. Please refer to sections 3.2, 3.3, 8.5 of the User Manual.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "CONFIDENTIAL" in large, bold letters. The letters are slightly distorted, giving them a textured appearance. The word is oriented horizontally across the image.
The Inspiration ... Ventilator System consists of two major systems, the pneumatic system and the electrical/electronic system, which combine under software control to deliver respiratory support at operator-determined parameters. The pneumatic system, under microprocessor control, supplies air and oxygen to the patient system, external to the device. The electricallelectronic system inputs and supervises one of three alternative power sources to the unit and provides electronic control of the ventilator's pneumatic system components.
The Inspiration™ Ventilator System can operate with a range of standard ventilator accessories, such as breathing circuits, nebulizer and humidifier.
### Section 16.h Certification Statement
In accordance with the requirements of 21 CFR 807.87(j), the following certification is provided:
eVent Medical believes that all data and information submitted in this premarket notification are truthful and accurate and no material fact has been omitted.
Mike Griffiths Chief Executive Officer eVent Medical Ltd.
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
ood and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Mike Griffiths Chief Executive Officer eVent Medical Limited 800 Grand Avenue, Suite C3 Carlsbad, California 92008
Re: K021112
Trade/Device Name: Inspiration™ Ventilator System Regulation Number: 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: CBK Dated: August 20, 2002 Received: August 26, 2002
Dear Mr. Griffiths:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Griffiths
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (301) 594-4646 . Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Timothy A. Ulatowski
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
CONFIDENTIAL
#### INDICATIONS FOR USE STATEMENT 2.g
Applicant:
eVent Medical Ltd
510(k) Number:
Device Name:
InspirationTM Ventilator System
Indications for Use:
The InspirationTM Ventilator System is indicated for use with a wide range of patients from infant through adult, requiring respiratory support for a wide range of clinical conditions in hospital-type facilities and intra-hospital transport.
Yes (Per 21 CFR 801.109) Prescription Use:
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of CDRH, Office of Device Evaluation (ODE) (Per 21 CFR 801.109) (Optional Format 1-2-96)
FAUch
(Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices
KOZIIIZ 510(k) Number ._
Prescription Use
or
OTC Use
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.