NELLCOR PURITAN BENNETT, 760 VENTILATOR SYSTEM

K984379 · Puritan Bennett Corp. · CBK · Dec 10, 1998 · Anesthesiology

Device Facts

Record IDK984379
Device NameNELLCOR PURITAN BENNETT, 760 VENTILATOR SYSTEM
ApplicantPuritan Bennett Corp.
Product CodeCBK · Anesthesiology
Decision DateDec 10, 1998
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 868.5895
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The 760 Ventilator System is used to provide continuous ventilation to patients requiring respiratory support. This device is used for a wide range of patients from pediatric to adult and for a wide variety of clinical conditions. The 760 Ventilator is intended for use in hospitals and hospital type facilities which provide respiratory care for patients requiring respiratory support. The 760 Ventilator may be used during hospital and hospital type facility transport. The 760 Ventilator is not to be used in the presence of flammable anesthetics. The 760 Ventilator is intended for sale by or on the order of a physician only. This product is intended for operation by trained and qualified clinicians only and is intended for servicing by trained and qualified persons only.

Device Story

760 Ventilator is a critical care ventilator providing continuous respiratory support to pediatric and adult patients; used in hospitals and during facility transport. Device delivers mandatory and spontaneous breaths; adds Pressure Control Ventilation (PCV) and adjustable rise time/exhalation sensitivity in Pressure Support Ventilation (PSV) compared to predecessor. Performs respiratory mechanics calculations (auto-PEEP, patient resistance, compliance) via expiratory/inspiratory pause maneuvers. Operated by trained clinicians; output displayed on user interface for clinical decision-making regarding patient ventilation settings. Pneumatic design and breath delivery control algorithms largely inherited from 740 Ventilator; modifications implemented via software and additional user interface keypads.

Clinical Evidence

Bench testing only. Performance testing conducted per ASTM F1100-90 standard. Software verification and validation performed per FDA guidance. Environmental testing conducted per FDA reviewer guidance. No clinical data presented.

Technological Characteristics

Critical care ventilator; pneumatic breath delivery system. Features: PCV, PSV with adjustable rise time/exhalation sensitivity, respiratory mechanics calculation. Connectivity: User interface panel with keypads. Standards: ASTM F1100-90, EN60601-1:1990, EN 60601-1-2:1993, CAN/CSA C22.2 No. 601-1M90:1994, UL 2601-1:1994, EN 794-1. Software-controlled breath delivery and calculation algorithms.

Indications for Use

Indicated for pediatric and adult patients (tidal volume 0.04 - 2 L) requiring continuous invasive or non-invasive respiratory support in hospital or hospital-type settings. Contraindicated for use in the presence of flammable anesthetics.

Regulatory Classification

Identification

A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows a series of handwritten numbers and letters. The characters are 'K984379' and they are written in a slightly slanted and somewhat messy style. The contrast between the characters and the background is high, making them easily distinguishable. | SUBMITTER: | Nellcor Puritan Bennett Ireland,<br>a subsidiary of Mallinckrodt Inc., | |-------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------| | on behalf of: | Puritan-Bennett Corp.,<br>a subsidiary of Mallinckrodt Inc. | | DATE: | December 2, 1998 | | COMMON NAME: | Continuous Ventilator | | PROPRIETARY NAME: | 760 Ventilator System (760 Ventilator) | | CONTACT: | Robbie Walsh, Regulatory Affairs Manager<br>Nellcor Puritan Bennett Ireland,<br>Mervue,<br>Galway,<br>Ireland.<br>Tel: + 353 91 753771<br>Fax: + 353 91 770156 | | CLASSIFICATION: | Class II per 21 CFR 868.5895<br>Continuous Ventilator | ## PREDICATE DEVICES: Puritan-Bennett Corp. is claiming substantial equivalence to the following three predicate medical devices: | Predicate Device | 510(k) Number | Classification | |--------------------------------------------------|---------------|-----------------------------------------------------------| | Puritan Bennett Corp.,<br>740 Ventilator | K964540 | Class II,<br>Continuous Ventilator<br>per 21 CFR 868.5895 | | Puritan Bennett Corp.,<br>7200 Series Ventilator | K902506B | Class II,<br>Continuous Ventilator<br>per 21 CFR 868.5895 | | Puritan Bennett Corp.,<br>840 Ventilator | K970460 | Class II,<br>Continuous Ventilator<br>per 21 CFR 868.5895 | {1}------------------------------------------------ ## Device Description: A The device is a low-cost, critical care ventilator intended to provide continuous ventilation for pediatric and adult patients. The 760 Ventilator is the second in a planned series of ventilators to be developed by Puritan-Bennett Corp., the "700 Series Ventilators", with the first being the 740 Ventilator. As such the 760 Ventilator encompasses all the basic functionality and modes of operation of the 740 Ventilator. The major difference between the two ventilators is that the 760 Ventilator adds Pressure Control Ventilation (PCV) as a mandatory breath type and offers adjustable rise time factor and exhalation sensitivity setting in Pressure Support Ventilation (PSV). Additionally, the 760 Ventilator also offers the ability to perform respiratory mechanics calculations and maneuvers as a standard feature using the EXPIRATORY PAUSE (expiratory pause) key to calculate auto-PEEP and the INSPIRATORY PAUSE (inspiratory pause) key to calculate patient resistance and compliance. These modifications are implemented on the 760 Ventilator through additional functionality in software and by adding keypads on the User Interface panel. The pneumatic design, breath delivery control algorithms and the electrical circuitry (apart from the User Interface printed circuit board) have remained unchanged from 740 to 760 models. - Intended Use: B The intended use has remained unchanged from 740 to 760 Ventilators and is provided below; Purpose and function of device: - · The 760 Ventilator is intended to provide continuous ventilation to patients requiring respiratory support. - · This product is intended for a wide range of patients from pediatric to adult and for a wide variety of clinical conditions. Intended patient population: - The intended patient population includes pediatric and adult patients (tidal volume 0.04 - 2 L) who require continuous respiratory support. - . Intended for patient who require either invasive or non-invasive ventilation. Intended environment of use: - · The 760 Ventilator is intended for use in hospitals and hospital type facilities which provide respiratory care for patients requiring respiratory support. {2}------------------------------------------------ - · The 760 Ventilator may be used during hospital and hospital type facility transport - · The 760 Ventilator is not to be used in the presence of flammable anesthetics - · The 760 Ventilator is intended for sale by or on the order of a physician only. - This product is intended for operation by trained and qualified clinicians only and is intended for servicing by trained and qualified persons only. ## C Substantial Equivalence The intended use of the 760 Ventilator is the same as that for standard, currently marketed critical care ventilators. The materials and design of this device are similar to those of the predicate devices. The technical characteristics of the 760 Ventilator do not introduce new questions regarding safety or effectiveness of critical care ventilators. Furthermore, the labeling associated with the 760 Ventilator provides similar information as the predicate devices. Information provided in the 510(k) submission supports the determination of substantial equivalence. Software design and development, (including verification and validation testing, test and software quality procedures) was conducted using FDA's Guidance for the Content of Premarket Submissions for Software contained in medical devices, dated May 29 1998, as a guidance and per internal company requirements. Environmental testing was conducted using FDA's Reviewers Guidance for Premarket Notification Submissions, Nov. 1993 draft as a guideline and per internal company requirements. Performance testing was conducted using the ASTM F1100-90 standard as a guidance and per internal, company requirements. The 760 Ventilator device design and testing are also compliant with various voluntary, international standards including: EN60601-1:1990, EN 60601-1-2:1993, CAN/CSA C22.2 No. 601-1M90:1994, UL 2601-1:1994, EN 794-1 and 93/42/EEC Medical Device Directive. The combined testing and analysis of results provides assurance that the device meets its specifications and is safe and effective for its intended use. In summary Puritan-Bennett Corp. has demonstrated the 760 Ventilator to be safe and effective. This device is considered to be substantially equivalent to currently marketed devices which have been previously cleared by FDA. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image contains the text "Public Health Service". The text is in a simple, sans-serif font and is horizontally aligned. The text is likely part of a document or sign, indicating an affiliation with a public health organization. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 DEC 1 0 1998 Ms. Robbie Walsh Puritan-Bennett Corporation c/o Nellcor Puritan Bennett Ireland Mervue, Galway Ireland Re: K984379 760 Ventilator System Regulatory Class: II (two) Product Code: 73 CBK December 4, 1998 Dated: December 7, 1998 Received: Dear Ms. Walsh: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 A substantially equivalent determination assumes compliance to 895. with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {4}------------------------------------------------ Page 2 - Ms. Robbie Walsh This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Dittman J. Callahan Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## INDICATION FOR USE | 510(k) Number: | K984379 | |---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name: | 760 Ventilator System | | Indication for Use: | The 760 Ventilator System is used to provide continuous<br>ventilation to patients requiring respiratory support. This<br>device is used for a wide range of patients from pediatric to<br>adult and for a wide variety of clinical conditions. | Prescription Use: Yes (Per 21 CFR 801 109) Mark Kramer (Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices 1984379 510(k) Number _
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...