NELLCOR PURITAN BENNETT, 840 VENTILATOR SYSTEM (MODEL 840)
K970460 · Puritan Bennett Corp. · CBK · Apr 30, 1998 · Anesthesiology
Device Facts
| Record ID | K970460 |
| Device Name | NELLCOR PURITAN BENNETT, 840 VENTILATOR SYSTEM (MODEL 840) |
| Applicant | Puritan Bennett Corp. |
| Product Code | CBK · Anesthesiology |
| Decision Date | Apr 30, 1998 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 868.5895 |
| Device Class | Class 2 |
| Attributes | Therapeutic, Pediatric |
Intended Use
The 840 Series Ventilator is intended to provide continuous ventilation to patient’s requiring respiratory support. This product is intended for a wide range of patients from infant to adult and for a wide variety of clinical conditions. The 840 Ventilator System is used to provide continuous ventilation to patient's requiring respiratory support. This device is used for a wide range of patients from infant to adult and for a wide variety of clinical conditions.
Device Story
Critical care ventilator providing continuous respiratory support to infant, pediatric, and adult patients in hospital settings. Inputs include air and oxygen gas sources; user settings via touch-screen LCDs and rotary knob. Dual-microprocessor architecture (Breath Delivery Unit and GUI) ensures independent control and monitoring to prevent single-fault failure. Device delivers mandatory (volume or pressure-controlled) or spontaneous breaths (with/without pressure support) via proportional solenoid valves. Features three ventilation modes (A/C, SPONT, SIMV) and four breathing types (VC, PC, PS, NONE). Hierarchical 'smart' alarms monitor patient and technical status. Clinicians use GUI to view patient data and adjust settings. Benefits include precise respiratory support and integrated safety monitoring.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included biocompatibility analysis, software verification and validation per FDA guidance, environmental and electrical safety testing, and performance testing against international standards (EN60601-1, ISO 10651-1, etc.).
Technological Characteristics
Critical care ventilator; dual-microprocessor control (Breath Delivery Unit and GUI); proportional solenoid valves for gas delivery. Supports invasive/non-invasive ventilation. Connectivity: optional 804 Compressor and 802 Backup Power Source. Standards: EN60601-1, EN 60601-1-2, ISO 9703-1/2/3, EN 475, prEN 794-1, ISO 10651-1. User interface: dual touch-screen LCDs and rotary knob.
Indications for Use
Indicated for infant, pediatric, and adult patients (tidal volume 25-2500 mL) requiring continuous invasive or non-invasive respiratory support for various clinical conditions. Prescription use only.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
Predicate Devices
- Puritan-Bennett Corp. 7200 Series Ventilator (K902506/B)
- Bear Medical Systems Inc. Bear 1000 Ventilator (K912619)
- Siemens Corp. Servo 300 Ventilator (K902859)
- Dräger Evita 4 Ventilator (K961687)
Related Devices
- K181170 — Trilogy Evo Universal · Respironics, Inc. · Jul 18, 2019
- K964540 — NELLCOR PURITAN BENNETT, 740 VENTILATOR SYSTEM · Puritan Bennett Corp. · Nov 14, 1997
- K131252 — PURITAN BENNETT 980 SERIES VENTILATOR SYSTEM · Covidien · Feb 24, 2014
- K222644 — Nihon Kohden NKV-440 Ventilator System · Nihon Kohden Orangemed, Inc. · May 31, 2023
- K041775 — GE DATEX-OHMEDA ENGSTROM CARESTATION · Datex-Ohmeda, Inc. · Oct 26, 2004
Submission Summary (Full Text)
{0}
K970460
APR 30 1998
# SUMMARY OF SAFETY AND EFFECTIVENESS
SUBMITTER: Puritan-Bennett Corp.
DATE: January 15, 1997
COMMON NAME: Continuous Ventilator
PROPRIETARY NAME: 840 Ventilator System (840 Ventilator)
CONTACT: Ann-Marie Butler
Regulatory Affairs Project Manager
Puritan-Bennett Corp.
2200 Faraday Ave.
Carlsbad, CA 92008
(619) 603-5300 (phone)
(619) 603-5901 (fax)
CLASSIFICATION: Class II per 21 CFR 868.5895
Continuous Ventilator
## PREDICATED DEVICES:
Puritan-Bennett Corp. is claiming substantial equivalence to the following predicate medical devices:
| Predicate Device | 510(k) Number | Classification |
| --- | --- | --- |
| Puritan-Bennett Corp.
7200 Series Ventilator | K902506/B (7200ae) | Continuous, Critical
Care Ventilator |
| Bear Medical Systems Inc.
Bear 1000 Ventilator | K912619 | Continuous, Critical
Care Ventilator |
| Siemens Corp.
Servo 300 Ventilator | K902859 | Continuous, Critical
Care Ventilator |
| Dräger
Evita 4 Ventilator | K961687 | Continuous, Critical
Care Ventilator |
{1}
# 1. DEVICE DESCRIPTION:
The device is a critical care ventilator intended to provide continuous ventilation for infant, pediatric, and adult patients.
The 840 Ventilator's graphic user interface (GUI) includes ventilator status keys and indicators, symbols and abbreviations with associated definitions, ventilator settings, and patient data. User input to the 840 Ventilator is provided by means of two monochrome or color touch-screen LCDs and a rotary knob. Feedback to the user is accomplished by the LCD displays and LED indicators.
The alarms associated with the 840 Ventilator meet and exceed alarm standards for modern critical care ventilators and have been developed in compliance with ISO 9703-1, ISO 9703-2, ISO 9703-3, and EN 475. These alarms are within the classification of "smart" alarms as per the ISO 9703 series of standards. The 840 Ventilator provides both non-technical (patient related) and technical (ventilator related) alarms. These alarms are arranged in a hierarchical structure with high, medium, and low urgency categories.
The 840 Ventilator includes two microprocessors: 1) the breath delivery unit (BDU) microprocessor which controls ventilation and 2) the GUI microprocessor which manages the user interface and monitors ventilator and patient data. Each microprocessor verifies that the other's instructions are being carried out properly. Using two independent microprocessors in this fashion prevents a single fault from causing a simultaneous failure of controlling and monitoring functions.
Oxygen and air connect directly to the BDU, supplying gas to each of two proportional solenoid (PSOL) valves. The optional 804 Compressor provides pressurized air to the BDU and can be used in place of wall or bottled air. The optional 802 Backup Power Source provides DC power to the BDU power supply in the event that AC power is lost.
The 840 Ventilator supplies mandatory or spontaneous breaths with a preset oxygen concentration. A mandatory (or assisted) breath can be pressure or volume controlled. Volume controlled breaths provide the patient with a preset tidal volume, peak flow, waveform, and oxygen concentration. Pressure controlled breaths provide preset pressure, inspiratory time and $\% \mathrm{O}_2$. A spontaneous breath allows inspiratory flows of up to $200\mathrm{L / min}$, with or without pressure support.
The 840 Ventilator offers three modes of ventilation:
- Assist/control (A/C), which consists entirely of mandatory breaths.
- Spontaneous (SPONT), which consists entirely of spontaneous breaths.
- Synchronous intermittent mandatory ventilation (SIMV), which can include both mandatory and spontaneous breaths.
{2}
The 840 Ventilator offers four breathing types:
- Volume-Controlled (VC), in which the ventilator delivers a preset tidal volume at a preset peak flow, flow waveform, and $\% \mathrm{O}_2$.
- Pressure-Controlled (PC) in which the ventilator delivers a preset pressure, inspiratory time and $\% \mathrm{O}_2$.
- Pressure Support (PS), in which spontaneous effort is augmented by a preset PS value and $\% \mathrm{O}_2$.
- No Pressure Support (NONE), in which spontaneous effort is not augmented by a preset PS value, although $\% \mathrm{O}_2$ is preset.
## 2. INTENDED USE:
**Purpose and function of device:**
- The 840 Series Ventilator is intended to provide continuous ventilation to patient’s requiring respiratory support.
- This product is intended for a wide range of patients from infant to adult and for a wide variety of clinical conditions.
**Intended patient population:**
- The intended patient population includes infant, pediatric, and adult patients (tidal volume 25 - 2500 mL) who require continuous respiratory support.
- Intended for patient who require either invasive or non-invasive ventilation.
**Intended environment of use:**
- The 840 Ventilator is intended for use in hospitals and hospital type facilities which provide respiratory care for patients requiring respiratory support.
- The 840 Ventilator may be used during hospital and hospital type facility transport provided that electrical power and compressed gas are supplied.
- The 840 Ventilator is not to be used in the presence of flammable anesthetics.
- The 840 Ventilator is intended for sale by or on the order of a physician only.
- This product is intended for operation by trained and qualified clinicians only and is intended for servicing by trained and qualified persons only.
{3}
# 3. SUBSTANTIAL EQUIVALENCE:
The intended use of the 840 Ventilator is the same as that for standard, currently marketed critical care ventilators. The materials and design of this device are similar to those of the predicate devices. The technical characteristics of 840 Ventilator do not introduce new questions regarding safety or effectiveness of critical care ventilators. Furthermore, the labeling associated with the 840 Ventilator provides similar information as the predicate devices.
Information provided in the 510(k) submission supports the determination of substantial equivalence. Biocompatibility analysis and laboratory testing demonstrate Puritan-Bennett's 840 Ventilator product to be safe for its intended use. Software design and development, (including verification and validation testing, test and software quality procedures) was conducted using FDA's Reviewers Guidance of Medical Device Software Submissions, 15 Dec. 1995 draft as a guidance and per internal company requirements. Environmental and electrical testing was conducted using FDA's Reviewers Guidance for Premarket Notification Submissions, Nov. 1993 draft as a guideline. Performance testing was conducted using FDA's Reviewer Guidance for Ventilators draft as a guidance and per internal, company requirements. The 840 Ventilator device design and testing are also compliant with various voluntary, international standards including: EN60601-1:1990, EN 60601-1-2:1993, CAN/CSA C22.2 No. 601-1M90:1994, UL 2601-1:1994, prEN 794-1, ISO 10651-1, and 93/42/EEC MDD.
The combined testing and analysis of results provides assurance that the device meets its specifications and is safe and effective for its intended use.
In summary Puritan-Bennett Corp. has provided evidence that shows the 840 Ventilator to be safe and effective. This device is considered to be substantially equivalent to currently marketed devices which have been previously cleared by FDA.
{4}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
APR 30 1998
Ms. Ann-Marie Butler
Puritan-Bennett Corp.
2200 Faraday Avenue
Carlsbad, CA 92008
Re: K970460
840 Ventilator System
Regulatory Class: II (two)
Product Code: 73 CBK
Dated: March 6, 1998
Received: April 1, 1998
Dear Ms. Butler:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}
Page 2 - Ms. Ann-Marie Butler
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,

Thomas J. Callahan, Ph.D.
Director
Division of Cardiovascular,
Respiratory, and Neurological Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{6}
# INDICATION FOR USE
510(k) Number: K970460
Device Name: 840 Ventilator System
Indication for Use: The 840 Ventilator System is used to provide continuous ventilation to patient's requiring respiratory support. This device is used for a wide range of patients from infant to adult and for a wide variety of clinical conditions.
Prescription Use: Yes
(Per 21 CFR 801.109)

(Division Sign-Off)
Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number
Puritan-Bennett Corp.
840 Ventilator System
510(K) Submission