K031982 · Birdsong Medical Devices, Inc. · CAF · Apr 2, 2004 · Anesthesiology
Device Facts
Record ID
K031982
Device Name
BIRDSONG LEVERHALER, MODEL LH-1000
Applicant
Birdsong Medical Devices, Inc.
Product Code
CAF · Anesthesiology
Decision Date
Apr 2, 2004
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5630
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The LeverHaler Spacer is a spacer device that is designed to effectively deliver respirable medical aerosols produced by most MDI's to a patient during inhalation. This product is intended for use in the treatment of lung disease. The device has been designed for use by all patients who have been prescribed MDI treatment by their physician. It is intended for use in both the hospital and homecare environments.
Device Story
LeverHaler Spacer is an injection-molded polycarbonate mixing chamber for FDA-approved metered dose inhalers (MDIs); facilitates aerosol delivery for lung disease treatment. Features a lever actuator to assist patients with limited hand strength, coordination, or dexterity; allows manual MDI activation or lever-assisted actuation. Used in hospitals and homecare environments by patients under physician prescription. Device provides effective aerosol concentration for inhalation; does not generate aerosols itself. No software, alarms, or detection capabilities. Benefits include improved ease of use for patients with physical limitations and consistent aerosol delivery.
Clinical Evidence
No clinical testing performed. Bench testing only. In vitro comparative study performed per FDA 1993 guidance for nebulizers/MDIs/spacers. Evaluated particle size distribution, dose output, MMAD, GSD, plume analysis, and life testing using three drugs. Results showed LeverHaler performance equivalent or superior to predicate devices.
Technological Characteristics
Injection-molded polycarbonate housing; silicone mouthpiece and valve. Mechanical operation; no energy source. Mixing chamber volume ~170 ml. Non-sterile. Single-patient use. No software or connectivity.
Indications for Use
Indicated for patients of all ages prescribed MDI treatment for lung disease. Used in hospital and homecare settings.
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the text "K031982" in a handwritten style. The text is oriented diagonally, with the characters slightly tilted to the right. The handwriting appears somewhat casual and the numbers are clearly distinguishable.
# Birdsong Medical LeverHaler Spacer 510(k) Summary:
In accordance with 21 CFR section 807.92 BirdSong Medical is submitting the following safety and effectiveness summary.
- 1) Submitter Information
Eric Fugelsang President BirdSong Medical Devices, Inc. 1456 Rt. 22, Suite A103 Brewster, NY 10509
- 2) Name of Device
Proprietary Name: LeverHaler Spacer Common Name is Handheld Spacer Classification Name: Nebulizer Accessory
- 3) Substantially equivalent to: OptiHaler (Healthscan Products, Inc.), K911807.
- 4) Device Description and System Overview:
The LeverHaler Spacer (Spacer) is an injection molded polycarbonate spacer device intended for use with FDA approved metered dose inhalers (MDI's). The MDI canister fits through a slot in a lever located on the top of the spacer. Activation of the MDI canister can be done by manually pressing down on the MDI (typical activation), or the MDI may be actuated by depressing the Lever Actuator down towards the top of the spacer body.
Image /page/0/Picture/11 description: The image shows a diagram of a medical device, specifically an inhaler. The different parts of the inhaler are labeled, including the mouthpiece, lever actuator, MDI (metered dose inhaler), air inlets, air outlets, housing, and chamber. The diagram provides a clear visual representation of the inhaler's components and their arrangement.
Figure 1: Diagram of LeverHaler Spacer
{1}------------------------------------------------
The device is made from polycarbonate injection molded plastic and is designed to The device is made from polycurson injoning. When not in use, up to two MDI canisters can be stored inside of the LeverHaler Chamber.
The product is sold non-sterile, and is a single-patient device intended to be used The product is sold non-sichie, and to a sing or cleaning and use of the device. All for up to one year. Easoning couring of this device are identical to a defined predicate device.
This device does not generate aerosols. Its purpose is to provide an effective mixing chamber for the aerosols produced by a metered dose inhaler to assure milking chamber for the doneentration for inhalation by the patient. This is a prescription device.
The device has no detection capabilities. It is mechanical and has no alarm functions or capabilities. There is no software integrated or used in conjunction with this device.
Product testing has been completed according to the "Reviewer Guidance for Nebulizers, Metered Dose Inhalers, Spacers and Actuators (10-01-93)." Product testing performed by Birdsong Medical provides clear proof that the LeverHaler Spacer is superior to the MDI and is comparable or superior to the predicate devices tested. The premarket notification submitted to FDA contains a full discussion of product testing, which includes delivered aerosol potency, MMAD, GSD, particle distribution, retained aerosols (Plume Analysis) and life testing. All tests included comparative testing of the two predicate devices and the LeverHaler Spacer.
### Design Considerations and Operation of the Device:
The device is intended to be used with FDA approved MDI drugs for treatment of lung disease. The Lever Actuator is designed for patients with limited hand strength, coordination or dexterity. The MDI can be actuated easily with the Lever or may be actuated by depressing the MDI downwards (typical MDI activation).
#### Testing:
All product and comparative in vitro testing performed was based upon the requirements of the "Guidance for Nebulizers, Metered Dose Inhalers, Spacers and Actuators: Reviewer Guidance 10-01-93." Testing was conducted by an independent medical device manufacturer and test laboratory contracted by Birdsong Medical. All tests were performed according to documented test protocol.
The following tests were completed for the LeverHaler Spacer, and the predicate devices: OptiHaler and AeroChamber (specified in this premarket notification):
- Particle size distribution (using 3 drugs) .
- Dose output testing (drug quantity and total mass using 3 drugs) .
- Mean median aerodynamic diameter (MMAD) .
- Geometric standard deviation (GSD) .
10
{2}------------------------------------------------
- Plume Analysis Testing .
- Single patient use testing (life testing) .
No clinical testing was performed on this product.
Software Validation: Not applicable: there is no software in this product.
Sterilization Validation: Not applicable: this product is sold and used as a non-sterile product.
Biocompatibility: All materials used in this device are incorporated in other predicate devices, as well as nebulizers, connectors and mouthpieces currently sold in the uevices, as well as nobles on appropriate for the intended use described herein.
## Comparative Product Matrix
## Table 2: 510(k) Summary - Substantial Equivalence to Predicate Devices
| item: | Characteristic: | Birdsong<br>Medical | Healthscan<br>Products, Inc. | Monaghan<br>Medical |
|-------|--------------------------------|----------------------------------|----------------------------------|----------------------------------|
| | Model: | LeverHaler<br>Spacer | OptiHaler | AeroChamber |
| 4. | Spacer Material | Clear<br>Polycarbonate | Clear<br>Polycarbonate | Clear Polyester |
| 5. | Mouthpiece and Valve | Silicone | Polycarbonate | Polyethelene<br>and Silicone |
| 5. | Single Patient Use? | Yes | Yes | Yes |
| 6. | Sterility? | Non-Sterile | Non-Sterile | Non-Sterile |
| 7. | Ventilator Use? | No | No | Yes |
| 8. | Maximum Length of Use: | Single patient<br>up to 365 days | Single patient<br>Not stipulated | Single patient<br>Not stipulated |
| 8. | Effective Size of mixing area: | Approximately<br>170 ml | Approximately<br>90 ml | Approximately<br>140 ml |
| 9. | Method of Operation: | Mechanical | Mechanical | Mechanical |
| 10. | Prescription Device? | Yes | Yes | Yes |
| 11. | Provided as Kit? | No | No | No |
| 12. | Port Protection? | Yes | Yes | Yes |
| | Premarket Submission: | Pending | (K911807) | (K930574)<br>(K884803) |
The Birdsong Medical Spacer met all design requirements Conclusions of all Testing: and passed all validation and comparative product testing. The device is manufactured from the identical materials used in a predicate device noted in the 510(k) submission. LeverHaler performance in all in vitro tests was equivalent or superior to the comparative predicate device noted in the table above.
Based upon these results, it is our conclusion that the LeverHaler Spacer is as safe, as effective and performs as well as or better than the legally marketed predicate OptiHaler Spacer device used in comparative product testing.
75
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a snake winding around it, overlaid on three horizontal bars. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR = 2 2004
Mr. Eric Fugelsang President Birdsong Medical Devices, Incorporated 1456 Rt. 22. Suite A103 Brewster, NY 10509
Re: K031982
Trade/Device Name: Leverhaler Spacer Regulation Number: 868,5630 Regulation Name: Nebulizer Regulatory Class: II Product Code: CAF Dated: February 13, 2004 Received: February 13, 2004
Dear Mr. Fugelsang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Fugelsang
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### INDICATIONS FOR USE
[510(k)] Number: K031982
Device Name: Birdsong Medical Devices LeverHaler Spacer
Indications For Use:
The LeverHaler Spacer is a spacer device that is designed to effectively deliver respirable medical aerosols produced by most MDI's to a patient during inhalation. This product is intended for use in the treatment of lung disease. The device has been designed for use by all patients who have been prescribed MDI treatment by their physician. It is intended for use in both the hospital and homecare environments.
Prescription Use V (Part 21 CFR 801 Subpart D)
AND/OR
Over-the-Counter Use (21 CFR 807 Subpart €)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
fthuh
Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Denta
510(k) Number: k031982
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.