INOMAX DSIR (DELIVERY SYSTEM)

K110344 · Ino Therapeutics · MRN · Jun 24, 2011 · Anesthesiology

Device Facts

Record IDK110344
Device NameINOMAX DSIR (DELIVERY SYSTEM)
ApplicantIno Therapeutics
Product CodeMRN · Anesthesiology
Decision DateJun 24, 2011
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 868.5165
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The INOmax DS delivery system delivers INOmax® (nitric oxide for inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It may be used with most ventilators. The INOmax DS provides continuous integrated monitoring of inspired O2, NO2, and NO, and a comprehensive alarm system. The INOmax DS incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source. The INOmax DS includes a backup NO delivery capability that provides a fixed flow of 250 mL/min of NO which along with user supplied 10 L/min of oxygen provides 20 ppm in the gas flow to a patients breathing circuit. It may also use the INOblender for backup. The target patient population is controlled by the drug labeling for INOmax and is currently neonates. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates.

Device Story

Device delivers nitric oxide (NO) gas into patient breathing circuits; tracks ventilator waveforms via injector module to provide synchronized, proportional dosing. Dual-channel design: first channel manages delivery (CPU, flow controller, injector); second channel monitors gas concentrations (NO, NO2, O2) and manages alarms. Includes backup delivery mode (fixed 250 mL/min NO flow). Used in NICU and transport; operated by clinicians. Provides continuous monitoring of inspired gases; battery-powered for 6 hours. Output informs clinician of gas concentrations and delivery status; enables precise titration of NO therapy to benefit neonates with respiratory conditions. This submission adds compatibility with Vapotherm Precision Flow system.

Clinical Evidence

No clinical studies were required to support substantial equivalence. Evidence consists of bench testing validating compatibility with the Vapotherm Precision Flow system, confirming O2 dilution, NO delivery accuracy, and NO2 generation levels remain within specifications.

Technological Characteristics

Dual-channel architecture (delivery CPU/flow controller vs. monitoring CPU/gas cells). Components include delivery unit, blender, stand/cart, and NO gas tanks. Battery-powered (6 hours). Connectivity includes ventilator waveform tracking. Complies with FDA guidance for Nitric Oxide Delivery Apparatus (January 24, 2000).

Indications for Use

Indicated for neonates requiring nitric oxide inhalation therapy. Used in NICU and transport settings.

Regulatory Classification

Identification

The nitric oxide administration apparatus is a device used to add nitric oxide to gases that are to be breathed by a patient. The nitric oxide administration apparatus is to be used in conjunction with a ventilator or other breathing gas administration system.

Special Controls

The special control for this device is FDA's “Guidance Document for Premarket Notification Submissions for Nitric Oxide Administration Apparatus, Nitric Oxide Analyzer, and Nitrogen Dioxide Analyzer.”

The special control developed by the agency is a guidance document, entitled Guidance Document for Premarket Notification Submissions for Nitric Oxide Delivery Apparatus, Nitric Oxide Analyzer and Nitrogen Dioxide Analyzer. This guidance document identifies the risks associated with these types of devices and contains information that will help manufacturers address those risks. This document is available on FDA's website at http://www.fda.gov/cdrh/ode/1157.pdf.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Appendix A – K110344 Section 5; Summary Additional Information Response. :: · . : . . :: # 510(k) Summary JUN 2 4 2011 | In accordance with 21 CFR 807.92 the following summary of information is provided: | | |------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | <b>Date:</b> | December 16, 2010 | | <b>Submitter:</b> | INO Therapeutics/Ikaria | | | 2902 Dairy Drive | | | Madison, Wisconsin 53718 | | <b>Primary Contact Person:</b> | Larry Lepley | | | Associate Director, Regulatory Affairs | | | INO Therapeutics/Ikaria | | | T: 608-226-3415 | | | F: 608-226-3402 | | <b>Secondary Contact Person:</b> | David Trueblood | | | Director, Regulatory Affairs | | | INO Therapeutics/Ikaria | | | T: 608-395-3910 | | | F: 608-226-3402 | | <b>Device: Trade Name:</b> | INOmax DS (Delivery System) | | <b>Common/Usual Name:</b> | Nitric Oxide Administration Apparatus (primary) | | | Nitric Oxide Administration Apparatus, Back-up System | | | Nitric Oxide Analyzer | | | Nitrogen Dioxide Analyzer | | <b>Classification Names:</b> | Apparatus, Nitric Oxide Delivery, or Apparatus, Nitric Oxide<br>Backup Delivery, Class II - 21 CFR 868.5165 | | <b>Product Code:</b> | MRN (Primary), MRQ, MRP | | <b>Predicate Device(s):</b> | INOmax DSIR K061901, K070867, K071516, K080484,<br>K081691, K090958, K092545, K093922 | | <b>Device Description:</b> | The INOmax DSIR uses a "dual-channel" design to ensure the<br>safe delivery of INOmax. The first channel has the delivery CPU<br>the flow controller and the injector module to ensure the accurate<br>delivery of NO. The second channel is the monitoring system,<br>which includes a separate monitor CPU, the gas cells (NO, NO2,<br>and O2 cells) and the user interface including the display and<br>alarms. The dual-channel approach to delivery and monitoring<br>permits INOmax delivery independent of monitoring but also<br>allows the monitoring system to shutdown INOmax delivery if it<br>detects a fault in the delivery system such that the NO<br>concentration could become greater than 100 ppm. | {1}------------------------------------------------ Appendix A - K110344 Section 5; Summary Additional Information Response. Intended Use: The INOmax DS delivery system delivers INOmax® (nitric oxide of inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It may be used with most ventilators. The INOmax DS provides continuous integrated monitoring of inspired O2, NO2, and NO, and a comprehensive alarm system. The INOmax DS incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source. The INOmax DS includes a backup NO delivery capability that provides a fixed flow of 250 mL/min of NO which along with user supplied 10 L/min of oxygen provides 20 ppm in the gas flow to a patients breathing circuit. It may also use the INOblender for backup. The target patient population is controlled by the drug labeling for INOmax and is currently neonates. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates. Technology: All revisions of INOmax DSIR utilize component technology to deliver Nitric Oxide gas to the patient. The components consist of the Delivery System unit, the blender, a stand/cart and the NO gas tanks. Determination of Summary of Non-Clinical Tests: Substantial Equivalence: The testing concluded four requirements necessary for the operation of the INOmax DSrg and Vapotherm Precision Flow system to be compatible: O2 dilution The measured O2 was within ± 4% v/v of the expected value (see Pass/Fail Table on page 8; does not include tolerance of the Precision Flow). INOmax DSm delivery accuracy The measured values were + 15% of setting or 1.3 ppm whichever is greater which is within specification of the INOmax DSIR (see Pass/Fail Table on page 8). NO2 generation The maximum generated NO2 was at 80 ppm and 100% oxygen, as would be expected. NO2 levels did not exceed 1.3 ppm on any of the flow settings (per INOmax label < 3 ppm on 100% FiO2 and any dose setting). The maximum generated NO2 at 40 ppm {2}------------------------------------------------ Appendix A - K110344 Section 5; Summary Additional Information Response, and 60% oxygen was 0.4 ppm (worse case per FDA guideline of < 1 ppm on 60% FiO2 at 40 ppm dose setting). #### Backup Delivery The delivered NO during backup delivery was within 5% (see Pass/Fail table, NO Measured section) of the INOmax DSIR specification at flow rates above 5 L/min. Measured NO2 levels did not exceed 0.2 ppm on any settings tested and 100% oxygen (per INOmax label < 3 ppm on 100% FiO2 and any dose setting). The measured O2 was within + 3% v/v of the expected value (see Pass/Fail table 100% Oxygen section of test report). The INOmax DS revision complies to "Guidance Document for Premarket Notification Submissions for Nitric Oxide Delivery Apparatus, Nitric Oxide Analyzer and Nitrogen Dioxide Analyzer" (January 24, 2000) FDA-1157 The results of testing validated that the additional interface to the Vapotherm Precision Flow system did not alter the safety or effectiveness of the INOmax DS operation. ### Summary of Clinical Tests: The subject of this premarket submission, INOmax DSJR interfaced to Vapotherm Precision Flow did not require clinical studies to support substantial equivalence. Comparison of the Similarities and Differences with the Predicate Device. Other than the item listed below, all functional/technological characteristics remain the same between the reviewed INOmax DSIR and the predicate INOmax DSJR. #### Similarities: No functional or technological change was made to INOmax DSJR for this 510(k) submission. The INOmax DS18 submitted for this review retains the same as the predicate device(s). #### Differences: Labeling change; listing and connection criteria for Vapotherm Precision Flow as an accepted interface for INOmax DSIR to be added in the INOmax DSIR Instructions for Use (IFU). #### Conclusion: INO Therapeutics/Ikaria considers the INOmax DSB to be as safe, as effective, and performance is substantially equivalent to the predicate device. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract image of three human profiles facing to the right, with a wavy line underneath them. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Mr. Larry Lepley Associate Director, Regulatory Affairs INO Therapeutics 2902 Dairy Drive Madison, Wisconsin 53718 JUN 2 4 2011 Re: K110344 Trade/Device Name: INOmax DS Regulation Number: 21 CFR 868.5165 Regulation Name: Nitric Oxide Administration Apparatus Regulatory Class: II Product Code: MRN Dated: June 2, 2011 Received: June 3, 2011 Dear Mr. Lepley: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Mr. Lepley Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Anthony D. Watson Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## 510(k) Number (if known): INOmax DS Device Name: Indications for Use: The INOmax DS delivery system delivers INOmax® (nitric oxide for inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It may be used with most ventilators. The INOmax DS provides continuous integrated monitoring of inspired 02, NO2 and NO, and a comprehensive alarm system. The INOmax DS incorporates a battery that provides up to 6 hours of uninterrupted NO delivery in the absence of an external power source. The INOmax DS includes a backup NO delivery capability that provides a fixed flow of 250 mL/min of NO which along with user supplied 10 L/min of oxygen provides 20 ppm in the gas flow to a patients breathing circuit. It may also use the INOblender for backup. The target patient population is controlled by the drug labeling for INOmax and is currently neonates. The primary targeted clinical setting is the Neonatal Intensive Care Unit (NICU) and secondary targeted clinical setting is the transport of neonates. Prescription Use_X____________________________________________________________________________________________________________________________________________________________ AND/OR Use (Part 21 CFR 801 Subpart D) Over-The-Counter (Part 21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) L. Schulther (Division Sign-Uff) (Division Sign-On) Division of Anesthesiology, General Hospital Division of Anesthesion of Boxicos ാസ്ട്രിംസ് 6. Infection Control, Dental Devices 510(k) Number: K110344
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...