KYOLING CPR Mask with Oxygen Port and Without Oxygen Port
Applicant
Hangzhou Jinlin Medical Appliances Co., Ltd.
Product Code
CBP · Anesthesiology
Decision Date
Jul 9, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5870
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CPR Mask with Oxygen Port is designed to assist in providing immediate life support (mouth to mask ventilation) to health emergency victims requiring oxygen support (inhalation) or cardiopulmonary resuscitation (CPR) rescue techniques. The CPR Mask with Oxygen Port is for prescription use. The CPR Mask without Oxygen Port is designed to assist in providing immediate life support (mouth to mask ventilation) to health emergency victims requiring cardiopulmonary resuscitation (CPR) rescue techniques. The resuscitator without oxygen port is for over-the-counter use.
Device Story
Device is a manual CPR mask used in emergency situations to provide mouth-to-mask ventilation. Components include a transparent dome, one-way filtered valve, head strap, and optional oxygen port. Device acts as a physical barrier between rescuer and victim to prevent direct contact while facilitating an airtight seal for simultaneous mouth and nose ventilation. Operated by rescuers (laypeople or professionals) in emergency settings. Output is the delivery of rescue breaths to the victim. Benefits include reduced risk of cross-contamination and improved ventilation efficiency during resuscitation.
Clinical Evidence
Bench testing only. No clinical or animal studies were conducted. Performance testing verified inspiratory and expiratory resistance against AS-4259-1995 standards, with all results <5 cmH2O. Biocompatibility testing performed per ISO 10993-5 and ISO 10993-10.
Technological Characteristics
Materials: PVC (mask), silicone (one-way valve), non-woven (strap). Dimensions: 99.5mm x 122mm. Connectivity: None. Sterilization: Not specified. Standards: ISO 5356-1:2004 (connectors), AS 4259-1995 (resuscitation performance), ISO 10993-5/10 (biocompatibility).
Indications for Use
Indicated for health emergency victims requiring oxygen support or cardiopulmonary resuscitation (CPR) rescue techniques. Prescription version for oxygen support/CPR; OTC version for CPR.
Regulatory Classification
Identification
A nonrebreathing valve is a one-way valve that directs breathing gas flow to the patient and vents exhaled gases into the atmosphere.
Genuine First Aid CPR Face Mask without oxygen port (K112126)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol featuring three stylized human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 9, 2015
Hangzhou Jinlin Medical Appliances Co., Ltd. c/o Mr. Mike Gu Regulatory Affairs Manager OSMUNDA Medical Device Consulting Co., Ltd No.9, 16th Ave. Hangzhou Economic and Technological Development Zone, Hangzhou, Zhejiang Province 310018. CHINA
Re: K142764
Trade/Device Name: KYOLING CPR Mask with Oxygen Port, KYOLING CPR Mask without Oxygen Port Regulation Number: 21 CFR 868.5870 Regulation Name: Non-Rebreathing Valve Regulatory Class: II Product Code: CBP Dated: June 1, 2015 Received: June 8, 2015
Dear Mr. Gu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Page 2 - Mr. Mike Gu
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K142764
Device Name
KYOLING CPR Mask With Oxygen Port
### Indications for Use (Describe)
The CPR Mask with Oxygen Port is designed to assist in providing immediate life support (mouth to health emergency victims requiring oxygen support (inhalation) or cardiopulmonary resuscitation (CPR) rescue techniques. The CPR Mask with Oxygen Port is for prescription use.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{4}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known) K142764
### Device Name
KYOLING CPR Mask Without Oxygen Port
### Indications for Use (Describe)
The CPR Mask without Oxygen Port is designed to assist in providing immediate life support (mouth to health emergency victims requiring cardiopulmonary resuscitation (CPR) resuscitator without oxygen port is for overthe-counter use.
Type of Use (Select one or both, as applicable)
] Prescription Use (Part 21 CFR 801 Subpart D)
|× Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{6}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{7}------------------------------------------------
# 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
#### ـــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــ SUBMITTER
Hangzhou Jinlin Medical Appliances Co., Ltd.
No.9, 16th Ave. Hangzhou Economic and Technological
Development Zone, 310018, Hangzhou, Zhejiang Province, China
Phone: +86-571-86911905
Fax: +86-571-86840897
| Primary Contact Person: | Mike Gu |
|----------------------------|-----------------------------------------------------|
| | Regulatory Affairs Manager |
| | OSMUNDA Medical Device Consulting Co., Ltd |
| | Tel: (+86) 20-6232 1333 |
| | Fax: (+86) 20-8633 0253 |
| Secondary Contact Person: | Ms. Lydia HE |
| | Hangzhou Jinlin Medical Appliances Co., Ltd |
| Date Prepared: | June 01, 2015 |
| II. DEVICE | |
| Device trade name: | KYOLING CPR Mask With Oxygen Port(Prescription Use) |
| | KYOLING CPR Mask Without Oxygen Port(OTC Use) |
| Common/Usual Name: | Emergency CPR Mask |
| Classification Names | Valve, Non-Rebreathing |
| Regulation classification: | 21 CFR 868.5870 |
| Panel: | Anesthesiology |
| Regulation Class: | II |
{8}------------------------------------------------
СВР Product Code:
III. Prescription Use: PREDICATE DEVICE
Brand MedSource CPR Mask, MedSource International, LLC, K081516
OTC Use:
Genuine First Aid CPR Face Mask without oxygen port,, Genuine First Aid LLC, K112126
These predicates have not been subject to a design-related recall.
No reference devices were used in this submission.
#### IV. DEVICE DESCRIPTION
The KYOLING CPR mask is used during cardiopulmonary resuscitation procedures, which are used in emergency situations to supply oxygen and produce blood flow in the heart and lungs.
The CPR mask is made of PVC and one-way valve made of medical grade K-resin, the mask is used for mouth-to-mask breathing, it provides a physical barrier between the rescuer and victim, eliminating direct contact of the rescuer's lips with the unknown subject; and also it promotes an airtight seal to the face allowing ventilation through both the mouth and nose simultaneously. The KYOLING CPR Mask includes transparent dome, universal breathing tube, one-way filtered valve, head strap, and with Oxygen Port or without Oxygen Port.
# Specification
CPR Mask with Oxygen Port: The physical size for the device is 99.5mm in height* 122mm in length;
- Inspiratory resistance: <5 cmH2O (at 50 L/min);
- Expiratory resistance: <5 cmH₂O (at 50 L/min)
CPR Mask without Oxygen Port: The physical size for the device is 99.5mm in height * 122mm in length;
- Inspiratory resistance: <5 cmH2O (at 50 L/min);
- Expiratory resistance: <5 cmH2O (at 50 L/min)
{9}------------------------------------------------
#### V. INDICATIONS FOR USE
The CPR Mask with Oxygen Port is designed to assist in providing immediate life support (mouth to mask ventilation) to health emergency victims requiring oxygen support (inhalation) or cardiopulmonary resuscitation (CPR) rescue techniques. The CPR Mask with Oxygen Port is for prescription use.
The CPR Mask without Oxygen Port is designed to assist in providing immediate life support (mouth to mask ventilation) to health emergency victims requiring cardiopulmonary resuscitation (CPR) rescue techniques. The resuscitator without oxygen port is for over-the-counter use.
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The sponsor identified the similarities and differences of the proposed CPR MASK device to the legally marketed predicate MedSource CPR Mask K081516 to which substantial equivalency is claimed.
| Specification | Predicate Device | Proposed Device |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | MedSource International, LLC | Hangzhou Jinlin Medical Appliances Co., Ltd |
| Device name | MedSource (Brand) CPR Mask | KYOLING CPR Mask With Oxygen Port(Prescription Use) |
| K number | K081516 | -- |
| Indications<br>for Use | The MedSource CPR Mask is designed to<br>assist in providing immediate life support<br>(mouth to mask ventilation) to health<br>emergency victims requiring oxygen<br>support (inhalation) or cardiopulmonary<br>resuscitation (CPR) rescue techniques. | The CPR Mask with Oxygen Port is designed to<br>assist in providing immediate life support (mouth<br>to mask ventilation) to health emergency victims<br>requiring oxygen support (inhalation) or<br>cardiopulmonary resuscitation (CPR) rescue<br>techniques. The CPR Mask with Oxygen Port is for<br>prescription use. |
| Inspiratory<br>Resistance | 2.84-2.87 cmH2O | 3.09-3.17 cmH2O |
| Expiratory<br>Resistance | 2.96-3.01 cmH2O | 3.21-3.26 cmH2O |
| Raw<br>materials | CPR one-way valve: PVC/Silicon valve<br>Face Mask: PVC<br>Strap: Non-Woven | CPR one-way valve: Silicon<br>Face Mask: PVC<br>Strap: Non-Woven |
| Dimensions | Length: 100±3mm<br>Width: 82±3mm | |
{10}------------------------------------------------
| Standards | ISO 5356-1:2004 Anaesthetic and respiratory equipment -- Conical connectors -- Part 1: Cones<br>and sockets; |
|-----------|-----------------------------------------------------------------------------------------------------------------------|
| | AS 4259-1995 Ancillary devices for expired air resuscitation; |
| | ISO 10993-5:2009 Biological evaluation of medical devices -- Part 5: Tests for in vitro<br>cytotoxicity; |
| | ISO 10993-10:2010 Biological evaluation of medical devices -- Part 10: Tests for irritation and<br>skin sensitization |
Though the Inspiratory Resistance and Expiratory Resistance are slightly different, both the subject and predicate device meet the requirements for AS-4259-1995 Standard.
The sponsor identified the similarities and differences of the proposed CPR MASK device to the legally marketed predicate Genuine First Aid LLC K112126 to which substantial equivalency is claimed.
| Specification | Predicate Device | Proposed Device |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Genuine First Aid LLC | Hangzhou Jinlin Medical Appliances Co., Ltd |
| Device name | Genuine First Aid CPR Face Mask<br>without oxygen port | KYOLING CPR Mask Without Oxygen Port(Over-<br>the-Counter Use) |
| K number | K112126 | -- |
| Indications<br>for Use | The CPR mask (without oxygen port) is<br>single use designed for mouth to mask<br>ventilation to health emergency victims<br>requiring cardiopulmonary resuscitation<br>(CPR) rescue techniques of a<br>nonbreathing adult. It is also used as a<br>barrier that will direct expired air<br>from the patient away from the user. | The CPR Mask without Oxygen Port is designed to<br>assist in providing immediate life support (mouth<br>to mask ventilation) to health emergency victims<br>requiring cardiopulmonary resuscitation (CPR)<br>rescue techniques. The resuscitator without<br>oxygen port is for over-the-counter use. |
| Inspiratory<br>Resistance | 1.94@50L/min | 3.09-3.17 cmH2O |
| Expiratory<br>Resistance | 2.04@50L/min | 3.21-3.26 cmH2O |
| Raw<br>materials | CPR one-way valve: K-Resin, Silicone<br>Face Mask: PVC<br>Strap: Non-Woven | CPR one-way valve: Silicon<br>Face Mask: PVC<br>Strap: Non-Woven |
| Dimensions | Length: 100±3mm<br>Width: 82±3mm | |
{11}------------------------------------------------
| Standards | ISO 5356-1:2004 Anaesthetic and respiratory equipment -- Conical connectors -- Part 1: Cones<br>and sockets; |
|-----------|-----------------------------------------------------------------------------------------------------------------------|
| | AS 4259-1995 Ancillary devices for expired air resuscitation; |
| | ISO 10993-5:2009 Biological evaluation of medical devices -- Part 5: Tests for in vitro<br>cytotoxicity; |
| | ISO 10993-10:2010 Biological evaluation of medical devices -- Part 10: Tests for irritation and<br>skin sensitization |
Though the Inspiratory Resistance and Expiratory Resistance are slightly different, both the subject and predicate device meet the requirements for AS-4259-1995 Standard.
### VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
# Biocompatibility testing:
The biocompatibility evaluation for the KYOLING CPR Mask was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing," May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The battery of testing included the following tests:
- . ISO 10993-5: 2009-Cytotoxicity
- ISO 10993-10: 2010-Sensitization
- ISO 10993-10: 2010-Irritation
The cushion and valve is considered surface contacting for duration of less than 24 hours.
# Bench testing:
Expiratory resistance and inspiratory resistance testing was for Kyoling CPR mask with oxygen port done using the test methods described in AS-4259-1995 Ancillary Devices for Expired Air Resuscitations
| Parameters | Standard | Proposed device |
|-----------------------|----------|------------------|
| Inpiratory resistance | <5cmH2O | 3.19~3.24 cm H2O |
{12}------------------------------------------------
| the country of the country of the country of the country of the country<br>Expiratory resistance | <5cmH30 | 3.31~3.37cmH₂O |
|--------------------------------------------------------------------------------------------------|---------|----------------|
|--------------------------------------------------------------------------------------------------|---------|----------------|
Expiratory resistance and inspiratory resistance testing was for Kyoling CPR mask without oxygen port done using the test methods described in AS-4259-1995 Ancillary Devices for Expired Air Resuscitations
| Parameters | Standard | Proposed device |
|-----------------------|----------|--------------------|
| Inpiratory resistance | <5cmH₂O | 3.19~3.24 cm H₂O |
| Expiratory resistance | <5cmH₂O | 3.31~3.37cmH $₂$ O |
Animal and clinical study:
The subject of this premarket submission, KYOLING CPR Mask, does not require clinical studies to support substantial equivalence.
### VIII. CONCLUSIONS
Hangzhou Jinlin Medical Appliances Co., Ltd considers the Kyoling CPR mask with oxygen port is as safe and effective as the predicate device. It has the same intended use, indications for use, technological characteristics, and principles of operation as those of the predicate device. The minor differences between the Kyoling CPR mask with oxygen port and its predicate device raise no new issues of safety or effectiveness. Thus, the Kyoling CPR mask with oxygen port is substantially equivalent to its predicate device.
Hangzhou Jinlin Medical Appliances Co., Ltd considers the Kyoling CPR mask without oxygen port is as safe and effective as the predicate device. It has the same intended use, indications for use, technological characteristics, and principles of operation as those of the predicate device. The minor differences between the Kyoling CPR mask without oxygen port and its predicate device raise no new issues of safety or effectiveness. Thus, the Kyoling CPR mask without oxygen port is substantially equivalent to its predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.