MEDSOURCE CPR MASK WITH & WITHOUT OXYGEN PORT

K081516 · Medsource International, LLC · CBP · Aug 26, 2008 · Anesthesiology

Device Facts

Record IDK081516
Device NameMEDSOURCE CPR MASK WITH & WITHOUT OXYGEN PORT
ApplicantMedsource International, LLC
Product CodeCBP · Anesthesiology
Decision DateAug 26, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5870
Device ClassClass 2
AttributesTherapeutic, 3rd-Party Reviewed

Indications for Use

The MedSource CPR Mask is designed to assist in providing immediate life support (mouth to mask ventilation) to health emergency victims requiring oxygen support (inhalation) or cardiopulmonary resuscitation (CPR) rescue techniques. The CPR Mask with Oxygen Port is for prescription use. The MedSource CPR Mask is designed to assist in providing immediate life support (mouth to mask ventilation) to health emergency victims requiring cardiopylmonary resuscitation (CPR) resue techniques.

Device Story

Single-use, full-size PVC mask system; includes universal breathing tube, one-way filtered valve, and head strap. Used for emergency mouth-to-mask ventilation; provides oxygen support via optional port. Operated by rescuers in emergency settings. Device facilitates rescue breathing; protects rescuer via one-way valve; delivers supplemental oxygen when port is utilized. Benefits patient by enabling immediate life support during respiratory or cardiac arrest.

Clinical Evidence

Bench testing only. Performance testing conducted per AS-4259-1995 standards compared inspiratory and expiratory resistance against Spiracle CPR mask. Results: MedSource mask (1.94 cm H2O inspiratory, 2.04 cm H2O expiratory) vs. Spiracle mask (2.04 cm H2O inspiratory, 2.14 cm H2O expiratory) at 50 lpm flow. Biocompatibility testing (cytotoxicity, sensitization, irritation) passed per ISO 10993 requirements.

Technological Characteristics

One-piece PVC mask; silicone one-way valve; non-woven strap. Dimensions: full-size. Connectivity: none. Sterilization: non-sterile. Materials: PVC, silicone, non-woven fabric. Compliant with ISO 10993-1 for biocompatibility. Mechanical operation via one-way valve for respiratory assistance.

Indications for Use

Indicated for health emergency victims requiring mouth-to-mask ventilation, cardiopulmonary resuscitation (CPR), or oxygen support. No specific age or gender contraindications listed.

Regulatory Classification

Identification

A nonrebreathing valve is a one-way valve that directs breathing gas flow to the patient and vents exhaled gases into the atmosphere.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## ubmitter: K081516 MedSource International, LLC 5346 Shoreline Drive Mound, Minnesota 55364 952-472-0131 ## Contact/Consultant: AUG 2 6 2008 Howard Cooper, Quality Consultant EQACT, INC. 317-826-4398 (O) 317-826-4398(F) 317-523-2314 (C) htc@eqact.com ## General Information . - Common or Usual Name: Emergency CPR Mask . - Proprietary Name: MedSource CPR Mask With Oxygen Port (Prescription Use) - MedSource CPR Mask Without Oxygen Port (Over-the-Counter Use) . - Product Code:CBP . - . Panel: Anesthesiology - Classification: . Class II - Regulatory Reference: 21 CFR §868.5870 . - Single Use: Yes . - Sterile: No . - Packaging Materials: PE & PET Polymer film and medical package paper using heat as . the sealing method ## Indications for Use (with Oxygen Port): The MedSource CPR Mask is designed to assist in providing immediate life support (mouth to mask ventilation) to health emergency victims requiring oxygen support (inhalation) or cardiopulmonary resuscitation (CPR) rescue techniques. # Indications for Use (without Oxygen Port): The MedSource CPR Mask is designed to assist in providing immediate life support (mouth to mask ventilation) to health emergency victims requiring cardiopylmonary resuscitation (CPR) resue techniques. {1}------------------------------------------------ ## Tevice Characteristics: The MedSource CPR Mask has the following characteristics: A one piece, single use, full-size, light-weight PVC mask system which includes: - A Universal breathing tube. . - A One-way filtered valve. . - A Head strap. . - With Oxygen Port or Without Oxygen Port. . - Packaged for easy portability and quick access. . ## Patient Contact | Device Component | Material of Construction | Patient<br>Contact | Contact Time | |-------------------|--------------------------|--------------------|--------------| | CPR one-way valve | PVC/Silicon valve | Mouth | 3-60 minutes | | Face Mask | PVC | Skin | 3-60 minutes | | Strap | Non-Woven | Skin/Hair | 3-60 minutes | | Suggestively Equivalent<br>Comparison Point | Predicate Devices | Result of Comparison | |-----------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------| | Intended Use | Laerdal® Pocket Mask With and<br>without oxygen port K861401<br>Foremount Pocket Size Resuscitator<br>K042727 | Substantially equivalent | | Technological Characteristics<br>(Materials of Construction,<br>Dimensions, Performance in<br>expiratory and inspiratory<br>resistance) | Spiracle Pocket Size Resuscitator<br>K042727<br>Laerdal® Pocket Mask K861401 | Materials of construction-<br>same or similar<br>Performance-minor<br>variations<br>Dimensions-minor variations | | Instructions for Use | Spiracle Pocket Size Resuscitator<br>K042727<br>Laerdal® Pocket Mask K861401 | Very similar | #### Recognized Consensus Standards: | Standard | Title | Status of Compliance | |----------------------------------|-------------------------------------------------------------------------------------------|-------------------------------------------------------| | ISO 5356-1:2004 | Anaesthetic and Respiratory Equipment -<br>Conical Connectors: Part 1: Cones and Sockets. | Used as a guide but did not do<br>performance testing | | IS 4259-1995 | Ancillary Devices for Expired Air Resuscitation. | Used as a guide but did not do<br>performance testing | | ISO10993-1<br>edition-1997-12-15 | (2nd Biological Evaluation Of Medical Devices—Part<br>1: Evaluation and Testing | Compliant to applicable sections | {2}------------------------------------------------ ## 510(K) Summary for the MedSource CPR Mask #### Performance Testing Expiratory And Inspiratory Resistance Performance testing was done using the test methods described in AS-4259-1995, Ancillary Devices for Expired Air Resuscitation. Some minor modifications were made to the e methods | Device Tested | Inspiratory Resistance | Expiratory Resistance | |--------------------|---------------------------|---------------------------| | MedSource CPR Mask | 1.94 cm H2O @ 50 lpm flow | 2.04 cm H2O @ 50 lpm flow | | Spiracle CPR Mask | 2.04 cm H2O @ 50 lpm flow | 2.14 cm H2O @ 50 lpm flow | #### Discussion of Results: - 1. The results in the above table demonstrate that the MedSource CPR Mask performance exceeds the performance of the predicate device. - 2. For both inspiratory and expiratory resistance, the difference between the MedSource CPR mask and the Spiracle foldable mask is 5%. This demonstrates that the MedSource CPR Mask performance and the Spiracle foldable mask perform substantially the same test conditions assuming there are inherent variations in the test methods... Accuracy and repeatability are not discussed in the AS 4259-1995 standard. #### iocompatibility Raw Material Biocompatibility Testing | Biocompatibility<br>Test | ISO 10993 Requirement | Testing results | |--------------------------------------------------|-----------------------|-----------------| | Cytotoxicity | Required | Passes | | Sensitization | Required | Passes | | Irritation<br>or<br>Intracutaneous<br>reactivity | Required | Passes | | Pyrogen | Not required | Passes | | Haemocompatibility | Not required | Passes | | Systemic Toxicity | Not required | Passes | #### Conclusion: The potential for causing irritation is very remote for the following reasons: - 1. The raw materials have been tested beyond the requirements of ISO 10993 for skin contact. - 2. The time of contact and percentage of body exposure is very low. - 3. The injection molding process is not significantly alter the raw material characteristics {3}------------------------------------------------ Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MedSource International, LLC C/O Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW Buffalo, Minnesota 55313 AUG 2 6 2008 Re: K081516 Trade/Device Name: MedSource CPR Mask with Oxygen Port / Model PM103 MedSource CPR Mask / Model PM104 Regulation Number: 21 CFR 868.5870 Regulation Name: Nonrebreathing Valve Regulatory Class: II Product Code: CBP Dated: August 20, 2008 Received: August 21, 2008 Dear Mr. Job: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Mr. Job Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Ove Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosures {5}------------------------------------------------ # Indications for Use 510(k) Number (if known): Device Name: MedSource CPR Mask with Oxygen Port / Model PM103 Indications for Use: The MedSource CPR Mask is designed to assist in providing immediate life support (mouth to mask ventilation) to health emergency victims requiring oxygen support (inhalation) or cardiopulmonary resuscitation (CPR) rescue techniques. The CPR Mask with Oxygen Port is for prescription use. Prescription Use X (Part 21 CFR 801 Subpart D) OR Over-the-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Vhls (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: K081516 Page 1 of 1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...