POCKETSPACER

K992038 · Ferraris Medical, Inc. · CAF · Sep 9, 1999 · Anesthesiology

Device Facts

Record IDK992038
Device NamePOCKETSPACER
ApplicantFerraris Medical, Inc.
Product CodeCAF · Anesthesiology
Decision DateSep 9, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5630
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The PocketSpacer is intended to assist with the delivery of aerosolized medications when used in conjunction with commercialized MDI canisters with the associated activator elbow. The spacer device is indicated to assist in the treatment of patients with lung disease such as, but not limited to, asthma, emphysema, chronic bronchitis, and cystic fibrosis.

Device Story

PocketSpacer is a cylindrical aerosol holding chamber (10 cm x 4.5 cm) used with commercial MDI canisters; prevents off-center actuations. Device features an open mesh mouthpiece with concentrically placed circular orifices; acts as a filter to remove non-respirable particles (throat dose) while allowing respirable particles to pass for inhalation. Used by patients to improve medication delivery efficiency compared to MDI use alone. No valves present. Bench testing confirms performance comparability to predicate in total dose, respirable dose, and mass median aerodynamic diameter.

Clinical Evidence

Bench testing only. Cascade Impactor testing compared the PocketSpacer to the MicroChamber and MDI alone across four drug categories. Results showed the PocketSpacer performed comparably to the predicate in total dose, respirable dose, and mass median aerodynamic diameter, while providing finer, more respirable aerosols than MDI alone.

Technological Characteristics

Cylindrical tube, 10 cm x 4.5 cm. Features an open mesh mouthpiece with varying circular orifices for particle size selection. No valves. Passive mechanical device; no energy source or software.

Indications for Use

Indicated for patients with lung disease, including asthma, emphysema, chronic bronchitis, and cystic fibrosis, requiring assistance with aerosolized medication delivery via MDI canisters.

Regulatory Classification

Identification

A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ SEP d 1999 #### 510(k) SUMMARY # Ferraris Medical Inc. PocketSpacer # Submitter's Name. Address. Telephone Number. Contact Person and Date Prepared Mr. David R. Malys President Ferraris Medical Inc. 9681 Wagner Road Post Office Box 344 Holland. New York 14080 Jonathan S. Kahan. Esq. Hogan & Hartson L.L.P. 555 Thirteenth Street, N.W. Washington, DC 20004-1109 Telephone: (202) 637-5794 Facsimile: (202) 637-5910 ### Date Prepared: June 16, 1999 # Name of Device and Name/Address of Sponsor Ferraris Medical Inc. PocketSpacer Mr. David R. Malys President Ferraris Medical Inc. 9681 Wagner Road Post Office Box 344 Holland, New York 14080 Common or Usual Name Spacer; Aerosol Holding Chamber Classification Name Accessory to a Nebulizer #### Predicate Devices Respiratory Delivery System's MicroChamber ### Intended Use The PocketSpacer is intended to assist with the delivery of aerosolized medications when used in conjunction with commercialized MDI canisters with the associated activator elbow. The spacer device is indicated to assist in the treatment of patients with lung disease such as, but not limited to, asthma, emphysema, chronic bronchitis, and cystic fibrosis. {1}------------------------------------------------ # Technological Characteristics and Substantial Equivalence The PocketSpacer is a cylindrical tube, approximately 10 cm long by 4.5 cm wide, and accepts the MDI manufacturer's mouthpiece. The MDI receptacle has an orifice of size and shape similar to the MDI mouthpiece. The device is designed to prevent off-center actuations. An open mesh with varying size circular orifices, but no valve, is in the round mouthpiece. The mesh with concentrically placed circular orifices removes the throat dose by permitting only respirable particles to be inhaled by the user. The PocketSpacer is substantially equivalent to the MicroChamber. #### Performance Data Ferraris performed the Cascade Impactor Test comparing the PocketSpacer with the MicroChamber and an MDI actuator without a spacer device for four categories of drugs. The PocketSpacer performed comparably to the MicroChamber in terms of total dose, respirable dose, and mass median aerodynamic diameter. In addition, the PocketSpacer device with a commercial MDI provided finer, more highly respirable aerosols compared to the aerosols generated with the MDIs alone. {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES.USA Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 SEP g 1999 Mr. Jonathan S. Kahan Ferraris Medical, Inc. c/o Hogan & Hartson L.L.P. 555 Thirteenth Street, N.W. Washington, DC 20004-5600 Re: K992038 PocketSpacer Metered Dose Inhaler Spacer Requlatory Class: II (two) 73 CAF Product Code: Dated: June 16, 1999 Received: June 16, 1999 Dear Mr. Kahan: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 A substantially equivalent determination assumes compliance to 895. with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Druq Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in In addition, FDA may publish further announcements regulatory action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3}------------------------------------------------ #### Page 2 - Mr. Jonathan S. Kahan This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Thomas J. Callahon Thomas J. Callehan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Ferraris Medical Inc.'s PocketSpacer Device Name: Indications for Use: The PocketSpacer is intended to assist with the delivery of aerosolized medications when used in conjunction with commercialized MDI canisters with the associated activator elbow. The spacer device is indicated to assist in the treatment of patients with lung disease such as, but not limited to, asthma, emphysema, chronic bronchitis, and cystic fibrosis. 1992038 (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Cardiovas and Neurological De 510(k) Numbe Prescription Use (Per 21 C.F.R. 801.109) OR Over-The-Counter Use_ (Optional Format 1-2-96)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...