BIRDSONG LEVERHALER, MODEL LH-1000

K031982 · Birdsong Medical Devices, Inc. · CAF · Apr 2, 2004 · Anesthesiology

Device Facts

Record IDK031982
Device NameBIRDSONG LEVERHALER, MODEL LH-1000
ApplicantBirdsong Medical Devices, Inc.
Product CodeCAF · Anesthesiology
Decision DateApr 2, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5630
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The LeverHaler Spacer is a spacer device that is designed to effectively deliver respirable medical aerosols produced by most MDI's to a patient during inhalation. This product is intended for use in the treatment of lung disease. The device has been designed for use by all patients who have been prescribed MDI treatment by their physician. It is intended for use in both the hospital and homecare environments.

Device Story

LeverHaler Spacer is an injection-molded polycarbonate mixing chamber for FDA-approved metered dose inhalers (MDIs); facilitates aerosol delivery for lung disease treatment. Features a lever actuator to assist patients with limited hand strength, coordination, or dexterity; allows manual MDI activation or lever-assisted actuation. Used in hospitals and homecare environments by patients under physician prescription. Device provides effective aerosol concentration for inhalation; does not generate aerosols itself. No software, alarms, or detection capabilities. Benefits include improved ease of use for patients with physical limitations and consistent aerosol delivery.

Clinical Evidence

No clinical testing performed. Bench testing only. In vitro comparative study performed per FDA 1993 guidance for nebulizers/MDIs/spacers. Evaluated particle size distribution, dose output, MMAD, GSD, plume analysis, and life testing using three drugs. Results showed LeverHaler performance equivalent or superior to predicate devices.

Technological Characteristics

Injection-molded polycarbonate housing; silicone mouthpiece and valve. Mechanical operation; no energy source. Mixing chamber volume ~170 ml. Non-sterile. Single-patient use. No software or connectivity.

Indications for Use

Indicated for patients of all ages prescribed MDI treatment for lung disease. Used in hospital and homecare settings.

Regulatory Classification

Identification

A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the text "K031982" in a handwritten style. The text is oriented diagonally, with the characters slightly tilted to the right. The handwriting appears somewhat casual and the numbers are clearly distinguishable. # Birdsong Medical LeverHaler Spacer 510(k) Summary: In accordance with 21 CFR section 807.92 BirdSong Medical is submitting the following safety and effectiveness summary. - 1) Submitter Information Eric Fugelsang President BirdSong Medical Devices, Inc. 1456 Rt. 22, Suite A103 Brewster, NY 10509 - 2) Name of Device Proprietary Name: LeverHaler Spacer Common Name is Handheld Spacer Classification Name: Nebulizer Accessory - 3) Substantially equivalent to: OptiHaler (Healthscan Products, Inc.), K911807. - 4) Device Description and System Overview: The LeverHaler Spacer (Spacer) is an injection molded polycarbonate spacer device intended for use with FDA approved metered dose inhalers (MDI's). The MDI canister fits through a slot in a lever located on the top of the spacer. Activation of the MDI canister can be done by manually pressing down on the MDI (typical activation), or the MDI may be actuated by depressing the Lever Actuator down towards the top of the spacer body. Image /page/0/Picture/11 description: The image shows a diagram of a medical device, specifically an inhaler. The different parts of the inhaler are labeled, including the mouthpiece, lever actuator, MDI (metered dose inhaler), air inlets, air outlets, housing, and chamber. The diagram provides a clear visual representation of the inhaler's components and their arrangement. Figure 1: Diagram of LeverHaler Spacer {2}------------------------------------------------ The device is made from polycarbonate injection molded plastic and is designed to The device is made from polycurson injoning. When not in use, up to two MDI canisters can be stored inside of the LeverHaler Chamber. The product is sold non-sterile, and is a single-patient device intended to be used The product is sold non-sichie, and to a sing or cleaning and use of the device. All for up to one year. Easoning couring of this device are identical to a defined predicate device. This device does not generate aerosols. Its purpose is to provide an effective mixing chamber for the aerosols produced by a metered dose inhaler to assure milking chamber for the doneentration for inhalation by the patient. This is a prescription device. The device has no detection capabilities. It is mechanical and has no alarm functions or capabilities. There is no software integrated or used in conjunction with this device. Product testing has been completed according to the "Reviewer Guidance for Nebulizers, Metered Dose Inhalers, Spacers and Actuators (10-01-93)." Product testing performed by Birdsong Medical provides clear proof that the LeverHaler Spacer is superior to the MDI and is comparable or superior to the predicate devices tested. The premarket notification submitted to FDA contains a full discussion of product testing, which includes delivered aerosol potency, MMAD, GSD, particle distribution, retained aerosols (Plume Analysis) and life testing. All tests included comparative testing of the two predicate devices and the LeverHaler Spacer. ### Design Considerations and Operation of the Device: The device is intended to be used with FDA approved MDI drugs for treatment of lung disease. The Lever Actuator is designed for patients with limited hand strength, coordination or dexterity. The MDI can be actuated easily with the Lever or may be actuated by depressing the MDI downwards (typical MDI activation). #### Testing: All product and comparative in vitro testing performed was based upon the requirements of the "Guidance for Nebulizers, Metered Dose Inhalers, Spacers and Actuators: Reviewer Guidance 10-01-93." Testing was conducted by an independent medical device manufacturer and test laboratory contracted by Birdsong Medical. All tests were performed according to documented test protocol. The following tests were completed for the LeverHaler Spacer, and the predicate devices: OptiHaler and AeroChamber (specified in this premarket notification): - Particle size distribution (using 3 drugs) . - Dose output testing (drug quantity and total mass using 3 drugs) . - Mean median aerodynamic diameter (MMAD) . - Geometric standard deviation (GSD) . 10 {3}------------------------------------------------ - Plume Analysis Testing . - Single patient use testing (life testing) . No clinical testing was performed on this product. Software Validation: Not applicable: there is no software in this product. Sterilization Validation: Not applicable: this product is sold and used as a non-sterile product. Biocompatibility: All materials used in this device are incorporated in other predicate devices, as well as nebulizers, connectors and mouthpieces currently sold in the uevices, as well as nobles on appropriate for the intended use described herein. ## Comparative Product Matrix ## Table 2: 510(k) Summary - Substantial Equivalence to Predicate Devices | item: | Characteristic: | Birdsong<br>Medical | Healthscan<br>Products, Inc. | Monaghan<br>Medical | |-------|--------------------------------|----------------------------------|----------------------------------|----------------------------------| | | Model: | LeverHaler<br>Spacer | OptiHaler | AeroChamber | | 4. | Spacer Material | Clear<br>Polycarbonate | Clear<br>Polycarbonate | Clear Polyester | | 5. | Mouthpiece and Valve | Silicone | Polycarbonate | Polyethelene<br>and Silicone | | 5. | Single Patient Use? | Yes | Yes | Yes | | 6. | Sterility? | Non-Sterile | Non-Sterile | Non-Sterile | | 7. | Ventilator Use? | No | No | Yes | | 8. | Maximum Length of Use: | Single patient<br>up to 365 days | Single patient<br>Not stipulated | Single patient<br>Not stipulated | | 8. | Effective Size of mixing area: | Approximately<br>170 ml | Approximately<br>90 ml | Approximately<br>140 ml | | 9. | Method of Operation: | Mechanical | Mechanical | Mechanical | | 10. | Prescription Device? | Yes | Yes | Yes | | 11. | Provided as Kit? | No | No | No | | 12. | Port Protection? | Yes | Yes | Yes | | | Premarket Submission: | Pending | (K911807) | (K930574)<br>(K884803) | The Birdsong Medical Spacer met all design requirements Conclusions of all Testing: and passed all validation and comparative product testing. The device is manufactured from the identical materials used in a predicate device noted in the 510(k) submission. LeverHaler performance in all in vitro tests was equivalent or superior to the comparative predicate device noted in the table above. Based upon these results, it is our conclusion that the LeverHaler Spacer is as safe, as effective and performs as well as or better than the legally marketed predicate OptiHaler Spacer device used in comparative product testing. 75 {4}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a snake winding around it, overlaid on three horizontal bars. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol. Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 APR = 2 2004 Mr. Eric Fugelsang President Birdsong Medical Devices, Incorporated 1456 Rt. 22. Suite A103 Brewster, NY 10509 Re: K031982 Trade/Device Name: Leverhaler Spacer Regulation Number: 868,5630 Regulation Name: Nebulizer Regulatory Class: II Product Code: CAF Dated: February 13, 2004 Received: February 13, 2004 Dear Mr. Fugelsang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ Page 2 - Mr. Fugelsang Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely vours. Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ #### INDICATIONS FOR USE [510(k)] Number: K031982 Device Name: Birdsong Medical Devices LeverHaler Spacer Indications For Use: The LeverHaler Spacer is a spacer device that is designed to effectively deliver respirable medical aerosols produced by most MDI's to a patient during inhalation. This product is intended for use in the treatment of lung disease. The device has been designed for use by all patients who have been prescribed MDI treatment by their physician. It is intended for use in both the hospital and homecare environments. Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR Over-the-Counter Use (21 CFR 807 Subpart €) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) fthuh Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Denta 510(k) Number: k031982
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...