PLEUR-EVAC SAHARA PLUS CONTINUOUS REINFUSION AUTOTRANSFUSION SYSTEM
K130043 · Teleflexmedical, Inc. · CAC · Feb 12, 2013 · Anesthesiology
Device Facts
Record ID
K130043
Device Name
PLEUR-EVAC SAHARA PLUS CONTINUOUS REINFUSION AUTOTRANSFUSION SYSTEM
Applicant
Teleflexmedical, Inc.
Product Code
CAC · Anesthesiology
Decision Date
Feb 12, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 868.5830
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
AUTOTRANSFUSION 1. For the collection of autologuous blood from the patient's pleural cavity or mediastinal area for reinfusion purposes in trauma and post-operative situations CHEST DRAINAGE 1. To evacuate air and/or fluid from the chest cavity or mediastinum 2. To help prevent air and/or fluid from re-accumulating in the chest cavity or mediastinum. 3. To help re-establish and maintain normal intra-thoracic pressure gradients. 4. To facilitate complete lung re-expansion to restore normal breathing dynamics. The Pleur-evac® Autotransfusion Bag is indicated as a sterile, single use device used for collection and reinfusion of autologous blood from the thoracic cavity when attached to a Pleur-evac® System. The fluid path is non-pyrogenic.
Device Story
Sterile, single-use autotransfusion and chest drainage system. Collects autologous blood from pleural or mediastinal cavities for reinfusion; functions as chest drainage unit post-autotransfusion. Operates via vacuum-assisted collection and reinfusion. Used in clinical settings (trauma/post-operative) by healthcare professionals. Modification involves material change in indirect patient-contacting component due to supplier obsolescence. System maintains intra-thoracic pressure gradients and facilitates lung re-expansion.
Clinical Evidence
Bench testing only. Performance verified via pressure equalization rate testing, filter position visual inspection, and microbial filtration efficiency (ASTM F1608-00(2009)) with LRV > 4.
Technological Characteristics
Autotransfusion apparatus; sterile, single-use. Patient-contacting materials compliant with ISO 10993-1. Features High Negative Relief Valve (HNRV) assembly with integrated filter. Mechanical operation for vacuum-based fluid collection and reinfusion. No electronic or software components.
Indications for Use
Indicated for trauma and post-operative patients requiring autologous blood collection from pleural or mediastinal cavities for reinfusion, or chest drainage to evacuate air/fluid, prevent re-accumulation, maintain intra-thoracic pressure, and facilitate lung re-expansion. Contraindicated in patients with pericardial/mediastinal/systemic infections, pulmonary/respiratory infections, malignant neoplasms, coagulopathies, suspected thoraco-abdominal injuries with enteric contamination, impaired renal function, or intraoperative use of specific topical hemostatic/antiseptic agents.
Regulatory Classification
Identification
An autotransfusion apparatus is a device used to collect and reinfuse the blood lost by a patient due to surgery or trauma.
K031554 — S-1150 PLEUR-EVAC SAHARA PLUS CONTINUOUS REINFUSION AUTOTRANSFUSION SYSTEM · Genzyme Biosurgery · Jul 25, 2003
K120953 — PLEUR-EVAC SAHARA PLUS CONTINOUS REINFUSION AUTOTRANSFUSION SYSTEM · Tlelflex Medical, Inc. · Dec 10, 2012
K963850 — PLEUR-EVAC SAHARA PLUS MODEL S-1150 CONTINUOUS REINFUSION AUTOTRANSFUSION SYSTEM · Deknatel, Inc. · Apr 22, 1997
K140197 — PLEUR-EVAC PLUS CONTINUOUS REINFUSION AND AUTOTRANSFUSION SYSTEM · Teleflexmedical, Inc. · Oct 21, 2014
K140205 — PLEUR-EVAC SAHARA PLUS CONTINUOUS REINFUSION AND AUTOTRANSFUSION SYSTEM · Teleflexmedical, Inc. · Oct 23, 2014
Submission Summary (Full Text)
{0}------------------------------------------------
# R 130043
Special 510(k) Pleur-evac Saḥara® Plus Continuous Reinfusion Autotransfusion System Section 8 - Summary of Safety and Effectiveness
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
## Pleur-evac Sahara® Plus Continuous Reinfusion Autotransfusion System
#### A. Name, Address, Phone and Fax Number of Applicant
Teleflex Medical, Incorporated 2917 Weck Drive Research Triangle Park, NC 27709 USA Phone: 919-433-8049 919-433-4996 Fax:
#### B. Contact Person
Natalie Smith Regulatory Affairs Specialist
Lorraine DeLong Manager RA/QE Surgical
# C. Date Prepared
January 7, 2013
#### D. Device Name
Trade Name: Pleur-evac Sahara® Plus Continuous Reinfusion Autotransfusion System
Common Name: Autotransfusion Apparatus
Classification Name: Autotransfusion Apparatus
#### E. Device Description
The Pleur-evac Sahara® Plus Continuous Reinfusion Autotransfusion System is provided as a sterile unit intended for single patient use. The fluid path is non-pyrogenic. The Pleur-evac Sahara Plus System is used for the collection and continuous reinfusion of autologous blood. By attaching the Pleur-evac Sahara Autotransfusion Bag, the Pleurevac Sahara Plus System serves as a bag reinfusion system. When autotransfusion is completed, the Pleur-evac Sahara Plus System can serve as a chest drainage collection unit.
Teleflex Medical, Inc.
Page 8.1
{1}------------------------------------------------
#### F. Indications for Use
#### AUTOTRANSFUSION
- 1. For the collection of autologuous blood from the patient's pleural cavity or mediastinal area for reinfusion purposes in trauma and post-operative situations
#### CHEST DRAINAGE
- 1. To evacuate air and/or fluid from the chest cavity or mediastinum
- 2. To help prevent air and/or fluid from re-accumulating in the chest cavity or mediastinum.
- 3. To help re-establish and maintain normal intra-thoracic pressure gradients.
- 4. To facilitate complete lung re-expansion to restore normal breathing dynamics.
The Pleur-evac® Autotransfusion Bag is indicated as a sterile, single use device used for collection and reinfusion of autologous blood from the thoracic cavity when attached to a Pleur-evac® System. The fluid path is non-pyrogenic.
#### G. Contraindications
Pleur-evac Sahara® Plus Continuous Reinfusion Autotransfusion System is contraindicated for:
- · Pericardial, mediastinal, or systemic infections
- · Pulmonary and respiratory infection or infestation
- · Presence of malignant neoplasms
- · Coagulopathies
- Suspected thoraco-abdominal injuries with possible enteric contamination
- Impaired renal function
- Intraoperative thoracic or mediastinal cavity use of topical thrombin, microfibrillar hemostatic agents or providine-iodine antiseptic gels or solutions and non I.V. compatible antibiotics
#### H. Substantial Equivalence
The proposed Pleur-evac® Plus Continuous Autotransfusion System is substantially equivalent to the predicate devices:
| Predicate Device | Manufacturer | 510(k) No. | Date Cleared |
|----------------------------------------|---------------------------|------------|-------------------|
| Pleur-evac® Autotransfusion<br>Systems | Teleflex Medical,<br>Inc. | K120953 | December 10, 2012 |
## I. Comparison To Predicate Devices
The proposed Pleur-evac Sahara® Plus Continuous Reinfusion Autotransfusion System has the same technology, indications for use and functional characteristics as
Teleflex Medical, Inc.
{2}------------------------------------------------
the predicate system. The proposed modification is a change in material to an indirect patient contacting component, which was driven by supplier obsolescence.
#### J. Materials
All patient contacting materials are in compliance with ISO10993-1.
#### K. Technological Characteristics
A comparison of the technological characteristics of the proposed Pleur-evac Sahara® Plus Continuous Reinfusion Autotransfusion Systems and the predicate has been performed. The results of this comparison demonstrate that the Pleur-evac® High Negative Relief Valve filter is equivalent to the marketed predicate devices in performance characteristics.
#### L. Performance Data
The bench testing has been performed to verify that the performance of the proposed Pleur-evac Sahara® Plus Continuous Reinfusion Autotransfusion Systems is substantially equivalent to the predicate device. The proposed filter material was validated according to test specifications and is summarized below.
| Performance Data Results Summary | | |
|----------------------------------|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Response Type | Test Specification | Result |
| Quantitative | Pressure Equalization<br>Rate | PASS<br>All values tested allowed the pressure in a full<br>collection chamber to rise from 17 cm H2O<br>vacuum to 2 cm H2O vacuum between 1<br>second and 22 seconds. |
| Qualitative<br>(Pass, Fail) | Filter Position Visual | PASS<br>All covers tested contacted the filter material<br>on the entire cover retaining surface. |
| Qualitative<br>(Pass, Fail) | Filter Visual | PASS<br>All Pleur-evac units tested contained a filter in<br>the HNRV assembly. |
| Quantitative | ASTM F1608-<br>00(2009) | PASS<br>All Pleur-evac units tested had an LRV > 4. |
#### L. Conclusion
Based upon the comparative test results, the proposed Pleur-evac Sahara® Plus Continuous Reinfusion Autotransfusion System is substantially equivalent in performance to the predicate devices cleared to market via 510(k) K120953. The modification made to the proposed Pleur-evac Sahara® Plus Continuous Reinfusion Autotransfusion System does not introduce any new issues of safety and effectiveness.
Teleflex Medical, Inc.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name arranged in a circular fashion around a stylized graphic. The graphic consists of three curved lines that resemble a person embracing a stylized globe. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
FEB 1 2 2013
Teleflex Medical Incorporated c/o Ms. Natalie Smith Regulatory Affairs Specialist 2917 Weck Drive Research Triangle Park, NC 27709
Re: K130043
Pleur-evac Sahara® Plus Continuous Reinfusion Autotransfusion System Regulation Number: 21 CFR 868.5830 Regulation Name: Autotransfusion Apparatus Regulatory Class: Class II Product Code: CAC Dated: January 22, 2013 Received: January 23, 2013
#### Dear Ms. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Natalie Smith
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
# Matthew G. Hillebrenner
for
Bram D. Zuckerman, M.D. · Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
Page 1 of 1
510(k) Number:
1413004380
Device Name:
Pleur-evac Sahara® Plus Continuous Reinfusion Autotransfusion System
Indications for Use:
#### AUTOTRANSFUSION
- 1. For the collection of autologuous blood from the patient's pleural cavity or mediastinal area for reinfusion purposes in trauma and post-operative situations
### CHEST DRAINAGE
- 1. To evacuate air and/or fluid from the chest cavity or mediastinum
- 2. To help prevent air and/or fluid from re-accumulating in the chest cavity or mediastinum.
- 3. To help re-establish and maintain normal intra-thoracic pressure gradients.
- 4. To facilitate complete lung re-expansion to restore normal breathing dynamics.
The Pleur-evac® Autotransfusion Bag is indicated as a sterile, single use device used for collection and reinfusion of autologous blood from the thoracic cavity when attached to a Pleur-evac® System. The fluid path is non-pyrogenic.
Prescription Use XX (Part 21 CFR 801 Subpart D) AND/OR
Over-the-counter use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Aflelel
(Division Sign-Off)
Division of Cardiovascular Devices
510(k)
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.