PLEUR-EVAC SAHARA PLUS CONTINUOUS REINFUSION AND AUTOTRANSFUSION SYSTEM
K140205 · Teleflexmedical, Inc. · CAC · Oct 23, 2014 · Anesthesiology
Device Facts
Record ID
K140205
Device Name
PLEUR-EVAC SAHARA PLUS CONTINUOUS REINFUSION AND AUTOTRANSFUSION SYSTEM
Applicant
Teleflexmedical, Inc.
Product Code
CAC · Anesthesiology
Decision Date
Oct 23, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5830
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
For the collection of autologous blood from the patient's pleural cavity or mediastinal area for reinfusion purposes in trauma and post-operative situations, To evacuate air and/or fluid from the chest cavity or mediastinum, To help prevent air and/or fluid from re-accumulating in the chest cavity or mediastinum, To help re-establish and maintain normal intra-thoracic pressure gradients, To facilitate complete lung re-expansion to restore normal breathing dynamics.
Device Story
Three-chamber collection/reinfusion system; collects autologous blood from pleural/mediastinal cavities; functions as dry seal/dry suction chest drainage unit post-autotransfusion. Used in clinical settings (trauma/post-operative); operated by healthcare professionals. Input: blood/air/fluid from chest cavity via drainage tubes. Transformation: gravity or suction-assisted collection into chambers; blood transfer via accessory bag for reinfusion. Output: collected blood for reinfusion; evacuated air/fluid. Benefits: restores normal intra-thoracic pressure/breathing dynamics; provides autologous blood source.
Clinical Evidence
Bench testing only. Testing included inter-chamber leak tests, leak integrity tests (at -60 cm H2O), burst strength tests (minimum 5 psig), and reinfusion tube separation force tests (minimum 10 lbs). All tests passed.
Technological Characteristics
Three-chamber system; dry seal/dry suction mechanism. Materials compliant with ISO 10993-1:2009 and FDA Bluebook G95-1. Sterile, single-patient use. Non-pyrogenic fluid path. Mechanical operation; no electronic components or software.
Indications for Use
Indicated for trauma and post-operative patients requiring autologous blood collection from pleural/mediastinal cavities, evacuation of air/fluid from chest/mediastinum, prevention of air/fluid re-accumulation, maintenance of intra-thoracic pressure, and lung re-expansion. Contraindicated for patients with infections (pericardial, mediastinal, systemic, pulmonary), malignant neoplasms, coagulopathies, suspected thoraco-abdominal injuries with enteric contamination, impaired renal function, or those exposed to specific topical agents (thrombin, microfibrillar hemostatic agents, povidone-iodine, non-IV compatible antibiotics) in the thoracic/mediastinal cavity.
Regulatory Classification
Identification
An autotransfusion apparatus is a device used to collect and reinfuse the blood lost by a patient due to surgery or trauma.
Predicate Devices
S-1150-08LF Pleur-evac Sahara Plus Continuous Reinfusion Autotransfusion System (K130043)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure in profile, with three overlapping faces suggesting community and support.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 23, 2014
Teleflex Medical, Inc. % Ms. Holly Hallock Regulatory Affairs Specialist 2917 Weck Dr. P.O. Box 12600 Research Triangle Park, North Carolina 27709
Re: K140205
> Trade/Device Name: Pleur-evac Sahara Plus Continuous Reinfusion And Autotransfusion System Regulation Number: 21 CFR 868.5830 Regulation Name: Autotransfusion Apparatus Regulatory Class: Class II Product Code: CAC Dated: September 9, 2014 Received: September 10, 2014
Dear Ms. Hallock,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
M. A. Hillebrand
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Page 1 of 1
#### 510(k) Number:
K140205
Device Name:
Pleur-evac Sahara® Plus Continuous Reinfusion Autotransfusion System
#### Indications for Use:
- . For the collection of autologous blood from the patient's pleural cavity or mediastinal area for reinfusion purposes in trauma and post-operative situations,
- To evacuate air and/or fluid from the chest cavity or mediastinum,
- To help prevent air and/or fluid from re-accumulating in the chest cavity or ● mediastinum,
- To help re-establish and maintain normal intra-thoracic pressure gradients, ●
- o To facilitate complete lung re-expansion to restore normal breathing dynamics.
Prescription Use XX (Part 21 CFR 801 Subpart D)
AND/OR
Over-the-counter use ___ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
# 510(k) SUMMARY
#### Pleur-evac Sahara® Plus Continuous Reinfusion Autotransfusion System
### A. Name, Address, Phone and Fax Number of Applicant
Teleflex Medical, Incorporated 2917 Weck Drive Research Triangle Park, NC 27709 USA Phone: 919-433-4918 Fax: 919-433-4996
#### B. Contact Person
Holly Kornegay Regulatory Affairs Specialist
Lorraine DeLong RA/QE Manager - Surgical
### C. Date Prepared
January 24, 2014
# D. Device Name
| Trade Name: | Pleur-evac Sahara® Plus Continuous Reinfusion<br>Autotransfusion System |
|----------------------------|-------------------------------------------------------------------------|
| Common Name: | Autotransfusion Apparatus |
| Regulatory Classification: | Class II |
| Regulation Number: | 21 CFR 868.5830 |
| Product Code: | CAC |
# E. Device Description
Provided as a sterile unit and intended for single patient use, the S-1150-08LF Pleurevac Sahara® Plus Continuous Reinfusion Autotransfusion System is a threechamber, collection/reinfusion system used for the collection and continuous reinfusion of autologous blood. By attaching a blood transfer bag, which is available as an accessory item, the S-1150-08LF Pleur-evac Sahara® Plus serves as a bag reinfusion system with a non-pyrogenic fluid path. When autotransfusion is complete, the S-1150-08LF Pleur-evac Sahara® Plus can serve as a dry seal/dry suction chest drainage collection unit.
{4}------------------------------------------------
# F. Indications for Use
Pleur-evac Sahara® Plus Continuous Reinfusion Autotransfusion Systems are indicated for:
- For the collection of autologous blood from the patient's pleural cavity or . mediastinal area for reinfusion purposes in trauma and post-operative situations,
- To evacuate air and/or fluid from the chest cavity or mediastinum,
- To help prevent air and/or fluid from re-accumulating in the chest cavity or . mediastinum,
- To help re-establish and maintain normal intra-thoracic pressure gradients,
- To facilitate complete lung re-expansion to restore normal breathing dynamics.
### G. Contraindications
Pleur-evac Sahara® Plus Continuous Reinfusion Autotransfusion Systems are contraindicated for:
- · Pericardial, mediastinal, or systemic infections,
- · Pulmonary and respiratory infection or infestation,
- · Presence of malignant neoplasms,
- · Coagulopathies,
- · Suspected thoraco-abdominal injuries with possible enteric contamination,
- · Impaired renal function,
- · Intraoperative thoracic or mediastinal cavity use of topical thrombin, microfibrillar hemostatic agents or providine-iodine antiseptic gels or solutions and non I.V. compatible antibiotics.
# H. Substantial Equivalence
The proposed Pleur-evac Sahara® Plus Continuous Reinfusion Autotransfusion System is substantially equivalent to the predicate device:
| Predicate Device | Manufacturer | 510(k) No. | Date Cleared |
|------------------------------------------------------------------------------------------|------------------|------------|-------------------|
| S-1150-08LF Pleur-evac<br>Sahara® Plus Continuous<br>Reinfusion Autoransfusion<br>System | Teleflex Medical | K130043 | February 12, 2013 |
# I. Comparison to Predicate Devices
The Pleur-evac Sahara® Plus Continuous Reinfusion Autotransfusion System has the same indication for use and functional characteristics as the predicate system. The proposed modifications are changes in the material of the faceplate and the universal connector.
{5}------------------------------------------------
### J. Materials
All patient contacting materials are in compliance with ISO 10993-1:2009 and FDA Bluebook Memorandum G95-1.
### K. Technological Characteristics
A comparison of the technological characteristics of the proposed Pleur-evac Sahara® Plus Continuous Reinfusion Autotransfusion System and the predicate has been performed. The results of this comparison demonstrate that the Pleur-evac Sahara® Plus Continuous Reinfusion Autotransfusion System faceplate and universal connector are equivalent to the marketed predicate device in technological characteristics. A summary of these comparisons is included in the table below. For a complete comparison chart, please refer to Section 16.
| Technological<br>Characteristics | Predicate Device,<br>S-1150-08LF | Proposed Device,<br>S-1150-08LF | Comparison |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Indications for<br>Use | For the collection of<br>autologous blood from<br>the patient's pleural<br>cavity or mediastinal<br>area for reinfusion<br>purposes in trauma and<br>post-operative<br>situations, To evacuate air and/or<br>fluid from the chest<br>cavity or mediastinum, To help prevent air<br>and/or fluid from re-<br>accumulating in the<br>chest cavity or<br>mediastinum, To help re-establish<br>and maintain normal<br>intra-thoracic pressure<br>gradients, To facilitate complete<br>lung re-expansion to<br>restore normal<br>breathing dynamics. | For the collection of<br>autologous blood from the<br>patient's pleural cavity or<br>mediastinal area for<br>reinfusion purposes in<br>trauma and post-operative<br>situations, To evacuate air and/or fluid<br>from the chest cavity or<br>mediastinum, To help prevent air and/or<br>fluid from re-accumulating<br>in the chest cavity or<br>mediastinum, To help re-establish and<br>maintain normal intra-<br>thoracic pressure gradients, To facilitate complete lung<br>re-expansion to restore<br>normal breathing dynamics. | Same |
| Collection | Yes, the predicate featured a<br>collection compartment for<br>either gravity or suction<br>drainage and collection. | Yes, the proposed S-1150-08LF<br>features a collection<br>compartment for either gravity<br>or suction drainage and<br>collection. | Same |
| Reinfusion | Yes, when connected to an<br>accessory bag, the predicate | Yes, when connected to an<br>accessory bag, the proposed S- | Same |
Teleflex Medical, Inc.
{6}------------------------------------------------
#### L. Performance Data
Teleflex has performed bench testing to verify that the performance of the proposed Pleur-evac Sahara® Plus Continuous Reinfusion Autotransfusion System is substantially equivalent to the predicate device, and that the Pleur-evac Sahara® Plus Continuous Reinfusion Autotransfusion System is seamlessly interchangeable with the predicate device. Various functionality tests were performed to ensure that the faceplate and universal connector will perform as intended. The test results are summarized below. For more details, please refer to Section 23
| Product<br>Description | Quantity | Test Parameters | Results |
|---------------------------------------------|----------|------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Inter- Chamber<br>Leak Test | 60 | All compartments must<br>hold the indicated volumes<br>with liquid spillover<br>occurring only at the<br>spillover level. | Pass |
| Leak Integrity Test | 60 | The unit shall be airtight<br>when operated at negative<br>pressure 60 cm H20 | Pass |
| Burst Test | 60 | The unit shall have a<br>minimum burst strength of<br>5 psig (PSI Gauge) at<br>welded joints with all<br>openings to atmosphere<br>sealed | Pass |
| Reinfusion Tube<br>Separation Force<br>Test | 60 | The separation force of the<br>tube and spring from the<br>reinfusion port and the<br>tube from the spike port<br>shall be a minimum of 10<br>lbs axially. | Pass |
#### M. Conclusion
Based upon the comparative test results, the proposed Pleur-evac Sahara® Plus Continuous Reinfusion Autotransfusion System is substantially equivalent to the predicate device cleared to market via 510(k) K130043. The modifications made to the proposed Pleur-evac Sahara® Plus Continuous Reinfusion Autotransfusion System do not introduce any new issues of safety and effectiveness.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.