AirLife Adult Heated Wire BiPAP/NIV Circuit is intended for use with adult population that requires mechanical ventilation. The AirLife Adult Heated Wire BiPAP/NIV Circuit is used with spontaneously breathing individuals that benefit from high flow therapy. The product is single use device, non-sterile and used in professional healthcare environments under a doctor's supervision and by skilled clinicians. The AirLife Adult Heated Wire BiPAP/NIV Circuit is designed to work with noninvasive ventilators and compatible to the Fisher & Paykel MR850 humidifier.
Device Story
AirLife Adult Heated Wire BiPAP/NIV Circuit is a single-use, non-sterile corrugated plastic tube with internal spiral resistance wires. It functions as an airway conduit between a noninvasive ventilator and a patient. The device connects to a Fisher & Paykel MR850 humidifier to deliver warmed, humidified breathing gases. Resistance wires generate heat to maintain gas temperature and humidity, minimizing breathing effort. Used in hospital environments for up to 7 days, operated by skilled clinicians. Output is the delivery of conditioned gas to the patient's airway. Benefits include maintained ventilation and reduced work of breathing.
Clinical Evidence
Bench testing only. No clinical data. Performance testing included ISO 5367 (breathing sets), ISO 8185 (humidifiers), ISO 5356-1 (connectors), and electrical safety/EMC (IEC 60601-1, IEC 60601-1-2). Biocompatibility testing performed per ISO 10993 series for prolonged contact (cytotoxicity, sensitization, irritation, implantation, genotoxicity, extractables/leachables). All tests passed.
Technological Characteristics
Corrugated plastic tube with internal spiral resistance wires (Copper/Ni alloy). Single-limb design with 22mm adapters. Max power 36W. Compatible with Fisher & Paykel MR850 humidifier. Operates at 3-60 LPM flow. Standards: ISO 5367, ISO 8185, ISO 5356-1, AAMI/ANSI ES60601-1, AAMI/ANSI/IEC 60601-1-2.
Indications for Use
Indicated for adult patients requiring mechanical ventilation, including spontaneously breathing individuals benefiting from high flow therapy. Used in professional healthcare environments under clinician supervision. Compatible with noninvasive ventilators and Fisher & Paykel MR850 humidifier.
Regulatory Classification
Identification
A breathing system heater is a device that is intended to warm breathing gases before they enter a patient's airway. The device may include a temperature controller.
{0}------------------------------------------------
Public Health Service
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 14, 2017
Vyaire Medical Colleen Watson (O'Keeffe) Acting Director, Regulatory Affairs 75 North Fairway Drive Vernon Hills, Illinois 60061
Re: K170378
Trade/Device Name: AirLife Adult Heated Wire BiPAP/NIV Circuit Regulation Number: 21 CFR 868.5270 Regulation Name: Breathing system heater Regulatory Class: Class II Product Code: BZE Dated: August 11, 2017 Received: August 14, 2017
Dear Colleen Watson (O'Keeffe):
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Tara A. Ryan -S
Lori Wiggins, MPT, CLT Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170378
Device Name
AirLife Adult Heated Wire BiPAP/NIV Circuit
Indications for Use (Describe)
AirLife Adult Heated Wire BiPAP/NIV Circuit is intended for use with adult population that requires mechanical ventilation. The AirLife Adult Heated Wire BiPAP/NIV Circuit is used with spontaneously breathing individuals that benefit from high flow therapy.
The product is single use device, non-sterile and used in professional healthcare environments under a doctor's supervision and by skilled clinicians. The AirLife Adult Heated Wire BiPAP/NIV Circuit is designed to work with noninvasive ventilators and compatible to the Fisher & Paykel MR850 humidifier.
Type of Use (*Select one or both, as applicable*)
| <div> <span> <b> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </b> </span> </div> |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <b> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </b> </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Vyaire Medical. The logo is in blue and consists of the word "Vyaire" in a stylized font, with the word "MEDICAL" in smaller letters underneath. Above the word "Vyaire" are three light blue curved lines.
75 N. Fairway Dr.
Vernon Hills, IL 60061
Phone: 847 362-8107
Vyaire Medical.com
# 510(k) Summary
#### 1. SUBMITTER
Carefusion/Vyaire Medical, Inc. 26125 Riverwoods Blvd Mettawa, IL 60045 Phone: (224) 706-6818
Contact Person: Colleen O'Keeffe Acting Director, Regulatory Affairs
Date Prepared: September 11, 2017
#### 2. Device
| Product Name: | AirLife Adult Heated Wire BiPAP/NIV Circuit |
|-------------------------------|---------------------------------------------------------------------------|
| Trade or proprietary<br>name: | AirLife™ |
| Device Name: | Heated Breathing Circuit |
| Common Name: | Heater, breathing system w/wo controller (not humidifier or<br>nebulizer) |
| Classification Name: | Breathing system heater (21 CFR 868.5270) |
| Regulatory Class: | II |
| Product Code: | BZE |
#### 3. Predicate Device
RT 219 BiLevel /CPAP Circuit was cleared under 510(k) K983112 MR850 Respiratory Humidifier from Fisher and Paykel. The RT219 is an accessory of F&P 850 System. This predicate device has not been subject to a design-related recall.
#### 4. Device Description
The AirLife Adult Heated Wire BiPAP/NIV Circuit is intended to deliver and warm breathing gases before they enter the patient's airway. The heated wire circuit is intended for use for adult population that requires non-invasive (NIV) mechanical ventilation. It is intended for spontaneously breathing individuals who require mechanical ventilation. The duration of use is up to 7 days in a hospital environment. The aim is to maintain adequate ventilation and minimize the effort of breathing.
- 5. Principle of Operation
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Vyaire Medical. The word "Vyaire" is written in a large, bold, blue font. Below the word "Vyaire" is the word "MEDICAL" written in a smaller, thinner, blue font. Above the word "Vyaire" is a blue, wave-like design.
75 N. Fairway Dr.
Vernon Hills, IL 60061
Phone: 847 362-8107
Vyaire Medical.com
The AirLife Adult Heated Wire BiPAP/NIV Circuit is a corrugated plastic tube with a spiral resistance wires within the tubing that generate heat to maintain temperatures and humidity. It is intended to warm gases before they enter a patient's airway. The device delivers humidity to patients requiring active non-invasive humidification, acting as a conduit transporting gases between ventilators, humidifiers and the patient. The AirLife Adult Heated Wire BiPAP/NIV Circuit is designed to operate at a minimum flow rate of 3 LPM and a maximum flow rate of 60 LPM.
#### Indication for use 6.
The AirLife Adult Heated Wire BiPAP/NIV Circuit is intended for use with adult only population that requires mechanical ventilation. The AirLife Adult Heated Wire BiPAP/NIV Circuit is used with spontaneously breathing individuals that benefit from high flow therapy.
The product is single use device, non-sterile and used in professional healthcare environments under a doctor's supervision and by skilled clinicians. The AirLife Adult Heated Wire BiPAP/NIV Circuit is designed to work with noninvasive ventilators and compatible with the Fisher & Paykel MR850 humidifier.
### 7. Comparison of technological characteristics with the predicate device
The fundamental scientific technology is the same for both proposed and predicate device. It is based on acting as an airway conduit between a breathing machine and the patient. The AirLife Adult Heated Wire BiPAP/NIV Circuit is substantially equivalent to the predicate device RT 219 BiLevel /CPAP Circuit regarding safety, effectiveness, design (technology), materials and intended use. The proposed AirLife Adult Heated Wire BiPAP/NIV Circuit is designed to operate at the same minimum flow rate of 3 LPM and maximum 60 LPM. Successful test results (specific enthalpy, humidity output and surface temperature) ensured that the flow rate of the proposed device does not raise any concerns regarding safety and effectiveness. The similarities and differences between the proposed device and the predicate device are outlined below:
| Element of comparison | Proposed Device | Predicate Device |
|---------------------------|-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|
| Intended Use | Intended to warm<br>breathing gases before<br>they enter a patient's<br>airway | Intended to warm breathing<br>gases before they enter a<br>patient's airway |
| Principle of Operation | Resistance wires within the<br>tubing generate heat to<br>maintain temperatures and<br>humidity | Resistance wires within the<br>tubing generate heat to<br>maintain temperatures and<br>humidity |
| Limb Length (Inspiratory) | 5 ft | 5 ft |
| Dryline Length | 1.7 ft | 1.7 ft |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Vyaire Medical. The logo is in blue and features the word "Vyaire" in a stylized font. Below the word "Vyaire" is the word "MEDICAL" in a smaller, sans-serif font. Above the word "Vyaire" are three curved lines in a lighter shade of blue.
75 N. Fairway Dr.
Vernon Hills, IL 60061
Phone: 847 362-8107
Vyaire Medical.com
| Element of comparison | Proposed Device | Predicate Device |
|----------------------------------|-------------------------------------------------|-------------------------------------------------|
| Unheated Extension Length | 1 ft | 1 ft |
| Intended Patient Use | Adult | Adult |
| Usage | Disposable | Disposable |
| Design | Single Limb | Single Limb |
| Flow rate | Max 60 LPM | Max 70 LPM |
| Nominal Inside Diameter | 30 mm | 30 mm |
| Design | Corrugated Breathing Tube<br>with 22mm Adapters | Corrugated Breathing Tube<br>with 22mm Adapters |
| Circuit Design | Single | Single |
| Maximum Power of Heater Base | 60W | 60W |
| Maximum Power of Circuit | 36W | 60W |
| Min circuit resistance per Limb | 20.5 Ω | 12.0 Ω |
| Maximum Power/foot of Limb | 7.2W/ft | 12W/ft |
| Conductor | Copper/Ni alloy | Copper alloy |
| Disposable Exhalation Port (DEP) | Present | Present |
| Compatible Humidifiers | Fisher and Paykel MR850 | Fisher and Paykel MR850 |
#### 8. Performance Data
The proposed device was tested to ensure compliance to the following standards:
# 8.1 Performance Testing
| Performance Characteristic | Standard |
|--------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|
| Anaesthetic and respiratory equipment – Breathing sets<br>and connectors | ISO 5367, 5th Ed: 2014-<br>10-15 |
| Respiratory tract humidifiers for medical use. Particular<br>requirements for respiratory humidification system | ISO 8185, 3rd Ed: 2008-<br>06-15 |
| Anaesthetic and respiratory equipment. Conical<br>connectors. Cones and sockets | ISO 5356-1: 3rd Ed;<br>2004-05-15 |
| Medical Electrical Equipment - Part 1: General<br>Requirements For Basic Safety And Essential<br>Performance (iec 60601-1:2005, mod) | AAMI / ANSI ES60601-<br>1:2005/(R)2012 And<br>A1:2012, |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Vyaire Medical. The logo is in blue and features the word "Vyaire" in a stylized font. Below "Vyaire" is the word "MEDICAL" in a smaller, sans-serif font. Above the word "Vyaire" are two light blue curved lines.
75 N. Fairway Dr. Vernon Hills, IL 60061 Phone: 847 362-8107 Vyaire Medical.com
| Performance Characteristic | Standard |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------|
| | C1:2009/(R)2012 And<br>A2:2010/(R)2012<br>(Consolidated Text) |
| Medical Electrical Equipment -- Part 1-2: General<br>Requirements For Basic Safety<br>And Essential Performance -- Collateral Standard:<br>Electromagnetic Disturbances -- Requirements And Tests | AAMI / ANSI / IEC<br>60601-1-2:2014 |
| Performance Characteristic | Results |
|--------------------------------------------|---------|
| Inspiratory Limb | Pass |
| Extension Line | Pass |
| Dry Line | Pass |
| System Leak Test | Pass |
| Resistance to Flow | Pass |
| % Increase in Flow Resistance with Bending | Pass |
| Compliance of Breathing Tubes | Pass |
| Security of Attachment | Pass |
| Conical Connector | Pass |
| Resistance to Melt | Pass |
| Temperature Sensor Leakage | Pass |
| Temperature Sensor Compatibility | Pass |
| Breathing Tube Surface Temperature | Pass |
| Specific Enthalpy | Pass |
| Humidity Output | Pass |
| ANSI/AAMI IEC 60601-1:2005 | Pass |
| IEC 60601-1-2, Ed. 4.0, 2014-02 | Pass |
#### 8.2 Biocompatibility
Tests for an externally communicating, tissue by way of gas path and direct mucosal contact with prolonged contact (greater than 24 hours but less than 30 days): Cytotoxicity, Sensitization, Irritation, Muscle Implantation, Genotoxicity and Extractables/Leachables
| Performance Characteristic | Standard |
|----------------------------------------------------------------------------------------------------------------------------|------------------------------------|
| Biological Evaluation of Medical Devices-Part 1:<br>Evaluation and Testing | AAMI/ANSI/ISO 10993-1:2009 |
| Biological evaluation of medical devices – Part 3: Tests<br>for genotoxicity, carcinogenicity and reproductive<br>toxicity | AAMI/ANSI/ISO 10993-3:2009 (R2014) |
| Biological Evaluation of Medical Devices-Part 5: Tests<br>for In Vitro Cytotoxicity | AAMI/ANSI/ISO 10993-5:2009 (R2014) |
| Biological evaluation of medical devices – Part 6: Tests<br>for local effects after implantation | AAMI/ANSI/ISO 10993-6:2009 (R2014) |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for Vyaire Medical. The logo consists of the word "Vyaire" in a bold, sans-serif font, with the word "MEDICAL" in a smaller, lighter font underneath. Above the word "Vyaire" are two curved lines in a light blue color, which appear to be stylized representations of air waves.
75 N. Fairway Dr. Vernon Hills, IL 60061
Phone: 847 362-8107 Vyaire Medical.com
| Performance Characteristic | Standard |
|--------------------------------------------------------------------------------------------------------------------|----------------------------------------|
| Biological Evaluation of Medical Devices-Part 10: Tests<br>for Irritation and Skin Sensitization. | AAMI/ANSI/ISO 10993-10:2010<br>(R2014) |
| Biological evaluation of medical devices – Part 11:<br>Tests for systemic toxicity | AAMI/ANSI/ISO 10993-11:2006<br>(R2010) |
| Biological evaluation of medical devices – Part 12:<br>Sample preparation and reference materials | AAMI/ANSI/ISO 10993-12: 2012 |
| Biological Evaluation of Medical Devices Part 17:<br>Establishment of allowable limits for leachable<br>substances | AAMI/ANSI/ISO 10993-17:2002 |
| Biological Evaluation of Medical Devices Part 18:<br>Chemical characterization of materials | AAMI/ANSI/ISO 10993-18:2005 |
# 9. Conclusion
The non-clinical data demonstrates that the AirLife Adult Heated Wire BiPAP/NIV Circuit is as safe and as effective as the predicate and therefore substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.