The myAir app is indicated for patients: • prescribed with a compatible ResMed Air11 platform device to simulate therapy prior to using their device with their prescribed settings, and to configure their settings to support therapy. It is an optional software accessory to allow patients to acclimate to and operate their therapy device. • prescribed a NightOwl wearable device to provide the user interface to operate the connected device and aid in the home sleep testing process. The device is intended for home and hospital use for: • new and existing patients of ResMed Air10 and Air11 PAP therapy devices and • new users who are prescribed a compatible NightOwl home sleep test (HST).
Device Story
myAir is a mobile medical application serving as a patient-facing interface for ResMed PAP therapy devices and NightOwl home sleep test (HST) sensors. The app connects via Bluetooth to compatible hardware to enable patient self-monitoring, therapy usage tracking, and device control. Users operate the app to start/stop therapy and adjust comfort settings on connected PAP devices. It also facilitates data transfer between the hardware and cloud-based backends (Machine Cloud Service or EctoSense server). The device is intended for use by patients in home or hospital settings. By providing a user interface for device operation and acclimation, the app aims to support therapy adherence and simplify the home sleep testing process. It does not perform diagnostic analysis or display diagnostic results.
Clinical Evidence
Bench testing only. No clinical data provided. Verification and validation testing included non-functional requirements and end-to-end functional testing to confirm performance requirements for the added device control software functions.
Technological Characteristics
Mobile software application (app). Connectivity: Bluetooth and HTTPS (cellular/wireless). Platform: Smartphone. Interoperability: ResMed Air11 PAP devices and NightOwl HST. Software functions: Device control (start/stop therapy, comfort settings), data communication, and therapy simulation. No patient-contacting components.
Indications for Use
Indicated for patients >30kg prescribed compatible ResMed Air11 PAP devices for therapy simulation and configuration, or prescribed NightOwl wearable devices for home sleep testing. Intended for home and hospital use by new and existing PAP patients and new HST users.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
K211155 — Auto CPAP System · BMC Medical Co., Ltd. · Mar 9, 2022
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 28, 2025
ResMed Corp
Rose Malonzo
Senior Specialist, Regulatory Affairs
9001 Spectrum Center Blvd.
San Diego, California 92123
Re: K250624
Trade/Device Name: myAir
Regulation Number: 21 CFR 868.5905
Regulation Name: Noncontinuous Ventilator (IPPB)
Regulatory Class: Class II
Product Code: BZD
Dated: February 28, 2025
Received: March 3, 2025
Dear Rose Malonzo:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K250624 - Rose Malonzo
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K250624 - Rose Malonzo
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Rachana Visaria -S
Rachana Visaria, Ph.D.
Assistant Director
DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K250624 | |
| Device Name myAir | |
| Indications for Use (Describe) The myAir app is indicated for patients: • prescribed with a compatible ResMed Air11 platform device to simulate therapy prior to using their device with their prescribed settings, and to configure their settings to support therapy. It is an optional software accessory to allow patients to acclimate to and operate their therapy device. • prescribed a NightOwl wearable device to provide the user interface to operate the connected device and aid in the home sleep testing process. The device is intended for home and hospital use for: • new and existing patients of ResMed Air10 and Air11 PAP therapy devices and • new users who are prescribed a compatible NightOwl home sleep test (HST). | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
ResMed
myAir
Traditional 510(k)
# 510(k) SUMMARY
[As required by 21 CFR 807.92(c)]
Date of Submission: 28 February 2025
Company Name/Owner: ResMed Corp
9001 Spectrum Center Blvd
San Diego, CA 92123
USA
Official Contact: Mrs. Rose Malonzo
Senior Specialist, Global Product Regulatory Affairs
ResMed Corp
9001 Spectrum Center Blvd
San Diego, CA 92123 USA
Tel: 858-285-5670
rose.malonzo@resmed.com
Device Trade Name: myAir
Device Common Name: Ventilator, Non-Continuous (Respirator)
Classification and Classification Name: Noncontinuous ventilator (IPPB)
Product Code: BZD
MNR (secondary)
Predicate Device: myAir
K241216
Submission Reason: Modified device
510(k) Summary
{5}
ResMed
myAir Traditional 510(k)
# Device Description
myAir is a companion mobile medical application ("app") that serves as a patient self-monitoring, therapy usage tracking, and engagement platform for patients prescribed with compatible ResMed PAP therapy devices and the NightOwl home sleep test (HST) sensors. The app allows the patient to connect via Bluetooth to a compatible hardware device for control of their prescribed PAP or HST device and to allow self-tracking of device usage data. myAir can also be used as a communication pathway using the Bluetooth connection with the compatible device in order to send or receive data. Analysis of patient diagnostic data or display of diagnostic results are not in scope of myAir features.
The myAir subject device for this premarket submission adds device software functions related to starting and stopping therapy, and control of comfort settings (collectively referred to as "device control"). The addition of these device software functions within the myAir subject device is a modification to the K241216 myAir predicate device, in addition to sustaining previously cleared medical device functions.
# Indications for Use
The myAir app is indicated for patients:
- prescribed with a compatible ResMed Air11 platform device to simulate therapy prior to using their device with their prescribed settings, and to configure their settings to support therapy. It is an optional software accessory to allow patients to acclimate to and operate their therapy device.
- prescribed a NightOwl wearable device to provide the user interface to operate the connected device and aid in the home sleep testing process.
The device is intended for home and hospital use for:
- new and existing patients of ResMed Air10 and Air11 PAP therapy devices and
- new users who are prescribed a compatible NightOwl home sleep test (HST).
# Non-clinical testing
Non-clinical verification and validation testing completed for device software functions under review introduced with the modifications to the myAir app demonstrated that the device met all intended performance requirements. Testing included:
- Software verification and validation, including non-functional requirements and end-to-end functional testing.
The following FDA guidance were conformed to:
- Content of Premarket Submissions for Device Software Functions: June 2023
- Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions: September 2023
510(k) Summary
{6}
ResMed
myAir Traditional
510(k)
# Substantial Equivalence
The subject device and predicate device have a similar intended use, and similar technological characteristics and operating principle.
The predicate for this premarket submission is myAir (K241216) as this predicate has a similar intended use and technological characteristics to the subject device myAir. Similar technological characteristics include:
- Bluetooth connectivity and communication
- Executing Test Drive workflow (device function cleared under K200656)
- Executing HST workflow (device function cleared under K241216)
The technological difference between the predicate and subject device is:
- The subject device adds device control functions (device functions under review)
510(k) Summary
Page 3 of 7
{7}
ResMed
myAir
Traditional 510(k)
| Characteristic | Predicate Device: myAir
**Manufacturer:** ResMed Corp
**510(k) Number:** K241216 | Subject Device: myAir with device control functions
**Manufacturer:** ResMed Corp
**510(k) Number:** K250624 | Substantial Equivalence |
| --- | --- | --- | --- |
| Indications for Use | The myAir app is indicated for patients:
• prescribed with a compatible ResMed Air11 platform device to simulate therapy prior to using their device with their prescribed settings. It is an optional software accessory to allow patients to acclimate to their therapy device.
• prescribed a NightOwl wearable device to provide the user interface to operate the connected device and aid in the home sleep testing process.
The device is intended for home and hospital use for:
• new and existing patients of ResMed Air10 and Air11 PAP therapy devices and
• new users who are prescribed a compatible NightOwl home sleep test (HST). | The myAir app is indicated for patients:
• prescribed with a compatible ResMed Air11 platform device to simulate therapy prior to using their device with their prescribed settings and to configure their settings to support therapy. It is an optional software accessory to allow patients to acclimate to and operate their therapy device.
• prescribed a NightOwl wearable device to provide the user interface to operate the connected device and aid in the home sleep testing process.
The device is intended for home and hospital use for:
• new and existing patients of ResMed Air10 and Air11 PAP therapy devices and
• new users who are prescribed a compatible NightOwl home sleep test (HST). | The proposed myAir indications for use adds language to support device control functions. This device modification uses the same technological characteristics to execute the added device software functions. The difference in the indications for use does not raise new questions of safety or efficacy for the subject device. |
| Regulation Number | 21 CFR §868.5905 | 21 CFR §868.5905 | Identical |
510(k) Summary
Page 4 of 7
{8}
ResMed
myAir
Traditional 510(k)
| Characteristic | Predicate Device: myAir
Manufacturer: ResMed Corp
510(k) Number: K241216 | Subject Device: myAir with device control functions
Manufacturer: ResMed Corp
510(k) Number: K250624 | Substantial Equivalence |
| --- | --- | --- | --- |
| Classification Name | Noncontinuous ventilator (IPPB) | Noncontinuous ventilator (IPPB) | Identical |
| FDA Product Code | BZD | BZD | Identical |
| Prescription Use | Yes | Yes | Identical |
| Patient Population | Patients weighing more than 66 lb (30kg) | Patients weighing more than 66 lb (30kg) | Identical |
| Intended Environment of Use | Home | Home | Identical |
| Patient Contacting | No, myAir is software. | No, myAir is software. | Identical |
| Display Type | Smartphone display | Smartphone display | Identical |
| Device Interoperability and V&V | ResMed Air11 platform devices and NightOwl HST | ResMed Air11 platform devices and NightOwl HST | Identical, however device software functions related to NightOwl HST are not impacted by this modification and are therefore not in scope for this premarket notification. |
| Device Connection Requirement | Yes. myAir app must be connected to a compatible AS11 or HST device to operate device functions as intended. | Yes. myAir app must be connected to a compatible AS11 or HST device to operate device functions as intended. | Identical, however device software functions related to NightOwl HST are not impacted by this modification and are therefore not in scope for this premarket notification. |
510(k) Summary
Page 5 of 7
{9}
ResMed
myAir
Traditional 510(k)
| Characteristic | Predicate Device:
myAir | Subject Device:
myAir with device control functions | Substantial Equivalence |
| --- | --- | --- | --- |
| | **Manufacturer:**
ResMed Corp | **Manufacturer:**
ResMed Corp | |
| | **510(k) Number:**
K241216 | **510(k) Number:**
K250624 | |
| Device Control | Yes. myAir can control the connected therapy or HST device. | Yes. myAir can control the connected therapy or HST device. | Identical |
| Communication Pathways Available | Bluetooth and HTTPS (cellular or wireless internet connection) | Bluetooth and HTTPS (cellular or wireless internet connection) | Identical |
| Device data to app communication pathway | Pathways:
1) Device data transfers to Machine Cloud Service (MCS) then to Galapagos.
2) Therapy device data transferred from device to app to MCS.
3) HST device data transfers from sensor through the myAir app to EctoSense backend server for analysis. | Pathways:
1) Device data transfers to Machine Cloud Service (MCS) then to Galapagos.
2) Therapy device data transferred from device to app to MCS.
3) HST device data transfers from sensor through the myAir app to EctoSense backend server for analysis. | Identical |
| App Store Availability | Google Play Store | Apple App Store | Difference in mobile app store availability does not raise new questions of safety or efficacy for the subject device. |
| Therapy Settings Changes | None | None | Identical |
| Adjust Comfort Settings | None | Yes. The subject device myAir can adjust comfort settings on the connected AS11 therapy device. | Difference in functionality does not raise new questions of safety or efficacy for the subject device. |
| Software Level of Documentation | Basic | Basic | Identical |
510(k) Summary
Page 6 of 7
{10}
ResMed
myAir
Traditional 510(k)
# Substantial Equivalence Conclusion
The subject device is substantially equivalent to the predicate K241216 because:
- They have similar intended use;
- They have similar technological characteristics;
- They have similar performance characteristics;
- The differences do not raise any new questions of safety or effectiveness; and
- It is as safe and effective as the predicate device.
510(k) Summary
Page 7 of 7
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.