The ConchaTherm® Neptune™ is a respiratory humidifier designed to heat and humidify respiratory gases delivered via endotracheal tubes, nasal cannula or face masks to adult, pediatric, infant and neonatal patients. This system may be used with either conventional (non-heated wire) breathing circuits or compatible (21-volt) Hudson Respiratory Care Incorporated (RCI) heated-wire circuits. The ConchaTherm® Neptune™ can be used with ventilators, continuous flow systems, oxygen diluters and blenders, adjustable nebulizer adapters for aerosol therapy, or non-flammable anesthesia gases to help maintain patient body temperature.
Device Story
ConchaTherm® Neptune™ is a servo-controlled heated humidifier system; provides heat and moisture to respiratory gases. Components include humidifier unit, CONCHA-COLUMN® cartridge, sterile water reservoir, dual temperature probe, and breathing circuit (heated or non-heated). Device monitors proximal airway temperature; regulates heat to column and circuit. Used in clinical settings with ventilators, oxygen blenders, or anesthesia machines. Healthcare providers operate system to maintain patient body temperature and prevent airway drying. Benefits include improved patient comfort and respiratory gas conditioning.
Clinical Evidence
Bench testing only. Compliance verified against IEC/EN 60601-1 (General Safety), CAN/CSA C22.2 No. 601.1-M90, IEC/EN 60601-1-2 (Electromagnetic Compatibility), and ISO 8185 (Humidifiers for Medical Use).
Technological Characteristics
Servo-controlled heated humidifier; includes CONCHA-COLUMN® cartridge and dual temperature probe. Compatible with 21-volt heated-wire or non-heated breathing circuits. Standards: IEC/EN 60601-1, IEC/EN 60601-1-2, ISO 8185. Connectivity: Integrated with respiratory gas delivery systems (ventilators, blenders).
Indications for Use
Indicated for adult, pediatric, infant, and neonatal patients requiring heated and humidified respiratory gases delivered via endotracheal tubes, nasal cannula, or face masks during ventilation, continuous flow therapy, or anesthesia.
Regulatory Classification
Identification
A respiratory gas humidifier is a device that is intended to add moisture to, and sometimes to warm, the breathing gases for administration to a patient. Cascade, gas, heated, and prefilled humidifiers are included in this generic type of device.
K131912 — CONCHATHERM NEPTUNE HEATED HUMIDIFIER, COMFORT FLO HUMIDIFICATION SYSTEM, CONCHASMART COLUMN · Teleflex, Inc. · Mar 6, 2014
K162242 — Comfort Flo Humidification System · Teleflexmedical, Inc. · Jan 19, 2017
K993355 — HUDSON RCI CONCHA COLUMN, CAT. NO. 385-40 AND HUDSON RCI CIRCUIT, CAT. NO. 780-40, 780-41, 780-42, 780-43 · Hudson Respiratory Care, Inc. · May 24, 2000
K141940 — CONCHASMART COLUMN · Teleflex, Inc. · Nov 24, 2014
K071958 — THERA-HEAT HEATED HUMIDIFIER AND ADULT DUAL AND SINGLE LIMB HEATED WIRE VENTILATOR BREATHING CIRCUITS · Smiths Medical Asd, Inc. · Nov 21, 2007
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Teleflex Medical. The word "Teleflex" is in a bold, sans-serif font, with a stylized "T" that resembles a medical device. Below "Teleflex" is the word "MEDICAL" in a smaller, sans-serif font. The logo is in black and white.
eleflex Medical Incorporated 2917 Weck Drive Research Triangle Park, NC 27709 USA 866-246-6990 Phone: 919-361-3914 Fax: www.teleflex.com
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
APR 2 3 2007
# The ConchaTherm® Neptune™ Heated Humidifier
## A. Name, Address, Phone and Fax Number of Applicant
Teleflex Medical Incorporated 2917 Weck Drive Research Triangle Park, NC 27709 USA Phone: 919-361-3927 919-361-4061 Fax:
### B. Contact Person
Elizabeth (Betty) Landon Sr. Regulatory Affairs Specialist
#### C. Date Prepared
April 23, 2007
#### D. Device Name
Trade Name: The ConchaTherm® Neptune™ heated humidifier
Common Name: Humidification System
Classification Name: Respiratory Gas Humidifier (21 CFR 868.5450, Product Code BTT)
#### E. Device Description
The ConchaTherm® Neptune™ is designed to provide heat and moisture to the medical gases delivered to a patient via continuous flow, invasive ventilation or non-invasive ventilation. The ConchaTherm® Neptune™ heated humidifier is the main component of the Teleflex Medical ConchaTerm Humidification system, which allow gas to be delivered to a patient using various interfaces.
The ConchaTherm® Neptune™ heated humidification system consists of the ConchaTherm® Neptune™ heated humidifier, the CONCHA-COLUMN® humidifier cartridge, CONCHA sterile water reservoir, a Hudson RCI dual temperature probe, and either a non-heated wire breathing circuit or a Hudson RCI 21-volt heated-wire breathing circuit. The servo-controlled heated humidifier continuously monitors the proximal airway temperature and regulates the heat supplied to the column by the heating element, and if used, the heated-wire circuit.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Teleflex Medical. The logo is black and consists of the word "Teleflex" in a bold, sans-serif font. Below the word "Teleflex" is the word "MEDICAL" in a smaller, sans-serif font. The logo is simple and modern.
Teleflex Medical Incorporated 2917 Weck Drive Research Triangle Park, NC 27709 USA Phone: 866-246-6990 Fax: 919-361-3914 www.teleflex.com
## F. Indications for Use
The ConchaTherm® Neptune™ is a respiratory humidifier designed to heat and humidify respiratory gases delivered via endotracheal tubes, nasal cannula or face masks to adult, pediatric, infant and neonatal patients. This system may be used with either conventional (non-heated wire) breathing circuits or compatible (21-volt) Hudson Respiratory Care Incorporated (RCI) heated-wire circuits.
The ConchaTherm® Neptune™ can be used with ventilators, continuous flow systems, oxygen diluters and blenders, adjustable nebulizer adapters for aerosol therapy, or nonflammable anesthesia gases to help maintain patient body temperature.
## G. Substantial Equivalence
The device is similar in intended use, materials, design, and performance characteristics to the Teleflex Medical ConchaTherm IV® (K923946) and the Fisher & Paykel MR850 (K033710). The determination of substantial equivalence for this device was based on a detailed device description, performance testing, and conformance with voluntary performance standards.
### H. Summary of Testing
The ConchaTherm® Neptune™ heated humidifier has been tested and evaluated for compliance with the Standard for Safety of Medical Electrical Equipment, Part 1: General Requirements for Safety (IEC/EN 60601-1). Medical Electrical Equipment. Part 1: General Requirements for Safety, Collateral Standard, CAN/CSA C22.2 No. 601.1-M90 (IEC/EN 60601-1); Standard for Electromagnetic Compatibility and Tests (IEC/EN 60601-1-2) and Humidifiers for Medical Use-General Requirements for Humidification Systems (ISO 8185).
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS). The logo features a stylized caduceus, a symbol often associated with healthcare, consisting of a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Elizabeth Landon Senior Regulatory Affairs Specialist Teleflex Medical 2917 Weck Drive Research Triangle Park, North Carolina 27709
APR 2 3 2007
Re: K063758
.
Trade/Device Name: ConchaTherm® Neptune™ Heated Humidifier Regulation Number: 21 CFR 868.5450 Regulation Name: Respiratory Gas Humidifier Regulatory Class: Class II Product Code: BTT Dated: March 16, 2007 Received: March 19, 2007
Dear Ms. Landon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
#### Page 2 - Ms. Landon
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Susan Thorne
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known): K063758
Device Name: The ConchaTherm® Neptune™
Indications for Use:
The ConchaTherm® Neptune™ is a respiratory humidifier designed to heat and humidify respiratory gases delivered via endotracheal tubes, nasal cannula or face masks to adult, pediatric, infant and neonatal patients. This system may be used with either conventional (non-heated wire) breathing circuits or compatible (21-volt) Hudson Respiratory Care Incorporated (RCI) heated-wire circuits.
The ConchaTherm® Neptune™ can be used with ventilators, continuous flow systems, oxygen diluters and blenders, adjustable nebulizer adapters for aerosol therapy, or non-flammable anesthesia gases to help maintain patient body temperature.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Cullh
Page 1 of
Anesthesiology, General Hospital, on Control Demes
:(K) Number: K06375-8
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.