K162242 · Teleflexmedical, Inc. · BTT · Jan 19, 2017 · Anesthesiology
Device Facts
Record ID
K162242
Device Name
Comfort Flo Humidification System
Applicant
Teleflexmedical, Inc.
Product Code
BTT · Anesthesiology
Decision Date
Jan 19, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5450
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Comfort Flo Humidification System provides a continuous flow of heated and humidified gas to spontaneously breathing patients. It is indicated for single use by neonate/infant, pediatric, and adult patients in professional healthcare environments.
Device Story
Comfort Flo Humidification System delivers heated, humidified respiratory gases to spontaneously breathing patients. System components include heated wire breathing circuit, ConchaSmart column, and accessories. Operates in conjunction with Neptune Heated Humidifier using ConchaSmart technology. Humidification achieved via paper wick within sterile ConchaSmart column; water absorbed by wick directly in airflow; column warmed to increase water/air temperature for vapor creation. Used in professional healthcare environments (hospitals) by clinicians. Output is conditioned gas (near 37°C, 10mg/l humidity) at flow rates 1-60 LPM. Heated wire circuit controls condensation (rainout). Benefits include respiratory support for patients requiring humidified gas. Device is disposable/single-use.
Clinical Evidence
Bench testing only. No clinical data. Testing included ISO 8185 (respiratory humidifiers), ISO 5356-1 (connectors), ISO 5367 (breathing tubes), and comprehensive biocompatibility (ISO 10993 series). Performance testing verified enthalpy, resistance to flow, leakage, humidity output, warm-up time, temperature tracking, and rainout control post-aging and throughout useful life.
Technological Characteristics
Materials: Polypropylene, PVC, Polystyrene, Silicone, Engage, Cyclohexanone, Cyclohexane/Styrene Butadiene adhesive. Sensing/Actuation: Heated wire circuit, paper wick humidification. Energy: Heated humidifier base. Form factor: Breathing circuit with 10mm/15mm ID tubing. Connectivity: Standalone system. Sterilization: Non-sterile. Software: None.
Indications for Use
Indicated for neonate/infant, pediatric, and adult patients requiring continuous flow of heated and humidified gas in professional healthcare environments. For use with spontaneously breathing patients.
Regulatory Classification
Identification
A respiratory gas humidifier is a device that is intended to add moisture to, and sometimes to warm, the breathing gases for administration to a patient. Cascade, gas, heated, and prefilled humidifiers are included in this generic type of device.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles facing right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 19, 2017
Teleflex Medical, Inc. Brian Gall Senior Regulatory Affairs Specialist 2917 Weck Drive Research Triangle Park, NC 27709
Re: K162242 Trade/Device Name: Comfort Flo Humidification System Regulation Number: 21 CFR 868.5450 Regulation Name: Respiratory gas humidifier Regulatory Class: Class II Product Code: BTT Dated: December 16, 2016 Received: December 19, 2016
Dear Mr. Gall:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D. Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162242
Device Name Comfort Flo Humidification System
Indications for Use (Describe)
The Comfort Flo Humidification System provides a continuous flow of heated and humidified gas to spontaneously breathing patients. It is indicated for single use by neonate/infant, pediatic, and adult patients in professional healthcare environments.
| Type of Use (Select one or both, as applicable) |
|--------------------------------------------------------------------------------------|
| <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
| <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY
### A. Name, Address, Phone and Fax Number of Applicant
Teleflex Medical, Incorporated 2917 Weck Drive Research Triangle Park, NC 27709 USA Phone: 919-228-4350 919-361-3939 Fax:
### B. Contact Person
Brian Gall Senior Regulatory Affairs Specialist
## C. Date Prepared
16 December 2016
### D. Device Name
| Trade Name: | Comfort Flo Humidification System |
|-----------------------|-----------------------------------|
| Classification Name: | Respiratory Gas Humidifier |
| Product Code: | BTT |
| Regulation Number: | 868.5450 |
| Classification: | II |
| Classification Panel: | Anesthesiology |
## E. Predicate Device
This submission demonstrates substantial equivalence to the following predicate devices:
- Comfort Flo Humidification System K131912 ●
### F. Device Description
The Comfort Flo Humidification System is intended to deliver heated and humidified respiratory gases to spontaneously breathing patients. The Comfort Flo Humidification System is designed to be used in conjunction with the Neptune Heated Humidifier with ConchaSmart Technology for neonate/infant, pediatric, and adult populations. The Comfort Flo Humidification System includes a Heated Wire Breathing Circuit, a ConchaSmart Column, and various Class I exempt accessories.
### G. Indications for Use
The Comfort Flo Humidification System provides a continuous flow of heated and humidified gas to spontaneously breathing patients. It is indicated for single use by neonate/infant, pediatric, and adult patients in professional healthcare environments.
{4}------------------------------------------------
# H. Technological Characteristics Comparison to the predicate
| Comparative<br>Characteristics | Comfort Flo Humidification<br>System<br>K131912 | Proposed Comfort Flo<br>Humidification System |
|-------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended<br>Use/Indications<br>for Use | To provide a continuous flow of<br>heated and humidified gas to<br>spontaneously breathing patients. | The Comfort Flo Humidification<br>System provides a continuous flow of<br>heated and humidified gas to<br>spontaneously breathing patients. It<br>is indicated for single use by<br>neonate/infant, pediatric, and adult<br>patients in professional healthcare<br>environments. |
| Environment of<br>Use | Professional healthcare environment<br>(ex. Hospital) | Professional healthcare environment<br>(ex. Hospital) |
| Patient<br>Population | Neonate/infant, pediatric, Adult | Neonate/infant, pediatric, Adult |
| Gas Source<br>Compatibility | Non-Ventilator (continuous flow<br>system) | Non-Ventilator (continuous flow<br>system) |
| Compatible<br>Humidifiers | ConchaTherm Neptune Heated<br>Humidifier | ConchaTherm Neptune Heated<br>Humidifier |
| Compatible<br>Columns | ConchaSmart Column | ConchaSmart Column |
| Humidification<br>Method | Paper wick within the sterile Concha<br>Column. Water fills the bottom of the<br>column and is absorbed up the wick<br>which is directly in the air flow. The<br>column is warmed to increase water<br>and air temperature to aid the<br>creation and absorption of water<br>vapor into the airstream. | Paper wick within the sterile Concha<br>Column. Water fills the bottom of the<br>column and is absorbed up the wick<br>which is directly in the air flow. The<br>column is warmed to increase water<br>and air temperature to aid the creation<br>and absorption of water vapor into the<br>airstream. |
| Gas Delivery | Inspiratory limb only | Inspiratory limb only |
| Water Type | Sterile water sold separately | Sterile water sold separately |
| Enthalpy | Per ISO 8185:2007<br>Thermal overshoot to the patient<br>connection port limited to an energy<br>equivalent to 43°C and 100% relative<br>humidity when averaged over any 30<br>second period | Per ISO 8185:2007<br>Thermal overshoot to the patient<br>connection port limited to an energy<br>equivalent to 43°C and 100% relative<br>humidity when averaged over any 30<br>second period |
| Disposable vs.<br>Reusable | Disposable, single use | Disposable, single use |
| Sterility | Gamma sterilized ConchaSmart<br>column. Circuit and all other<br>accessories are non-sterile | Non-sterile |
| Patient Interface<br>Type | Nasal Cannula (sold separately, not<br>part of this 510(k)) | Nasal Cannula (sold separately, not<br>part of this 510(k)) |
| Comfort Flo<br>Humidification<br>System<br>Components | ● Heated Wire Breathing Circuit<br>● ConchaSmart Column<br>● Various Class 1 Exempt devices | ● Heated Wire Breathing Circuit<br>● ConchaSmart Column<br>● Various Class 1 Exempt devices |
| Comparative<br>Characteristics | Comfort Flo Humidification<br>System<br>K131912 | Proposed Comfort Flo<br>Humidification System |
| Heated Wire<br>Breathing Circuit | 10 mm ID smooth bore PVC tubing<br>(catalog# 2010 and 2414) | 10 mm ID smooth bore PVC tubing<br>(catalog# 2010 and 2414) |
| Tubing | 15 mm ID corrugated<br>Polypropylene/Engage tubing<br>(catalog# 2415 and 2416) | 15 mm ID corrugated<br>Polypropylene/Engage tubing<br>(catalog# 2415 and 2416) |
| Simulated Use | Circuit, column, and nasal cannula<br>assembly operate at 37°C (+/- 2°C)<br>for up to 37 days with visible<br>condensation, without loss of<br>functional integrity. | Circuit, column, and nasal cannula<br>assembly operate at 37°C (+/- 2°C)<br>for up to 21 days with visible<br>condensation, without loss of<br>functional integrity. |
| Humidity Output | System delivers gas conditioned at<br>or near 37°C and 10mg/l of humidity<br>to the patient interface from 1 to 60<br>LPM. | System delivers gas conditioned at or<br>near 37°C and 10mg/l of humidity to<br>the patient interface from 1 to 60 LPM. |
| Condensation<br>(rainout) | System controls amount of<br>condensation in the circuit using the<br>heated wire settings of the<br>ConchaTherm Neptune Heated<br>Humidifier from 1 to 60 LPM. | System controls amount of<br>condensation in the circuit using the<br>heated wire settings of the<br>ConchaTherm Neptune Heated<br>Humidifier from 1 to 60 LPM. |
| Shelf Life | 5 year shelf life on the Corrugated<br>Comfort Flo Kits (catalog# 2415 and<br>2416) | 1 year shelf life on the Corrugated<br>Comfort Flo Kits (catalog# 2415 and<br>2416) |
| Useful Life | 30 Days | 21 Days |
| | Polypropylene | Polypropylene (different blend) |
| | PVC | PVC |
| | Polypropylene | Polypropylene |
| | Polystyrene | Polystyrene |
| Breathing Circuit<br>Materials | Silicone | Silicone |
| | Cyclohexanone | Cyclohexanone |
| | Engage | Engage (different blend) |
| | Cyclohexane / Styrene Butadiene<br>(adhesive) | Cyclohexane / Styrene Butadiene<br>(adhesive) |
{5}------------------------------------------------
The following are the primary differences in the proposed Comfort Flo Humidification System:
- . The Indications for Use Statement has been updated per to better align with 21 CFR 814.20(b)(3)(i). There are no changes to the content or meaning of the Indications for Use from the predicate to the proposed device.
- The ConchaSmart column (part of the Comfort Flo Humidification System) included in the ● predicate device was sterilized by gamma radiation. The circuit and other accessories were non-sterile. The ConchaSmart column was made non-sterile in submission K141940. The proposed Comfort Flo Humidification system is completely non-sterile.
{6}------------------------------------------------
- . The shelf life and useful life of the predicate are different from the proposed device. The proposed shelf life and useful life are based on test data.
- There are two new materials in the proposed Heated Wire Breathing Circuit ● has undergone all necessary biocompatibility and performance tests
## I. Performance Data
The following testing was performed on the proposed device.
- . ISO 8185:2007 - Respiratory tract humidifiers for medical use -- Particular requirements for respiratory humidification systems, applicable sections ISO 5356-1:2004 - Anaesthetic and respiratory equipment -- Conical connectors -- Part 1: Cones and sockets, applicable sections
- ISO 5367:2000 (Section 4.4, and Annexes D and E, per ISO 8185) Breathing tubes ● intended for use with anaesthetic apparatus and ventilators, applicable sections
- . Biocompatibilitv
- o ISO 10993-1:2009 Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process
- o ISO 10993-5:2009 Biological evaluation of medical devices -- Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10:2010 Biological evaluation of medical devices -- Part 10: o Tests for irritation and skin sensitization
- ISO 10993-12:2012 Biological evaluation of medical devices -- Part 12: O Sample preparation and reference materials
- o ISO 10993-17:2002 Biological evaluation of medical devices -- Part 17: Establishment of allowable limits for leachable substances
- ISO 10993-18:2002 Biological evaluation of medical devices -- Part 18: o Chemical characterization of materials
- Shelf life ●
- . Useful life
## J. Summary of Non-Clinical Testing
| Test Description | Test Objective | Results |
|-------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Shelf life / Useful life<br>testing | The proposed Corrugated Comfort Flo Heated<br>Wire Breathing Circuit, when used with the<br>Hudson RCI Neptune Heated Humidifier must<br>perform as intended when subject to the following<br>test methods:<br>Pre- and Post-aging Visual Inspection Length per ISO 5367 Connection strength testing Maximum Temperature / Minimum flow testing ISO 8185 Section 101 Useful life testing Rainout control | PASS |
{7}------------------------------------------------
| Test Description | Test Objective | Result |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Enthalpy related to<br>flow variations,<br>Resistance to Flow,<br>Resistance to Flow<br>with a bend, and<br>Leakage, Humidity<br>Output, Enthalpy<br>related to use and<br>system errors, Warm<br>Up Time and<br>Temperature Tracking | The proposed Corrugated Comfort Flo Heated<br>Wire Breathing Circuit, when used with the<br>Hudson RCI Neptune Heated Humidifier must<br>perform as intended post accelerated aging and<br>useful life testing when subject to the test<br>methods from the following applicable sections of<br>ISO 8185:<br>Section 51.103 Section 56.101.2 and 56.101.3 Section 101 Section 51.103 with faults from 3.6 bb Section 6.8.2 a) 14 | PASS |
## K. Conclusion
The proposed Comfort Flo Humidification System is substantially equivalent to the predicate devices listed above in that the indications for use, the intended use, and fundamental scientific technology remain unchanged. The difference between the predicate device and the proposed device is the material of construction of the corrugated tubing on the Comfort Flo Heated Wire Breathing Circuit. The tubing material is being changed to a different grade of the same material. The proposed device was tested with equivalent biocompatibility and performance testing as the predicate, and met all of the acceptance criteria.
The device data and test results demonstrate that the device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.