FLUENCY PLUS TRACHEOBRONCHIAL STENT GRAFT

K050832 · Bard Peripheral Vascular, Inc. · JCT · Jul 7, 2005 · General, Plastic Surgery

Device Facts

Record IDK050832
Device NameFLUENCY PLUS TRACHEOBRONCHIAL STENT GRAFT
ApplicantBard Peripheral Vascular, Inc.
Product CodeJCT · General, Plastic Surgery
Decision DateJul 7, 2005
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.3720
Device ClassClass 2
AttributesTherapeutic

Intended Use

The FLUENCY® Plus Tracheobronchial Stent Graft is indicated for use in the treatment of tracheobronchial strictures produced by malignant neoplasms or in benign strictures after all alternative therapies have been exhausted.

Device Story

Self-expanding Nitinol stent encapsulated with carbon-impregnated ePTFE; pre-loaded on flexible pull-back delivery system. Used to maintain patency of tracheobronchial tree. Radiopaque tantalum markers facilitate placement. Physician-operated in clinical setting. Modifications include inner catheter tip and outer catheter material/configuration to reduce deployment force, minimize tip entanglement, and improve placement precision. Device provides mechanical support to strictured airways, benefiting patients by restoring airway patency.

Clinical Evidence

Bench testing only. Design verification and validation performed in accordance with FDA Guidance for the Content of Premarket Notifications for Esophageal and Tracheal Prostheses (April 28, 1998).

Technological Characteristics

Self-expanding Nitinol stent; carbon-impregnated ePTFE encapsulation; radiopaque tantalum markers. Flexible pull-back delivery system. Available in various lengths and diameters. Sterilization method not specified. No software or electronic components.

Indications for Use

Indicated for patients with tracheobronchial strictures caused by malignant neoplasms or benign strictures where alternative therapies have failed.

Regulatory Classification

Identification

The tracheal prosthesis is a rigid, flexible, or expandable tubular device made of a silicone, metal, or polymeric material that is intended to be implanted to restore the structure and/or function of the trachea or trachealbronchial tree. It may be unbranched or contain one or two branches. The metal tracheal prosthesis may be uncovered or covered with a polymeric material. This device may also include a device delivery system.

Special Controls

The special control for this device is FDA's “Guidance for the Content of Premarket Notification Submissions for Esophageal and Tracheal Prostheses.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K050832 pg. lot 3 ## JUL - 7 2005 ecial 510(k) LUENCY® Plus Tracheobronchial Stent Graft Page 78 ## FLUENCY® Plus Tracheobronchial Stent Graft 510(k) Summary of Safety and Effectiveness 21 CFR 807.92(a). As required by the Safe Medical Devices Act of 1990, coded under Section 513, Part (I)(3)(A) of the Food, Drug, and Cosmetic Act, a summary of the safety and effectiveness information upon which substantial equivalence determination is based as follows: #### General Information: | Submitter Name:<br>Address: | Bard Peripheral Vascular, Inc.<br>1625 W. Third Street<br>P.O. BOX 1740<br>Tempe, AZ 85280-1740 | |----------------------------------|-------------------------------------------------------------------------------------------------| | Telephone Number:<br>Fax Number: | (480) 894-9515<br>(480) 449-2546 | | Contact Person: | Dennis Salzmann, Ph.D.<br>Regulatory Affairs Technical Manager | #### Device Information: | Device Trade Name: | FLUENCY® Plus Tracheobronchial Stent Graft | |--------------------|--------------------------------------------| | Common/Usual Name: | Tracheal Prosthesis | | Classification: | Class II with Special Controls | The special controls for this device are compliant with FDA Guidance for the Content of Premarket Notifications for Esophageal and Tracheal Prostheses. April 28, 1998. Classification Panel: General & Plastic Surgery #### Predicate Device: FLUENCY® Tracheobronchial Stent Graft (K031041, cleared June 19, 2003) #### Summary of Change: The design modifications to the FLUENCY® Tracheobronchial Stent Graft are modification of the inner catheter tip, and modification to the outer catheter material and configuration. These changes result in a subject device that has a reduced deployment force, reduced risk of tip entanglement during withdrawal of the delivery system, allows Bard Peripheral Vascular, Inc TRADE SECRET/CONFIDENTIAL INFORMATIC Notify CR Bard Before Releasing this Docume Image /page/0/Picture/18 description: The image shows the word "BARD" in a bold, sans-serif font. The letters are outlined in black, giving them a distinct and clear appearance. The overall design is simple and straightforward, focusing on the legibility of the word. {1}------------------------------------------------ K050832 pg 2 of 3 for more precise stent graft placement during deployment, and provides more delivery system length options for physicians. All other aspects of the subject device remain the same as the predicate device. #### Device Description: The FLUENCY® Plus Tracheobronchial Stent Graft includes a self-expanding Nitinol stent encapsulated with carbon-impregnated ePTFE pre-loaded on a flexible "pull-back" delivery system. It is a single use prosthetic device designed to maintain the patency of the tracheobronchial tree in patients with tracheobronchial strictures. Highly radiopaque tantalum markers on the stent graft ends facilitate stent graft placement. The FLUENCY® Plus Tracheobronchial Stent Graft is available in various lengths and diameters. It is preloaded into various size delivery catheters, depending on the size of the stent graft and is available in various delivery system lengths. #### Intended Use of Device: The FLUENCY® Plus Tracheobronchial Stent Graft is indicated for use in the treatment of tracheobronchial strictures produced by malignant neoplasms or in benign strictures after all alternative therapies have been exhausted. ### Technological Comparison to Predicate Device: The technological characteristics of FLUENCY® Plus Tracheobronchial Stent Graft are substantially equivalent to those of the predicate FLUENCY® Tracheobronchial Stent Graft in terms of intended use, application, user population, basic design, performance, labeling, packaging, and sterilization method. #### Non-Clinical Performance Data: Design verification and validation of the modified device was done with conformance to or evaluated based on the following FDA guidance: Guidance for the Content of Premarket Notifications for Esophageal and Tracheal Prostheses. April 28, 1998. Image /page/1/Picture/15 description: The image shows the word "BARD" in a bold, sans-serif font. The letters are outlined in black, giving them a distinct appearance. The overall design is simple and clear, making the word easily readable. {2}------------------------------------------------ K050832 pg 3 of 3 All test results confirm the modified device to be substantially equivalent to the predicate device. #### Conclusions: The FLUENCY® Plus Tracheobronchial Stent Graft met all the predetermined performance criteria of design verification and validation as specified by applicable standards, guidance, test protocols and/or customer inputs. The FLUENCY® Plus Tracheobronchial Stent Graft is substantially equivalent to the legally marketed predicate device, the FLUENCY® Trachobronchial Stent Graft. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it, often used as a symbol of medicine. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA" is arranged in a circular fashion around the caduceus symbol. JAN 1 6 2007 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Bard Peripheral Vascular, Inc. % Mr. Joshua Smale Regulatory Affairs Specialist 1625 West 3rd Street P.O. Box 1740 Tempe, Arizona 85280 Re: K050832 Trade/Device Name: FLUENCY® Plus Tracheobronchial Stent Graft Regulation Number: 21 CFR 878.3720 Regulation Name: Tracheal prosthesis Regulatory Class: II Product Code: JCT Dated: June 6, 2005 Received: June 7, 2005 Dear Mr. Smale: This letter corrects our substantially equivalent letter of July 7, 2005. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set {4}------------------------------------------------ Page 2 - Mr. Joshua Smale forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to continue marketing your device as described in your Section 510(k) I mis letter will and in your he FDA finding of substantial equivalence of your device to a legally promatics noticate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you a the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Kio Mark N. Melkerson Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health {5}------------------------------------------------ # Indications for Use 510(k) Number (if known): K050832 Device Name: FLUENCY® Plus Tracheobronchial Stent Graft Indications for Use: The FLUENCY® Plus Tracheobronchial Stent Graft is indicated for use in the treatment of tracheobronchial strictures produced by malignant neoplasms. Prescription Use X (Part21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CORH, Office of Device Evaluation (ODE) Signature 100-01 Division of General, Restorative, and Neurological Devices 310(k) Number /<050f32 Image /page/5/Picture/17 description: The image shows the word "BARD" in a stylized, blocky font. The letters are bold and outlined in black, giving them a three-dimensional appearance. The overall design is simple and geometric.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from the tree

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...