VASCULAR ARCHITECTS ASPIRE COVERED STENT AND 718 DELIVERY SYSTEM
K031641 · Vascular Architects, Inc. · JCT · Aug 1, 2003 · General, Plastic Surgery
Device Facts
Record ID
K031641
Device Name
VASCULAR ARCHITECTS ASPIRE COVERED STENT AND 718 DELIVERY SYSTEM
Applicant
Vascular Architects, Inc.
Product Code
JCT · General, Plastic Surgery
Decision Date
Aug 1, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3720
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Vascular Architects aSpire® Covered Stent and 718™ Delivery System are indicated for use in the treatment of tracheobronhial strictures produced by malignant neoplasms.
Device Story
The aSpire® Covered Stent is a spiral stent composed of Nickel Titanium encapsulated in a covering; used to treat tracheobronchial strictures caused by malignant neoplasms. The 718™ Delivery System allows a physician to expand the stent within the stricture and confirm positioning before release. The device is used in a clinical setting to maintain airway patency, providing mechanical support to the airway wall to alleviate obstruction caused by tumors. The stent remains in the patient to provide long-term structural support.
Clinical Evidence
Bench testing only. Simulated use and performance testing were conducted to evaluate mechanical integrity and device performance, demonstrating substantial equivalence to predicate devices.
Technological Characteristics
Spiral stent constructed from Nickel Titanium, fully encapsulated in a covering. Delivery system designed for stent expansion and positioning confirmation. Class II device, product code JCT.
Indications for Use
Indicated for treatment of tracheobronchial strictures caused by malignant neoplasms in patients requiring airway stenting.
Regulatory Classification
Identification
The tracheal prosthesis is a rigid, flexible, or expandable tubular device made of a silicone, metal, or polymeric material that is intended to be implanted to restore the structure and/or function of the trachea or trachealbronchial tree. It may be unbranched or contain one or two branches. The metal tracheal prosthesis may be uncovered or covered with a polymeric material. This device may also include a device delivery system.
Special Controls
The special control for this device is FDA's “Guidance for the Content of Premarket Notification Submissions for Esophageal and Tracheal Prostheses.”
Predicate Devices
Vascular Architects aSpire® Covered Stent and Delivery Catheter (K003173)
Vascular Architects aSpire® Covered Stent and Controlled Expansion® Delivery System (K012544)
Vascular Architects aSpire® Covered Stent and Controlled Expansion® Delivery System - 15cm stent length (K030567)
Related Devices
K030567 — MODIFICATION TO VASCULAR ARCHITECTS ASPIRE COVERED STENT AND CONTRILLED EXPANSION DELIVERY SYSTEM · Vascular Architects, Inc. · Mar 24, 2003
K220424 — Through the Scope Tracheal Stent System · Micro-Tech (Nanjing) Co., Ltd. · Jul 8, 2022
K012544 — VASCULAR ARCHITECTS ASPIRE COVERED STENT AND CONTROLLED EXPANSION DELIVERY SYSTEM · Vascular Architects, Inc. · Nov 15, 2001
K964121 — WALLSTENT TRACHEOBRONCHIAL ENDOPROSTHESIS · Boston Scientific Scimed, Inc. · Dec 4, 1996
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image contains the words "Public Health Service" in bold black font. The words are stacked on top of each other, with "Public Health" on the top line and "Service" on the bottom line. The text appears to be extracted from a document or sign.
Image /page/0/Picture/2 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized image of three human profiles facing to the right, stacked on top of each other. The profiles are connected and form a single, abstract shape. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" is arranged in a circular pattern around the image.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV - 3 2006
The Wallace Enterprises, Inc. DBA Vascular Architects % Mr. Kevin F. MacDonald Regulatory Consultant 229 Marvilla Circle Pacifica, California 94044
Re: K031641
Trade/Device Name: Vascular Architects aSpire® Covered Stent and 718™ Delivery System
Regulation Number: 21 CFR 878.3720 Regulation Name: Tracheal prosthesis Regulatory Class: II Product Code: JCT Dated: May 27, 2003 Received: May 27, 2003
Dear Mr. MacDonald:
This letter corrects our substantially equivalent letter of August 1, 2003.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{1}------------------------------------------------
Page 2 - Mr. Kevin F. MacDonald
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Herbert Lemons
1 Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
Vascular Architects aSpire® Covered Stent Amendment to K012544, K030567, K031641 July 28, 2006
# Indications for Use
510(k) Number (if known):
Device Name:
K031641
718TM Delivery System
Indications for Use:
The Vascular Architects aSpire® Covered Stent and 718™ Delivery System are indicated for use in the treatment of tracheobronhial strictures produced by malignant neoplasms.
Vascular Architects aSpire® Covered Stent and
Prescription Use X
OR
Over-The-Counter Use
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Helmut Lehmann
(Per 21 CFR 801.109)
(Division ptional Format 1-2-96) Division of General, Restorative, and Neurological Devices
**510(k) Number** K031641
Page 8
{3}------------------------------------------------
03/641 p. iota
Vascular Architects aSpire® Covered Stent Amendment to K012544, K030567, K031641 July 28, 2006
# 510(k) Summary of Safety and Effectiveness
This summary of 510(k) safety and effectiveness information is being submitted in I his summary of 5 ro(it) sareems of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is K031641.
## General Information
| Date Amended: | July 28, 2006 |
|----------------|------------------------------------------------------------------------------------------------|
| Classification | Class II, Tracheal Prosthesis per 21 CFR § 878.3720 |
| Product Code | JCT |
| Common Name: | Tracheal Stent |
| Trade Name | Vascular Architects aSpire® Covered Stent and 718TM<br>Delivery System |
| Submitter | Wallace Enterprises, Inc. DBA Vascular Architects<br>1650 Elm Hill Pike<br>Nashville, TN 37210 |
| Contact | Kevin F. MacDonald<br>Regulatory Consultant<br>Tel 415 600 0875 |
#### Intended Use
The Vascular Architects aSpire® Covered Stent and 718" Delivery System are indicated The Vascular Arenteets applies Connial strictures produced by malignant neoplasms.
#### Predicate Devices
- K003173 Vascular Architects aSpire® Covered Stent and Delivery Catheter
- K012544 Vascular Architects aSpire® Covered Stent and Controlled Expansion® Delivery System
- K030567 Vascular Architects aSpire® Covered Stent and Controlled Expansion® Delivery System - 15cm stent length
#### Performance Standards
Performance Standards have not yet been established by the FDA under section 514 of the Federal Food, Drug and Cosmetic Act.
#### Device Description
The Vascular Architects aSpire® Covered Stent is a spiral stent made from Nickel The Vasculal Arentects aspires OOVE ou other of Steen Steen at the proximal and
{4}------------------------------------------------
K03164/ p. 2 of 2
Vascular Architects aSpire® Covered Stent Amendment to K012544, K030567, K031641 July 28, 2006
distal ends completely encapsulating the Nickel Titanium. The 718TM Delivery Catheter is designed to allow the user to expand the stent within the stricture and confirm positioning prior to release.
## Comparison To Predicate Device
The Vascular Architects aSpire® Covered Stent shares the same intended use as the predicate device and is identical in mode of action. The materials used in the subject device are similar to those used in the predicate device. The only difference between the subject device and the predicate device is in the length of the stent.
## Testing Summary
Simulated use and performance testing was conducted on the Vascular Architects aSpire® Covered Stent and 718™ Delivery System. Results of the bench testing performed demonstrate the mechanical integrity and device performance of the subject device is substantially equivalent to that of the predicate device.
# Statement of Substantial Equivalence
The Vascular Architects aSpire® Covered Stent and 718™ Delivery System is substantially equivalent to the legally marketed Vascular Architects aSpire™ Covered Stent size Vascular Architects aSpire® Covered Stent and Controlled Expansion® Delivery System.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.