K220424 · Micro-Tech (Nanjing) Co., Ltd. · JCT · Jul 8, 2022 · General, Plastic Surgery
Device Facts
Record ID
K220424
Device Name
Through the Scope Tracheal Stent System
Applicant
Micro-Tech (Nanjing) Co., Ltd.
Product Code
JCT · General, Plastic Surgery
Decision Date
Jul 8, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3720
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Through the Scope Tracheal Stent System is in the treatment of tracheobronchial strictures caused by malignant neoplasms.
Device Story
Device is a flexible delivery system preloaded with a self-expanding Nitinol metallic stent; stent is fully covered with silicone membrane and Parylene N coating to restrict tumor in-growth. Used in clinical settings by trained healthcare professionals; deployed through an endoscope working channel under direct vision. Delivery system features coaxial sheaths and inner core for desheathing, deployment, and repositioning. Stent maintains tracheal luminal patency; benefits patient by relieving airway obstruction caused by malignant neoplasms.
Clinical Evidence
No clinical data. Bench testing only, including biocompatibility (ISO 10993-1), dimension, expansion force, compression/recoil, deployment force/accuracy, dislodgement, migration, tensile strength, repositioning, corrosion, fatigue, austenite finish temperature, shelf-life (ASTM F1980-16), sterilization (ISO 11135), and MR compatibility (ASTM F2052, F2182, F2119, F2213, F2503).
Indicated for treatment of tracheobronchial strictures caused by malignant neoplasms in patients requiring airway patency maintenance.
Regulatory Classification
Identification
The tracheal prosthesis is a rigid, flexible, or expandable tubular device made of a silicone, metal, or polymeric material that is intended to be implanted to restore the structure and/or function of the trachea or trachealbronchial tree. It may be unbranched or contain one or two branches. The metal tracheal prosthesis may be uncovered or covered with a polymeric material. This device may also include a device delivery system.
Special Controls
The special control for this device is FDA's “Guidance for the Content of Premarket Notification Submissions for Esophageal and Tracheal Prostheses.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
Micro-Tech (Nanjing) Co., Ltd. Sally He RA Engineer No.10 Gaoke Third Road, Nanjing National Hi-Tech Industrial Development Zone Nanjing, Jiangsu Province 210032 China
Re: K220424
Trade/Device Name: Through the Scope Tracheal Stent System Regulation Number: 21 CFR 878.3720 Regulation Name: Tracheal Prosthesis Regulatory Class: Class II Product Code: JCT Dated: June 9, 2022 Received: June 13, 2022
Dear Sally He:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Brandon Blakely, Ph.D. Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory, and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220424
Device Name Through the Scope Tracheal Stent System
Indications for Use (Describe)
The Through the Scope Tracheal Stent System is in the treatment of tracheobronchial strictures caused by malignant neoplasms.
Type of Use (Select one or both, as applicable)
× Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a logo with the letters "MT" in a stylized, three-dimensional block font. The letters are a solid blue color, and the logo has a registered trademark symbol (®) in the upper right corner. The logo has a modern and corporate feel.
## 510K Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
The assigned 510(k) Number: K220424
## 1. Date of Preparation: 2022-07-07
## 2. Sponsor Identification
## Micro-Tech (Nanjing) Co., Ltd.
No.10 Gaoke Third Road, Nanjing National Hi-Tech, Industrial Development Zone, Nanjing,
Jiangsu Province, PRC
Establishment Registration Number: 3004837686
Contact Person: Sally He
Position: RA Engineer
Tel: +86-25-58646395
Fax: +86-25-58350006
Email: RA.Micro-Tech@outlook.com
### 3. Identification of Proposed Device
Trade Name: Through the Scope Tracheal Stent System
Common Name: Tracheal Stent
## Regulatory Information
Classification Name: Tracheal Prosthesis
Classification: 2
Product Code: JCT
Regulation Number: 878.3720
Review Panel: Anesthesiology
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a logo with the letters "MT" in a stylized, three-dimensional design. The letters are primarily blue, with white highlights that create a sense of depth and dimension. A registered trademark symbol is located in the upper right corner of the logo.
#### Identification of Predicate Device 4.
510(k) Number: K202204 Product Name: Tracheal Stent System (OTW) Manufacturer: Micro-Tech (Nanjing) Co., Ltd.
#### 5. Indications for Use
The Through the Scope Tracheal Stent System is indicated for use in the treatment of tracheobronchial strictures caused by malignant neoplasms.
### Device Description 6.
The Through the Scope Tracheal Stent System consists of a flexible delivery system preloaded with a self-expanding implantable metallic stent. The stent is made of Nitinol wire weaved in a tubular mesh shape. This structure may make the stent more flexible, compliant and self-expanding. The stent is fully covered with silicone membrane and a polymer coating. The Parylene N is added on the surface of silicone membrane to restrict tumor in-growth through the wire mesh. A retrieval loop made of PE&PP is threaded through the proximal and distal ends of the stent and is intended to aid in removal during the stent placement procedure. The stent has different dimension with the diameters of 10mm, 12mm, 16mm, 18mm, with the lengths of 20mm, 30mm, 40mm, 50mm, 60mm, 80mm.
The stent is deployed through the endoscopy working channel and under direct vision of endoscopy. The delivery system allows for desheathing to deploy and reposition the stent during the placement procedure. The delivery system consists of two coaxial sheaths and one inner core. The outer sheath serves to constrain the stent until being retracted during the stent deployment. The middle sheath serves to support the delivery system. The round tip acts as a guide when the delivery system enters the body. The front handle is used for deploying the stent. The seal ring, locking ring, and safe lock work to lock the device and prevent the stent from being exposed. The decoration nut connects with the back handle.
The device is supplied sterile, intended for single use only, and is available for prescription use only. Use of this device is restricted to a trained healthcare professional.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a blue logo with the letters "MT" in a stylized, blocky font. The letters are interconnected and have a three-dimensional appearance, suggesting depth. A small, circled "R" symbol, indicating a registered trademark, is positioned in the upper right corner of the logo.
## 7. Comparison of Technological Characteristics
The Through the Scope Tracheal Stent System incorporates substantially equivalent device materials, design, configuration, packaging fundamental technology, sterilization process and intended use as those featured in the predicate device Tracheal Stent System (OTW).
| | Proposed Device | Predicate Device | | | |
|----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| Item | Through the Scope Tracheal | Tracheal Stent System (OTW) | Remark | | |
| | Stent System | (K202204) | | | |
| Product Code | JCT | JCT | Same | | |
| Regulation No. | 878.3720 | 878.3720 | Same | | |
| Class | 2 | 2 | Same | | |
| Supplied in Sterile | Yes | Yes | Same | | |
| Configuration | Stent and delivery system | Stent and delivery system | Same | | |
| Diameter of Stent<br>(mm) | 10, 12, 14, 16, 18 | 10, 12, 14, 16, 18, 20, 22 | Similar | | |
| Length of Stent (mm) | 20, 30, 40, 50, 60, 80 | 20, 30, 40, 50, 60, 70, 80, 100 | Similar | | |
| Maximum OD (D) of<br>Delivery System<br>(mm) | 2.7 | 4, 6 | Different | | |
| Working Length<br>(mm) | 1200 | 650 | Different | | |
| Covering | Fully Covered | Partially Covered,<br>Fully Covered | Similar | | |
| Main Stent material | Nitinol | Nitinol | Same | | |
| Main Introduction<br>system materials | PTFE, Pebax, Peek | PTFE, Pebax, Peek | Same | | |
| Compatible<br>endoscopy working<br>channel | ≥2.8mm<br>Olympus bronchoscope larger<br>than 2.8mm working channel is<br>recommended. | N/A, the device does not pass<br>through the working channel of<br>endoscopy. | Different | | |
| Surgical Technique | The Through the Scope Tracheal<br>Stent System can be directly<br>implanted into the airways<br>through the flexible bronchoscopy | Over the Wire, insert the delivery<br>system through the guidewire | Different | | |
| Item | Proposed Device<br>Through the Scope Tracheal<br>Stent System | Predicate Device<br>Tracheal Stent System (OTW)<br>(K202204) | Remark | | |
| | working channel and the<br>deployment is under the direct<br>vision of bronchoscopy. | | | | |
| Indications for Use | The Through the Scope Tracheal<br>Stent System is indicated for use<br>in the treatment of<br>tracheobronchial strictures caused<br>by malignant neoplasms. | The Tracheal Stent System (OTW)<br>is indicated for use in the<br>treatment of tracheobronchial<br>strictures caused by malignant<br>neoplasms. | Same | | |
| | Stent function | Maintaining tracheal luminal<br>patency in tracheal strictures | Maintaining tracheal luminal<br>patency in tracheal strictures | Same | |
| | | Principle of operation | The proposed device consists of<br>the stent and delivery system. The<br>outer sheath of the delivery<br>system serves to constrain the<br> | The proposed device consists of<br>the stent and delivery system. The<br>outer sheath of the delivery system<br>serves to constrain the stent before<br>deployment. Loosen the safe lock,<br>then withdraw the front handle to<br>deploy the stent. | Same |
| | | | Single Use | Yes | Yes |
| | Packaging | | Single-use EO sterilized blister<br>with one device per blister | Single-use EO sterilized blister<br>with one device per blister | Same |
| Shelf Life | Two years | | Two years | Same | |
| Biocompatibility | Conform to ISO 10993-1 | | Conform to ISO 10993-1 | Same | |
| Sterilization | EO Sterilized, SAL:10-6 | | EO Sterilized, SAL:10-6 | Same | |
| Labeling | Conform to 21 CFR part 801 | Conform to 21 CFR part 801 | Same | | |
| MRI information | Comply with ASTM F 2503,<br>ASTM F 2052, ASTM F2119,<br>ASTM F2182. ASTM F2213 | Comply with ASTM F 2503,<br>ASTM F 2052, ASTM F2119,<br>ASTM F2182. ASTM F2213 | Same | | |
# Comparison to Predicate Devices:
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a blue logo with the letters 'MT' in a stylized, blocky font. The letters are interconnected and appear to be in a three-dimensional perspective. A registered trademark symbol (®) is located in the upper right corner of the logo.
The proposed device Through the Scope Tracheal Stent System is similar in design to Tracheal Stent System (OTW), which consists of a flexible delivery system preloaded with a self-expanding implantable metallic stent. Both stents are made of Nitinol wire by fabricating as a single, integral
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows a blue logo with the letters "MT" in a stylized, three-dimensional font. The letters are bold and appear to be interconnected, creating a sense of unity and strength. A small "R" in a circle, indicating a registered trademark, is positioned in the upper right corner of the logo.
framework tube and a covering is applied to the surface of the stent.
The dimensions of proposed device are covered within the range of that of the predicate device. All comparative non-clinical performance testing have been tested and have met the requirements of substantial equivalence to the predicate device. Both the proposed device and predicated device is EO sterilized and has a two-year shelf life. After EO sterilized and aging, the bench testing and sterility testing of the proposed device meet the requirements of substantial equivalence to the predicate device.
Therefore, the difference between proposed device and predicated device is considered not to affect substantial equivalence between the proposed and predicate devices concerning safety and effectiveness.
### 8. Performance Data
The biocompatibility evaluation for the Through the Scope Tracheal Stent System was conducted in accordance with ISO 10993-1: 2009 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process" and FDA's biocompatibility guidance, Use of International Standard ISO-10993-1, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a risk management process (issued on September 4, 2020,) the following tests were conducted:
Stent Biocompatibility Testing:
- a) Vitro Cytotoxicity
- Skin Sensitization b)
- Irritation c)
- Acute Systemic Toxicity d)
- Pyrogen e)
- f) Muscle Implant
- Chemical Characterization and Biological Risk Assessment g)
Delivery System Biocompatibility Testing:
- a) Vitro Cytotoxicity
- Skin Sensitization b)
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows a blue logo with the letters "MT" in a stylized, three-dimensional design. The letters are bold and appear to be interconnected, with the "M" slightly overlapping the "T". A registered trademark symbol (®) is located in the upper right corner of the logo.
Irritation c)
The device specific guidance document was consulted in preparing this premarket submission, "Guidance for the content of premarket notifications for esophageal and tracheal prostheses issued April 28th, 1998". The following tests were conducted and evaluated for the subject device:
- Dimension Testing a)
- Silicone Thickness Test b)
- c) Expansion Force Testing
- Compression and Recoil Testing d)
- Deployment Force Testing e)
- f) Deployment Accuracy and Damage Testing
- Dislodgement Testing g)
- Migration Force and Removability Testing h)
- i) Tensile Strength Testing
- j) Repositioning Testing
- k) Corrosion Testing
- 1) Fatigue testing
- m) Austenite Finish Temperature Testing:
Shelf-life testing and packaging integrity testing was conducted based on an accelerated aging test in accordance with ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices and ISO 11607-1:2019: Packaging for terminally sterilized medical devices -Part 1: Requirements for materials, sterile barrier systems and packaging systems and ISO 11607-2: 2019: Packaging for terminally sterilized medical devices - Part 2: Validation requirements for forming, sealing and assembly processes. Two-years aging test was performed to demonstrate longer stability and support the results of the accelerated aging test.
Sterilization validation was carried out in accordance with ISO 11135:2014+A1:2018 "Sterilization of Health Care products - Ethylene Oxide - Part 1: Requirements for Development, Validation, and
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows a blue logo with the letters "MT" in a stylized, three-dimensional font. The letters are interconnected, with the "M" on the left and the "T" on the right. A small "R" in a circle, indicating a registered trademark, is positioned in the upper right corner of the logo. The logo has a clean and modern design.
Routine Control of Sterilization processes for Medical Devices".
MR compatibility was evaluated in accordance with ASTM F 2052-15 Standard Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in the Magnetic Resonance Environment, ASTM F2182-19e2 Standard Test Method for Measurement of Radio Frequency Induced Heating On or Near Passive Implants During Magnetic Resonance Imaging, ASTM F2119-07(2013)Standard Test Method for Evaluation of MR Image Artifacts from Passive Implants, ASTM F2213-17 Standard Test Method for Measurement of Magnetically Induced Torque on Medical Devices in the Magnetic Resonance Environment, ASTM F2503 - 13 Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment and FDA guidance on Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment issued on May 20, 2021.
The results of all the performance testing demonstrated that the proposed device met the acceptance criteria and support substantial equivalence to the predicate device Tracheal Stent System (OTW) cleared under K202204.
### 9. Clinical Test Conclusion
No clinical study is included in this submission.
### Substantially Equivalent (SE) Conclusion 10.
Based on the indications for use, technological characteristics, and safety and performance testing, the Through the Scope Tracheal Stent System has been shown to be appropriate for its intended use and is considered to be substantially equivalent to the currently cleared predicate device Tracheal Stent System (OTW) cleared under K202204.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.