AcuPebble Ox100 is a wearable device intended for use in the recording, analysis, displaying, exporting, and storage of biophysical parameters to aid in the evaluation of adult patients with, or with suspected, obstructive sleep apnea (OSA). The device is primarily intended for home setting use (although can also be used in healthcare settings) under the direction of a Healthcare Professional (HCP).
Device Story
AcuPebble Ox100 is a wearable home sleep apnea test (HSAT) system. It consists of a neck-worn physiological sensor (identical to AcuPebble SA100), a finger-worn pulse oximetry sensor, a mobile app, and cloud-based analysis software. The neck sensor captures acoustic signals (heart sounds, respiration, snoring) and movement; the finger sensor captures PPG signals. The mobile app guides the patient, collects data, and uploads it to the cloud. Cloud software analyzes signals to derive OSA diagnostic parameters (AHI, ODI). Healthcare professionals access results via a web interface to evaluate patients. The device aids in OSA diagnosis by providing objective physiological data, potentially reducing the need for full polysomnography in appropriate cases.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including biocompatibility (ISO 10993), electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), home healthcare environment safety (IEC 60601-1-11), and software validation (IEC 62304). SpO2 performance is compliant with ISO 80601-2-61.
Technological Characteristics
Neck sensor: piezoelectric microphones; Finger sensor: LED/photodiode (PPG). Connectivity: Bluetooth to mobile app, then cloud-based analysis. Power: Rechargeable Li-poly (neck) and Li-ion (finger) batteries. Biocompatibility: ISO 10993-1. Software: Moderate level of concern. Non-sterile.
Indications for Use
Indicated for adult patients with or suspected of having obstructive sleep apnea (OSA). Intended for use in home or healthcare settings under the direction of a healthcare professional.
Regulatory Classification
Identification
A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 3, 2023
Acurable Limited Esther Rodriguez-Villegas Cso Finsgate, 5-7 Cranwood Street London, London EC1V 9EE United Kingdom
Re: K222950
Trade/Device Name: AcuPebble Ox100 Regulation Number: 21 CFR 868.2375 Regulation Name: Breathing Frequency Monitor Regulatory Class: Class II Product Code: MNR Dated: December 19, 2022 Received: December 27, 2022
Dear Esther Rodriguez-Villegas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
## Rachana Visaria -S
Rachana Visaria, Ph.D. Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known)
K222950
Device Name
AcuPebble Ox100
Indications for Use (Describe)
AcuPebble Ox100 is a wearable device intended for use in the recording, analysis, displaying, exporting, and storage of biophysical parameters to aid in the evaluation of adult patients with, or with suspected, obstructive sleep apnea (OSA). The device is primarily intended for home settings use (although can also be used in healthcare settings) under the direction of a Healthcare Professional (HCP).
Type of Use (Select one or both, as applicable)
| <div> <span> <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" stroke-width="1" width="16"></rect> <path d="M2,2 L14,14 M14,2 L2,14" stroke="black" stroke-width="2"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" stroke-width="1" width="16"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments reqarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An aqency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR §807.92.
| 510(k) Submitter: | Acurable Limited<br>Finsgate<br>5-7 Cranwood Street<br>London EC1V 9EE<br>United Kingdom |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact person: | Professor Esther Rodriguez-Villegas<br>CSO<br>Acurable Ltd<br>5th Floor, 21 Knightsbridge, London<br>SW1X 7LY, UK<br>esther@acurable.com<br>+442075946193 |
| Date summary prepared: | February 2, 2023 |
| Trade Name: | AcuPebble Ox100 |
| Product Classification Name: | Ventilatory effort recorder |
| Regulation number: | 21 CFR §868.2375 (Breathing Frequency<br>Monitor) |
| Product Code: | MNR |
| Classification Panel: | Anesthesiology |
| Device Classification: | Class II |
{4}------------------------------------------------
### 1. Intended Use / Indications for Use
AcuPebble Ox100 is a wearable device intended for use in the recording, analysis, displaying, exporting, and storage of biophysical parameters to aid in the evaluation of adult patients with, or with suspected, obstructive sleep apnea (OSA). The device is primarily intended for home setting use (although can also be used in healthcare settings) under the direction of a Healthcare Professional (HCP).
### 2. DEVICE DESCRIPTION
AcuPebble Ox100 is prescribed by a Health Care Professional for the patient to use in the home as a 'home sleep apnea test' (HSAT). AcuPebble Ox100 comprises: 1- A small physiological sensing device, identical to the previously cleared device AcuPebble SA100 (K210480) that is worn on the front of the neck around the suprasternal notch area (the "AcuPebble SA100 sensor"), and attaches with a single use biocompatible adhesive tape (same one as for "AcuPebble SA100" too); 2- Another physiological wireless sensing device ("AcuPebble finger oximetry sensor") worn on the finger; 3- A mobile device app (the "AcuPebble Ox100 app") that guides the patient through the steps of the test, collects the data from the sensor (continuously in the case of the "AcuPebble SA100 sensor", and at the end of the test in the case of the "AcuPebble finger oximetry sensor"), and uploads them in the cloud ; 4- A cloud-based software that analyses the signals from the "AcuPebble SA100 sensor" (already cleared as part of K210480); 5- A web app user interface for healthcare professionals where they can set up a study and see the results, including different physiological channel traces and OSA diagnostic parameters. Outputs of the device relevant to its intended use are further described in the comparison table below.
### 3. PREDICATE DEVICE
The predicate device is shown in the Table below:
| K number | Product Code | Class | Device Name |
|----------|--------------|-------|--------------------|
| K210480 | MNR | II | AcuPebble<br>SA100 |
{5}------------------------------------------------
## QCUI e
### 4. Summary of Comparison to Predicate and Reference
| | DEVICE SUBJECT OF<br>THIS APPLICATION<br>AcuPebble Ox100 by<br>Acurable Ltd | PREDICATE DEVICE<br>AcuPebble SA100,<br>K210480, by Acurable<br>Ltd | REFERENCE DEVICE<br>NightOwl, K191031,<br>by Ectosense | COMPARISON |
|--------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for use | AcuPebble Ox100 is a<br>wearable device<br>intended for use in the<br>recording, analysis,<br>displaying, exporting,<br>and storage of<br>biophysical parameters<br>to aid in the evaluation<br>of adult patients with,<br>or with suspected,<br>obstructive sleep<br>apnea (OSA). The<br>device is primarily<br>intended for home<br>setting use (although it<br>can also be used in<br>healthcare settings)<br>under the direction of<br>a Healthcare<br>Professional (HCP). | AcuPebble SA100 is<br>indicated to sense,<br>record, and interpret a<br>patient's physiological<br>signals (including<br>respiratory pattern)<br>during sleep for the<br>purpose of prescreening<br>patients for obstructive<br>sleep apnea (OSA)<br>syndrome. The device is<br>designed for use in<br>home-screening of adults<br>with suspected possible<br>sleep breathing disorders<br>(although it can also be<br>used in clinic). Results are<br>used to assist the<br>healthcare professional's<br>in the patient's<br>evaluation. The system is<br>not intended as a<br>substitute for full<br>polysomnography when<br>additional parameters<br>such as sleep stages, limb<br>movements, or EEG<br>activity are required. | The NightOwl is a<br>wearable device<br>intended for use in the<br>recording, analysis,<br>displaying, exporting,<br>and storage of<br>biophysical parameters<br>to aid in the evaluation<br>of sleep-related<br>breathing disorders of<br>adult patients suspected<br>of sleep apnea. The<br>device is intended for<br>the clinical and home<br>setting use under the<br>direction of a<br>Healthcare Professional<br>(HCP). | Substantially<br>equivalent to the<br>predicate and<br>identical to the<br>reference device.<br>This does not raise<br>additional issues of<br>safety or efficacy<br>since AcuPebble<br>Ox100 provides the<br>channels of<br>AcuPebble SA100<br>and complements<br>this with a PPG based<br>oximetry channel as<br>per the reference<br>device. In the<br>context of the<br>indications for use,<br>the additional<br>channels provided by<br>AcuPebble Ox100<br>when compared to<br>the reference device:<br>1- Were already<br>approved as part of<br>AcuPebble SA100; 2-<br>Complement<br>oximetry to aid in the<br>diagnosis of OSA |
| Regulation<br>number | 868.2375 | 868.2375 | 868.2375 | Same |
| Product Code | MNR | MNR | MNR | Same |
| Generic<br>Device Name | Ventilatory Effort<br>Recorder | Ventilatory Effort<br>Recorder | Ventilatory Effort<br>Recorder | Same |
| Prescription | Prescription | Prescription | Prescription | Same |
| Target | Adults | Adults | 22 years old and older | Same |
| population | | | | |
| Intended<br>environment<br>for use | Primarily home<br>environment, although<br>it can be used in<br>healthcare<br>environments | Primarily home<br>environment, although it<br>can be used in<br>healthcare<br>environments | Clinical and Home<br>environment | Same |
| Sensor<br>placement | Front of the neck and<br>finger | Front of the neck | Finger | Substantially<br>equivalent. The<br>sensor on the neck<br>allows sensing of<br>physiological signals<br>directly generated by<br>the cardiorespiratory<br>system, in the same<br>way as the predicate.<br>This is in addition to<br>the surrogate signals<br>(blood related<br>features) that are<br>sensed on the finger<br>as in the reference<br>device. |
| Physical<br>sensing<br>elements | Piezoelectric<br>(microphones)/<br>LED+photodiode | Piezoelectric<br>(microphones) | Piezoelectric<br>(accelerometer)/LED+photodiode | Substantially<br>equivalent. The<br>optical sensors do not<br>interfere with the<br>other ones and are<br>used to extract SpO2<br>as in the reference<br>device. |
| Patient<br>contact | Yes | Yes | Yes | Same |
| Means of<br>attachment | Adhesive (patient self-<br>applied) | Adhesive (patient self-<br>applied) | Adhesive (patient self-<br>applied) | Substantially<br>equivalent. They both<br>use adhesives and<br>they are both self-<br>applied |
| Sleep night<br>use | Multiple nights possible | Multiple nights possible | Multiple nights possible | Same |
| Portability | Yes (Wearable) | Yes (wearable) | Yes (wearable) | Same |
| Size and<br>weight | 29.5mm diameter x<br>16mm height, 7g weight<br>for neck sensor. 38mm x<br>30mm x38mm, 15g<br>weight for finger<br>sensor. | 29.5mm diameter x<br>16mm height, 7g weight<br>for neck sensor. 38mm x<br>30mm x38mm | 19x28x11mm, 6g | Substantially<br>equivalent. The<br>sensor on the neck is<br>identical to the one in<br>the predicate. The<br>finger sensor is not in<br>the predicate but the<br>reference device uses<br>a sensor on the finger |
| | | | | which is comparable<br>to the one in this<br>device in size and<br>weight. |
| Recording<br>device | Mobile device records<br>signals streamed from<br>the sensors and uploads<br>them into a cloud-based<br>server | Mobile device records<br>signals streamed from<br>the sensor and uploads<br>them into a cloud-based<br>server | Mobile device records<br>signals streamed from<br>the sensor and uploads<br>them into a cloud-<br>based server | Same |
| Output<br>Channels | SPO2, PPG, Pulse rate<br>(x2 from PPG and from<br>sounds), activity<br>(movement), heart<br>sounds, respiration<br>inhalation and<br>exhalation patterns,<br>airflow (Acoustic<br>derived), snoring | Pulse rate (from<br>sounds), activity<br>(movement), heart<br>sounds, respiration<br>inhalation and exhalation<br>patterns, airflow<br>(Acoustic derived),<br>snoring | SPO2, PPG, Pulse rate,<br>activity (movement) | Substantially<br>equivalent. The<br>additional channels of<br>AcuPebble Ox100<br>provide extra<br>information that<br>serves to further<br>aid/support in the<br>interpretation and<br>subsequent<br>intervention/manage<br>ment of the disease<br>by a healthcare<br>professional. These<br>channels already<br>appear in other<br> |
| Physical<br>sensing<br>elements | Piezoelectric<br>(microphones)/<br>LED+photodiode | Piezoelectric<br>(microphones) | Piezoelectric<br>(accelerometer)/<br>LED+photodiode | Substantially<br>equivalent. The<br>optical sensors do not<br>interfere with the<br>other ones and are<br>used in the same way<br>as in the predicate. |
| Types of<br>events used<br>for OSA<br>indexes | Apneas/hypopneas | Apneas/hypopneas | Apneas/hypopneas | Same |
| Calculated<br>OSA AASM<br>recommended<br>indexes | AcuPebble_AHI<br>AcuPebble_ODI | AcuPebble_AHI<br>AcuPebble_ODI | NightOwl's AHI | Same as the<br>predicate (the<br>indexes are obtained<br>from acoustic derived<br>physiological<br>biomarkers).<br>In the reference<br>device they are<br>calculated from<br>optically obtained<br>biomarkers (this<br>includes SpO2<br>obtained from the<br>PPG signal. |
| | | | | |
| Diagnostic<br>LR+ | 29.36 /32.29 | 29.36 /32.29 | 5.10 | Same |
| Diagnostic<br>LR- | 0.08 /0.04<br>(CI- 0.03 to 0.19/0.01 to<br>0.16) | 0.08 /0.04<br>(CI- 0.03 to 0.19/0.01 to<br>0.16) | 0.04<br>(CI- 0.01 to 0.17) | Same |
| Diagnostic<br>Sensitivity | 92.73% /95.92%<br>(CI-82.41% to 97.98%<br>/86.02% to 99.50%) | 92.73% /95.92%<br>(CI- 82.41% to 97.98%<br>/86.02% to 99.50%) | 96.55% (CI- 88.09% to<br>99.58%) | Same |
| Diagnostic<br>Specificity | 96.84% /97.03% | 96.84% /97.03% | 70.4% to 81.08% | Same |
| Pulse rate | 50-120bm (3.62bpm<br>validation accuracy)<br>when obtained from<br>heart sounds<br>PR from PPG meets the<br>requirement of ±2bpm<br>or ±2%. | 50-120bm (3.62bpm<br>validation accuracy)<br>when obtained from<br>heart sounds | 50-118bpm (2.26bpm<br>validation accuracy) | Same |
| SpO2<br>performance | Compliant with ISO<br>80601-2-61<br>80%-100%, 2% root-<br>mean-square<br>difference<br>70%-79%, 3% root-<br>mean-square<br>difference | Not applicable | Compliant with ISO<br>80601-2-61, i.e. SpO2<br>accuracy with a root-<br>mean-square difference<br>of less or equal to 4%<br>over 70-100% SaO2. | Same |
| Data Transfer | From the sensor to the<br>mobile phone using<br>Bluetooth and from the<br>phone to the cloud<br>(server) through a<br>smartphone by wireless<br>connection.…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.