K203343 · Wesper, Inc. · MNR · Dec 21, 2021 · Anesthesiology
Device Facts
Record ID
K203343
Device Name
Wesper Lab
Applicant
Wesper, Inc.
Product Code
MNR · Anesthesiology
Decision Date
Dec 21, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.2375
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The Wesper Lab is a digital recording device designed to be used under the direction of a physician or trained technician but may be applied by a layperson. Wesper Lab records multiple physiological parameters from a sleeping patient for the purpose of simultaneous or subsequent display of the parameters. The displayed data assists in the identification of sleep apnea by trained personnel. Wesper Lab is intended to be used for adult sleep studies at home or clinical environment. The body-worn component of the system is single-use, to be discarded nightly application. The device does not monitor or diagnose the patient and does not issue any alarms.
Device Story
Wesper Lab is a wearable, single-use digital recording system for adult sleep studies. It consists of two flexible, adhesive patches (thoracic and abdominal) and a mobile application. Patches use optical sensors to measure respiratory effort and solid-state accelerometers for body position. Data is transmitted via Bluetooth Low Energy (BLE) to a patient's mobile app, which also interfaces with an FDA-cleared pulse oximeter for SpO2 and pulse rate. The app uploads data to secure cloud storage. Healthcare professionals download data to a PC, execute the Study Output Module (SOM), and interpret the results to assist in identifying sleep apnea. The device is intended for use under physician direction; it does not provide real-time monitoring, diagnosis, or alarms. It benefits patients by enabling sleep studies in home environments.
Clinical Evidence
Prospective, multi-center clinical study (n=45) comparing Wesper Lab to in-lab polysomnography (PSG). Patients aged 22-76, diverse BMI (19.6-56.1) and skin tones. Primary endpoint: correlation between Wesper Lab and PSG AHI. Results showed 95.1% correlation (one-sided lower CI 91%). Slope and intercept not statistically different from 1 and 0. No adverse events reported.
Technological Characteristics
Wearable, flexible, single-use adhesive patches. Sensing: optical sensors for respiratory effort, solid-state accelerometer for body position. Connectivity: BLE to mobile app, cloud-based storage. Power: internal Li-ion battery. Biocompatibility: ISO 10993 compliant. Electrical safety: IEC 60601, IEC 60529 (IP Code). Software: moderate level of concern.
Indications for Use
Indicated for adult patients undergoing sleep studies in home or clinical environments to record physiological parameters (sleep position, respiratory effort, flow/pressure) to assist trained personnel in identifying sleep apnea. No contraindications specified.
Regulatory Classification
Identification
A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.
Dymedix Reusable Respiratory Effort Belt Model 6015 (K040605)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.
December 21, 2021
Wesper Inc. Amir Reuveny CEO 234 5th Avenue New York, New York 10001
Re: K203343
Trade/Device Name: Wesper Lab Regulation Number: 21 CFR 868.2375 Regulation Name: Breathing Frequency Monitor Regulatory Class: Class II Product Code: MNR Dated: November 19, 2021 Received: November 23, 2021
Dear Amir Reuveny:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Rachana Visaria, Ph.D. Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
Device Name Wesper Lab
#### Indications for Use (Describe)
The Wesper Lab is a digital recording device designed to be used under the direction of a physician or trained technician but may be applied by a layperson. Wesper Lab records multiple physiological parameters from a sleeping patient for the purpose of simultaneous or subsequent display of the parameters. The displayed data assists in the identification of sleep apnea by trained personnel. Wesper Lab is intended to be used for adult sleep studies at home or clinical environment. The body-worn component of the system is single-use, to be discarded nightly application. The device does not monitor or diagnose the patient and does not issue any alarms.
Type of Use (Select one or both, as applicable)
| <div> <span style="font-size:10pt"> <span style="font-family:Wingdings">✓</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span style="font-size:10pt"> <span style="font-family:Wingdings">□</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
### 1. SUBMITTER
| Company & Address: | Wesper Inc.<br>234 5th Avenue<br>New York, NY 10001<br>516-654-4166 |
|--------------------|---------------------------------------------------------------------|
| Contact Person: | Amir Reuveny |
| Date Prepared: | November 11, 2020 |
#### 2. DEVICE
| Name of Device: | Wesper Lab |
|-----------------------|------------------------------|
| Common or Usual Name: | Breathing frequency monitor. |
| Classification Name: | Breathing frequency monitor. |
| Regulation: | 21 CFR 868.2375 |
| Regulatory Class: | II |
| Product Code: | MNR |
#### 3. PREDICATE DEVICE
- Natus Medical Embletta MPR PG (K122516) (Predicate Device) ●
- Compumedics Summit IP (K040194) (Reference Device)
- Dymedix Reusable Respiratory Effort Belt Model 6015 (K040605) (Reference Device) .
Neither the predicate nor reference devices have been subject to a design-related recall.
#### 4. DEVICE DESCRIPTION
Wesper Lab ("the device") is indicated as a digital recording device designed to be used under the direction of a physician or trained technician but may be applied by a layperson. Wesper Lab records multiple physiological parameters from a sleeping patient for the purpose of simultaneous or subsequent display of the parameters. The device consists of an abdominal patch, a thoracic patch and a mobile application. The patches are single-use wearable, flexible, thin, and wireless, and are designed to record sleep data in adult patients. Both patches are identical and differ only in their anatomical designation at test setup time. The flexible fabric allows the patch to retract and expand as the patient breathes and moves during sleep.
The mobile application ("the app") resides on the patient's personal mobile device, relaying sleep data wirelessly to a secure remote storage location ("the cloud") for subsequent analysis by a healthcare professional.
The patches collect multiple physiological parameters related to sleep to be used by a healthcare professional. Specifically, the patches measure sleep position, respiratory effort, and Wesper-Sum- Flow / Wesper-Sum-Pressure.
{4}------------------------------------------------
Data from the patches is transmitted via Bluetooth low energy (BLE) throughout the night to the app, which uploads the data to the cloud. A third BLE port on the app connects to an FDA cleared pulse oximeter, which provides pulse rate and blood oxygen saturation measurements.
The data recorded by the patches is relayed to a remote secure storage, where it will be downloaded to a local PC. Then, Wesper staff will execute the Study Output Module (SOM) and save the data locally on the desktop computer. The data is then ready for interpretation by a healthcare provider.
## 5. INDICATIONS FOR USE
The Wesper Lab is a digital recording device designed to be used under the direction of a physician or trained technician but may be applied by a layperson. Wesper Lab records multiple physiological parameters from a sleeping patient for the purpose of simultaneous or subsequent display of the parameters. The displayed data assists in the identification of sleep apnea by trained personnel. Wesper Lab is intended to be used for adult sleep studies at home or clinical environment. The body-worn component of the system is single-use, to be discarded after its dedicated nightly application. The device does not monitor or diagnose the patient and does not issue any alarms.
#### 6. COMPARISON OF INTENDED USE
Both the Wesper Lab and the predicate device have the same intended use, specifically, for the recording of physiological parameters from a sleeping patient.
## 7. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Recording of multiple physiological parameters from a sleeping patient for the purpose of simultaneous or subsequent display of the parameters is the technical principle for both the subject and predicate devices. The displayed data assists in the identification of sleep apnea by trained personnel. The subject and the predicate device (K122516) are based on the following similar technological elements:
- Input from respiratory effort sensors
- . Coupled with an FDA- cleared pulse oximeter
- o Currently compatible with the Nonin WristOx Model 3150 BLE
- . Body position output
- Data storage
- . Study access via file
The following technological differences exist between the subject and predicate devices:
- Both devices are placed around or on the torso area and measure respiratory effort. The predicate device records respiratory effort measurements from compatible flexible belts using Respiratory Inductance Plethysmography (RIP) or Polyvinylidene Fluoride (PVDF) sensors, while Wesper Lab records respiratory effort measurements from flexible adhesive patches.
{5}------------------------------------------------
- . The Wesper Lab and predicate Embletta MPR PG are similar as they measure the flow of air and its pressure. However, they differ in that the Wesper Lab does not produce a direct measurement of airflow using a nasal cannula to sense physiological flow of air as the Embletta MPR PG does.
- . The reference devices demonstrate that there are other legally marketed Class II devices that utilize an indirect assessment of airflow (K040194) and a localized measurement of respiratory effort (K040605) and that these differences between the subject and the predicate device do not raise new questions of safety and effectiveness.
- . Both the subject and predicate devices record data on the device and then transmit the data to another location. The subject device transmits the data wirelessly to a mobile phone and cloud storage, whereas the predicate device uses wireless transmission to a proprietary relay.
| Device & Predicate Device(s): | Wesper Lab<br>(K203343) | Embletta MPR<br>PG (K122516) | Similarities<br>and<br>Differences | |
|-------------------------------|--------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------|------|
| Classification regulation | 21 CFR<br>868.2375 | 21 CFR<br>868.2375 | Same | |
| Product code | MNR | MNR | Same | |
| Intended Use | For the<br>recording of<br>physiological<br>parameters from<br>a sleeping<br>patient. | For the<br>recording of<br>physiological<br>parameters<br>from a sleeping<br>patient. | Same | |
| Intended environment | Home or clinical<br>environment | Home or<br>clinical<br>environment | Same | |
| Patient population | Adults | Adults and<br>pediatric | Similar;<br>substantially<br>equivalent | |
| Device Type | Wearable<br>Sensor | Wearable<br>Sensor | Same | |
| Main anatomical Site | Thorax,<br>Abdomen,<br>Finger | Head, Thorax,<br>Abdomen,<br>Finger | Similar;<br>substantially<br>equivalent | |
| Airflow | Indirect<br>measurement<br>using<br>respiratory<br>effort signals. | Direct<br>measurement<br>using<br>pressure<br>transducers and<br>thermistors. | Similar;<br>substantially<br>equivalent | |
| Respiratory Efforts | Thoracic and<br>abdominal<br>effort based on | Thoracic and<br>abdominal<br>effort based on | Similar;<br>substantially<br>equivalent | |
| | optical<br>sensor. | RIP / PVDF<br>sensors. | | |
| Body Position | Solid State<br>Accelerometer | Solid State<br>Accelerometer | Same | |
| Display type | Visual display<br>including LEDs<br>and device<br>specific visual<br>indicators. | Visual display<br>including<br>LEDs and<br>device specific<br>visual<br>indicators. | Similar;<br>substantially<br>equivalent | |
| Power source | Internally<br>powered using<br>li-ion<br>battery | 2 x AA<br>batteries,<br>standard<br>alkaline or<br>rechargeable | Similar;<br>substantially<br>equivalent | |
| Data storage | Internal memory<br>and secured<br>cloud storage | Internal<br>memory and<br>secured cloud<br>storage | Same | |
| Communication interface | Bluetooth low<br>energy (BLE) | Bluetooth low<br>energy (BLE)<br>and USB | Similar;<br>substantially<br>equivalent | |
| Access to recorded data | Output file<br>downloaded to a<br>dedicated PC. | Output file<br>downloaded to<br>a dedicated PC. | Same | |
| Device dimensions | 3.8 L x 2.1 W x<br>0.23 D | 3.9 L x 2.8 W<br>x 0.4 D | Similar;<br>substantially<br>equivalent | |
| Biocompatibility | Biocompatible<br>in accordance<br>with ISO 10993 | Biocompatible<br>in accordance<br>with ISO<br>10993 | Same | |
| Signal Recorded Channels | 1. Thoracic<br>effort<br>2. Abdominal<br>effort<br>3. Body position<br>4. Airflow<br>5. Pressure | 1. Gravity<br>2. Thoracic<br>effort<br>3. Abdominal<br>effort<br>4. Body<br>position<br>5. Airflow<br>6. Pressure<br>7. Sound<br>8. Bipolar ExG | Similar;<br>substantially<br>equivalent | |
| Pulse oximeter | Coupled with an<br>authorized<br>FDA-cleared<br>pulse oximeter | Coupled with<br>an authorized<br>FDA-cleared<br>pulse oximeter | Same | |
| | | to<br>measure pulse-<br>rate and SpO2. | to<br>measure pulse-<br>rate and SpO2. | |
| | Sterility | Non-sterile | Non-sterile | Same |
| | Electrical safety | IEC 60601 | IEC 60601 | Same |
Below is a table containing a comprehensive comparison of the subject and predicate devices:
{6}------------------------------------------------
{7}------------------------------------------------
#### 8. PERFORMANCE DATA
#### 8.1 Transit Testing
To show the packaging solution's safety in transit, Wesper tested it against ISTA 3A -Packaged- Products for Parcel Delivery System Shipments 70kg (150 lb) or Less (standard, small, flat or elongated). This procedure simulates the physical process of shipping the product through normal small parcel distribution channels, by running it through a series of steps, such as atmospheric conditioning, vibration, etc. Wesper Lab met all acceptance criteria.
#### 8.2. Biocompatibility Testinq
The biocompatibility evaluation for the Wesper Lab was conducted in accordance with the FDA Guidance "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process," September 4, 2020, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. The Wesper Lab is considered tissue contacting for less than 24 hours. The patient contacting materials of the Wesper Lab are biocompatible.
#### 8.3 Software
Software development and testing has been performed for Wesper Lab in accordance with IEC 62304:2006+AMD1:2015 CSV, Medical Device Software - Software Life-Cycle Processes, and 'General Principles of Software Validation - Final Guidance for Industry and FDA Staff'. Additionally, cyber security risks have been assessed in accordance with the FDA Guidance 'Content of Premarket Submissions for Management of Cybersecurity in Medical Devices'. The software for this device is considered as a "moderate" level of concern, a malfunction or a latent design flaw in the software device could lead to an erroneous diagnosis or a delay in delivery of appropriate medical care that would likely lead to minor injury.
#### 8.4. Electrical Safety / EMC
The Wesper Lab was tested in accordance with IEC 60601 - Medical Electrical Equipment and IEC 60529 ED. 2.2 B:2013 - Degrees of protection provided by enclosures (IP Code). Wesper Lab met all acceptance criteria.
#### 8.5. Bench Testing
Wesper's bench testing demonstrated substantial equivalence to the predicate for body position measurement. Acceptance criteria required that the devices agree to within a 45° interval for the entire range of degrees tested. The Wesper Lab met the acceptance criteria. Wesper conducted a bench test with a simulated breathing apparatus, comparing its respiratory effort sensors with the predicate's RIP technology over a range of breath frequencies amplitudes, and
{8}------------------------------------------------
perturbations (ex. BMI, body hair). The test showed Wesper Lab met or exceeded the predicate's performance in detecting clinically significant breathing events. Wesper conducted integration tests, to show the integrity of its interface with the Nonin WristOx.
#### 8.4. Clinical Testing
Wesper has conducted a prospective, multi-center clinical study for Wesper Lab to demonstrate equivalence to the gold standard - in-lab, polysomnographic (PSG) tests. Wesper Lab was applied to 45 patients undergoing an attended PSG study, with both systems recording their respective signals simultaneously from the patient. Study results showed a correlation between Wesper Lab and PSG AHI of 95.1%, with a onesided lower confidence interval of 91%. The slope and intercept were not statistically different from 1 and 0, respectively. No Adverse Events or Serious Adverse Events were reported. The study had a wide, well-distributed range of age (22-76 years), race (white, black and others), BMI (19.6-56.1) and skin tone (Fitzpatrick skin tone type II-VI), as well as a balanced mix of gender (60% male).
#### 9. CONCLUSION
The Wesper Lab and Embletta MPR PG have the same intended use and similar indications, technological characteristics and principles of operation. The technological differences between the Wesper Lab and the predicate device, as described above, do not present different questions of safety or effectiveness. Where differences exist, reference devices have been utilized that further demonstrate that there are no different questions of safety and effectiveness for the Wesper Lab.
The data provided in this 510(k) Premarket Notification support substantial equivalence of the Wesper Lab and Embletta MPR PG.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.