K122516 · Embla Systems · MNR · Feb 21, 2013 · Anesthesiology
Device Facts
Record ID
K122516
Device Name
EMBLETTA MPR
Applicant
Embla Systems
Product Code
MNR · Anesthesiology
Decision Date
Feb 21, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.2375
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Embletta MPR (Multiple Parameter Recorder) Sleep Data Recording System is a digital recording system designed to be used under the direction of a physician or trained technician or applied by a layperson under the direction of a physician or trained technician. The system records multiple physiological parameters for the purpose of simultaneous (on-line recording) or subsequent (ambulatory recording) uploading to a separate PC based data presentation software for graphical and numerical representation to allow trained personnel to identify sleep disorders. The data collected by this recording system will provide physicians with information to make a diagnosis of sleep disorders such as: Obstructive Sleep Apnea Syndrome, Central Alveolar Hypoventilation Syndrome, Central Sleep Apnea Syndrome, Restless Legs Syndrome, Periodic Limb Movement Disorder, Primary Snoring, Sleep-Related Neurogenic Tachypnea, Cheyne-Stokes-Breathing.
Device Story
Ambulatory sleep recording system; records physiological signals (e.g., pressure, oximetry, audio, EKG, EEG, EMG, respiration, thermistor) via sensors/electrodes. Data uploaded to PC-based software for graphical/numerical display. Used in hospitals, clinics, or home; operated by physicians, technicians, or patients. Embletta MPR (standalone) or Embletta MPR 'PG' (with wireless 'TX' unit) configurations. Provides diagnostic information for sleep disorders (e.g., OSA, CSA, RLS). No real-time monitoring or alarms. Benefits include facilitating sleep disorder diagnosis through ambulatory data collection.
Clinical Evidence
Bench testing only. Verification and validation testing performed on production-equivalent systems to confirm engineering specifications and design input requirements. Electrical safety and electromagnetic compatibility testing conducted per IEC 60601-1 and IEC 60601-1-2 standards. No clinical data provided.
Technological Characteristics
Non-metallic, battery-operated (2 AA) ambulatory recorder. Features 6-19 channel capability. Connectivity via USB and wireless communication. Derived channels (Actigraphy, Body Position, Snore) calculated in external software. Complies with IEC 60601-1 (safety) and IEC 60601-1-2 (EMC).
Indications for Use
Indicated for adult and pediatric (excluding neonatal and infant) patients to record physiological parameters during sleep to assist trained personnel in identifying sleep-related disorders. Not for use as a life monitor; no alarms.
Regulatory Classification
Identification
A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for "embla.". The logo consists of the word "embla" in a sans-serif font, with a small circle containing three stylized shapes above the letters. There is a horizontal line below the word "embla".
K1225lb
Global Headquarters 1 Hines Road Suite 202 Kanata, Ontario K2K 3C7 t: 888.662.7632 f: 613-270-0627 www.embla.com
# 510(K) SUMMARY
# FEB 2 1 2013
# SUBMITTER INFORMATION
Embla Systems A. Company Name: B. Company Address: Embla Systems 1 Hines Road, Suite 202 Kanata, Ontario K2K 3C7 Canada
(613) 254-8877 C. Company Phone:
D. Company Facsimile: (613) 270-0627
E. Company Contact:
Jennifer Armstrong Manager, Regulatory Affairs E-mail: Jennifer.armstrong(@embla.com
#### PREPARATION DATE
August 3, 2012 Revised: January 23, 2013
# DEVICE IDENTIFICATION
Embletta MPR Sleep Data Recording System A. Device Trade Name:
Embletta MPR B. Device Common Name:
Breathing frequency monitor C. Classification Name:
21 CFR 868.2375 D. Regulation Number:
E. Product Code: MNR
F. Device Class: Class II
#### PREDICATE DEVICES
A. Trade Name: Embla Compass, 510(k) Number: K041904
B. Trade Name: Embletta Gold, 510(k) Number: K073682
{1}------------------------------------------------
# DEVICE DESCRIPTION
The Embletta MPR Sleep Data Recording System is the third in the Embla series of small ambulatory sleep recorders. The first was the Embla Compass (also called the Embletta, K041904, Product Code MNR) and the second is the Embletta Gold (K073682, Product Code MNR).
The Embletta MPR Sleep Data Recording System records multiple physiological parameters for the purpose of simultaneous or subsequent uploading to a separate PC based data presentation software. The data collected by this recording system will provide physicians with information to make a sleep disorder diagnosis.
The Embletta MPR Sleep Data Recording System consists of three small devices that can be used in two separate configurations. The simplest configuration is the battery operated Embletta MPR recorder which is worn by the patient and used alone for small scale sleep studies (2 input + 3 derived channels). The other configuration is the battery operated Embletta MPR *PG' unit (similar to the Embletta MPR but with 6 input + 3 derived channels) which is worn by the patient and used with wireless communication to a separate mains powered Embletta 'TX' unit located remotely from the patient. The Embletta 'TX' unit can also record inputs from separate patient devices (such as CPAP devices).
Image /page/1/Picture/5 description: The image shows two Embla MPR devices. The device on the left is labeled "Embletta MPR" and the device on the right is labeled "Embletta MPR-TG". Both devices have a screen and three buttons below the screen. The buttons are green, black, and blue.
Embletta MPR
Embletta MPR 'PG'
{2}------------------------------------------------
TX Proxy
Image /page/2/Picture/2 description: The image shows a gray electronic device with a black front panel. The front panel has six DC input ports labeled 1 through 6, and two PDH ports labeled with plus and minus signs. The device has a rounded rectangular shape and appears to be made of plastic.
Each of these two configurations basically consists of a simple signal recording unit that can be connected with several electrodes and sensors non-invasively attached to a patient during sleep. The Embletta MPR and the Embletta MPR 'PG' are secured to the patient with a belt and holder.
### INTENDED USE
The Embletta MPR (Multiple Parameter Recorder) Sleep Data Recording System is a digital recording system designed to be used under the direction of a physician or trained technician or applied by a layperson under the direction of a physician or trained technician.
The system records multiple physiological parameters for the purpose of simultaneous (on-line recording) or subsequent (ambulatory recording) uploading to a separate PC based data presentation software for graphical and numerical representation to allow trained personnel to identify sleep disorders. The data collected by this recording system will provide physicians with information to make a diagnosis of sleep disorders such as:
| Disorder | Compass<br>(Embletta) | Embletta<br>Gold | Embletta<br>MPR |
|---------------------------------------------|-----------------------|------------------|-----------------|
| • Obstructive Sleep Apnea Syndrome | √ | √ | √ |
| • Central Alveolar Hypoventilation Syndrome | √ | √ | √ |
| • Central Sleep Apnea Syndrome | √ | √ | √ |
| • Restless Legs Syndrome | √ | √ | √ |
| • Periodic Limb Movement Disorder | √ | √ | √ |
| • Primary Snoring | √ | √ | √ |
| • Sleep-Related Neurogenic Tachypnea | √ | √ | √ |
| • Cheyne-Stokes-Breathing | √ | √ | √ |
{3}------------------------------------------------
| Recorded channels: | MPR<br>Used alone | MPR 'PG'<br>Used together | TX Proxy |
|-------------------------|-------------------|---------------------------|----------|
| • Pressure | √ | √ | |
| • Oximetry | √ | √ | |
| • Audio | √ | √ | |
| • EKG, EEG, EMG | | √ | |
| • Thermister | | √ | |
| • Respiration, Thorax | | √ | |
| • Respiration, Abdomen | | √ | |
| • DC channel | | 1 | 6 |
| • Therapy devices | | | √ |
| • Differential pressure | | | √ |
| Derived channels: | | | |
| • Snore | √ | √ | |
| • Actigraphy | √ | √ | |
| • Body Position | √ | √ | |
The system has 6 - 19 channel capability for recording signals:
The general intended environments are hospitals, institutions, sleep centers, the home and sleep clinics or the patient's home. The studies will be supervised by trained physicians, trained sleep technicians (RPSGT) or clinicians. The intended environments include any clean, dry, dust free environment suitable for a patient's relative comfort.
The Embletta MPR unit is intended for ambulatory or patient in-home studies while the Embletta MPR PG and TX Proxy units are intended for ambulatory, patient in-home studies and clinic on-line studies.
The Embletta MPR system is intended to be used for adult and pediatric (excluding neonatal and infant) studies.
The recorder does not provide any alarms and is not intended to be a life monitor.
### INDICATIONS FOR USE (Embletta MPR)
The Embletta MPR is a digital recording device designed to be used under the direction of a physician or trained technician but may be applied by a layperson. The Embletta MPR records multiple physiological parameters from a sleeping patient for the purpose of simultaneous or subsequent display of the parameters. The displayed data assists in the identification of sleep-related medical disorders by trained personnel.
The Embletta MPR is intended to be used for adult and pediatric (excluding neonatal and infant) studies. The device is not equipped with alarms and is not intended to be used as a monitor.
{4}------------------------------------------------
The intended environments include any clean, dry, dust free environment suitable for a patient's relative comfort.
The device does not monitor or diagnose the patient and does not issue any alarms.
### INDICATIONS FOR USE [Embla Compass (Embletta) predicate device]
The intended use of the Compass F10 system is to record physiological signals during sleep, scan the signals for abnormalities and represent the count of abnormal events in a form of a summary report. The results of the scan may be manually overwritten or corrected by the physician. The device is intended for use as a screening device to determine the need for clinical diagnosis and evaluation by polysomnography based on the patient's count of abnormal events. It is not intended for any diagnosis. It is not intended to be a monitor.
The Compass F10 device is intended to be used for adult and pediatric patients.
### INDICATIONS FOR USE (Embla Embletta Gold predicate device)
The Embletta Gold is a digital recording device designed to be used under the direction of a physician or trained technician but applied by a lay person. The Embletta Gold records multiple physiological parameters for the purpose of simultaneous or subsequent display of the parameters. The displayed data assists in the identification of sleep-related medical disorders by trained personnel.
The Embletta Gold is intended to be used for adult and pediatric (excluding neonatal and infant) studies. Note the recorder is not equipped with an alarm device and is not intended to be used as a life monitor.
The intended environments include any clean, dry, dust free environment suitable for a patient's relative comfort.
# COMPARISON TO PREDICATE DEVICES
The Embletta MPR Sleep Data Recording System is the third in a series of Embla ambulatory sleep data recorders. It is substantially equivalent in the following technological ways to the intended use and application to the two previous versions, the identified predicate devices;
| Note | Parameter | Embla Compass<br>(Embletta) | Embletta Gold | Embletta MPR |
|------|-------------------------------|-----------------------------------------------------------------------------------------|-----------------------------------------------------------------------|-----------------------------------------------------------------------|
| | 510(k) # | K041904 | K073682 | TBD |
| | Product Code | MNR | MNR | MNR |
| | | | | |
| 1 | Intended use* | Equivalent | Equivalent | Equivalent |
| 1 | Indications for Use | Equivalent | Equivalent | Equivalent |
| 2 | Target Population | Adult and pediatric<br>patient groups | Adult and pediatric<br>patient (excluding<br>neonatal and<br>infants) | Adult and pediatric<br>patient (excluding<br>neonatal and<br>infants) |
| | Where Used | Sleep clinics and<br>home | Sleep clinics and<br>home | Sleep clinics and<br>home |
| | Energy Used | Battery operated (2<br>AA) | Battery operated (2<br>AA) | Battery operated (2<br>AA) |
| | Human factors: | | | |
| 3 | Device setup<br>• | Similar (PC) | Similar (PC) | Similar (PC) |
| 4 | Electrode<br>•<br>application | Identical | Identical | Identical |
| 5 | Design | Signal sensing,<br>conditioning, noise<br>filtering,<br>amplification and<br>recording. | Identical | Identical |
| 6 | Performance: | | | |
| | Recording<br>Time<br>• | 72 hours | 48 hours | 24 hours (200 total<br>hrs) |
| | Data Interface<br>• | USB | USB | USB |
| | # of Channels<br>• | Depends on proxy | 8 + 3 derived | 16 + 3 derived |
| | EKG | √ | √ | √ |
| | EEG | √ | | √ |
| | EMG | √ | | √ |
| | EOG | √ | | |
| | Pulse | √ | | |
| | Respiration<br>(Ab) | √ | √ | √ |
| | Respiration<br>(Th) | √ | √ | √ |
| | Thermistor | | √ | √ |
| | Oximetry | √ | √ | √ |
| | Pressure | √ | √ | √ |
| | Dif. Pressure | | √ | √ |
| 7 | Actigraphy<br>(derived) | √ | √ | √ |
| 7 | Body Position<br>(derived) | √ | √ | √ |
| 7 | Snore<br>(derived) | √ | √ | √ |
| 8 | DC Auxiliary | | 1 | 7 |
| 9 | (C,V) PAP | | | |
{5}------------------------------------------------
. ·
{6}------------------------------------------------
| 10 | Audio | | | |
|----|-------------------|---------------------------|---------------------------|---------------------------|
| | Standards met | Same | Same | Same |
| | Materials | Non-metallic | Non-metallic | Non-metallic |
| | Biocompatibility | N/A, worn over<br>clothes | N/A, worn over<br>clothes | N/A, worn over<br>clothes |
| | Sterility | Not sterile | Not sterile | Not sterile |
| | Electrical Safety | IEC 60601-1 | IEC 60601-1 | IEC 60601-1 |
| | Handheld | Yes | Yes | Yes |
| | Weight | 146g | 190g | 136g |
#### NOTES:
- Indications for use and Intended use The Embla Compass (Embletta) and 1. Embletta Gold predicate devices are small palm-size portable devices that connect to one or more probes or sensors on the patient to record a variety of physiological signals. This data is then downloaded into a separate computer application software that presents the signals in a format that can be read by a trained technician or physician. This is an identical application of the Embletta MPR.
- 2. Target Population The target population for the Embletta MPR is identical to the Embla Compass (Embletta) device: Adult and pediatric patient groups. This is not an expansion of patient groups.
- 3. Device Setup All three devices are setup thru the use of a separate application software in an identical fashion. Channels are configured and identified, sampling rates are set, filtering options defined, recording timing (on/off times), etc.
- 4. Electrode Application All three devices use commercially available electrodes, cannulas, and thermistors. All three use the separate Embla XactTrace belts for thorax and abdomen respiration sensing. All three can use the separate Nonin XPOD oximetry devices.
- 5. Design All three have similar design strategies, different in the level of technology used (with age). All three include basic signal sensing, conditioning, noise filtering, amplification and recording.
- Performance Channel options are listed in the table. Channel specifications are 6. essentially identical to predicate device specifications.
- 7. Derived Channels The Actigraphy and Body Position channels are derived in the separate software application using the basic X, Y and Z position data from the integrated gravity sensors in all three devices. The Snore channel is derived in the separate software application using the pressure channel from the patient mask with frequency filtering in all three devices.
- DC Auxiliary Channels Due to increased user requests for channels to record 8. external +/- 5 Vdc signals, the Embletta MPR has additional (7) DC Auxiliary channels that can be identified and configured by the separate application software (usually flow, pressure, leak information, depending on the therapy systems additionally information about tidal volume, respiratory rate, inspiratory pressure, target pressure).
{7}------------------------------------------------
- 9. (C,V) PAP The Embletta MPR can accept data from therapy devices. This is a new feature for an ambulatory recording device not present in the two predicate devices.
- 10. Audio The Embletta MPR has an audio channel to permit communications from a control room location to the patient or a clinician with the patient. This is a new feature not present in the two predicate devices.
### TESTING AND PERFORMANCE DATA
The design input requirements for the Embletta MPR were created from, and based on predicate device specifications and functionality, adjusted and revised in accordance with new marketing requirements and current state of the art technology. Testing included the following:
- 1. Verification testing in the design stages. This testing consisted of verification of engineering specifications derived from the design input requirements. The document "Verification Protocols" and "Verification Report" have been submitted.
- 2. Validation testing after design completion. This testing consisted of validation of each design input requirement in a production equivalent system setting. The document "Validation Test Plan", "Validation Test Protocols" and "Validation Report" have been submitted.
- 3. Electrical Safety testing Safety tests have been performed to verify compliance with IEC 60601-1 and the applicable particular standards listed below to ensure
- that there are no potential hazards on patients, operators, or the surroundings.
- 4. Electromagnetic Compatibility testing
- a. Electromagnetic Compatibility tests have been performed to verify compliance with IEC 60601-1-2 to ensure that no intolerable electromagnetic disturbances are introduced.
- b. Immunity tests have been performed to verify compliance with IEC 60601-1-2 to ensure that the device operates satisfactorily in an electromagnetic environment.
| Medical Electrical Equipment: General Requirements for<br>Safety. | IEC 60601-1: 2001 2nd ed. |
|---------------------------------------------------------------------------------------------|-----------------------------|
| Electromagnetic Compatibility -Requirements and Tests | IEC 60601-1-2: 2001 2nd ed. |
| Particular Requirements for the Safety of<br>Electrocardiographs | IEC 60601-2-25:1993 |
| Particular Requirements for the Safety of<br>Electroencephalographs | IEC 60601-2-26:1994 |
| Particular Requirements for the Safety of Electromyographs<br>and Evoked Response Equipment | IEC 60601-2-40:1998 |
# CONCLUSION
We believe that the results of the performance testing for the Embletta MPR Sleep Data Recording System, as well as the substantial equivalence comparison to the two predicate devices, are adequate to support a conclusion for the safety and efficacy of the device,
{8}------------------------------------------------
Based on the test results and the substantial equivalence comparison, it is the conclusion of Embla Systems that the Embletta MPR Sleep Data Recording is substantially equivalent to devices already on the market and presents no new concerns about safety and effectiveness.
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of a caduceus, a symbol often associated with healthcare. The caduceus is depicted with a staff entwined by a serpent.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 21, 2013
Ms. Jennifer Armstrong Embla Systems Manager, Regulatory Affairs 1 Hines Road, Suite 202 Kanata, Ontario Canada K2K 3C7
Re: K122516
Trade/Device Name: Embletta MPR Sleep Data Recording System Regulation Number: 21 CFR 868.2375 Regulation Name: Breathing Frequency Monitor Regulatory Class: II Product Code: MNR Dated: January 24, 2013 Received: January 31, 2013
Dear Ms. Armstrong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{10}------------------------------------------------
Page 2 - Ms. Armstrong
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
http://www.fda.gov/MedicalDevices/ResourcestorYourIndustry/default.htm.
Sincerely yours.
Mary S. Runner
Susan Runner, DOSMA 2013.02.21
12:11:51 -05'00'
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation
Center for Devices and Radiological Health
#### Enclosure
{11}------------------------------------------------
#### Traditional 510(k) Notification for the Embletta MPR Sleep Data Recording System
January 23, 2013
#### SECTION 6
#### STATEMENT OF INDICATIONS OF USE
510(k) Number (if known); Unknown
Device Name: Embletta MPR Sleep Data Recording System
Indications For Use:
The Embletta MPR is a digital recording device designed to be used under the direction of a physician or trained technician but may be applied by a layperson. The Embletta MPR records multiple physiological parameters from a sleeping patient for the purpose of simultaneous or subsequent display of the parameters. The displayed data assists in the identification of sleep-related medical disorders by trained personnel.
The Embletta MPR is intended to be used for adult and pediatric (excluding infants and neonatal) studies. The device is not equipped with alarms and is not intended to be used as a monitor.
The intended environments include any clean, dry, dust free environment suitable for a patient's relative comfort.
The device does not monitor or diagnose the patient and does not issue any alarms.
Prescription Use 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C)
(Part
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) Intection Control, Dental Devices K122516 510(k) Number: Page 5 of 29
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.