K151361 · Nox Medical Ehf · MNR · Nov 6, 2015 · Anesthesiology
Device Facts
Record ID
K151361
Device Name
Nox RIP Belts & Cables
Applicant
Nox Medical Ehf
Product Code
MNR · Anesthesiology
Decision Date
Nov 6, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.2375
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Nox RIP Belts are intended for measuring of respiratory effort signals. They function as accessories for sleep/polysomnography (PSG) systems. The Nox RIP Belts are indicated for use on patients greater than 2 years of age. The intended environments are hospitals, institutions, sleep clinics, or other test environments, including the patient's home. The RIP Belt Cables are intended to interconnect Nox RIP belts (respiratory effort sensors) and Nox sleep devices, to allow measuring of respiratory effort signals. The RIP Belt Cables are indicated for use on patients greater than 2 years of age. The intended environments are hospitals, institutions, sleep clinics, or other test environments, including the patient's home. The Third Party RIP Belt Cables are intended to allow measuring of respiratory effort signals by interconnecting Nox RIP belts (respiratory effort sensors) and sleep devices with oscillation circuitry capable of measuring inductance between 1 and 5 µH. The Third Party RIP Belt Cables are indicated for use on patients greater than 2 years of age. The intended environments are hospitals, institutions, sleep clinics, or other test environments, including the patient's home.
Device Story
Nox RIP Belts and Cables are passive accessories for sleep/polysomnography (PSG) systems; they measure respiratory effort via Respiratory Inductance Plethysmography (RIP) technology. Belts are placed around patient thorax and abdomen; cables interconnect these sensors to sleep recorders. Third-party cables interface with recorders having oscillation circuitry (1-5 µH inductance). Used in hospitals, clinics, or homes; operated by healthcare professionals. Devices provide raw respiratory effort signals to the recorder; clinicians use these signals to assess respiratory effort, aiding in sleep disorder diagnosis. Benefits include non-invasive monitoring of respiratory effort during sleep.
Clinical Evidence
Bench testing only. Verification and validation performed per product requirements. Testing included signal integrity (SNR, range, bandwidth, linearity), EMC (IEC 60601-1-2), and electrical safety (AAMI/ANSI/ES 60601-1). Usability testing conducted per AAMI/ANSI/IEC 62366. Signal integrity compared to predicate demonstrated clinical equivalence.
Technological Characteristics
Passive respiratory effort sensors using RIP technology. Materials: Polyester/Dorlastan elastic, tin-plated copper wire, ABS/PC plastic, TPE/TPE-ET strain relief, gold-plated stainless steel snaps. Dimensions/form factor: wearable belts and cables. Connectivity: wired connection to sleep recorders. Standards: AAMI/ANSI/ES 60601-1, IEC 60601-1-2, ISO 14971, ISO 15223-1, AAMI/ANSI/IEC 62366, 21 CFR 898.
Indications for Use
Indicated for patients > 2 years of age requiring respiratory effort signal measurement during sleep studies in clinical or home environments.
Regulatory Classification
Identification
A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol featuring three stylized human profiles facing to the right, stacked on top of each other. The profiles are rendered in a dark color, contrasting with the white background.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 6, 2015
Nox Medical ehf C/O Kolbrun E Ottosdottir Director of Quality & Regulatory Katrinartuni 2 IS-105 Reykjavik ICELAND
Re: K151361
Trade/Device Name: Nox RIP Belts and Nox RIP Belt Cables Regulation Number: 21 CFR 868.2375 Regulation Name: Breathing frequency monitor Regulatory Class: Class II Product Code: MNR Dated: October 2, 2015 Received: October 5, 2015
Dear Mr. Ottosdottir:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
*Tejashri Purohit-Sheth, M.D.*
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indication for Use
510(k) Number (if known): K151361
Device Name: Nox RIP Belts
Indications for Use (Describe):
The Nox RIP Belts are intended for measuring of respiratory effort signals. They function as accessories for sleep/polysomnography (PSG) systems.
The Nox RIP Belts are indicated for use on patients greater than 2 years of age.
The intended environments are hospitals, institutions, sleep centers, sleep clinics, or other test environments, including the patient 's home.
Type of Use ( Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
# Indication for Use
510(k) Number (if known): K151361
Device Name: RIP Belt Cables
Indications for Use (Describe):
The RIP Belt Cables are intended to interconnect Nox RIP belts (respiratory effort sensors) and Nox sleep devices, to allow measuring of respiratory effort signals.
The RIP Belt Cables are indicated for use on patients greater than 2 years of age.
The intended environments are hospitals, institutions, sleep centers, sleep clinics, or other test environments, including the patient´s home.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{4}------------------------------------------------
# Indication for Use
510(k) Number (if known): K151361
Device Name: Third Party RIP Belt Cables
Indications for Use (Describe):
The Third Party RIP Belt Cables are intended to allow measuring of respiratory effort signals by interconnecting Nox RIP belts (respiratory effort sensors) and sleep devices with oscillation circuitry capable of measuring inductance between 1 and 5 µH.
The Third Party RIP Belt Cables are indicated for use on patients greater than 2 years of age.
The intended environments are hospitals, institutions, sleep centers, sleep clinics, or other test environments, including the patient's home.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{5}------------------------------------------------
## 510(k) Summary
### Submitter
Nox Medical ehf Katrinartuni 2 IS-105 Reykjavik Iceland Tel: +354 570 7170
Establishment Registration Number: 3007389703
#### Contact person
Kolbrun E Ottosdottir Katrinartuni 2 IS-105 Reykjavik Iceland Email: keo@noxmedical.com Direct tel: +354 570 7173
## Preparation Date
06 November 2015
## Device Details
| Trade Name: | Nox RIP Belts | Nox RIP Belt Cables |
|--------------------------|--------------------------------|-----------------------------|
| Product Groups | | |
| | RIP Belt Cables | Third Party RIP Belt Cables |
| Common Name: | Respiratory Effort<br>Sensor | Respiratory Effort Sensor |
| Classification<br>Name: | Ventilatory Effort<br>Recorder | Ventilatory Effort Recorder |
| Regulation<br>Number: | 868.2375 | 868.2375 |
| Product Code: | MNR | MNR |
| Device Class | Class II | Class II |
| Classification<br>Panel: | Anesthesiology | Anesthesiology |
{6}------------------------------------------------
#### Predicate Device
QDC-PRO AND NOX-RIP from Nox Medical ehf Product Code: MNR 510(k) Number: K124062
This predicate has not been subject to a design-related recall.
Note1: Please note that the Nox RIP Belts will be regarded as a subsystem of the predicate, QDC-PRO AND NOX-RIP. The QDC-PRO AND NOX-RIP includes a recorder and several accessories but only the RIP belts will be addressed in the comparison between the new device and the predicate.
Note2: Please note that the Nox RIP Belt Cables will be regarded as a subsystem of the predicate, QDC-PRO AND NOX-RIP. The QDC-PRO AND NOX-RIP includes a recorder and several accessories but only the RIP belt cable will be addressed in the comparison between the new device and the predicate.
{7}------------------------------------------------
### Device Description
The Nox RIP Belts are respiratory effort sensors that are intended to function as an accessory with sleep/polysomnography (PSG) systems. The RIP Belts measure respiratory effort signals based on Respiratory Inductance Plethysmography (RIP) technology, which is the gold standard technology for respiratory effort belts.
Two RIP belts are used to measure the respiratory effort of the patient. One belt is placed around the patient's abdomen and the other around the patient's thorax. Both abdomen and thorax belts are identical.
The Nox RIP Belt Cables are used to connect between the respiratory effort sensor (RIP belts) and the applicable sleep recorder/polysomnography (PSG) system.
There are two product groups for the Nox RIP Belt Cables; RIP Belt Cables and Third Party RIP Belt Cables.
The RIP Belt Cables are designed for use with Nox recorders only. Those are abdomen cables only because the thorax belt is attached directly to the Nox recorders.
The Third Party RIP Belt Cables are designed for use with third party recorders. The Third Party RIP Belt Cables come in pairs for abdomen and thorax.
{8}------------------------------------------------
## Intended Use
#### Nox RIP Belts
The Nox RIP Belts are intended for measuring of respiratory effort signals. They function as accessories for sleep/polysomnography (PSG) systems.
The Nox RIP Belts are indicated for use on patients greater than 2 years of age.
The intended environments are hospitals, institutions, sleep clinics, or other test environments, including the patient's home.
#### RIP Belt Cables
The RIP Belt Cables are intended to interconnect Nox RIP belts (respiratory effort sensors) and Nox sleep devices, to allow measuring of respiratory effort signals.
The RIP Belt Cables are indicated for use on patients greater than 2 years of age.
The intended environments are hospitals, institutions, sleep clinics, or other test environments, including the patient's home.
#### Third Party RIP Belt Cables
The Third Party RIP Belt Cables are intended to allow measuring of respiratory effort signals by interconnecting Nox RIP belts (respiratory effort sensors) and sleep devices with oscillation circuitry capable of measuring inductance between 1 and 5 µH.
The Third Party RIP Belt Cables are indicated for use on patients greater than 2 years of age.
The intended environments are hospitals, institutions, sleep clinics, or other test environments, including the patient's home.
## Technological Characteristics and Comparison with Predicate
#### Nox RIP Belts
The comparison table below is provided as a summary of the intended use and most relevant characteristics of the Nox RIP Belts relative to the predicate device.
Please note that the Nox RIP Belts will be regarded as a subsystem of the predicate. ODC-PRO AND NOX-RIP. The ODC-PRO AND NOX-RIP includes a recorder and several accessories but only the RIP belts will be addressed in the comparison between the new device and the predicate.
{9}------------------------------------------------
| Characteristic | QDC - PRO AND NOX-RIP (K124062) | Nox RIP Belts | Comparison Result |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|
| General | | | |
| Intended Use | The QDC-PRO device is a sensor unit intended for measuring of physiological signals during sleep. The signals measured are processed in the QDC-PRO device and the resulting signals made available at the output connector for acquisition by a third party Polysomnography (PSG)/sleep recorder. | The Nox RIP Belts are intended for measuring of respiratory effort signals. They function as accessories for sleep/polysomnography (PSG) systems. | Equivalent<br>See Note 1 below |
| Intended Environments | The intended environments are hospitals, institutions, sleep centers, sleep clinics, or other test environments. | The intended environments are hospitals, institutions, sleep centers, sleep clinics, or other test environments, including the patient´s home. | Equivalent<br>See Note 2 below |
| Patient Population | The QDC-PRO AND NOX-RIP are indicated for use in patients greater than 2 years of age. | The Nox RIP Belts are indicated for use on patients greater than 2 years of age. | Identical |
| Prescription Use | Yes | Yes | Identical |
| Type of Modules | RIP belts are disposable | RIP belts are disposable | Identical |
| Various Belt Sizes | Yes | Yes | Identical |
| Physical/Material | | | |
| RIP Belts Material | Elastic: Polyester/Dorlastan (not made with natural rubber latex), tin plated copper wire<br>Connector: ABS plastic | Elastic: Polyester/Dorlastan (not made with natural rubber latex), tin plated copper wire<br>Connector: ABS plastic | Identical |
| Disposable Components | Disposable respiratory effort sensor/RIP belts. | Disposable respiratory effort sensor/RIP belts. | Identical |
| Functional | | | |
| Connection to Patient | RIP Belts are attached to patient | RIP Belts are attached to patient | Identical |
| Type of Equipment to Be Connected to | Sleep recorder and RIP Belt Cable | Sleep recorder and RIP Belt Cable | Identical |
| Connector Type | RIP belts have connector for to connect to Nox proprietary RIP snaps | RIP belts have connector for to connect to Nox proprietary RIP snaps | Identical |
| Signals and Sensors | | | |
| Signals Measured | Respiratory Effort (Abdomen and Thorax) | Respiratory Effort (Abdomen and<br>Thorax) | Identical |
| Respiratory Effort<br>Technology | RIP (Respiratory Inductive Plethysmography)<br>technology | RIP (Respiratory Inductive<br>Plethysmography) technology | Identical |
| Signal Processing | No signal processing is done in RIP Belts | No signal processing is done in<br>RIP Belts | Identical |
| Power and Isolation | | | |
| Power Source | No power source. RIP Belts contain only<br>passive components | No power source. RIP Belts<br>contain only passive components | Identical |
| Isolation | No component included in the RIP Belts that<br>are relied on as means of protection | No component included in the<br>RIP Belts that are relied on as<br>means of protection | Identical |
| Standards | | | |
| AAMI/ANSI/ES<br>60601-1 | Yes | Yes | Identical |
| IEC 60601-1-2 | Yes | Yes | Identical |
Table 1 Comparison Summary of the New Device (Nox RIP Belts) and Predicate Device
{10}------------------------------------------------
Note 1: Nox RIP Belts are intended for measuring of respiratory effort signals and function as accessories for sleep systems. The predicate, ODC-PRO AND NOX RIP, is intended for measuring of physiological signals during sleep. Respiratory effort is one example of physiological signals and both the new device and the predicate measure the signals during sleep. Furthermore, the predicate RIP belts are intended for measuring respiratory effort signals as an accessory to the ODC-PRO recorder. Therefore the indication for use can be reqarded the same.
Note 2: The intended environment depends on the classification of the sleep recorder to be used with the Nox RIP Belts. The predicate device, ODC-PRO AND NOX-RIP, is intended to be used with a third party polysomnography (PSG) recorder. PSG studies are rarely performed in home environment and therefore it is not listed in the intended environments for the predicate. Simpler sleep recorders like polygraph (PG) recorders are commonly used at patients' home. The Nox RIP Belts may be used with any type of sleep recorders/PSG recorders including PG recorders. Therefore, the home environment was added for the Nox RIP Belts. If a sleep recorder is intended for home use, the type of RIP Belts does not change or influence the intended environment of the recorder. The home inclusion to the environment of use does therefore not pose any new safety and effectiveness concerns for the Nox RIP Belts. Additionally, risk analysis, verification and validation results demonstrate that the use of Nox RIP Belts in the home environment does not raise any new concerns regarding the intended use, safety or efficiency. Furthermore, these are prescription devices and are always used under supervision of a licensed medical practitioner. The intended environments are therefore deemed equivalent.
The result of the above comparison is that the indication for use, patient population and intended environments for the Nox RIP Belts can be regarded the same as that of the predicate QDC-PRO AND NOX-RIP device.
The comparison above demonstrates that all major technological characteristics, including features, materials, and principles of operation of the Nox RIP Belts are identical to the predicate QDC-PRO AND NOX RIP.
{11}------------------------------------------------
#### Nox RIP Belt Cables
The comparison table below is provided as a summary of the intended use and most relevant characteristics of the Nox RIP Belt Cables relative to the predicate device.
Please note that the Nox RIP Belt Cables will be regarded as a subsystem of the predicate, QDC-PRO AND NOX-RIP. The QDC-PRO AND NOX-RIP includes a recorder and several accessories but only the RIP belt cable will be addressed in the comparison between the new device and the predicate.
{12}------------------------------------------------
| Characteristic | QDC - PRO AND NOX-<br>RIP (K124062) | Nox RIP Belt Cables | | |
|-----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|
| | | RIP Belt Cables | Third Party RIP Belt<br>Cables | Comparison Results |
| | | General | | |
| Intended Use | The QDC-PRO device is<br>a sensor unit intended<br>for measuring of<br>physiological signals<br>during sleep. The signals<br>measured are processed<br>in the QDC-PRO device<br>and the resulting signals<br>made available at the<br>output connector for<br>acquisition by a third<br>party Polysomnography<br>(PSG)/sleep recorder. | The RIP Belt Cables are<br>intended to interconnect<br>Nox RIP belts<br>(respiratory effort<br>sensors) and Nox sleep<br>devices, to allow<br>measuring of respiratory<br>effort signals. | The Third Party RIP Belt<br>Cables are intended to<br>allow measuring of<br>respiratory effort signals<br>by interconnecting Nox<br>RIP belts (respiratory<br>effort sensors) and sleep<br>devices with oscillation<br>circuitry capable of<br>measuring inductance<br>between 1 and 5 μH. | Equivalent<br>See Note 3 below |
| Intended<br>Environments | The intended<br>environments are<br>hospitals, institutions,<br>sleep centers, sleep<br>clinics, or other test<br>environments. | The intended<br>environments are<br>hospitals, institutions,<br>sleep centers, sleep<br>clinics, or other test<br>environments, including<br>the patient's home. | The intended<br>environments are<br>hospitals, institutions,<br>sleep centers, sleep clinics,<br>or other test<br>environments, including<br>the patient's home. | Equivalent<br>See Note 4 below |
| Patient<br>Population | The QDC-PRO AND<br>NOX-RIP are indicated<br>for use in patients<br>greater than 2 years of<br>age. | The RIP Belt Cables are<br>indicated for use on<br>patients greater than 2<br>years of age. | The Third Party RIP Belt<br>Cables are indicated for<br>use on patients greater<br>than 2 years of age. | Identical |
| Prescription<br>Use | Yes | Yes | Yes | Identical |
| Type of Modules | RIP Belt Cable for<br>connection to recorder<br>and RIP belts | RIP Belt Cable for<br>connection to recorder<br>and RIP belts | RIP Belt Cables for<br>connection to recorder and<br>RIP belts | Identical |
| Physical / Material | | | | |
| RIP Cables<br>Material | ● PVC – wire jacket<br>● PC with 10% glass<br>fiber – belt end<br>● PC with 10% glass<br>fiber – device end<br>● Gold plated<br>stainless steel -<br>snaps<br>● TPE - strain relief | ● PVC – wire jacket<br>● ABS/PC - belt end<br>● ABS/PC - device<br>end<br>● Gold plated<br>stainless steel -<br>snaps<br>● TPE - strain relief | ● PVC – wire jacket<br>● ABS/PC - belts end<br>● TPE-ET - device end<br>strain relief<br>● Gold plated stainless<br>steel - snaps<br>● TPE-ET - strain relief<br>Note: TPE-ET is also<br>known as Riteflex® | Equivalent<br>See Note 5 below |
| Disposable | RIP Belt Cable is | RIP Belt Cable is | Third Party RIP Belt Cables | Identical |
| Components | reusable | reusable | are reusable | |
| Functional | | | | |
| Connection to<br>Patient | RIP Belt Cable is<br>attached to patient via<br>connection to RIP belts | RIP Belt Cable is<br>attached to patient via<br>connection to RIP belts | Third Party RIP Belt Cables<br>are attached to patient via<br>connection to RIP belts | Identical |
| Type of<br>Equipment to<br>Be Connected<br>to | Interconnect sleep<br>recorder and respiratory<br>effort sensor (RIP belt) | Interconnect sleep<br>recorder and respiratory<br>effort sensor (RIP belt) | Interconnect sleep<br>recorder and respiratory<br>effort sensor (RIP belt) | Identical |
| Belt Cable<br>Connectors -<br>Recorder End | Proprietary connector for<br>Nox RIP snaps | Proprietary connector for<br>Nox RIP snaps | Touch proof 1 mm 2 pin<br>female connector | Equivalent<br>See Note 6 below |
| Belt Cable<br>Connectors -<br>Belt End | Gold coated proprietary<br>metal snaps | Gold coated proprietary<br>metal snaps | Gold coated proprietary<br>metal snaps | Identical |
| Signals and Sensors | | | | |
| Signal<br>Processing | No signal processing<br>done in RIP Belt Cable | No signal processing<br>done in RIP Belt Cable | No signal processing done<br>in Third Party RIP Belt<br>Cables | Identical |
| Power and Isolation | | | | |
| Power Source | No power source. RIP<br>Belt Cable contains only<br>passive components | No power source. RIP<br>Belt Cable contains only<br>passive components | No power source. Third<br>Party RIP Belt Cables<br>contain only passive<br>components | Identical |
| Isolation | No component included<br>in the RIP Belt Cable<br>that is relied on as<br>means of protection | No component included<br>in the RIP Belt Cable<br>that is relied on as<br>means of protection | No component included in<br>the Third Party RIP Belt<br>Cables that is relied on as<br>means of protection | Identical |
| Standards | | | | |
| AAMI/ANSI/ES<br>60601-1 | Yes | Yes | Yes | Identical |
| IEC 60601-1-2 | Yes | Yes | Yes | Identical |
Table 2 Comparison Summary of the New Device (Nox RIP Belt Cables) and Predicate Device
{13}------------------------------------------------
Note 3: Nox RIP Belt Cables are intended to allow measuring of respiratory effort signals and function as accessories for sleep systems. The predicate, QDC-PRO AND NOX RIP, is intended for measuring of physiological signals during sleep. Respiratory effort is one example of physiological signals and both the new device and the predicate allow measuring of the signals during sleep. Furthermore, the predicate RIP belt cable is intended to interconnect RIP belts and the QDC-PRO recorder to allow measuring of respiratory effort signals. Therefore the indication for use can be regarded the same.
{14}------------------------------------------------
Note 4: The intended environment depends on the classification of the sleep recorder to be used with the Nox RIP Belt Cables. The predicate device, ODC-PRO AND NOX-RIP, is intended to be used with a third party polysomnography (PSG) recorder. Simpler sleep recorders like polygraph (PG) recorders are commonly used at patients' home. The Nox RIP Belt Cables may be used with any type of sleep recorders/PSG recorders including PG recorders. Therefore, the home environment was added for the Nox RIP Belt Cables. If a sleep recorder is intended for home use, the type of RIP belt cables does not change or influence the intended environment of the recorder. The home inclusion to the environment of use does therefore not pose any new safety and effectiveness concerns for the Nox RIP Belt Cables. Additionally, risk analysis, verification and validation results demonstrate that the use of Nox RIP Belt Cables in the home environment does not raise any new concerns regarding the intended use, safety or efficiency. Furthermore, these are prescription devices and are always used under supervision of a licensed medical practitioner. The intended environments are therefore deemed equivalent.
Note 5: The RIP Belt Cables and Third Party RIP Belt Cables are compliant to standard AAMI/ANSI ES/60601-1. Furthermore, verification testing and risk analysis show that minor difference in material does not raise new questions about safety and effectiveness. The material of the new devices and the predicate can thus be regarded as equivalent.
Note 6: All cables, both new and predicate, are compliant with standard AAMI/ANSI ES/60601-1. Furthermore, verification testing, signal integrity comparison and risk analysis show that different connectors do not raise new questions about safety and effectiveness. The recorder end connectors can thus be regarded as equivalent.
The result of the above comparison is that the indication for use, patient population and intended environments for the Nox RIP Belt Cables can be regarded the same as that of the predicate QDC-PRO AND NOX-RIP device.
Furthermore, the comparison above shows that there are no significant differences in major technological characteristics, including features, materials, and principles of operation of Nox RIP Belt Cables compared to the predicate. The minor differences between the new and predicate device do not raise any new questions regarding safety and effectiveness. It can therefore be determined that Nox RIP Belt Cables have the same major technological characteristics as the predicate device, QDC-PRO AND NOX RIP.
{15}------------------------------------------------
## Performance Testing Summary
Thorough internal testing has demonstrated that the Nox RIP Belts and Nox RIP Belt Cables are suitable for their intended use.
The design of the Nox RIP Belts and Nox RIP Belt Cables was tested, verified and validated throughout the design process according to product requirement specifications and intended use.
The Nox RIP Belts and Nox RIP Belt Cables comply with the following standards:
- ISO 14971 Second edition 2007, Medical devices Application of risk management to . medical devices
- . ISO 15223-1:2012, Medical devices – Symbols to be used with medical device labels, labelling and information to be supplied
- . AAMI/ANSI/ES 60601-1:2005, Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance
- . IEC 60601-1-2 Edition 3: 2007-03, Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests
- AAMI/ANSI/IEC 62366:2007/(R)2013, Medical devices Application of usability . engineering to medical devices
Additionally, the Third Party RIP Belt Cables comply with
- 21 CFR 898 FDA PERFORMANCE STANDARD FOR ELECTRODE LEAD WIRES AND . PATIENT CABLES:2014
#### Verification
Nox RIP Belts and Nox RIP Belt Cables were tested towards test protocols written to cover all product requirement specifications. The test results demonstrate that the design output conforms to design input and that the device fulfills all physical characteristics, performance criteria, functional requiremental requirements, interface requirements, packaging and labeling requirements as well as safety and reliability requirements defined.
#### Risk Analysis
Risk analysis was performed for RIP Belts and RIP Belt Cables according to the standard: ISO 14971 - Medical devices - Application of risk management to medical devices, appropriate measures have been implemented and their effectiveness verified and validated.
#### Standards Testing
Assessment has been conducted for RIP Belts and RIP Belt Cables towards the standard AAMI/ANSI/ES 60601-1:2005. The result of this assessment is that the RIP Belt Cables are in compliance with the standard AAMI/ANSI/ES 60601-1:2005.
#### EMC Testing
Assessment has been conducted for RIP Belts and RIP Belt Cables towards the electromagnetic compatibility standard IEC 60601-1-2 Edition 3: 2007-03 to ensure compatibility with requirements addressing electrostatic discharge (ESD), electromagnetic disturbances, magnetic fields, electrical fast transients and bursts, and surge. The result of this assessment is that the RIP Belts and RIP Belt Cables are in compliance with the standard IEC 60601-1-2 Edition 3: 2007-03.
{16}------------------------------------------------
#### Usability
The RIP Belts and RIP Belt Cables were designed to minimize use errors and use-associated risks according to the usability engineering standard AAMI/ANSI/IEC 62366:2007/(R)2013. Usability testing resulted in all usability goals passed.
#### Signal Quality and Comparison Testing
Signal integrity tests were conducted for the Nox RIP Belts and Nox RIP Belt Cables with focus on signal to noise ratio, signal range, bandwidth and linearity and the test results compared to the signal integrity test conducted for the predicate QDC-PRO AND NOX-RIP.
The results from these signal integrity tests and comparison demonstrate that the Nox RIP Belts and Nox RIP Belt Cables signals are clinically equivalent to that of the predicate QDC-PRO AND NOX RIP.
#### Conclusion
Based on the testing, risk analysis and verification activates described above and detailed comparison to the predicate device provided in Table 2 above, it is the conclusion of Nox Medical that the Nox RIP Belts and Nox RIP Belt Cables are substantially equivalent to device already on the market (cleared by the 510(k) process) and present no new concerns about safety and effectiveness.
The Nox RIP Belts and Nox RIP Belt Cables are therefore substantial equivalent to the predicate QDC-PRO AND NOX-RIP from Nox Medical (K124062).
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.