The QDC-PRO device is a sensor unit intended for measuring of physiological signals during sleep. The signals measured are processed in the QDC-PRO device and the resulting signals made available at the output connector for acquisition by a 3rd party polysomnography (PSG)/sleep recorder. The QDC-PRO device is indicated for use in patients greater than 2 years of age. The intended environments are hospitals, institutions, sleep clinics, or other test environments.
Device Story
QDC-PRO is a wearable sensor unit for polysomnography (PSG) studies; measures respiratory effort via respiratory inductive plethysmography (RIP) belts, nasal pressure/snoring via nasal cannula, body position, activity, and audio via internal microphone. Device processes signals internally, including Quantitative Diagnostic Calibration (QDC) of RIP signals to estimate tidal volume; outputs analog signals to 3rd party PSG amplifiers. Used in hospitals, sleep clinics, or test environments; operated by clinicians/technicians. Provides calibrated respiratory data to assist in diagnosis of sleep-disordered breathing. Powered by 1.5V AA battery; features status display and control buttons.
Clinical Evidence
Bench testing only. No clinical trials performed. Evidence includes signal comparison tests against predicate devices (NOX T3, Respitrace QDC) confirming signal quality and clinical equivalence. QDC function performance verified against Respitrace QDC. Compliance with IEC60601-1, IEC60601-1-2, and ISO14971 risk management standards documented.
Technological Characteristics
Materials: PC with 10% glass fiber case, polyester/steel clips, PVC/ABS/PC cables, latex-free polyester/Dorlastan RIP belts. Sensing: RIP technology, solid-state position/activity sensor, solid-state pressure sensor, microphone. Connectivity: Analog output to 3rd party PSG systems; USB for firmware updates. Power: 1.5V AA battery. Standards: IEC60601-1, IEC60601-1-2, IEC62304, IEC62366, ISO14971.
Indications for Use
Indicated for patients > 2 years of age requiring measurement of physiological signals during sleep for polysomnography studies. No specific contraindications defined.
Regulatory Classification
Identification
A breathing (ventilatory) frequency monitor is a device intended to measure or monitor a patient's respiratory rate. The device may provide an audible or visible alarm when the respiratory rate, averaged over time, is outside operator settable alarm limits. This device does not include the apnea monitor classified in § 868.2377.
{1}------------------------------------------------
# 510(k) Summary
JUL 1 0 2013
### Submitter
Nox Medical Vatnagordum 18 IS-104 Reykjavik Iceland
Tel: +354 570 7170
Establishment Registration Number: 3007389703
#### Contact person
Kolbrun E Ottosdottir Vatnagordum 18 IS-104 Reykjavik Iceland Email: keo@noxmedical.com Direct tel: +354 698 3555
### Preparation Date
December 24, 2012
Trade Name:
#### Device
#### ODC-PRO
Common Name: Classification Name: Requlation Number: Product Code: Device Class: Classification Panel:
Respiratory Sensor Unit Ventilatory Effort Recorder 868.2375 MNR Class II Anesthesiology
### Predicate Devices
NOX T3 from Nox Medical Product Code: MNR 510(k) Number: K082113
Respitrace QDC from SensorMedics Corporation Product Code: BZQ 510(k) Number: K903011
{2}------------------------------------------------
# Device Description
The ODC-PRO device is intended to function as an accessory for 3rd party PSG systems, delivering the respiratory signals needed in PSG studies along with position/activity signals.
Polysomnography (PSG) is a multi-parametric sleep study indicated for diagnosis of various sleep disorders. The ODC-PRO is therefore a part of a full PSG, the gold standard sleep study, and its main objectives is to provide signals used for diagnoses of sleep disordered breathing (SDB).
The signals measured by the ODC-PRO are provided as analog signals to a general 3rd party PSG amplifier that has DC inputs with characteristics matching the QDC-PRO device signal output specifications.
The output signals include:
- Abdomen respiratory effort (RIP) .
- . Thorax respiratory effort (RIP)
- . SUM of abdomen and thorax respiratory effort (RIP)
- Nasal pressure from nasal cannula .
- Snore signal from nasal cannula .
- . Body position
- . Activity
- Audio .
The respiratory effort is measured by the use of respiratory inductive plethysmography (RIP).
The QDC-PRO device provides calibration for the RIP signals by the use of Quantitative Diagnostic Calibration (QDC) technique. The calibrated RIP signals (Sum) represents the tidal volume of the respiration better than un-calibrated RIP signals.
The QDC-PRO contains a sensor unit, respiratory effort sensors (RIP belts) and cables. The QDC-PRO is worn by the patient. It measures signals from two respiratory effort sensors, audio via an inbuilt microphone, nasal pressure and snoring via a nasal cannula and patient 's position/activity data. The signals are processed within the device and the resulting signals made available at the output connector for acquisition by a 3rd party polysomnography (PSG)/sleep recorder.
The QDC-PRO is powered with one AA (1.5V) battery and has a display for status indication, signal integrity, and buttons for control.
{3}------------------------------------------------
### Intended Use
The QDC-PRO device is a sensor unit intended for measuring of physiological signals during sleep. The signals measured are processed in the QDC-PRO device and the resulting signals made available at the output connector for acquisition by a 3rd party polysomnography (PSG)/sleep recorder.
The QDC-PRO device is indicated for use in patients greater than 2 years of age.
The intended environments are hospitals, institutions, sleep clinics, or other test environments.
## Technological Characteristics
The comparison table below is provided as a summary of the most relevant characteristics of the QDC-PRO device relative to the predicate devices. The comparison table demonstrates that the QDC-PRO device has no significant differences from the predicate devices for the intended use that would adversely affect product safety and effectiveness.
{4}------------------------------------------------
| Characteristic | NOX T3 | RESPITRACE QDC | QDC - PRO | | | |
|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|
| | (K082113) | (K903011) | | | | |
| | | General | | | | |
| Intended Use | The Nox T3 device is intended<br>for ambulatory recording of<br>physiological signals during<br>sleep. The recorded signals<br>are then downloaded to a PC<br>where the signals can be<br>viewed and analyzed by use of<br>the Nox T3 application<br>(Noxturnal). | The RESPITRACE QDC multi-<br>channel waveform generator<br>is indicated for diagnostic<br>polysomnograpy. This device<br>generates respiratory<br>waveforms from RESPITRACE<br>technology. | The QDC-PRO device is a<br>sensor unit intended for<br>measuring of physiological<br>signals during sleep. The<br>signals measured are<br>processed in the QDC-PRO<br>device and the resulting<br>signals made available at the<br>output connector for<br>acquisition by a 3rd party<br>Polysomnography (PSG)/sleep<br>recorder. | | | |
| Contraindication<br>for Use | The Nox T3 system is NOT<br>intended for any patient<br>monitoring or automatic<br>diagnosis. | Not defined in labeling | The ODC-PRO device is NOT<br>intended for any patient<br>monitoring. | | | |
| Intended<br>Environments | The intended environments<br>are hospitals, institutions,<br>sleep centers, sleep clinics, or<br>other test environments,<br>including patient's home. | Not defined in labeling | The intended environments<br>are hospitals, institutions,<br>sleep centers, sleep clinics, or<br>other test environments. | | | |
| Patient<br>Population | The Nox T3 system is<br>indicated for use in patients<br>greater than 2 years of age | Not defined in labeling | The QDC-PRO device is<br>indicated for use in patients<br>greater than 2 years of age. | | | |
| Prescription Use | Yes | Yes | Yes | | | |
| | | Physical/Material | | | | |
| Material Case | ABS/PC | Painted metal housing | PC with 10% glass fiber | | | |
| Material Device<br>Clips | Polyester/Steel | Not defined in labeling | Polyester/Steel | | | |
| Material RIP<br>connection<br>cable | PVC and ABS/PC | Not defined in labeling | PVC and ABS/PC | | | |
| RIP Belts<br>Material | Polyester/Dorlastan, latex free | Not defined in labeling | Polyester/Dorlastan, latex free | | | |
| Dimension | 79mm (3.11") W<br>63mm (2.48")H<br>21mm(0.83") D | 195mm (76.8") W<br>275mm (108.3") H<br>110mm(43.3") D | 83mm (3.27") W<br>63mm (2.48")H<br>21mm(0.83") D | | | |
| Weight | 65g | 2000g | 71g | | | |
| Disposable<br>Components | Disposable respiratory effort<br>sensor/RIP belts.<br>Filter tube connector<br>Remaining portions require<br>cleaning. | Disposable respiratory effort<br>sensor/RIP belts.<br>Remaining components<br>require cleaning. | Disposable respiratory effort<br>sensor/RIP belts.<br>Remaining portions require<br>cleaning. | | | |
| Acquisition<br>units | One unit | One unit | One unit | | | |
| | Functional | | | | | |
| Number of<br>channels | Eight channels. | Three channels. | Eight channels. | | | |
Table 1 Comparison Summary of the New Device (QDC-PRO) and Predicate Devices
{5}------------------------------------------------
| Characteristic | NOX T3<br>(K082113) | RESPITRACE QDC<br>(K903011) | QDC - PRO |
|-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Operation time | 24 hours (depending on<br>batteries) | Undefined (mains operated) | 8 hours or longer (depending<br>on batteries) |
| Data Acquisition<br>Control | Data acquisition<br>microprocessor controlled | Not defined in labeling | Data acquisition<br>microprocessor controlled |
| Acquisition<br>parameter<br>Control | Acquisition parameters set<br>from application SW | Acquisition parameters<br>controlled from the device<br>software only | Acquisition parameters<br>controlled from the device<br>software only |
| Calibration of<br>RIP | No | Yes | Yes |
| Device Data<br>Storage | Yes | No | No |
| Data Interface | Via USB connection to PC | Via analog interface to 3rd<br>party sleep system<br>Via serial port to PC | Via analog interface to 3rd<br>party sleep system |
| Firmware<br>upgrade | Via USB connection to PC | Not defined in labeling | Via USB connection to PC |
| Connection to<br>Patient | RIP belts for respiratory effort<br>RIP belts for attaching of<br>device and clip straps to<br>secure position of device<br>Plastic tubing and cannula for<br>pressure measurement<br>Probes or Flexi Wrap for<br>oximetry<br>Touch proof electrode cables | RIP belts for respiratory effort | RIP belts for respiratory effort<br>RIP belts for attaching of<br>device and clip straps to<br>secure position of device<br>Cannula for pressure<br>measurement |
| Signals and Sensors | | | |
・
.
.
{6}------------------------------------------------
| Characteristic | NOX T3<br>(K082113) | RESPITRACE QDC<br>(K903011) | QDC - PRO |
|-------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| Signals outputs | Respiratory Effort (Abdomen<br>and Thorax) | Respiratory Effort (Abdomen<br>and Thorax) - calibrated<br>RIP SUM | Respiratory Effort (Abdomen<br>and Thorax) - calibrated<br>RIP SUM |
| | RIP SUM<br>Body position<br>Activity<br>Nasal /mask pressure<br>Airflow<br>Snore<br>Respiratory sound/audio<br>Oxygen Saturation<br>Pulse<br>EEG, EOG, EMG, ECG | | Body position<br>Activity<br>Nasal pressure<br>Snore<br>Respiratory sound /Audio<br>intensity |
| Sensor<br>Technology<br>used in/with<br>the system | Solid state position/activity<br>sensor<br>Respiratory Effort Sensors<br>(RIP technology)<br>Microphone<br>Solid state pressure sensor<br>Oximetry<br>Gold cup electrodes<br>Ag/AgCL electrodes | Respiratory Effort Sensors<br>(RIP technology) | Solid state position/activity<br>sensor<br>Respiratory Effort Sensors<br>(RIP technology)<br>Microphone<br>Solid state pressure sensor |
| | | Power and Isolation | |
| Power Source | 1.5V by 1 AA battery (Data<br>Acquisition)<br>Host PC (Data Transfer) | 115/230V AC (Data<br>Acquisition) | 1.5V by 1 AA battery (Data<br>Acquisition)<br>Host PC (FW upgrade) |
| Galvanic<br>Connection to<br>mains/Patient<br>Isolation | Device has no galvanic<br>connections to mains as it is a<br>battery operated device<br>Not possible to connect<br>auxiliary devices to the device | Yes via connection over mains<br>operated 3rd party amplifier<br>Isolation between mains and<br>applied part<br>Via isolation mains adapter<br>(mains operated) | Yes via connection over mains<br>operated 3rd party amplifier<br>Isolation between mains and<br>applied part |
| | | User Interface | |
| Data validation | The device has a display to<br>view signal integrity and<br>status of the device | No | The device has a display to<br>view signal integrity and<br>status of the device |
| Control on<br>device | Push buttons | On/off button | Push buttons |
| | | Standards | |
| IEC60601-1 or<br>equivalent FDA<br>recognized<br>standard | Yes | Yes | Yes |
| IEC60601-1-2<br>or equivalent<br>FDA recognized<br>standard | Yes | Yes | Yes |
{7}------------------------------------------------
### Performance Testing Summary
Thorough internal and external testing has demonstrated that the QDC-PRO device is effective and safe for its intended use.
The ODC-PRO device complies with the applicable EMC and patient safety standards:
- IEC60601-1:2007 Medical electrical equipment Part 1-2: General requirements for . basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
- AAMI/ ANSI ES60601-1:2005 Medical electrical equipment Part 1: General . requirements for basic safety and essential performance (IEC 60601-1).
The software in the QDC-PRO device was developed in compliance with the requirements listed in the standard: IEC62304 - Medical Device Software - Software Life Cycle Processes, which provides a framework of life cycle processes with activities and tasks necessary for the safe design and maintenance of medical device software, including testing, verification and validation.
The design of the QDC-PRO device was tested, verified and validated throughout the design process according to requirement specifications and intended use.
Risk analysis was performed according to the standard: ISO14971 - Medical devices -Application of risk management to medical devices, appropriate measures was implemented and their effectiveness verified and validated.
Signal comparison tests conducted for the QDC-PRO device confirm the quality of the signals being measured by the device. The results demonstrate that the QDC-PRO signals may be regarded as clinically equivalent to that of the predicates Respitrace QDC and NOX T3.
Quantitative Diagnostic Calibration (QDC) function comparison tests conclude that the QDC-PRO performance may be regarded as clinically equivalent to that of the predicate Respitrace QDC.
Usability testing compliant with the standard: IEC 62366 - Medical devices - Application of usability engineering to medical devices, demonstrates that the QDC-PRO device is simple and safe to operate and minimizes the likelihood of errors and lapses.
Clinical evaluation for the ODC-PRO device concludes that all of the signals measured and processed by the device are comparable to the already validated and publicly available Respitrace QDC and Nox T3 device, respectively.
{8}------------------------------------------------
# Conclusion
Based on the testing, risk analysis, verification and validation activates described above and Besed on the essang, how analysis) - chivices provided in Table 1 above, it is the conclusion of Nox Medical that the QDC-PRO device is substantially equivalent to the predicates NOX T3 from Nox Medical (K082113) and Respitrace QDC from SensorMedics Corporation (K903011) and presents no new concerns about safety and effectiveness.
{9}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avento Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 10, 2013
Nox Medical C/O Ms. Kolbrun E. Ottosdottir Quality/Regulatory Manager Katrinartuni 2 Reykjavik, Iceland IS-105
Re: K124062
・
Trade/Device Name: QDC-PRO Regulation Number: 21 CFR 868.2375 Regulation Name: Ventilatory Effort Recorder Regulatory Class: II Product Code: MNR Dated: June 5, 2013 Received: June 10, 2013
Dear Ms. Ottosdottir:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{10}------------------------------------------------
#### Page 2 - Ms. Ottosdottir
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.hun for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Kwame Ulmer, M.S Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{11}------------------------------------------------
# Indication for Use
510(k) Number (if known): Kl24062
Device Name: QDC-PRO
Indications For Use:
The QDC-PRO device is a sensor unit intended for measuring of physiological signals during sleep. The signals measured are processed in the QDC-PRO device and the resulting signals made available at the output connector for acquisition by a 3rd party polysomnography (PSG)/sleep recorder.
The QDC-PRO device is indicated for use in patients greater than 2 years of age.
The intended environments are hospitals, institutions, sleep centers, sleep clinics, or other test environments.
× Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use_ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Anya C. Harry
| Digitally signed by | Anya C. Harry |
|--------------------------------------|-----------------------------|
| DN: c=US, o=U.S. Government, ou=HHS, | |
| ou=FDA, ou=People, cn=Anya C. Harry, | |
| 0.9.2342.19200300.100.1.1=0011315590 | |
| Date: | 2013.07.09 16:24:57 -04'00' |
Page 1 of 1
(Division Sign-Off) (Division Sign-Oil)
Division of Anesthesio:ogy, General Hospital Infection Control, Dental Devices
510(k) Number: K124062
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.