The Pulse Oximeter is a handheld non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate of adult, adolescent, child and infant patients in hospitals, hospital-type facilities and homecare. The device is not intended to be used under motion and low perfusion. It is designed for finger thickness between 0.8cm and 2.2cm (0.3 inches to 0.9 inches).
Device Story
Handheld, battery-powered pulse oximeter (models MD300CN130, MD300CN356) for non-invasive spot-checking of SpO2 and pulse rate. Uses dual-wavelength LED (660nm red, 905nm infrared) and photodetector to measure light absorption in pulsating arteriolar vascular bed; calculates SpO2 based on systolic/diastolic absorption ratios. Operated by clinicians or patients in hospitals or homecare. Displays SpO2, pulse rate, pulse bar, and (for MD300CN356) perfusion index and waveforms. Provides immediate physiological data to assist in patient monitoring. Benefits include portable, non-invasive assessment of oxygenation status.
Clinical Evidence
Clinical study conducted per ISO 80601-2-61 Annex EE. 11 healthy adult volunteers (ages 20-42, diverse pigmentation) evaluated under steady-state, non-motion conditions. Induced hypoxia levels ranged from 100% to 70% SaO2. SpO2 accuracy compared against arterial blood co-oximeter reference. Results confirmed performance specifications met.
Technological Characteristics
ABS enclosure, silicone gel fingertip cushion. Dual-wavelength LED (660nm/905nm) sensing. Battery-powered (2xAAA). Dimensions accommodate finger thickness 0.8-2.2cm. Standards: IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2, ISO 80601-2-61, ISO 10993-5/10.
Indications for Use
Indicated for spot-checking SpO2 and pulse rate in adult, adolescent, child, and infant patients in hospital and homecare settings. Not for use under motion or low perfusion conditions.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 8, 2023
Beijing Choice Electronic Technology Co., Ltd. Haiying Zhao Quality Director 2nd Floor 3rd Floor and Room 410-412 4th Floor No. 2 Building No. 9 Shuangyuan Road Shijingshan Beijing, Beijing 100041 China
Re: K220101
Trade/Device Name: Pulse Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DOA Dated: February 21, 2023 Received: March 2, 2023
Dear Haiying Zhao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
James J. Lee -S
James J. Lee, Ph.D. Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K220101
Device Name Pulse Oximeter
Indications for Use (Describe)
The Pulse Oximeter is a handheld non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate of adult, adolescent, child and infant patients in hospital-type facilities and homecare. The device is not intended to be used under motion and low perfusion. It is designed for finger thickness between 0.8cm and 2.2cm (0.3 inches to 0.9 inches).
Type of Use (Select one or both, as applicable)
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section II 510(k) Summary
This summary of 510(k) is being submitted in accordance with the requirements of 21 CFR 807.92.
There is no prior submission for the device.
#### Submitter Information 1.
### Manufacturer
Establishment Registration Number: 3005569927
Beijing Choice Electronic Technology Co., Ltd.
2nd Floor, 3rd Floor and Room 410-412 4th Floor, No. 2 Building, No. 9 Shuangyuan Road, Shijingshan District BEIJING, 100041, P.R. China.
#### Contact Person
Haiying Zhao
Beijing Choice Electronic Technology Co., Ltd.
2nd Floor, 3rd Floor and Room 410-412 4th Floor, No. 9 Shuangyuan Road, Shijingshan District BEIJING, 100041, P.R. China.
Phone: +86-10-88204631
Fax: 861088204632
Email: cc@choicemmed.com
Date prepared: November 10, 2021
#### 2. Identification of Proposed Device
Device Common Name: Oximeter
Device Trade/Proprietary Name: Pulse Oximeter
Model: MD300CN130, MD300CN356
Classification Name: Oximeter
Regulation Number: 870.2700
Product Code: DQA
Regulatory Class: Class II
Panel: Anesthesiology
#### 3. Device Description
The proposed device, Pulse Oximeter, is a battery powered device, which can mainly detect and display the measured oxyhemoglobin saturation (SpO2) and pulse rate (PR) value. It available two models, MD300CN130 and MD300CN356. The model MD300CN130 is adopted LED screen to display SpO2 and Pulse Rate (PR) value, low power indication and pulse bar, the device has 2 display modes. The model MD300CN356 using LCD screen with 7 display modes, it can display SpO2, Pulse Rate (PR), Perfusion Index (PI), waveform, battery indicator, signal indicator and pulse bar, the brightness level can be adjusted
{4}------------------------------------------------
1-10 level.
The proposed device consists of power supply module, detector and emitter LED, signal collection and process module, display module, user interface and button control circuit.
The enclosure of the proposed device is made of ABS and the fingertip cushion is made of Silicone Gel.
The proposed device is not for life-supporting or life-sustaining, not for implant.
The device is not sterile, and the transducers are reusable and do not need sterilization.
The device is for prescription.
The device does not contain drug or biological products.
#### 4. Identification of Predicate Device
Trade/Proprietary Name: Fingertip Pulse Oximeter
Model: MD300CI218
Common Name: Pulse Oximeter
K-number: K181503
Regulation Number: 21 CFR 870.2700
Device Class: II
Product Code: DQA
Panel: Anesthesiology
Manufactured by: Beijing Choice Electronic Technology Co., Ltd.
{5}------------------------------------------------
#### ડ. Substantially Equivalent (SE) Comparison
Table 1 Comparison Table between the Proposed Device and Predicate Device
| Comparison Elements | Proposed Device | | Predicate Device<br>(K181503) | Remark | |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|
| | MD300CN130 | MD300CN356 | | | |
| Product Name | Pulse Oximeter | Pulse Oximeter | Fingertip Pulse Oximeter | / | |
| Model | MD300CN130 | MD300CN356 | MD300CI218 | / | |
| Regulation No. | 21 CFR 870.2700 | 21 CFR 870.2700 | 21 CFR 870.2700 | Same | |
| Regulatory Class | II | II | II | Same | |
| Classification Name | Oximeter | Oximeter | Oximeter | Same | |
| Product Code | DQA | DQA | DQA | Same | |
| Intended Use | The Pulse Oximeter is a handheld non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate of adult, adolescent, child and infant patients in hospitals, hospital-type facilities and homecare. The device is not intended to be used under motion and low perfusion. It is designed for finger thickness between 0.8cm and 2.2cm (0.3 inches to 0.9 inches). | | The Fingertip Pulse Oximeter MD300CI218 is a handheld non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate of adult, adolescent, child and infant patients in hospitals, hospital-type facilities and homecare environment. | Similar 1 | |
| Intended Population | Adult, adolescent, child and infant patients | | Adult, adolescent, child and infant patients | Same | |
| Components | Power supply module, detector and emitter LED, signal collection and processor module, display module, user interface and button control. | | Power supply module, detector and emitter LED, signal collection and processor module, display module, Bluetooth module, user interface and button control. | Similar, Analysis 2 | |
| Design Principle | The pulse oximeter works by applying a sensor to a pulsating arteriolar vascular bed. The sensor contains | | The pulse oximeter works by applying a sensor to a pulsating arteriolar vascular | Same | |
| Comparison Elements | Proposed Device | | Predicate Device<br>(K181503) | Remark | |
| | MD300CN130 | MD300CN356 | | | |
| | a dual light source and photo detector. The one<br>wavelength of light source is 660nm, which is red<br>light; the other is 905nm, which is infrared-red light.<br>Skin, bone, tissue and venous vessels normally<br>absorb a constant amount of light over time. The<br>photo detector in finger sensor collects and converts<br>the light into electronic signal which is proportional<br>to the light intensity. The arteriolar bed normally<br>pulsates and absorbs variable amounts of light during<br>systole and diastole, as blood volume increases and<br>decreases. The ratio of light absorbed at systole and<br>diastole is translated into an oxygen saturation<br>measurement. This measurement is referred to as<br>SpO2. | | bed. The sensor contains a dual light<br>source and photo detector. The one<br>wavelength of light source is 660nm,<br>which is red light; the other is 905nm,<br>which is infrared-red light. Skin, bone,<br>tissue and venous vessels normally absorb<br>a constant amount of light over time. The<br>photo detector in finger sensor collects<br>and converts the light into electronic<br>signal which is proportional to the light<br>intensity. The arteriolar bed normally<br>pulsates and absorbs variable amounts of<br>light during systole and diastole, as blood<br>volume increases and decreases. The ratio<br>of light absorbed at systole and diastole is<br>translated into an oxygen saturation<br>measurement. This measurement is<br>referred to as SpO2. | | |
| Measurement | Red | 660± 3nm | | 660± 3nm | |
| Wavelength | Infrared | 905± 10nm | | 905± 10nm | |
| Design | Display Type | LED | LCD | OLED | Different □ Analysis 3 |
| Features | Working Time | continuously operated<br>as long as 18 hours | continuously operated as<br>long as 20 hours | continuously operated as long as 18 hours | Similar □ Analysis 4 |
| | Display Mode | 2 display modes | 7 display modes | 2 display modes | Similar, Analysis 5 |
| Power supply | | 2*AAA alkaline batteries | | 2*AAA alkaline batteries | Same |
| Performance Data | | SpO2. PR | SpO2. PR. PI | SpO2. PR | Similar, Analysis 6 |
| Comparison Elements | Proposed Device | | Predicate Device<br>(K181503) | Remark | |
| | MD300CN130 | MD300CN356 | | | |
| SpO2 | SpO2 Display Range | 0%~100% | 0%~100% | | |
| | SpO2 Measurement Range | 70%~100% | 70%~100% | | |
| | SpO2 Accuracy | 70%~100%, ± 2%;<br>0~69% no definition | 70%~100%, ± 2%;<br>0~69% no definition | | |
| | SpO2 Resolution | 1% | 1% | | |
| PR | PR Display Range | 30 bpm~250 bpm | 30 bpm~250 bpm | | |
| | PR Measurement Range | 30 bpm~250 bpm | 30 bpm~250 bpm | | |
| | PR Accuracy | 30 bpm~99 bpm, ± 2 bpm;<br>100 bpm~250 bpm, ± 2% | 30 bpm~99 bpm, ± 2 bpm;<br>100 bpm~250 bpm, ± 2% | | |
| | PR Resolution | 1 bpm | 1 bpm | | |
| PI | Display range | NA | 0.1%~20% | NA | |
| | Measure range | NA | 0.3~20.0% | NA | |
| | Resolution | NA | 0.1% | NA | |
| Transmitter | | NA | Bluetooth Compliance: Version 4.0…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.