The Fingertip Pulse Oximeter MD300CI218 is a handheld non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate of adult, adolescent, child and infant patients in hospital-type facilities and homecare environment.
Device Story
Handheld, battery-powered fingertip pulse oximeter; uses dual-wavelength LED (660nm red, 905nm infrared) and photodetector to measure light absorption in pulsating arteriolar vascular bed; calculates SpO2 and pulse rate based on systolic/diastolic absorption ratios; displays data on OLED screen; transmits data via Bluetooth 4.0 to mobile app for tracking; used in hospitals and homecare by clinicians or patients; provides real-time physiological monitoring to assist in health status assessment.
Clinical Evidence
Clinical study conducted per ISO 80601-2-61:2011 with 12 healthy adult volunteers (ages 18-45, diverse skin tones). Primary endpoint was SpO2 accuracy during steady-state conditions (70-100% range). Results demonstrated an Arms of 1.66%. No adverse events reported.
Technological Characteristics
Battery-powered (2xAAA); OLED display; Bluetooth 4.0 connectivity; dual-wavelength sensor (660nm/905nm). Materials: ABS enclosure, medical silicone gel cushion. Standards: IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2, ISO 80601-2-61. Non-sterile, reusable.
Indications for Use
Indicated for spot-checking SpO2 and pulse rate in adult, adolescent, child, and infant patients in hospital and homecare settings.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.
Beijing Choice Electronic Technology Co., Ltd. Haiying Zhao Quality Director No.9 Shuangyuan Road, Badachu Hi-tech Zone, Shijingshan District Beijing, 100041 China
Re: K181503
Trade/Device Name: Fingertip Pulse Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: September 4, 2018 Received: September 4, 2018
#### Dear Haiying Zhao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Todd D. Courtney -S
Tina Kiang, Ph.D. for Acting Director Division of Anesthesiology, General Hospital. Respiratory. Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Section II Indications for Use Statement
# Indications for Use
510(k) Number (if known): K181503
Device Name: _Fingertip Pulse Oximeter MD300CI218
Indications for Use:
The Fingertip Pulse Oximeter MD300CI218 is a handheld non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate of adult, adolescent, child and infant patients in hospital-type facilities and homecare environment.
| Prescription Use <span style="text-decoration: overline;">√</span> | AND/OR | Over-The-Counter Use <span style="text-decoration: overline;"></span> |
|--------------------------------------------------------------------|--------|-----------------------------------------------------------------------|
| (Part 21 CFR 801 Subpart D) | | (21 CFR 801 Subpart C) |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE
(1) OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
# Section III 510(k) Summary
This summary of 510(k) is being submitted in accordance with the requirements of 21 CFR 807.92.
There is no prior submission for the device.
# 3.1 Submitter Information
#### ● Manufacturer Name:
Establishment Registration Number: 3005569927 Beijing Choice Electronic Technology Co., Ltd. Room 4104, No. A12 Yuquan Road Haidian District 100143 Beijing, P.R.China
#### . Contact Person:
Lei Chen Beijing Choice Electronic Technology Co., Ltd. North Building 3F, No. 9 Shuangyuan Road, Badachu Hi-tech Zone, Shijingshan District Beijing China 100041 Phone: +86-10-88798300 Ext 6020 Fax:215-4052545 Email: cc@choicemmed.com
- . Date prepared: June 04, 2018
## 3.2 Proposed Device Information
Device Common Name: Pulse Oximeter Device Trade/Proprietary Name: Fingertip Pulse Oximeter Model: MD300CI218 Classification Name: Oximeter Regulation Number: 870.2700 Product Code: DQA
Class: II
Panel: Anesthesiology
{4}------------------------------------------------
### 3.3 Predicate Device
510(k) Number: K142888 Common Name: Pulse Oximeter Device Trade/Proprietary Name: Fingertip Pulse Oximeter Model: MD300C318T2 Classification Name: Oximeter Product Code: DQA Regulation Number: 870.2700
Device Class: II
Panel: Anesthesiology
Manufacturer: Beijing Choice Electronic Technology Co., Ltd.
Intended Use: The MD300C318T2 is intended for spot checking in measuring and displaying functional arterial hemoglobin (SpO2) and pulse rate. It is intended for adult, adolescent, child and infant users in hospitals, hospital facilities and home healthcare environments.
# 3.4 Device Description
The proposed device Fingertip Pulse Oximeter MD300CI218 is a battery powered device. It can detect and display the measured %SpO2 and pulse rate value and will automatically power off when there is no signal for longer than 8 seconds. The proposed device is adopted colorful color OLED screen to display SpO2, PR and waveform which can be displayed in 2 directions. And it is designed with the battery indicator function to warn the user that the battery power may be low.
The proposed device is normally applied to adult, adolescent, child and infant patients in hospitals, hospital facilities and homecare environment. And it can transmit the measurements to Smart Device installed the mobile APP via Bluetooth 4.0 to help the users to organize and track their health information.
The proposed device consists of power supply module, detector and emitter LED, signal collection and processor module, display module, Bluetooth module, user interface and button control.
The fingertip pulse oximeter works by applying a sensor to a pulsating arteriolar vascular bed. The sensor contains a dual light source and photo detector. The one wavelength of light source is 660nm, which is red light; the other is 905nm, which is infrared-red light. Skin, bone, tissue and venous vessels normally absorb a constant amount of light over time. The photo detector in finger sensor collects and converts the light into electronic signal which is proportional to the light intensity. The arteriolar bed normally pulsates and absorbs variable amounts of light during systole and diastole, as blood volume increases and decreases. The ratio of light absorbed at systole and diastole is translated into an
{5}------------------------------------------------
oxygen saturation measurement. This measurement is referred to as SpO2.
The proposed device is not for life-supporting or life-sustaining, not for implant.
The device is not sterile and the transducers are reusable and do not need sterilization and re-sterilization.
The device is for prescription.
The device does not contain drug or biological products.
The device is software-driven and the software validation is provided in software.
{6}------------------------------------------------
# 3.5 Comparison list of the technological characteristics
| Comparison Elements | Proposed Device | Predicate Device | |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | | |
| Product Name | Fingertip Pulse Oximeter | Fingertip Pulse Oximeter | |
| Model | MD300CI218 | MD300C318T2 | |
| Regulation No. | 21 CFR 870.2700 | 21 CFR 870.2700 | |
| Classification | II | II | |
| Classification Name | Oximeter | Oximeter | |
| Product Code | DQA | DQA | |
| Indications for Use | The MD300CI218 is a handheld non-invasive<br>device intended for spot-checking of oxygen<br>saturation of arterial hemoglobin (SpO2) and<br>Pulse Rate of adult, adolescent, child and<br>infant<br>patients in hospitals, hospital-type<br>facilities and homecare environment. | The MD300C318T2 is intended for spot-<br>checking in measuring and displaying<br>functional arterial hemoglobin (SpO2) and<br>pulse rate. It is intended for adult, adolescent,<br>child and infant users in hospitals, hospital<br>facilities and home healthcare environments. | |
| Comparison Statement | The proposed device and the predicated device have the same intended use and<br>classification | | |
| Components | Power supply module, detector and emitter<br>LED, signal collection and processor module,<br>display module,<br>Bluetooth<br>module,<br>user<br>interface and button control. | Power supply module, detector and emitter<br>LED, signal collection and process module,<br>display module, Bluetooth module, indicator<br>module, user interface. | |
| Design Principle | The fingertip pulse oximeter works by applying a sensor to a pulsating arteriolar vascular bed. The sensor contains a dual light source and photo detector. The one wavelength of light source is 660nm, which is red light; the other is 905nm, which is infrared-red light. Skin, bone, tissue and venous vessels normally absorb a constant amount of light over time. The photo detector in finger sensor collects and converts the light into electronic signal which is proportional to the light intensity. The arteriolar bed normally pulsates and absorbs variable amounts of light during systole and diastole, as blood volume increases and decreases. The ratio of light absorbed at systole and diastole is translated into an oxygen saturation measurement. This measurement is referred to as SpO2. | | Principle of the oximeter is as follows: A mathematical formula is established making use of Lambert Beer Law according to Spectrum Absorption Characteristics of Reductive hemoglobin (RHb) and Oxyhemoglobin (HbO2) in red and near-infrared zones. Operation principle of the instrument: Photoelectric Oxyhemoglobin Inspection Technology is adopted in accordance with Capacity Pulse Scanning and Recording Technology, so that two beams of different wavelength of lights (660nm glow and 940nm near infrared light) can be focused onto a human nail tip through a clamping finger-type sensor. A measured signal obtained by a photosensitive element, will be shown on the oximeter's display through process in electronic circuits and microprocessor. |
| Measurement | Red | 660±3nm | 660nm |
| Wavelength | Infrared | 905±10nm | 940nm |
| Comparison Statement | | The proposed device and the predicate device have the same design principle and similar measurement wavelength. The only difference is the wavelength of the infrared LED emitter, and we can verify that which will not effect the essential performance of the proposed device. | |
| Display Type | | OLED | OLED |
| | | | |
| User Interface | 2 display directions | 2 display directions | |
| Power supply | 2*AAA alkaline batteries | 2*AAA alkaline batteries | |
| Display Data | SpO2, PR | SpO2, PR | |
| SpO2 Display Range | 0~100% | 0%~100% | |
| SpO2 Measurement Range | 70%~100% | 70%~100% | |
| SpO2 Accuracy | 70%~100%, ±2%;<br>0~69% no definition | 70%~100%, ±2%;<br>0~69% no definition | |
| SpO2 Resolution | 1% | 1% | |
| PR Display Range | 30bpm~250bpm | 0bpm~235bpm | |
| PR Measurement Range | 30bpm~250bpm | 30bpm~235bpm | |
| PR Accuracy | 30bpm~99bpm, ±2bpm;<br>100bpm~250bpm, ±2% | 30bpm~99bpm, ±2bpm;<br>100bpm~235bpm, ±2% | |
| PR Resolution | 1% | 1% | |
| Wireless Transmission Range | 0~10m | 0~10m | |
| Antenna Type | Internal | Internal | |
| Transmitter | Bluetooth Compliance: Version 4.0 | Bluetooth Compliance: Version 2.1 | |
Table 3-1 Performance Specification Comparison Table between the Proposed Device and Predicate Device
{7}------------------------------------------------
{8}------------------------------------------------
{9}------------------------------------------------
| Premarket Notification 510(k) Submission-Section III 510(k) Summary |
|---------------------------------------------------------------------|
|---------------------------------------------------------------------|
| Operating<br>Temperature | 5°C~40°C | 5°C~40°C | |
|---------------------------------------------|--------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Relative Humidity | 15% ~93%, no condensation in operation;<br>≤93% no condensation in storage | 15% ~93%, no condensation in operation;<br>≤93% no condensation in storage | |
| Atmosphere<br>Pressure | 70kPa~106kpa | 86kPa~106kpa | |
| Comparison Statement | The proposed device has similar product specification as predicate device. | | |
| Contacting<br>Material | Battery Cover<br>Enclosure<br>Power Button | ABS<br>ABS | |
| | Fingertip Cushion | Medical Silicone Gel<br>Medical Silicone Gel | |
| Comparison Statement<br>Performance Testing | The contacting materials of the proposed device are same to those of the predicate device. | | |
| | Laboratory Testing | The laboratory tests include SpO2 and PR<br>accuracy Test, Weak Perfusion Test, High and<br>Low Temperature and Humidity Test,<br>Performance Test After Cleaning and<br>ISO80601-2-61<br>The laboratory tests include SpO2 and PR<br>accuracy Test, Weak Perfusion Test, High and<br>Low Temperature and Humidity Test,<br>Performance Test After Cleaning and<br>ISO80601-2-61 | |
| | Electrical Safety | Conformed to IEC60601-1, IEC 60601-1-11<br>Conformed to IEC60601-1, IEC 60601-1-11 | |
| EMC and Electrical Safety | Electromagnetic Compatibility | Conformed to IEC60601-1-2<br>Conformed to FCC certification | Conformed to IEC60601-1-2<br>Conformed to FCC certification |
| | | | |
| | Software and Cybersecurity | Moderate level of concern | Moderate level of concern |
| | | Compliance with:<br>FDA Guidance for the content of Premarket<br>Submissions for Software Contained in<br>Medical Devices and,<br>FDA Guidance for Content of Premarket<br>Submission for Management of Cybersecurity<br>in Medical Device<br>Risk Management in Compliance with<br>ISO14971:2007 | Compliance with FDA Guidance for the<br>content of Premarket Submissions for<br>Software Contained in Medical Devices<br>Risk Management in Compliance with<br>ISO14971:2007 |
| | | | |
| | |…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.