MASIMO SET AND MASIMO RAINBOW SET PULSE OXIMETERS AND SENSORS
Applicant
Masimo Corporation
Product Code
DQA · Cardiovascular
Decision Date
Sep 19, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K120657 · Sep 19, 2012
MASIMO SET AND MASIMO RAINBOW SET PULSE OXIMETERS AND SENSORS
Masimo Corporation
Prospective clinical studies of newborn screening in maternity centers
The sponsor used data from two large-scale clinical studies to support labeling changes for newborn CCHD screening, demonstrating the device's performance in routine clinical nursery environments.
Prospective clinical study (Sweden); Prospective clinical study
Newborn subjects; Sample Size: 39,821; Number of Sites: 5
Not applicable for this study
Detection of CCHD
Prospective clinical study (UK); Prospective clinical study
Newborn subjects; Sample Size: 20,055; Number of Sites: 6
Not applicable for this study
Detection of CCHD
Indications for Use
The Masimo Rad 5 Pulse Oximeter is indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor). The Masimo Rad 5 Pulse Oximeter is indicated for use with adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments.
Device Story
Masimo SET and rainbow SET pulse oximeters/CO-oximeters use noninvasive optical sensors to measure arterial oxygen saturation, pulse rate, and (in rainbow models) carboxyhemoglobin, methemoglobin, total hemoglobin, and respiration rate. Devices utilize motion-tolerant signal processing to provide data in clinical settings (hospitals, nurseries, home). This submission updates labeling to include newborn screening protocols for critical congenital heart disease (CCHD) based on clinical studies. Clinicians use output to screen newborns for low blood oxygen saturation, facilitating early detection of CCHD. Devices are intended for use by healthcare professionals in well-infant and intermediate care nurseries.
Clinical Evidence
No new clinical trials were conducted for this submission. Evidence relies on two cited prospective clinical studies: one involving 39,821 newborns using Masimo SET technology and another involving 20,055 newborns using Masimo rainbow SET technology. These studies demonstrated the efficacy of pulse oximetry screening for detecting critical congenital heart disease (CCHD) in well-infant and intermediate care nurseries.
Technological Characteristics
Devices utilize Masimo SET (Signal Extraction Technology) and rainbow SET technology for noninvasive optical monitoring. Features include motion-tolerant signal processing and validation for low perfusion conditions. Form factors include handheld and bedside monitors with compatible optical sensors. Connectivity includes data output to multi-parameter displays. No changes to hardware or materials were made in this submission.
Indications for Use
Indicated for continuous or non-continuous noninvasive monitoring of functional SpO2, pulse rate, and (for rainbow SET models) SpCO, SpMet, SpHb, and RRa. Applicable to adult, pediatric, infant, and neonatal patients, including those with motion or poor perfusion, in hospital, mobile, and home settings.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Masimo ReSposable SpO2 Series Oximetry Sensors (K111621)
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(k) Summary: K120657
| Submitted by: | Masimo Corporation<br>40 Parker<br>Irvine, CA 92618<br>Phone: (949) 297-7000<br>FAX: (949) 297-7592 |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact | Patricia Milbank<br>Vice President, Regulatory Affairs |
| Date Summary Prepared | September 14, 2012 |
| Trade Name | Masimo SET® Rad 5 Pulse Oximeter<br>Masimo SET® Rad 5v Pulse Oximeter<br>Masimo SET® Rad-8 Pulse Oximeter<br>Masimo rainbow SET® Rad 57 Pulse CO-Oximeter<br>Masimo rainbow SET® Radical 7 Pulse CO-Oximeter<br>Masimo rainbow SET® Rad 87 Pulse CO-Oximeter<br>Masimo LNOP Oximetry Sensors<br>Masimo LNCS/M-LNCS Oximetry Sensors<br>Masimo LNOP/M-LNCS/LNCS Multisite-L Oximetry Sensors<br>Masimo LNOP/M-LNCS/LNCS Y-I Oximetry Sensors<br>Masimo ReSposable SpO2 Series Oximetry Sensors |
| Common Name | Pulse Oximeter and Oximeter Sensor |
| Regulation Number/Class | 21 CFR 870.2700 / Class II |
| Product Code | DQA |
| Substantially<br>Equivalent Devices: | Masimo SET Rad 5 Pulse Oximeter (K033296) |
| | Masimo SET Rad 5v Pulse Oximeter (K040214)<br>Masimo SET Rad 8 Pulse Oximeter (K092838)<br>Masimo rainbow SET® Rad 57 Pulse CO-Oximeter (K080238)<br>Masimo rainbow SET® Radical 7 Pulse CO-Oximeter (K110280<br>Masimo rainbow SET® Rad 87 Pulse CO-Oximeter (K091241)<br>Masimo LNOP Oximetry Sensors (K111888)<br>Masimo LNCS/M-LNCS Oximetry Sensors (K051212 and<br>K101896)<br>Masimo LNOP/M-LNCS/LNCS Multisite-L Oximetry Sensors<br>(K111888)<br>Masimo LNOP/M-LNCS/LNCS Y-I Oximetry Sensors (K012992)<br>Masimo ReSposable SpO2 Series Oximetry Sensors (K111621) |
SEP 19 2012
{1}------------------------------------------------
K120657 Pg 20f2
### Description of the Device
Masimo SET® and Masimo rainbow SET® Pulse Oximeter instruments and sensors provide noninvasive monitoring of arterial oxygen saturation (SpO2) and pulse rate (PR). The Masimo rainbow SET technology also provides nonitoring of carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (g/dl SpHb), and/or respiration rate (RRa).
This 510(k) is being submitted to support modifications to the device labeling to incorporate the results of clinical trials described in peer-reviewed publications regarding the use of Masimo pulse oximeters and sensors intended to screen newborn patients for critical congenital heart disease (CCHD). These modifications include the published newborn screening protocol for CCHD recommended by a work group selected by SACHDNC, AAP, ACCF and AHA (the "CCHD Workgroup"), and instructions regarding proper use of the subject Masimo devices to implement this protocol.
These labeling changes are being made in direct response to action taken by HHS in September 2011 adding pulse oximeter screening of newborns for CCHD to the Federal Recommended Uniform Screening Panel (RUSP) Guidelines. These guidelines recommend use of devices that a) are motion-tolerant, b) report functional oxygen saturation, c) have been validated in low perfusion conditions; and d) have been cleared by the FDA for use in newborns, criteria that are fully met by the subject Masimo devices.
## Clinical Summary
In reaching their recommendations, the CCHD Workgroup relied upon two independent prospective clinical studies finding "... sufficient evidence to begin screening for low blood oxygen saturation through the use of pulse-oximetry monitoring to detect CCHD in well-infant and intermediate care nurseries."1
In both of these studies, the investigators selected the Masimo SET technology and the Masimo rainbow SET technology for newborn screening as follows:
- A prospective clinical study of 39,821 newborn subjects at 5 maternity centers in Sweden . using the Masimo Radical pulse oximeter.3 The Masimo Rad 5, Rad 5v and Rad 8 systems incorporate the same Masimo SET technology used in this study.
- A prospective clinical study of 20,055 newborn subjects at 6 maternity centers in the UK ● using the Masimo Radical 7 pulse CO-Oximeter.4 The Masimo Rad 57 and Rad 87 systems incorporate the same Masimo rainbow SET technology used in this study.
#### Non-Clinical Summary
No non-clinical studies were required to support the proposed changes to the labeling regarding the use of the Masimo SET and Masimo rainbow SET technologies to perform screening of newborn subjects for CCHD. The Masimo pulse oximeter devices comply with the requirements
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
SEP 19 2012
Masimo Corporation c/o Patricia Milbank, JD Vice President, Regulatory Affairs 40 Parker Irvine, CA 92618
Re: K120657
Trade/Device Name: Masimo SET and Masimo rainbow SET Pulse Oximeters and Sensors Regulation Number: 21 CFR 8870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: September 14, 2012 Received: September 17, 2012
#### Dear Ms. Milbank:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Patricia Milbank, JD
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number: K120657
Device Name: Masimo SET® Rad 5 Pulse Oximeter
Indications for Use:
The Masimo Rad 5 Pulse Oximeter is indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor). The Masimo Rad 5 Pulse Oximeter is indicated for use with adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments.
Prescription Use X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Cardiovascular Devices 51 C(k) Number K12065
{5}------------------------------------------------
510(k) Number: K120657
Device Name: Masimo SET® Rad 5v Pulse Oximeter
Indications for Use:
The Masimo Rad 5v Pulse Oximcter is indicated for the non-continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor). The Masimo Rad 5v Pulse Oximeter is indicated for use with adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments.
Prescription Use X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
e
(Division Sign-C Division of C 510(k) Number
{6}------------------------------------------------
510(k) Number: K120657
Device Name: Masimo SET® Rad 8 Pulse Oximeter
Indications for Use:
The Masimo Rad 8 Pulse Oximeter and accessories are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor). The Masimo Rainbow Rad 8 Pulse Oximeter and accessories are indicated for use with adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospital-type facilities, mobile, and home environments.
Prescription Use X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) Indications for Use
WE
(Division Slan-Off) 510(k) Number
{7}------------------------------------------------
#### 510(k) Number: K120657
Device Name: Masimo Rainbow SET® Rad 57 Pulse CO-Oximeter
Indications for Use:
The Masimo Rad 57 Pulse CO-Oximeter and accessories are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor) and total hemoglobin concentration (measured by an SpCO/SpMet/SpHb sensor). The Masimo Rad 57 Pulse CO-Oximeter and accessories are indicated for use with adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospital-type facilities, mobile, and home environments.
Prescription Use X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
he
്ഷം Sign-Off) : Asion of Cardiovascular Devices 510(k) Number K12045
{8}------------------------------------------------
510(k) Number: K120657
Device Name: Masimo rainbow SET® Radical 7 Pulse CO-Oximeter
Indications for Use:
The Masimo Radical 7 Pulse CO-Oximeter and accessories are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RRa). The Masimo Radical 7 Pulse CO-Oximeter and accessories are indicated for use with adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments. In addition, the Masimo Radical 7 Pulse CO-Oximeter and accessories are indicated to provide the continuous non-invasive monitoring data obtained from the Masimo Radical 7 Pulse CO-Oximeter end accessories of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate to multi-parameter devices for the display of those devices.
Prescription Use X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Sign Sign-Off) sion of Cardiovascular Devlaes 510{k} Number
{9}------------------------------------------------
510(k) Number: K120657
Device Name: Masimo Rainbow SET® Rad 87 Pulse CO-Oximeter
Indications for Use:
The Masimo Rad 87 Pulse CO-Oximeter and accessories are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RR). The Masimo Rad 87 Pulse CO- . Oximeter and accessories are indicated for use with adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments.
Prescription Use X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
{10}------------------------------------------------
510(k) Number: K120657
Device Name: Masimo LNOP Oximetry Sensors
#### Indications for Use:
The Masimo LNOP Oximetry Sensors are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor) for use with adult, pediatric, infant, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospitaltype facilities, mobile, and home environments.
Prescription Usc X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) 510(k) Number
{11}------------------------------------------------
510(k) Number: K120657
Device Name: Masimo LNCS/M-LNCS Oximetry Sensors
Indications for Use:
The Masimo LNCS/M-LNCS Oximetry Sensors are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor) for use with adult, pediatric, infant, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
:
Sign-Off)
on of Cardiovascular Devices
510(k) Number K120657
{12}------------------------------------------------
510(k) Number: K120657
Masimo LNOP/M-LNCS/LNCS Multisite-L Oximetry Sensors Device Name:
Indications for Use:
The Masimo LNOP/M-LNCS/LNCS Multisite-L Oximetry Sensors are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor) for use with adult, pediatric, infant, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments.
Prescription Use X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
We
(Division Sigh-Off) Division of Cardiovascular Devices 510(k) Number
{13}------------------------------------------------
510(k) Number: K120657
Masimo LNOP/M-LNCS/LNCS Y-I Oximetry Sensors Device Name:
Indications for Use:
The Masimo LNOP/M-LNCS/LNCS Y-I Oximetry Sensors are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor) for use with adult, pediatric, infant, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments.
Prescription Use X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
... . on Sidn-Off Division of C SIO(k) Numb
{14}------------------------------------------------
510(k) Number: K120657
Device Name: Masimo ReSposable SpO2 Series Oximetry Sensors
Indications for Use:
The Masimo ReSposable SpO2 Series Oximetry Sensors are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor) for use with adult, pediatric, infant, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments.
Prescription Use X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
(Division Sign-Off) Division of Cardiovascular Devices 47065 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.