The Masimo Rainbow SET® Rad 87 Pulse CO-Oximeter and accessories are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RR). The Masimo Rainbow SET® Rad 87 Pulse CO-Oximeter and accessories are indicated for use with adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospital-type facilities, mobile, and home environments.
Device Story
The Rad 87 is a pulse CO-oximeter using Masimo Rainbow SET technology to monitor arterial blood parameters. Inputs include multi-wavelength optical signals from finger sensors for SpO2, SpCO, SpMet, and SpHb, and vibratory signals from neck sensors for respiratory rate. The device processes these signals via an internal MX board to calculate physiological values and indices like Perfusion Index (PI), Pleth Variability Index (PVI), and Total Arterial Oxygen Content (SpOC). It is operated by clinicians in hospitals, mobile, or home settings. Output is displayed on the device screen and can be transmitted via RS-232 or Nurse Call interfaces. Clinicians use this real-time data to assess patient status and guide clinical interventions. The device benefits patients by providing continuous, noninvasive monitoring of critical blood parameters and respiratory status, reducing the need for invasive blood sampling.
Clinical Evidence
Clinical validation included studies on healthy adult volunteers (light to dark skin) and neonatal NICU patients. SpO2, SpCO, and SpMet accuracy were validated against laboratory CO-oximeters during induced hypoxia. SpHb accuracy was validated in adults and surgical patients. Pulse rate and low perfusion accuracy were validated via bench testing against simulators. Motion accuracy was validated in human blood studies with induced motion (2-5 Hz).
Indicated for continuous noninvasive monitoring of SpO2, pulse rate, SpCO, SpMet, SpHb, and respiratory rate in adult, pediatric, and neonatal patients. Suitable for use in hospital, mobile, and home environments, including patients with motion or poor perfusion.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
Masimo Rainbow SET® Rad 87 Pulse CO-Oximeters and Accessories (K080238)
Oridion Capnography Inc., Capnostream 20 with Integrated Pulmonary Index (K082268)
{0}------------------------------------------------
KO91241
| Submitted by: | Masimo Corporation<br>40 Parker<br>Irvine, CA 92618<br>Phone: (949) 297-7000<br>FAX: (949) 297-7001<br>NOV - 6 2009 |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Marguerite Thomlinson, Manager of Regulatory Affairs |
| Date Summary Prepared: | August 14, 2009 |
| Trade Name | Masimo Rainbow SET® Rad 87 CO-Oximeter and Accessories |
| Common Name | Pulse Oximeter and Sensor |
| Regulation Number: | 21 CFR 870.2700 |
| Regulation Name: | Oximeter |
| Regulation Class: | Class II |
| Product Code | DQA, BZQ, DPZ, JKS |
| Substantially Equivalent Devices | Masimo Rainbow SET® Rad 87 Pulse CO-Oximeters and Accessories, 510(k) Number – K080238<br><br>Oridion Capnography Inc., Capnostream 20 with Integrated<br>Pulmonary Index, 510(k) Number – K082268 |
Andromed Inc., Biological Sound Monitor (BSM) Sensor 510(k) Number - K021389
## Description of the Device
The Rainbow SET® Rad 87 Pulse CO-Oximeter and accessories (Rad 87) include the MX board with Masimo Rainbow SET technology.
The Rad 87 provides noninvasive monitoring of arterial oxygen saturation (%SpO₂), pulse rate, carboxyhemoglobin saturation (%SpCO), methemoglobin saturation (%SpMet), total hemoglobin concentration (g/dl SpHb), and/or respiratory rate (RR). Other information displayed by the Rad 87 includes: Low Signal IQ (Low SIQ), Perfusion Index (PI), Pleth Variability Index (PVI), Total Arterial Oxygen Content (SpOC), Respiratory Signal Quality (RSQ), alarm status, alarm silence, battery life, sensor status, and trends. The Rad 87 has output interfaces include: Nurse Call analog output, and RS-232 serial output.
The Rad 87 in this filing are the Rad-87 in the K080238 filing, but with the addition of respiratory rate monitoring.
## Intended Use/Indications for Use
The Masimo Rainbow SET® Rad 87 Pulse CO-Oximeter and accessories are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO₂), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RR). The Masimo Rainbow SET® Rad 87 Pulse CO-
{1}------------------------------------------------
Oximeter and accessories are indicated for use with adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospital-type facilities, mobile, and home environments.
## Principles of Operation
## SpO2 and Pulse Rate
Pulse oximetry is governed by the principles that oxyhemoglobin (oxygenated blood), deoxyhemoglobin (non-oxygenated blood), carboxyhemoglobin (blood with carbon monoxide content), and methemoglobin (blood with oxidized hemoglobin content) species differ in their absorption of visible and infrared light. The amount of arterial blood in tissue changes with the pulse (photoplethysography). Therefore the amount of light, absorbed by the varying quantities of arterial blood, changes accordingly.
## SpCO. SpMet, and SpHb General Description
The Rad-87 include the Masimo Rainbow SET technology for SpCO, SpMet and SpHb measurements, based on the same principles of pulse oximetry. The Masimo Rainbow SET technology uses a multiwavelength sensor to distinguish between oxygenated blood, deoxygenated blood, blood with carbon monoxide, blood with oxidized hemoglobin and blood plasma.
Once the Masimo Rainbow SET technology receives the signal from the sensor, it calculates the patient's functional oxygen saturation (SpO2), fractional concentration of carboxyhemoglobin (SpCO), fractional concentration of methemoglobin (SpMet), total hemoglobin concentration (SpHb) and pulse rate.
## Respiratory Rate General Description
The Masimo Rainbow SET technology also provides respiratory rate measurements, based on vibratory signals from respiratory sounds.
## Method of Operation
## SpO2, SpCO, SpMet, and SpHb
The instrument (Rad 87) is turned on. A sensor is attached to a patient's finger. The other end of the sensor is attached to a patient cable. The other end of the patient cable is connected to the Dual Channel cable or directly to the instrument. The connection of the patient cable to the Dual Channel cable is only needed when pulse CO-oximetry monitoring is concurrent with respiratory rate monitoring.
The instrument will begin continuously displaying the patient's pulse rate and SpO2 value. Depending on the type and/or configuration of the instrument, monitoring information would also include SpCO, SpMet, SpHb, PVI, and/or SpOC. The practitioner can then use the information to help assess the condition of the patient and as an aide in determining if any intervention is required by the practitioner.
Once the practitioner determines the patient no longer requires monitoring, the cable is disconnected from the sensor, the sensor is removed (and disposed of if it is a single use device), and the power to the instrument is turned off.
## Respiratory Rate
The instrument (Rad 87) is turned on. A sensor is attached to a patient's neck. The other end of the sensor is connected to a patient cable. The other end of the cable is connected to the Dual Channel cable. The Dual Channel cable is then connected to the instrument.
The instrument will begin continuously display the patient's respiratory rate. The practitioner can then use the information to help assess the condition of the patient and as an aide in determining if any intervention is required by the practitioner. Once the practitioner determines the patient no longer requires monitoring, the patient cable is disconnected from the sensor, the sensor is disposed and the power to the instrument is turned off.
{2}------------------------------------------------
## Specifications
The specifications for the Rad 87 are the following:
| FEATURES | SPECIFICATIONS |
|-----------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Display Ranges | Oxygen Saturation (SpO2): 0-100%<br>Pulse Rate (beat per minute or bpm): 25-240 bpm<br>Carboxyhemoglobin Saturation (SpCO): 0-99%<br>Methemoglobin Saturation (SpMet): 0-99.9%<br>Total Hemoglobin (SpHb): 0-25 g/dL<br>Respiratory Rate (RR): 0-150 breaths per minute<br>Total Oxygen Concentration (SpOC): 0-35 ml/dl<br>Perfusion Index: 0.02-20%<br>Pleth Variability Index: 0-100% |
| Accuracy | See Footnotes 1, 2, 3, 4, 5, 6, and 7 |
| SpO2, No Motion | 60-80 ± 3%, adults/pediatrics/infants |
| SpO2, Motion | 70-100 ± 2%, adults/pediatrics/infants; ± 3%, neonates |
| SpO2, Low Perfusion | 70-100 ± 2%, adults/pediatrics/infants/neonates |
| Pulse Rate, No Motion | 25-240 ± 3 bpm, adults/pediatrics/infants/neonates |
| Pulse Rate, Motion | 25-240 ± 5 bpm, adults/pediatrics/infants/neonates |
| Pulse Rate, Low Perfusion | 25-240 ± 3 bpm, adults/pediatrics/infants/neonates |
| SpCO | 1-40 ± 3%, adults/pediatrics/infants |
| SpMet | 1-15 ± 1%, adults/pediatrics/infants/neonates |
| SpHb | 8-17 ±1 g/dl (arterial or venous), adults/pediatrics/infants |
| RR | 4-70 ± 1 breath per minute, adults (> 30kg) |
| General | |
| Resolution | SpO2: 1%<br>Pulse Rate: 1 bpm<br>SpCO: 1%<br>SpMet: 0.1%<br>SpHb: 0.1 g/dl<br>RR: 1 breath per minute |
| Measurements | Low Signal IQ<br>Perfusion Index (PI)<br>Total Oxygen Concentration (SpOC)<br>Pleth Variability Index (PVI)<br>Respiratory Signal Quality (RSQ) |
| Electrical | |
| Power (AC) | Voltage Input Range: 100-240 VAC, 47-63 Hz |
| Batteries | Rechargeable |
| Circuitry | Microprocessor controlled<br>Automatic self-test of pulse CO-oximeter when powered on<br>Automatic setting of default parameters<br>Automatic alarm messages<br>Trend data output |
| Firmware | Rainbow SET technology, MX Board/Circuitry |
| Mechanical | |
| Material | Polycarbonate/ABS Blend |
| FEATURES | SPECIFICATIONS |
| Environmental | |
| Operating Temperature | 32 to 122°F (0 to 50°C) |
| Storage Temperature | -40 to 158°F (-40 to 70°C) |
| Relative Storage Humidity | 10 to 95% non-condensing |
| Operating Altitude | Pressure: 500-1,060 mbar<br>Altitude: -1,000-18,000 ft (-304-5,486m) |
| Mode & Sensitivity | |
| SpO2 Averaging Mode | 2, 4, 6, 8, 10, 12 and 16 seconds; FastSat |
| SpO2 Sensitivity | APOD, Normal, Maximum |
| Alarms | |
| Volume Level Adjustment: Pulse/Tone | OFF; 25% to 100% in 4 increments |
| Alarm Silence | 120 seconds delay; All mute: continuous silence |
| Out of Limit Alarms: SpO2, Pulse Rate, SpCO, SpMet, SpHb, RR, PI, PVI | High/low alarms |
| Sensor Condition Alarm | No Sensor; Sensor Off; Sensor Defect |
| System | System failure |
| Battery Alarm | Low battery |
| Display and Indicators | |
| Data Display | SpO2 (%)<br>Pulse rate (bpm)<br>SpCO (%)<br>SpMet (%)<br>SpHb (g/dl)<br>SpHbv(g/dl)<br>RR<br>SpOC(ml/dl)<br>Perfusion Index-PI (%)<br>Pleth Variability Index-PVI (%)<br>Signal IQ<br>Respiratory Signal Quality (RSQ)<br>Sensitivity indicator<br>Sensor status<br>Sensor time<br>Status messages<br>Alarm status<br>Battery status |
| Output Interface | |
| Analog output | Nurse Call |
| Serial Port<br>(RS-232 connector) | PC/printer connection<br>Philips Vuelink<br>RadNet<br>Patient Safety Net<br>Trends |
| Compliance | |
| EMC Compliance | EN 60601-1-2, Class B |
| Electrical Safety | IEC 60601-1, 2nd Edition; UL 60601-1 |
| Radio | 802.11 a/b/g |
| Type of Protection (AC Power) | Class 1 |
| Type of Protection (battery power) | Internally Powered |
| Degree of Protection-Patient Cable | Type BF-Applied Part |
| Liquid Ingress-Degree of Protection | IPX1 |
| Mode of Operation | Continuous |
:
{3}------------------------------------------------
{4}------------------------------------------------
#### Footnotes
- SpO2, SpCO and SpMet accuracy was determined by testing on healthy adult volunteers in the 1 range of 60-100% SpO2, 0-40% SpCO, and 0-15% SpMet against a laboratory CO-Oximeter. SpO2 and SpMet accuracy was determined on 16 neonatal NICU patients ranging in age from 7-135 days old and weighing between 0.5-4.25 kg. Seventy-nine (79) data samples were collected over a range of 70-100% SaO2 and 0.5-2.5% MetHb with a resultant accuracy of 2.9% SpO2 and 0.9% SpMet.
- ನ The Masimo sensors have been validated for no motion accuracy in human blood studies on healthy adult male and female volunteers with light to dark skin pigmentation in induced hypoxia studies in the range of 70-100% SpO2 against a laboratory CO-Oximeter and ECG monitor. This variation equals plus or minus one standard deviation. Plus or minus one standard deviation encompasses 68% of the population.
- The Masimo sensors have been validated for motion accuracy in human blood studies on healthy ന adult male and female volunteers with light to dark skin pigmentation in induced hypoxia studies while performing rubbing and tapping motions, at 2 to 4 Hz at an amplitude of 1 to 2 cm and a non-repetitive motion between 1 to 5 Hz at an amplitude of 2 to 3 cm in induced hypoxia studies in the range of 70-100% SpO- against a laboratory CO-oximeter and ECG monitor. This variation equals plus or minus one standard deviation which encompasses 68% of the population.
- The Masimo SET Technology has been validated for low perfusion accuracy in bench top testing য against a Biotek Index 2 simulator and Masimo's simulator with signal strengths of greater than 0.02% and transmission of greater than 5% for saturations ranging from 70 to 100%. This variation equals plus or minus one standard deviation which encompasses 68% of the population:
- The Masimo sensors have been validated for pulse rate accuracy for the range of 25-240 bpm in ട് bench top testing against a Biotek Index 2 simulator. This variation equals plus or minus one standard deviation which encompasses 68% of the population.
- SpHb accuracy has been validated on healthy adult male and female volunteers and on surgical റ patients with light to dark skin pigmentation in the range of 8-17 g/dl SpHb against a laboratory CO-oximeter. This variation equals plus or minus one standard deviation which encompasses 68% of the population. The SpHb accuracy has not been validated with motion or low perfusion.
- The following substances may interfere with pulse CO-oximetry measurements: 7
- · Elevated levels of Methemoglobin (MetHb) may lead to inaccurate SpO2 and SpCO measurements
- · Elevated levels of Carboxyhemoglobin (COHb) may lead to inaccurate SpO₂ measurements.
- · Very low arterial Oxygen Saturation (SpO2) levels may cause inaccurate SpCO and SpMet measurements
- · Severe anemia may cause erroneous SpO2 readings.
- · Dyes, or any substance containing dyes, that change usual blood pigmentation may cause erroneous readings.
- · Elevated levels of total bilirubin may lead to inaccurate SpO2, SpMet, SpCO and SpHb readings
{5}------------------------------------------------
## Test Summary
The Rad 87 comply with the voluntary standards as detailed in Section 9 of this submission. The following quality assurance measures were applied to the development of the Rad 87:
- . Risk Analysis
- Design Reviews .
- Biocompatibility Testing .
- . Performance Testing
- . Safety Testing
- Environmental Testing .
- Clinical Testing .
## Conclusions
The information in this 510(k) submission demonstrates that the Rad 87 are substantially equivalent to the predicate devices, with respect to safety, effectiveness, and performance.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo features the department's name arranged in a circular fashion around an emblem. The emblem is a stylized representation of an eagle.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002
Ms. Marguerite Thomlinson Manager of Regulatory Affairs Masimo Corporation 40 Parker Irvine, California 92618
M-M-M
Re: K091241
Trade/Device Name: Masimo Rainbow SET Rad 87 Pulse CO-Oximeter and Accessories Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA, BZQ,DPZ, JKS Dated: October 29, 2009 Received: October 30, 2009
NOV - 6 2009
Dear Ms. Thomlinson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
## Page 2- Ms. Thomlinson
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours
Susan Rose
Susan Runner, D.D.S., M.A. Acting Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
## Enclosure
{8}------------------------------------------------
## Section 4 - Indications for Use
510(k) Number (if known):
Masimo Rainbow SET Rad 87 Pulse CO-Oximeter and Accessories Device Name:
Indications For Use:
The Masimo Rainbow SET® Rad 87 Pulse CO-Oximeter and accessories are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RR). The Masimo Rainbow SET® Rad 87 Pulse CO-Oximeter and accessories are indicated for use with adult, pediatric, and neonatal patients during both no motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments.
Prescription Use X (Per 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (Per 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CD/RH, Office of Device Evaluation (ODE)
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K091241
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number:
K091241-Rad 87 w/ ARRM Masimo Response, 8/14/09 Page 7 of 180
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.