Convenience arterial blood samples from neonatal patients in clinical settings
Clinical data from convenience arterial blood samples collected from neonates in a clinical environment were used to verify the form, fit, and function of the RD SET Disposable sensors on the neonatal population.
Neonatal; Convenience samples; Clinical environment; Arterial blood gas
Neonatal subjects (≤ 1 month, ≤ 4.5 kg); Sample Size: 42 subjects (aggregated); Number of Sites: 3
Reference co-oximeter SaO2 measurements
SpO2 accuracy (ARMS)
Indications for Use
The Masimo Rad-97 and Accessories can communicate with network systems for supplemental remote viewing and alarming (e.g., at a central station). In addition, the Masimo Rad-97 and Accessories are indicated to provide the continuous non-invasive monitoring data obtained from the Masimo Rad-97 and Accessories for functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) to multiparameter devices for the display on those devices. The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) of adult, pediatric, and neonatal patients during both no motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments. The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of carboxyhemoglobin saturation (SpCO) of adult, pediatric, and infant patients during no motion conditions in hospitals and hospital-type facilities. The Masimo Rad-97 and Accessories are not intended to be used as the sole basis for making diagnosis or treatment decisions related to suspected carbon monoxide poisoning; it is intended to be used in conjunction with additional methods of assessing clinical signs and symptoms. The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of methemoglobin saturation (SpMet) of adult, pediatric, and neonatal patients during no motion conditions in hospitals and hospital-type facilities. The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of total hemoglobin concentration (SpHb) of adult and pediatric patients during no motion conditions in hospitals and hospital-type facilities. The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of respiratory rate (RRa) for adult, pediatric, and neonatal patients during no motion conditions in hospitals, hospital-type facilities, home environments, and transport within healthcare facilities. The optional Nomoline Capnography product family is intended to be connected to other medical backboard devices for monitoring of breath rate and CO2. The Nomoline Capnography product family is intended to be connected to a patient breathing circuit for monitoring of inspired/expired gases during anesthesia, recovery and respiratory care. The environment is the operating suite, intensive care unit and patient room. The intended patient population is adult, pediatric and infant patients. The optional non-invasive blood pressure (NIBP) module is indicated for the noninvasive measurement of arterial blood pressure in hospital-type facilities, mobile, and home environments. The NIBP module is designed to measure blood pressure for patient population described in the following table: [Table omitted]
Device Story
Rad-97 system; touchscreen patient monitor; provides continuous non-invasive monitoring of SpO2, PR, Pi, PVi, SpCO, SpMet, SpHb, SpOC, RRa, CO2, and NIBP. Inputs: optical signals from sensors (red/infrared light absorption); capnography gas sampling; NIBP cuff pressure. Processing: Masimo rainbow SET technology; calculates physiological parameters from plethysmographic waveforms and gas/pressure data. Outputs: numeric values, waveforms, alarm status, and status messages on integrated color LCD. Used in hospitals, ICUs, home, and transport; operated by clinicians. Integrates with Patient SafetyNet and EMR via wired/wireless interfaces. Supports remote viewing/alarming. Benefits: continuous, real-time physiological monitoring; aids clinical decision-making by providing early detection of changes in patient status.
Clinical Evidence
Clinical testing performed on 42 neonates (including 22 pre-term) across 3 sites using convenience arterial blood samples. 70 data pairs compared SpO2 sensor readings to reference co-oximeter SaO2. Results showed ARMS of 3.185% for all data (70-100% SaO2). Testing verified safe form, fit, and function of RD SET Disposable sensors on neonates.
Indicated for continuous non-invasive monitoring of SpO2, PR, SpCO, SpMet, SpHb, RRa, CO2, and NIBP in adult, pediatric, infant, and neonatal patients. Used in hospitals, hospital-type facilities, mobile, and home environments. SpCO/SpMet/SpHb/RRa/CO2 monitoring restricted to specific patient populations and conditions (e.g., no motion). NIBP indicated for neonates (birth-1mo), infants (1mo-2yrs), children (2-12yrs), adolescents (12-21yrs), and adults (21+).
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it.
Masimo Corporation Linus Park Vice President, Regulatory 52 Discovery Irvine. California 92618
Re: K191059
Trade/Device Name: Masimo Rad-97 and Accessories Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: MWI, DQA, DPZ, DXN, CCK, BZQ, FLL Dated: November 26, 2019 Received: November 27, 2019
Dear Linus Park:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For Todd Courtney Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K191059
Device Name Masimo Rad-97 and Accessories
### Indications for Use (Describe)
The Masimo Rad-97 and Accessories can communicate with network systems for supplemental remote viewing and alarming (e.g., at a central station, the Masimo Rad-97 and Accessories are indicated to provide the continuous non-invasive monitoring data obtained from the Masimo Rad-97 and Accessories for functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) to multiparameter devices for the display on those devices.
The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) of adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospital-type facilities, mobile, and home environments.
The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of carboxyhemoglobin saturation (SpCO) of adult, pediatric, and infant patients during no motion conditions in hospitals and hospital-type facilities. The Masimo Rad-97 and Accessories are not intended to be used as the sole basis for making diagnosis or treatment decisions related to suspected carbon monoxide poisoning; it is intended to be used in conjunction with additional methods of assessing clinical signs and symptoms.
The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of methemoglobin saturation (SpMet) of adult, pediatal patients during no motion conditions in hospitals and hospital-type facilities.
The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of total hemoglobin concentration (SpHb) of adult and pediatic patients during no motion conditions in hospital-type facilities.
The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of respiratory rate (RRa) for adult, pediatric, and neonatal patients during no motion conditions in hospitals, hospital-type facilities, home environments, and transport within healthcare facilities.
The optional Nomoline Capnography product family is intended to other medical backboard devices for monitoring of breath rate and CO2. The Nomoline Capnography product family is intended to a patient breathing circuit for monitoring of inspired gases during anesthesia, recovery and respiratory care. The environment is the operating suite, intensive care unit and patient population is adult, pediativ and infant patients.
The optional non-invasive blood pressure (NIBP) module is indicated for the noninvasive measurement of arterial blood pressure in hospitals, hospital-type facilities, mobile, and home environments. The NIBP module is designed to measure blood pressure for patient population described in the following table:
| Patient Population | Approximate Age Range |
|--------------------|---------------------------|
| Newborn (neonate) | Birth to 1 month of age |
| Infant | 1 month to 2 years of age |
| Child | 2 to 12 years of age |
{3}------------------------------------------------
Adolescent 12-21 years of age Adult 21 years of age and older
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of the Masimo symbol on the left, followed by the word "Masimo" in a bold, sans-serif font. To the right of the name is the text "MASIMO CORPORATION" in a smaller font, followed by the address "52 Discovery, Irvine, CA 92618".
| Submitter and Address of<br>Manufacturing Facility: | Masimo Corporation<br>52 Discovery<br>Irvine, CA 92618<br>Phone: (949) 297-7000<br>FAX: (949) 297-7592 |
|-----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date: | December 27, 2019 |
| Contact: | Linus Park<br>Vice President, Regulatory Affairs<br>Masimo Corporation<br>Phone: (949) 297-7337 |
| Trade Name: | Masimo Rad-97 and Accessories |
| Common Name: | Patient Monitor |
| Classification Regulation/<br>Product Code: | 21 CFR 870.2300, Class II/MWI |
| Additional Product Code: | 21 CFR 870.2700, Class II/DQA<br>21 CFR 868.1400, Class II/CCK<br>21 CFR 862.3220, Class II/JKS<br>21 CFR 868.2375, Class II/BZQ<br>21 CFR 870.1130, Class II/DXN<br>21 CFR 880.2910, Class II/FLL |
| Establishment Registration<br>Number: | 2031172 |
| Reason for Premarket<br>Notification: | Apply the previously cleared SpO2 performance specification to<br>the neonate population |
| Predicate Device: | K180046 – Masimo Rad-97 and Accessories |
| Performance Standards | No performance standards for the above device have been<br>promulgated pursuant to Section 514. |
### 5.1. Device Description
The subject device, Masimo Rad-97 System and Accessories (Rad-97 product family), features a touchscreen display that continuously displays numeric values for the measured monitoring parameters. The Rad-97 product family can be operated on AC power or internal rechargeable battery.
The subject device (Rad-97 product family) is substantially the same as the predicate (Rad-97 product family) cleared under K180046, and has the same indications for use. The Rad-97 comprises of the same measurement technologies as cleared in the predicate, which includes the Masimo rainbow SET technology, capnography technology, and noninvasive blood pressure (NIBP) technology. These technologies enable the Rad-97 product family to provide noninvasive monitoring of functional oxygen
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white checkmark inside, followed by the word "Masimo" in black, bold letters. To the right of the logo is the text "MASIMO CORPORATION" followed by the address "52 Discovery, Irvine, CA 92618".
saturation of arterial hemoglobin (SpO2), pulse rate (PR), Perfusion Index (Pi), Pleth Variability Index (PVi), carboxyhemoglobin (SpCO), methemoglobin (SpMet), total hemoglobin (SpHb), oxygen content (SpOC), acoustic respiration rate (RRa), and/or optional capnography parameters or optional noninvasive blood pressure (NIBP) parameters.
The subject of this submission is the update to the SpO2 performance specification for the RD SET Disposable sensors for the neonate population to match that of the cleared 1.5% Arms for adults (K180046).
| Feature | Rad-97 Specification |
|--------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Display | |
| Display type | Touchscreen, Color LCD (Backlit Active Matrix TFT LCD) |
| Measurement range | Functional Oxygen Saturation (SpO2): 0-100% |
| | Pulse Rate (PR): 25-240 beats per minute (bpm) |
| | Perfusion Index (Pi): 0.02-20% |
| | Pleth Variability Index (PVi): 0-100% |
| | Respiration Rate (RRa): 0-70 breaths per minute |
| | Carboxyhemoglobin Saturation (SpCO): 0-99% |
| | Methemoglobin Saturation (SpMet): 0-99.9% |
| | Total Hemoglobin (SpHb): 0-25g/dL; 0-250g/L(0-15.52 mmol/dL) |
| | Oxygen Concentration (SpOC): 0-35ml of O2/dL |
| | Carbon Dioxide (CO2); 0 vol%-15 vol % |
| | End Tidal CO2 (EtCO2): 0-25 %, 0-32.5 kPa, 0-244 mmHg |
| | Fractional Inspired CO2 (FiCO2): 0-25 %, 0-32.5 kPa, 0-244<br>mmHg |
| | Respiratory Rate (RR): 0-150 breaths/min |
| | <i>Adult</i> , Systolic: 40-260 mmHg, Diastolic: 20-200 mmHg, MAP 26-<br>220mmHg |
| | <i>Pediatric</i> , Systolic: 40-230 mmHg, Diastolic: 20-160 mmHg, MAP<br>26-183 mmHg |
| | <i>Neonatal</i> , Systolic: 40-130 mmHg, Diastolic: 20-100 mmHg, MAP<br>26-110mmHg |
| Display resolution | SpO2: 1% |
| | PR: 1 bpm |
| | RRa: 1breath per minute |
| | SpCO: 1% |
| | SpMet: 0.1% |
| | SpHb: 0.1 g/dL |
| | SpOC: 0.1 ml/dL |
| Accuracy (ARMS)* | Masimo Rainbow SET/ Masimo SET Parameters |
| SpO2, no motion | 70-100%, 1.5%, adults/pediatrics/infants/neonates ** |
| SpO2, motion | 70-100%, 1.5% adults/pediatrics/infants/neonates ** |
| Feature | Rad-97 Specification |
| SpO2, low perfusion | 70-100%, 2%, adults/pediatrics/infants/neonates |
| Pulse rate, no motion | 25-240 bpm, 3 bpm, adults/pediatrics/infants/neonates |
| Pulse rate, motion | 25-240 bpm, 5 bpm, adults/pediatrics/infants/neonates |
| Pulse rate, low perfusion | 25-240 bpm, 3 bpm, adults/pediatrics/infants/neonates |
| RRa | 4-70 breaths per minute, 1 breath per minute, adults/pediatrics |
| SpCO | 1-40%, 3%, adults/pediatrics/infants |
| SpMet | 1-15%, 1%, adults/pediatrics/infants/neonates |
| SpHb | 8-17 g/dL, 1 g/dL, adults/pediatrics |
| Accuracy | Optional Nomoline Capnography or Optional NIBP Parameters |
| CO2 | Single dry gasses at 22+ 5°C and 1013 + hPa:<br>0-15 volume % +0.2 volume% +2% or reading<br>All conditions: 0.3 kPa + 4% of reading |
| Respiration rate | 0-150 breaths/min + 1 breaths/min |
| NIBP | 0-300 mmHg, +3 mmHg |
| Displays/ Indicators | |
| Data displayed | SpO2, PR, SpCO, SpMet, SpHb, Rra, Pi, PVi, SpOC, NIBP, CO2,<br>pleth waveform, alarm status, status messages, sensor status, Signal<br>IQ, APOD, FastSat |
| Alarm | Alarm limits (audible and visual), 3D alarm, alarm silence, sensor<br>condition alarms, system failure alarms, low battery alarms |
| Averaging time | SpO2: 2-4, 4-6, 8, 10, 12, 14 and 16 seconds;<br>RRa: Trending, No Averaging, Fast, Medium, or Slow<br>PVi: Short or Long<br>Pi: Short or Long<br>SpHb: Short, Medium, or Long |
| Sensitivity mode | Normal, APOD, and MAX |
| Sleep study mode | 10 seconds |
| Electrical | |
| AC power | Input voltage: 100-240 VAC, 47-63 Hz; input power: 60 VA |
| Internal battery power | Rechargeable lithium ion battery |
| Input/Output Interface | |
| Patient applied part connection | M-20 connector<br>Nomoline Capnography input connector<br>NIBP Nib |
| Analog output | Nurse call output |
| USB interface | SatShare module and external devices (e.g. barcode reader/scanner)<br>connection |
| Wired (Ethernet) and wireless (Wi-<br>Fi/Bluetooth) interface | • Wired/wireless connection with networked Patient SafetyNet<br>and/or EMR systems<br>• Wireless connection with presence tag, Kite, external devices<br>(e.g. Caregiver thermometer) |
| Feature | Rad-97 Specification |
| Audio | Microphone supporting video/audio modes for surveillance/conferencing capabilities with Patient SafetyNet |
| Mechanical | |
| Enclosure Material | Thermoplastic |
| Dimensions | 9×4×6.5 inch (22.9×10.2×16.5cm) |
| Weight | 0.92 kg. (2.03 lbs.) without Capnography and NIBP |
| Environmental | |
| Operating Temperature | 0 to 35 °C (32 to 95 °F) |
| Storage/Transport Temperature | -20 to 60 °C (-4 to 140 °F) |
| Operating Humidity<br>Non NiBP/Nomoline Capnography | 10% to 95%, non-condensing |
| Operating Humidity<br>NiBP/Nomoline Capnography | 15% to 95%, non-condensing |
| Storage/Transport Humidity<br>Non NiBP/Nomoline Capnography | 10% to 90%, non-condensing |
| NiBP/Nomoline Capnography | 15% to 90% non-condensing |
| Operating Atmospheric Pressure<br>Non-NiBP/Nomoline<br>Capnography | 540 mbar to 1,060 mbar (540 hPa to 1060 hPa) |
| Nomoline Capnography | 525 mbar to 1,200 mbar (525 hPa to 1200 hPa) (corresponding to a max altitude of 4572 m / 15 000 feet) |
| Compliance | |
| Electrical Safety/EMC | IEC 60601 compliant |
| Type of Protection | Class I, when on AC power; internally powered, when on battery power |
| Degree of Protection | Defibrillation proof, BF-applied part |
| Protection against harm from<br>liquid ingress | IPX1, protection vertically falling water drops<br>IPX2, protection against falling water drops when enclosure is tilted at 15 degrees |
| Mode of Operation | Continuous operation |
The specifications for Rad-97 are as follows:
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of the word "Masimo" in bold, black letters, with a red circle containing a white checkmark to the left of the word. To the right of the logo is the text "MASIMO CORPORATION" followed by the address "52 Discovery, Irvine, CA 92618".
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white checkmark inside, followed by the word "Masimo" in black, bold letters. To the right of the word "Masimo" is the text "MASIMO CORPORATION" followed by the address "52 Discovery, Irvine, CA 92618".
*Аямs ассигасу is a statistical calculation of the difference between device measurements and reference measurements. Approximately two-thirds of the device measurements fell within +/-ARMs of the reference measurements in a controlled study.
** Applicable to compatible sensors. Arws validated in healthy adult volunteer subjects. See Section 5.5.5 for real world performance data collected in neonatal patients.
Note: RD SET Disposable sensors are same as RD SET Adhesive sensors. In order to be consistent with the clearance obtained under K180046 and also the current branding (Directions for Use) for the sensors,
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of the word "Masimo" in a bold, sans-serif font, with a red circle containing a white checkmark to the left of the word. To the right of the logo is the text "MASIMO CORPORATION" in a smaller, sans-serif font, followed by the address "52 Discovery, Irvine, CA 92618".
510k summary has been revised to reflect the subject of the submission as RD SET Disposable sensors in place of RD SET Adhesive sensors.
### 5.2. Intended Use/ Indications for Use
The Masimo Rad-97 and Accessories can communicate with network systems for supplemental remote viewing and alarming (e.g., at a central station). In addition, the Masimo Rad-97 and Accessories are indicated to provide the continuous non-invasive monitoring data obtained from the Masimo Rad-97 and Accessories for functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) to multiparameter devices for the display on those devices.
The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) of adult, pediatric, and neonatal patients during both no motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments.
The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of carbox yhemoglobin saturation (SpCO) of adult, pediatric, and infant patients during no motion conditions in hospitals and hospital-type facilities. The Masimo Rad-97 and Accessories are not intended to be used as the sole basis for making diagnosis or treatment decisions related to suspected carbon monoxide poisoning; it is intended to be used in conjunction with additional methods of assessing clinical signs and symptoms.
The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of methemoglobin saturation (SpMet) of adult, pediatric, and neonatal patients during no motion conditions in hospitals and hospital-type facilities.
The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of total hemoglobin concentration (SpHb) of adult and pediatric patients during no motion conditions in hospitals and hospital-type facilities.
The Masimo Rad-97 and Accessories are indicated for the continuous non-invasive monitoring of respiratory rate (RRa) for adult, pediatric, and neonatal patients during no motion conditions in hospitals, hospital-type facilities, home environments, and transport within healthcare facilities.
The optional Nomoline Capnography product family is intended to be connected to other medical backboard devices for monitoring of breath rate and CO2. The Nomoline Capnography product family is intended to be connected to a patient breathing circuit for monitoring of inspired/expired gases during anesthesia, recovery and respiratory care. The environment is the operating suite, intensive care unit and patient room. The intended patient population is adult, pediatric and infant patients.
The optional non-invasive blood pressure (NIBP) module is indicated for the noninvasive measurement of arterial blood pressure in hospital-type facilities, mobile, and home environments. The NIBP module is designed to measure blood pressure for patient population described in the following table:
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white checkmark inside, followed by the word "Masimo" in a bold, sans-serif font. To the right of the word "Masimo" is the text "MASIMO CORPORATION" in a smaller, lighter font, followed by the address "52 Discovery, Irvine, CA 92618".
| Patient Population | Approximate Age Range |
|--------------------|---------------------------|
| Newborn (neonate) | Birth to 1 month of age |
| Infant | 1 month to 2 years of age |
| Child | 2 to 12 years of age |
| Adolescent | 12-21 years of age |
| Adult | 21 years of age and older |
## 5.3. Technological Characteristics
## Principle of Operation
The principle of operations for the measurement capabilities of the Rad-97 product family does not change from the predicate (K180046) for the update to the RD SET Disposable sensor specifications for the neonate population. The principles of operation are still based upon the use of red and infrared wavelengths of light provided by an emitter and the detection of the signal from the light absorption of oxygenated blood and deoxygenated blood to provide functional oxygen saturation of hemoglobin (SpO2).
## Mechanism of Action for Achieving the Intended Effect
The mechanism of action did not change from the predicate. The sensor is applied in the same way and to the same measurement sites as the predicate.
## 5.4. Summary of Technological Characteristics of Subject Device Compared to Predicate Device
The only difference between the subject device and the predicate is the SpO2 performance specification is updated to match that as it was previously cleared for the adult population as part of the predicate. There are no other difference between the predicate RD SET Disposable sensors to that of the subject device.
The subject device, RD SET Adhesive sensors with updated SpO2 specifications for the neonate population and the predicate device, have the following key similarities:
- · Both devices have the same indicated populations;
- · Both sensors have the same form factor and components;
- · Both sensors have the same measurement application sites
The subject device, RD SET Adhesive sensor with update SpO2 performance specifications for the neonate population and the predicate device, have the following difference:
- · Subject device has an updated SpO2 performance specification of 1.5% Arms. The SpO2 specification is being updated to match the adult population SpO2 performance specification previously cleared as part of the predicate.
#### 5.5. Performance Data
As part of this submission, there were no physical changes to the RD SET Disposable sensor. The
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white checkmark inside, followed by the word "Masimo" in large, bold black letters. To the right of the word "Masimo" is the text "MASIMO CORPORATION" in smaller, gray letters, followed by the address "52 Discovery, Irvine, CA 92618".
changes was to the update of the labeling to align the performance specification for the neonate population to that of the adult population.
#### Biocompatibility Testing 5.5.1.
As there were no changes associated with this submission, no additional biocompatibility testing was required to support the substantial equivalence.
#### Electrical Safety and EMC 5.5.2.
As there were no hardware changes associated with this submission, no additional electrical safety or EMC testing was required to support the substantial equivalence.
#### 5.5.3. Software Verification and Validation Testing
As there were no software changes associated with this submission, no additional software verification and validation testing was required to support the substantial equivalence.
## 5.5.4. Mechanical Testing
As there were no mechanical changes associated with this submission, no additional mechanical testing was required to support the substantial equivalence.
### 5.5.5. Clinical Testing
As part of the submission, there was no changes made to the devices from the previous clearance in which the accuracy performance of the RD SET Disposable sensors was established to be 1.5% Arms validated on adults. As part of this submission, Masimo performed clinical testing on neonates using convenience arterial blood samples in accordance with the FDA guidance for Pulse Oximeters to verify the form. fit. and function of the RD SET Disposable sensors on neonate population. Although the RD SET Disposable sensors are not new or significantly modified the convenience samples were collected on neonatal population in order to provide greater assurance of the safe form, fit, and function on neonates.
As part of the study, convenience blood samples were collected from an arterial line of 19 neonatal subject's that were ≤ 1 month and weighing ≤ 4.5 kg. The SaO2 measurements from the blood samples using a reference co-oximeter were compared to the SpO2 measurements using a RD SET Disposable sensor. As the convenience blood samples were collected in a clinical environment dependent upon the availability of subjects, the results were subject to greater amounts of testing variability than would be expected in a controlled laboratory setting.
The analysis of the clinical data further supported the safe form, fit, and function of the RD SET Disposable Sensor on neonates. The Arms performance was calculated for reference purposes. See summary of Arms performance in table 5.5.1 below.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for Masimo Corporation. The logo consists of a red circle with a white checkmark inside, followed by the word "Masimo" in black, bold letters. To the right of the logo is the text "MASIMO CORPORATION" in gray, followed by the address "52 Discovery, Irvine, CA 92618" also in gray.
| Table 5.5.1 | | | | |
|-------------|-------|-----------|------|----------|
| SaO2 | Bias | Precision | RMS | Subjects |
| 70-100% | 0.08 | 2.58 | 2.59 | 19 |
| 85-100% | -0.09 | 1.39 | 1.40 | 12 |
| 70-85% | 0.37 | 3.83 | 3.85 | 7 |
To further support the safe form, fit, and function of the RD SET Disposable sensor on neonates, additional convenience arterial blood samples collected on neonates using the subject sensor from prior studies were aggregated with the data from the above study resulting in 70 data pairs collected on 42 subjects (including 22 Pre-term neonates with gestational ages ≤ 37 weeks) from 3 different sites. See summary of the observed results in table 5.5.2 below.
| Table 5.5.2 | | | | | | | |
|-------------|---------------|----------------------|-----------------|---------|-------|-----------|-------|
| Description | Age<br>(days) | Gest. Age<br>(weeks) | No.<br>Subjects | Samples | Bias | Precision | ARMS |
| All Data | 1-31 | 27-41 | 42 | 70 | 0.129 | 3.182 | 3.185 |
The results of the clinical testing supported the subject device did not raise any different questions of safety and effectiveness as compared to the predicate device.
#### 5.6 Conclusion
The data supports the substantial equivalence of the subject device with the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.