The Dolphin Medical 2150 Handheld Pulse Oximeter is indicated for use for spot checking and/or continuous noninvasive monitoring of fractional oxygen saturation of arterial hemoglobin, pulse rate, and patient temperature. The 2150 is for use with adult pediatric and neonatal patients in hospitals and hospital type facilities.
Device Story
Handheld pulse oximeter; utilizes Dolphin ONE OEM 701 module; inputs include light absorption signals from two-LED sensors and temperature data from YSI 400 series thermistor probes. Digital signal processing (128 MHz RISC processor) calculates SpO2, pulse rate, and temperature. Output displayed on LCD; includes infrared printer interface. Used in acute-care settings (OR, Recovery, ICU) by physicians, nurses, or respiratory therapists. Provides real-time patient data for clinical monitoring; supports decision-making regarding oxygenation and thermal status. Benefits include noninvasive, continuous assessment of patient vitals across diverse perfusion and motion conditions.
Clinical Evidence
Clinical validation performed via breathe-down protocol at VA Hospital of Wisconsin. Sample included conscious volunteers with SaO2 ranging 70-100%. Primary endpoint: accuracy (ARMS) compared to functional SaO2. Results: <2.0% ARMS for adults, pediatrics, and infants; <3.5% ARMS for neonates. Testing confirmed performance during motion and no-motion conditions, and across varying perfusion levels.
Technological Characteristics
Handheld form factor; 128 MHz RISC processor; OEM 701 oximetry engine. Interfaces with YSI 400 series thermistor probes. Power: internal rechargeable battery or medical-grade AC supply. Connectivity: infrared printer interface. Complies with EN 60601-1, EN 60601-1-2, ASTM 1415:1992, and E1112-00. Biocompatibility testing performed for prolonged skin contact.
Indications for Use
Indicated for spot checking and continuous noninvasive monitoring of SpO2, pulse rate, and patient temperature in adult, pediatric, and neonatal patients in hospital settings.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
Dolphin Stand Alone Pulse Oximeter and Accessories (K024235)
Dolphin Stand-Alone Pulse Oximeter and Accessories (K020075)
Aristo Medical Model 2101 Patient Monitor (K953817)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the text "K030629" in a handwritten style. The text is oriented diagonally, starting from the bottom left and rising to the upper right. The characters are bold and slightly distorted, giving them a casual, handwritten appearance.
# 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
# Dolphin Medical 2150 Handheld Pulse Oximeter
## Submitting by
Dolphin Medical Inc. 12525 Chadron Avenue Hawthorne, CA 90250
### Submission Correspondent
Bill Curnan Regulatory Specialist 9433 S. Morning Glory Lane Highlands Ranch, CO 80130
Phone: 720-939-6482 Fax: 786-551-8221
## Common, Classification & Proprietary Names
| Common Name: | oximeter |
|----------------------|----------------------------------------------|
| Classification Name: | oximeter |
| Proprietary Name: | Dolphin Medical 2150 Handheld Pulse Oximeter |
# Predicate Devices
The Dolphin Medical 2150 Handheld Pulse Oximeter is substantially equivalent to the following devices:
| K024235 | Dolphin Stand Alone Pulse Oximeter and Accessories |
|---------|----------------------------------------------------|
| K020075 | Dolphin Stand-Alone Pulse Oximeter and Accessories |
| K953817 | Aristo Medical Model 2101 Patient Monitor |
# Device Description
The Dolphin Medical 2150 Pulse Oximeter consists of the Dolphin ONE™ OEM 701 Module technology, and works with existing Dolphin ONE extension cables, and oximetry sensors to non-invasively calculate the functional saturation of arterial hemoglobin (SpO2) and pulse rate. It also includes a thermistor port which accepts off the shelf approved YSI 400 series patient temperature sensors for the reading of patient temperature. It features an easy-to-read display that presents patient data and status information: an LCD (liquid crystal display) display that shows the SpO2, pulse rate values, patient temperature, and other messages as appropriate.
{1}------------------------------------------------
A Dolphin ONE Extension Cable connects the sensor to the 2150. The cable is available in two configurations, three foot or eight in length.
The oximeter can be operated off either an internal rechargeable battery or with use of a medical grade AC power supply which is fumished with the unit.
The Dolphin Medical Model 2150 Pulse Oximeter will be used for continuous noninvasive monitoring and spot checking of SpO2, pulse rate, and patient temperature for adult, pediatric, infant and neonatal patients in hospital-type facilities. Typical acute-care uses are the Operating Room, Recovery Room and the Adult and Neonatal Intensive Care Units. The Dolphin Medical Model 2150 Pulse Oximeter and accessories will be used for adult, pediatric and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused. The user responsible for the operation of the Dolphin Medical Model 2150 will normally be a licensed clinical professional such as a physician, nurse, or respiratory therapist.
## Intended Use
The Dolphin Medical 2150 Handheld Pulse Oximeter is indicated for use for spot checking and/or continuous noninvasive monitoring of fractional oxygen saturation of arterial hemoglobin, pulse rate, and patient temperature. The 2150 is for use with adult pediatric and neonatal patients in hospitals and hospital type facilities.
### Technological Characteristics Comparison
The Dolphin 2150Handheld Pulse is a redesign of the Dolphin 2100 Stand Alone Pulse Oximeter (K002036) into a handheld configuration with the addition of a temperature sensor port.
The Dolphin Model 2150 Pulse Oximeter incorporates the Dolphin OEM 701 Oximeter Engine in a handheld Oximeter with patient temperature monitoring capability and an infra red printer interface.
Dolphin Medicals OEM 701 Module is a complete pulse oximeter engine, based on a patented Digital Signal Processing technology with a high speed 128 MHz RISC processor. This circuit board interfaces directly with the Dolphin ONE™ family of sensors and the host platform to calculate the SpO2 and pulse rate. It provides the required sensor drive currents, accepts the sensor data, runs the algorithms to calculate SPO2. Pulse Rate, and patient temperature and provides a digital output to the host unit for display. Additionally, the board contains all necessary support for direct connection to off the shelf YSI 400 series thermistor patient temperature probes.
The OEM 701 is a redesign of the OEM 601 which is the oximetry engine for the Dolphin Model 2100(k002036).
The Dolphin ONE Sensors measure light absorption of blood from two light emitting diodes. Oxygen saturated blood absorbs light differently as compared to unsaturated blood. The amount of light absorbed by the blood is used to calculate the ratio of oxygenated hemoglobin to total hemoglobin in arterial blood.
{2}------------------------------------------------
# Performance Testing
#### Biocompatibility t
Biocompatibility tests, appropriate for skin-contacting devices for prolonged exposure, have been performed on all Dolphin ONE sensors
#### Electrical Safety 트
The Dolphin Medical 2150 Handheld Pulse Oximeter was designed to comply with the following standards:
- 1. CSA C22.2 No. 601.1
- 2. EN 60601-1, and Amendments 1 and 2
- 3. EN 60601-1-2, 2002
- 4. EN 475 1995
- 5. EN 865: 1997
- 6. FDA Guidance Document for Pulse Oximeters: 977/1992
- 7. ASTM 1415:1992.
- 8. UL2601-1: Second Edition, 1997
- 9. IEC 6068-2-6
- 10. IEC 6068-2-27
- 11. IEC 6068-2-64
- 12. ISTA Procedure 2A
- 13. E1112-00
# Clinical Testing
The 2150 was validated using Dolphin ONE Sensors in breathe-down protocols at the VA Hospital of Wisconsin - Milwaukee, (Dr. Phillip Clifford, MD.). Scientific accuracy was demonstrated by statistically comparing Dolphin 2150 SpO2 values to functional SaO2 values. Volunteers participated in the breathe-down protocol while fully conscious at SaO2 values ranging from 70-100%. Data was analyzed to determine the ARMS for each probe. Clinical Validation for the Dolphin ONE Reusable, Adult disposable, Neonatal disposable resulted in an accuracy determination of less then 2.0% Arws in the range of 70-100% Sa02 for adults, pediatrics, and infants and less than 3.5% Arms in the range of 70-100 for Neonates.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name encircling a stylized eagle-like emblem. The emblem consists of three abstract human profiles facing right, with flowing lines beneath them, possibly representing water or movement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP - 4 2003
Mr. Bill Curnan Regulatory Specialist Dolphin Medical Incorporated 9433 S. Morning Glory Lane Highlands Ranch, Colorado 80130
Re: K030629
Trade/Device Name: Dolphin ONE 2150 Pulse Oximeter Regulation Number: 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: June 17, 2003 Received: June 18, 2003
Dear Mr. Curnan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
#### Page 2 - Mr. Curnan
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Susan Runner, DDS, MA
Interim Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### 4. STATEMENT OF INDICATIONS FOR USE (FDA FORM)
510(k):
KO30629
Device:
Dolphin ONE 2150 Pulse Oximeter
# Indications for Use:
The Dolphin Medical 2150 Handheld Pulse Oximeter is indicated for use for spot checking and/or continuous noninvasive monitoring of fractional oxygen saturation of arterial hemoglobin, pulse rate, and patient temperature. The 2150 is for use with adult pediatric and neonatal patients in hospitals and hospital type facilities.
(Division Sign-Off)
Division of Anesthesiology, General Hospital,
Infection Control, Dental Devices
510(k) Number:_
Prescription device \/
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.