K960972 · Marquest Medical Products, Inc. · CCK · Oct 18, 1996 · Anesthesiology
Device Facts
Record ID
K960972
Device Name
DISPOSABLE GAS SAMPLING INTERFACE KITS
Applicant
Marquest Medical Products, Inc.
Product Code
CCK · Anesthesiology
Decision Date
Oct 18, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1400
Device Class
Class 2
Intended Use
Indicated use - When a sample of a patient's expired gas is to be analyzed. Targeted population - Any patient requiring monitoring of expired gases. Environment of use - Hospital, anesthesia, ICU, PACU, respiratory therapy
Device Story
Small-bore tubing with luer fittings; facilitates transport of patient expired gas samples to external gas analyzers. Used in hospital environments (anesthesia, ICU, PACU, respiratory therapy) by clinicians. Device features various lengths, diameters, and optional in-line filters. Single-use disposable design.
Clinical Evidence
Bench testing only; performance testing of luer fittings confirms equivalence to predicate.
Technological Characteristics
Small-bore tubing with luer fittings; materials include PVC, PC, ABS, Acrylic, and PE. Optional in-line filter. Single-use, disposable.
Indications for Use
Indicated for patients requiring expired gas monitoring in hospital settings including anesthesia, ICU, PACU, and respiratory therapy.
Regulatory Classification
Identification
A carbon dioxide gas analyzer is a device intended to measure the concentration of carbon dioxide in a gas mixture to aid in determining the patient's ventilatory, circulatory, and metabolic status. The device may use techniques such as chemical titration, absorption of infrared radiation, gas chromatography, or mass spectrometry.
Predicate Devices
Gibeck, Inc. disposable gas sampling lines (K841722)
Related Devices
K133516 — CO2 MONITORING LINE, CO2 MONITORING LINE WITH FILTER · Flexicare · Apr 30, 2014
K251216 — 0184 CO2 Sampling line; 0184T CO2 Sampling line; 0139 CO2 Sampling line; 0182 CO2 Sampling line · Barbaras Development, Inc. · Jun 4, 2025
K171678 — Gas Sampling Line, Coextruded, 2m/7ft, 10/pkg, Disposable/ 100/pkg, Disposable, Gas Sampling Line, Coextruded, 2.5m/8ft, 10/pkg, Disposable, Gas Sampling Line, Coextruded, 3m/10ft, 10/pkg, Disposable/ 100/pkg, Disposable, Gas Sampling Line, Coextruded, 6m/20ft, 5/pkg, Disposable · Vyaire Medical, Inc. · Apr 30, 2018
K143150 — CO2 monitoring line with and without in-line filter · Besmed Health Business Corp · Feb 5, 2015
K023579 — GAS SAMPLING LINES · Promedic, Inc. · Jan 28, 2003
Submission Summary (Full Text)
{0}
The New Marq of Excellence
K900972
OCT 18 1996
Marquest Medical Products, Inc.
11039 East Lansing Circle
Englewood, Colorado 80112 U.S.A.
303 790 4835 phone
303 799 0210 fax
Non-Confidential Summary of Safety and Effectiveness
March 1, 1996
page 1 of 2
Marquest Medical Products, Inc.
11039 East Lansing Circle
Englewood, CO 80112
Tel - (303) 790-4835
Fax - (303) 799-0210
Official contact: Tom Dielmann, Director, Regulatory Affairs / Quality Assurance
Proprietary or Trade Name: Gas Sampling Interface Kits
Common/Usual Name: Disposable Gas Sampling Lines
Classification Name: Analyzer, Gas, Carbon Dioxide, Gaseous Phase
Intended device: Gas Sampling Interface Kits
Predicate devices: Gibeck, Inc. - disposable gas sampling lines - K841722
Device description: The intended product is a small bore tube with luer fittings which permit a sample of patient's expired gas to be delivered to an analyzer. The device may be of various lengths, tubing diameters and have an in-line filter.
Intended use:
Indicated use - When a sample of a patient's expired gas is to be analyzed.
Targeted population - Any patient requiring monitoring of expired gases.
Environment of use - Hospital, anesthesia, ICU, PACU, respiratory therapy
Comparison to predicate devices:
| Attribute | Marquest | Dryden (Gibeck) |
| --- | --- | --- |
| Use | | |
| Indicated for use with expired gas sampling monitors | Yes | Yes |
page 3 of 21
{1}
Non-Confidential Summary of Safety and Effectiveness (continued)
March 1, 1996
page 2 of 2
Comparison to predicate devices: (continued)
| Attribute | Marquest | Dryden (Gibeck) |
| --- | --- | --- |
| Use (continued) | | |
| Intended for use in anesthesia | Yes | Yes |
| Intended for use in respiratory therapy | Yes | Yes |
| Indicated for single use | Yes | Yes |
| Design | | |
| Luer fittings | Yes | Yes |
| Clear | Yes | Yes |
| Various diameters of tubing | Yes | Yes |
| Various lengths of tubing | Yes | Yes |
| Maybe supplied with filter | Yes | Yes |
| Materials | | |
| Luers made of PVC, PC, ABS, Acrylic | Yes | Yes |
| Tubing - PVC, PE, PVC/PE | Yes | Yes |
| Performance Testing | | |
| Luer fittings | Yes | Yes |
## B. Differences
Any differences that do exist would not have a significant effect on the safety or effectiveness of the device.
page 4 of 21
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.