The SpiroSphere is a compact device to measure inspiratory and expiratory lung function parameters in adults and children aged 4 years and older. It can be used by physicians in the office or hospital.
Device Story
SpiroSphere is a compact, battery-powered spirometry system consisting of a main unit and a sensor unit; used by physicians in clinical or hospital settings. Device measures inspiratory and expiratory lung function parameters (FEV1, FVC, PEF, FEF25-75, VC, IC, ERV) using a pneumotach transducer (Lilly type). Inputs are patient respiratory flows/volumes; device performs pressure-to-flow conversion and calculates pulmonary parameters per 2005 ATS/ERS standards. Outputs include numerical and graphical data displayed on a high-resolution color touch screen; data can be stored, printed via external printer, or transferred via USB, Wi-Fi, 3G, or Ethernet. Device aids clinicians in assessing lung function and monitoring respiratory health.
Clinical Evidence
Bench testing only. No clinical data presented. Performance verified against 2005 ATS/ERS spirometry standards for accuracy and precision. Software verified/validated per IEC 62304; electrical safety per IEC 60601-1; EMC per IEC 60601-1-2; biocompatibility per ISO 10993-1 and ISO 18562; usability per IEC 62366.
Technological Characteristics
Pneumotachograph (Lilly type) sensing; pressure-to-flow conversion. Main unit (battery/power supply) and sensor unit (battery). Connectivity: Bluetooth (Silicon Labs BT121), Wi-Fi (TI WL18xxMOD), 3G (Sierra Wireless HL 8548), Ethernet, USB. High-resolution color touch screen. Complies with 2005 ATS/ERS standards. Software developed per IEC 62304. Biocompatibility per ISO 10993-1/ISO 18562. Sterilization/reprocessing per 2015 FDA guidance.
Indications for Use
Indicated for measuring inspiratory and expiratory lung function parameters in adults and children aged 4 years and older. Used by physicians in office or hospital settings.
Regulatory Classification
Identification
A predictive pulmonary-function value calculator is a device used to calculate normal pulmonary-function values based on empirical equations.
K160253 — Vitalograph Model 6300 micro · Vitalograph (Ireland) , Ltd. · Aug 26, 2016
K022276 — KOKOMATE · Ferraris Respiratory, Inc. · Oct 4, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font size below.
February 15, 2019
eReserchTechnology GmbH % Dawn Tibodeau Third Party 510(k) Project Coordinator Tuv Sud America Inc. 1775 Old Highway 8 Nw New Brighton, Minnesota 55112-1891
Re: K173937
Trade/Device Name: SpiroSphere Regulation Number: 21 CFR 868.1890 Regulation Name: Predictive pulmonary-function value calculator Regulatory Class: Class II Product Code: BTY Dated: December 21, 2018 Received: December 31, 2018
Dear Dawn Tibodeau:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# James J. Lee -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K173937
Device Name SpiroSphere
Indications for Use (Describe)
The SpiroSphere is a compact device to measure inspiratory lung function parameters in adults and children aged 4 years and older.
It can be used by physicians in the office or hospital.
Tyne of Use (Select one or hoth, as annlicable)
| | <span style="font-size:10px">Type of Use (Select one or both, as applicable)</span> |
|--|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | <span> <span style="font-size:10px">☒</span> <span style="font-size:10px">Prescription Use (Part 21 CFR 801 Subpart D)</span> </span> <span> <span style="font-size:10px">☐</span> <span style="font-size:10px">Over-The-Counter Use (21 CFR 801 Subpart C)</span> </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the ERT logo. The logo consists of a circular graphic with three overlapping shapes in green and blue, followed by the letters "ERT" in blue. Below the letters is the text "Confidence at every turn" in a smaller font.
# GENERAL INFORMATION
| 1 | Type of Submission | Traditional 510(k) Submission/Third Party Review |
|---|--------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
| | Submission date: | 07/28/2017 |
| 2 | Submitter | |
| | Name: | eResearch Technology GmbH |
| | Address: | Sieboldstrasse3<br>D-97230 Estenfeld<br>Germany |
| | Contact person in Germany:<br>(Official Correspondent) | Johannes Thalhammer |
| | Address: | eResearch Technology GmbH<br>Sieboldstrasse 3, 97230 Estenfeld<br>Germany |
| | Phone:<br>FAX:<br>E-mail | +49 9305 720 6376<br>+49 9305 720 60<br><a href="mailto:johannes.thalhammer@ert.com">johannes.thalhammer@ert.com</a> |
| | Contact person in the U.S.: | Richard Miller |
| | Address | eResearch Technology<br>1818 Market Street, Suite 1000<br>Philadelphia, PA 19103 |
| | Phone/Fax:<br>E-mail | 215-2825588<br><a href="mailto:RMiller@ert.com">RMiller@ert.com</a> |
| 3 | Establishment Registration Number | 3008505660 |
| 4 | Common Name or Classification Name | Predicted pulmonary-function value calculator<br>(CFR 868.1890, Product Code BTY) |
| 5 | Trade Name | SpiroSphere |
| 6 | Device Classification | This is a Class II device |
| 7 | Classification Panel | 73 Anesthesiology Part 868 Code BTY |
{4}------------------------------------------------
#### 005 510(k) Summary
Image /page/4/Picture/1 description: The image is a logo for ERT. The logo consists of a circular graphic on the left and the text "ERT" on the right. The circular graphic is divided into three sections, with the top section being light green, the bottom left section being light blue, and the bottom right section being dark blue. The text "ERT" is in dark blue, and below it is the tagline "Confidence at every turn" in a smaller font.
#### 8 Reason for Premarket Notification
Introducing a (finished) device into commercial distribution (marketing) in the U.S. for the first time.
#### 9 Legally predicate marketed device
SpiroPro K000648 Code BTY
Reference Device SpiroPro with BT Code BTY, Code DQA K092324
#### Predicate Device Company 10
eResearchTechnology GmbH (device was formally listed by Erich Jaeger GmbH and Carefusion Germany 234 GmbH)
#### 11 Device Description
The SpiroSphere is a compact spirometry device. It's Sensor Unit is battery-powered. The Main Unit can be powered by battery or power supply. The SpiroSphere is used to measure inspiratory and expiratory lung function parameters in adults and children 4 years and older. The measured data is saved into the device and can be read out at any time.
A printer can be connected with the SpiroSphere and all needed data (e.g. reports, Screenshots) can be printed. Moreover it is possible to transfer data by USB, Wifi, 3G, and Ethernet.
Pulmonary functions
- Slow Spirometry
- Forced Spirometry ●
- Flow-Volume loop and Volume-Time tracing, pre/post tests ●
- Trending capabilities
#### 12 Indications for Use
The SpiroSphere is a compact device to measure inspiratory and expiratory lung function parameters in adults and children aged 4 years and older.
It can be used by physicians in the office or hospital.
#### 13 Required Components
Power Supply (medical grade) Pneumotach with Mouthpiece (ERT PT)
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the ERT logo. The logo consists of a circular graphic on the left and the text "ERT" on the right. The circular graphic is made up of three curved shapes in green and blue. Below the text "ERT" is the tagline "Confidence at every turn".
Instruction for Use NoseClips
#### Summary Table of Comparison 14
Comparison with SpiroPro (K000648) and SpiroPro with BT (K092324)
| | Predicate Device<br>SpiroPro (K000648) | Reference Device<br>SpiroPro with BT<br>(K092324) | SpiroSphere |
|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use /<br>Indications for<br>Use | The SpiroPro is a portable,<br>battery operated device<br>and can be used by<br>physicians in the office or<br>hospital, in occupational<br>medicine or by patients at<br>home.<br>The SpiroPro measures<br>inspiratory and expiratory<br>lung function parameter in<br>adults and children from 4<br>years on according the<br>ATS recommendations for<br>diagnostic devices.<br>Results are displayed<br>graphical and numerical on<br>the display of the device.<br>Optional the results can be<br>printed on an external<br>printer or transferred to a<br>PC. SpiroPro can save all<br>data in the internal non-<br>volatile memory for later<br>retrieval, print-out or<br>transfer to a PC. | The SpiroPro is a portable,<br>battery operated device<br>and can be used by<br>physicians in the office or<br>hospital, in occupational<br>medicine or by patients in<br>the home. The SpiroPro<br>measures inspiratory and<br>expiratory lung function<br>parameters in adults and<br>children 4 years and older.<br>In addition to the<br>pulmonary function<br>measurements, oxygen<br>saturation and heart rate<br>can be recorded. | The SpiroSphere is a<br>compact device to<br>measure inspiratory and<br>expiratory lung function<br>parameters in adults and<br>children aged 4 years and<br>older.<br>It can be used by<br>physicians in the office or<br>hospital. |
| Patient<br>population | Adults and children 4<br>years and older | Adults and children 4 years<br>and older | Adults and children 4<br>years and older |
| Operation<br>principle | - Measurement of<br>inspiratory and<br>expiratory flows and<br>volumes with<br>pneumotach transducer<br>- Calculation of lung<br>function parameters<br>- Results are displayed<br>and stored, they can be<br>printed and exported | - Measurement of<br>inspiratory and<br>expiratory flows and<br>volumes with<br>pneumotach transducer<br>- Calculation of lung<br>function parameters<br>- Results are displayed<br>and stored, they can be<br>printed and exported | - Measurement of<br>inspiratory and<br>expiratory flows and<br>volumes with<br>pneumotach transducer<br>- Calculation of lung<br>function parameters<br>- Results are displayed<br>and stored, they can be<br>printed and exported |
| Measurements | FEV1, FVC, PEF, FEF25-<br>75, VC, IC, ERV and | FEV1, FVC, PEF, FEF25-<br>75, VC, IC, ERV and | FEV1, FVC, PEF, FEF25-<br>75, VC, IC, ERV and |
| | others acc. to ATS* | others acc. to ATS* | others acc. to ATS* |
| *ATS conformity<br>(criteria) | 1994 ATS/ERS Spirometry<br>Standards | 2005 ATS/ERS Spirometry<br>Standards | 2005 ATS/ERS<br>Spirometry Standards |
| Fundamental<br>scientific<br>technology | Pneumotachograph,<br>pressure to flow<br>conversion technique | Pneumotachograph,<br>pressure to flow<br>conversion technique | Pneumotachograph,<br>pressure to flow<br>conversion technique |
| | (Lilly Type<br>Pneumotachograph) | (Lilly Type<br>Pneumotachograph) | (Lilly Type<br>Pneumotachograph) |
| | Mouthpiece (single patient<br>use) | Mouthpiece (single patient<br>use) | Mouthpiece (single patient<br>use) |
| Components | Pneumotach (single<br>patient use) | Pneumotach (single<br>patient use) | Pneumotach (single<br>patient use) |
| | Medical Grade Power<br>Supply | Medical Grade Power<br>Supply | Medical Grade Power<br>Supply |
| Screen Display | 54 x 35 mm | 54 x 35 mm | 162 x 122 mm |
| Interface | RS232 Interface | USB, RS232 Interface and<br>Bluetooth interface | Bluetooth<br>USB<br>Wifi/Ethernet<br>3G |
| Energy type | Li-Ion Battery | Li-Ion Battery | Main Unit:<br>Li-Ion Battery and Power<br>Supply<br>Sensor Unit:<br>Unit: Li-Ion Battery |
| Bluetooth<br>interface | None | WML-C46 (Mitsumi) | Sensor Unit:<br>Silicon Laboratories,<br>BT121 |
| Wi-Fi/BT interface | None | None | Main Unit:<br>Texas Instruments<br>WL18xxMOD |
| Ethernet | None | None | 10/100 MBit Ethernet |
| 3G Interface | None | None | Sierra Wireless HL 8548 |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image is a logo for ERT. The logo consists of a circular design with three sections in green and blue, along with the text "ERT" in a bold, blue font. Below the text, there is a tagline that reads "Confidence at every turn" in a smaller font.
#### 15 Comparison Summary
## Intended Use
The intended use of the SpiroSphere device is the same as in the predicate device SpiroPro.
# Technological Characteristics
The predicate device SpiroPro and the reference device SpiroPro with BT are presently in commercial distribution globally including the United States of America. The SpiroSphere has the same fundamental scientific
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the ERT logo. The logo consists of a circular graphic on the left and the text "ERT" on the right. The circular graphic is made up of three curved shapes in green and blue. Below the text "ERT" is the tagline "Confidence at every turn".
technology and is substantial equivalent in function and application to the predicate device.
# Biocompatibility
No new issues of biocompatibility have been raised with regard to the SpiroPro.
# Differences
- The SpiroSphere consists of a main unit for display and a sensor ● unit for measurement.
- The SpiroSphere provides a high resolution, color touch screen instead of a high-resolution graphical LCD touch screen
- . The SpiroSphere provides additional state of the art interfaces (Bluetooth, 3G, Wi-Fi, Ethernet, USB) to enable communication.
- The SpiroSphere provides additional Predicted Values (GLI 2012)
- SpiroSphere complies to the 2005 ATS/ERS Spirometry Standards ●
#### 16 Summary of Device Testing
The following practices were followed and monitored for development of the SpiroSphere:
- The device was developed and tested according to GMP Standard Operating Procedures for Medical Devices.
- Software verification and validation was done in accordance with IEC 62304:2006.
- Risk analysis of the SpiroSphere was performed according to ISO ● 14971:2007.
- Tests were performed to confirm that the SpiroSphere meets the . recommendations for accuracy and precision for Spirometry of the American Thoracic Society (ATS) according to ATS/ERS standards 2005.
- . The electrical safety testing was performed according to IEC 60601-1:2012 to demonstrate conformance with the requirements for basic safety and essential performance.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the ERT logo. The logo consists of a circular graphic on the left and the text "ERT" on the right. The circular graphic is made up of three curved shapes in green and blue. Below the text "ERT" is the tagline "Confidence at every turn".
- . The Electro Magnetic Compatibility testing was performed according to IEC 60601-1-2:2014.
- The FDA Guidance "Radio Frequency Wireless Technology in . Medical Devices" from 2013 was considered for the 3G/BT/WIFI functions and all requirements are fulfilled.
- . Human Factors/Usability Engineering validation was done according to IEC 62366 Edition 1.1: 2014.
- . Biocompatibility was evaluated in accordance with ISO 10993-1:2009 and ISO 18562:2017
- The FDA Guidance "Content of Premarket Submissions for . Management of Cybersecurity in Medical Devices" from 2014 has been considered in the device design and all requirements are fulfilled.
- . Final Guidance for Industry and FDA Staff: Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling from 2015 has been considered.
#### 17 Conclusion
Based on the intended use of the SpiroSphere and the results of the electronic safety testing and performance testing provided in the 510(k), the SpiroSphere is found to be substantially equivalent to the predicate device SpiroPro.
The non-clinical data as well as the hardware and software verification and validation demonstrate that the SpiroSphere device should perform as intended in the specified use conditions.
eResearchTechnology GmbH believes that the SpiroSphere is substantially equivalent to the current legally marketed device SpiroPro. The SpiroSphere has the same technological characteristics and was determined to be as safe and as effective as the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.