K183089 · Nuvoair AB · BZG · Jan 2, 2020 · Anesthesiology
Device Facts
Record ID
K183089
Device Name
Air Next
Applicant
Nuvoair AB
Product Code
BZG · Anesthesiology
Decision Date
Jan 2, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1840
Device Class
Class 2
Attributes
AI/ML, Pediatric
Indications for Use
Air Next is intended to be used by: Healthcare professionals trained to perform spirometry tests on patients of age ≥ 5 years old, > 10 kg and ≥ 110 cm. Air Next is intended to perform basic lung function and spirometry testing. The actual diagnosis shall be done by a healthcare professional in hospital and clinical settings.
Device Story
Hand-held spirometer (75g) measures lung function via disposable turbine; user exhales into turbine; airflow rotates propeller; digital infrared interruption sensor counts rotations; firmware converts rotations to airflow (L/s). Device connects via Bluetooth to mobile application (iOS/Android) for display of flow-time curves, effort, and test quality feedback. Used by trained healthcare professionals in clinical settings to assist in diagnosis of asthma, COPD, and cystic fibrosis. Benefits include portable, accurate lung function assessment compliant with ATS/ERS guidelines.
Clinical Evidence
No clinical data. Bench testing only. Device performance validated against ATS/ERS 2005 guidelines for spirometry accuracy (volume accuracy ±3% or ±0.050 L; flow accuracy ±5% or 200 mL/s). Electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and biocompatibility (ISO 10993-5, -10) testing passed.
Technological Characteristics
Hand-held spirometer (98x62x26 mm, 75g). Powered by 2x AAA 1.5V alkaline batteries. Sensing: Bidirectional turbine with infrared interruption. Connectivity: Bluetooth. Software: Mobile app-based display. Biocompatibility: ISO 10993-1 compliant. Safety: IEC 60601-1, IEC 60601-1-2, IEC 62366. Sterilization: Disposable turbine (single-use).
Indications for Use
Indicated for patients ≥ 5 years old, > 10 kg, and ≥ 110 cm for basic lung function and spirometry testing (FEV1, FVC) to detect, assess, and monitor lung diseases like asthma, COPD, and cystic fibrosis. For use by trained healthcare professionals in hospital/clinical settings.
Regulatory Classification
Identification
A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.
{0}------------------------------------------------
January 2, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
NuvoAir AB % Yolanda Smith Consultant Smith Associates 1468 Harwell Ave Crofton, Maryland 21114
Re: K183089
Trade/Device Name: Air Next Regulation Number: 21 CFR 868.1840 Regulation Name: Diagnostic Spirometer Regulatory Class: Class II Product Code: BZG Dated: December 20, 2019 Received: December 30, 2019
Dear Yolanda Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Michael Rvan Division Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K183089
Device Name Air Next
Indications for Use (Describe)
Air Next is intended to be used by:
Healthcare professionals trained to perform spirometry tests on patients of age ≥ 5 years old, > 10 kg and > 110 cm.
Air Next is intended to perform basic lung function and spirometry testing. The actual diagnosis shall be done by a healthcare professional in hospital and clinical settings.
Type of Use (Select one or both, as applicable)
| <span> <span style="font-size:16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|-------------------------------------------------------------------------------------------------------|
| <span> <span style="font-size:16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
#### SPONSOR
| Company Name:<br>Company Address | NuvoAir AB<br>17 Riddargatan<br>Stockholm 114 57<br>Sweden |
|----------------------------------|------------------------------------------------------------|
| Telephone: | +46 725538233 |
| Contact Person: | Vanessa Caldeira |
Summary Preparation Date: December 31, 2019
### DEVICE NAME
| Trade Name: | Air Next |
|----------------------|-----------------------|
| Common/Usual Name: | Spirometer |
| Classification Name: | Diagnostic Spirometer |
| Regulation Number: | 21 CFR 868.1840 |
| Product Code: | BZG |
| Device Class: | Class II |
| Reviewing Panel: | Anesthesiology |
### PREDICATE DEVICE
Legally Marketed Equivalent Device
| Predicate | K Number | Brand Name | Company |
|-----------|----------|--------------|------------------------------------|
| Primary | K072979 | Spirobank G | MIR Medical International Research |
| Reference | K061712 | Spirobank II | MIR Medical International Research |
## DEVICE DESCRIPTION
Air Next is intended to perform basic lung function and spirometry testing. It measures parameters such as the forced expiratory volume in 1 sec (FEV1) and the forced vital capacity (FVC) in a forced expiratory maneuver. These measures can be used for detection, assessment and monitoring of diseases affecting the lung function, such as bronchial Asthma, COPD and Cystic Fibrosis.
Air Next is a hand-held spirometer, weighing only 75g and it is powered by 2 AAA alkaline 1.5V batteries. lt consists of 3 main components: the Air Next device, the NuvoAir disposable turbine (delivered in one package) and the Air Next mobile application downloadable from Apple's and Google's Play Stores. The package includes one Air Next device, one NuvoAir disposable turbine through which the user can start using the Air Next, two AAA 1.5V alkaline batteries, one user manual and one cotton bag to carry the device. Dimensions of the Air Next are 98 x 62 x 26 mm, which can be seen in the engineering drawing in the attachments.
## DEVICE INDICATIONS FOR USE
Air Next is intended to be used by:
Healthcare professionals trained to perform spirometry tests on patients of age ≥ 5 years old, > 10 kg and ≥ 110 cm.
Air Next is intended to perform basic lung function and spirometry testing. The actual diagnosis shall be done by a healthcare professional in hospital and clinical settings.
CAUTION: Federal (USA) law restricts this device to sale by or on the order of a dentist or physician.
{4}------------------------------------------------
# COMPARISON OF TECHNICAL CHARACTERISTICS
Predicate Product Comparison Table
| Feature | Subject Device | Primary Predicate<br>K072979 | Similarities and<br>Differences |
|----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | NuvoAir | M.I.R. Medical<br>International Research | Same |
| Product Code | BZG | BZG | Same |
| Regulation No, | 21 CFR 868.1840 | 21 CFR 868.1840 | Same |
| Indications for<br>Use | Air Next is intended to be<br>used by:<br>Healthcare professionals<br>trained to perform spirometry<br>tests on patients of age<br>$\ge$ 5 years old, > 10 kg<br>and $\ge$ 110 cm.<br>Air Next is intended to<br>perform basic lung function<br>and spirometry testing.<br>The actual diagnosis shall be<br>done by a healthcare<br>professional in hospital and<br>clinical settings. | The Spirobank G<br>spirometer is intended to<br>be used by a physician or<br>by a patient under the<br>instruction of a physician<br>or paramedic. The device<br>is intended to test lung<br>function and can make<br>spirometry testing in<br>people of all ages,<br>excluding infants and<br>neonates. It can be used<br>in any setting. | Similar<br>indications for<br>the use. The only<br>difference<br>between subject<br>device and<br>predicate device<br>is that the<br>Indications for<br>Use in the subject<br>device includes<br>also weight and<br>height while<br>Predicate device<br>only uses age as a<br>limitation. Subject<br>device is not for<br>use in home<br>setting. |
| Use<br>Environment | Hospital and clinical<br>settings. | All settings that<br>include clinical and<br>home use | Different<br>Use is limited to<br>clinical settings<br>only. |
| Intended User | Trained health care<br>professionals perform<br>spirometry testing on<br>Adults and pediatric patients<br>over 5 years old | People of all ages,<br>excluding infants and<br>neonates. | Different |
| Prescription<br>only | Yes | Yes | Same |
| Feature | Subject Device | Primary Predicate K072979 | Similarities and Differences |
| Principle of Operation | The Air Next is designed to work with NuvoAir disposable turbine. When performing a spirometry test, the user exhales into the turbine. The airflow generated is forcing a propeller to rotate inside the turbine. The Air Next registers the speed of the spinning propeller by counting the rotations with a digital infrared interruption sensor. The algorithm in the firmware inside the Air Next device then converts the rotations into airflow measured in liters per second.<br>The device is also tested against B.T.P.S. (body temperature and pressure with saturated water vapor) conditions as prescribed by ATS guidelines and the results are well within range. | Spirobank G is a powerful and compact measurement device, intended for use by a respiratory specialist or by a suitably trained generalist.<br>Two different types of turbine sensors can be used with the device, one is reusable and one is single-patient disposable. A disposable mouthpiece is required in order to connect a subject to the spirometer.<br>The flow and volume measurement sensor is a digital turbine, based on the infrared interruption principle.<br>Cleaning of the disposable turbine is not required, as it is supplied clean in a sealed plastic bag. It must be disposed of after use.<br>Spirometry test interpretation is based on the Forced Vital Capacity (FVC) test and is based on the ATS standard. | Same |
| Flow Sensor | Bidirectional turbine with infrared interruption | Bidirectional turbine with infrared | Same |
| Physical Characteristics | | | |
| Energy Type | Two AAA 1.5V Alkaline batteries | 9V Alkaline battery | Different |
| Display | Touchscreen on smartphone or tablet | Touchscreen / LCD and membrane | Different |
| Operating Environment | T: min +10°C/max +40° C<br>RH: min 10 %/max 95 % ALT: max 2000 m | Temperature: MIN + 10 °C, MAX + 40 °C;<br>Humidity: MIN 10% RH; MAX 95%RH | Same |
| Physical Size | Light hand-held device | Light hand-held device | Same |
| Feature | Subject Device | Primary Predicate<br>K072979 | Similarities and<br>Differences |
| Weight | 75g | 160g | Different |
| Technical Specifications | | | |
| Volume<br>accuracy | ±3% of reading or<br>±0.050 L, whichever is greater | ±3% of reading or<br>±0.050 L, whichever is<br>greater | Same |
| Volume range | Up to 10 L | Up to 10 L | Same |
| Flow range | 0-15/s | 0-16 L/s | Similar |
| Flow accuracy | ±5% or 200 mL/s | ±5% or 200 mL/s | Same |
| Flow resistance | <0.5 cmH2O/L/s | <0.5 cmH2O/L/s | Same |
| Connectivity | Bluetooth | Bluetooth of USB | Same -<br>Bluetooth |
| Measured parameter | | | |
| Forced vital<br>capacity | FVC | FVC | Same |
| Volume expired<br>in the first | FEV0.75, FEV1, FEV3, FEV6 | FEV0.75, FEV1, FEV3, FEV6 | Same |
| Ratio between<br>volume expired<br>in a certain<br>time period<br>and FVC | FEV/FVC (FER) for 0.75 /1<br>/3 / 6 | FEV/FVC (FER) for 0.75<br>/1 /3 / 6 | Same |
| Peak expiratory<br>flow | PEF | PEF | Same |
| Forced<br>expiratory flow<br>between the<br>first 25% and<br>75% of the FVC | FEF25-75 (MEF) | FEF25-75 (MEF) | Same |
| Volume<br>inspired in the<br>first second of<br>the test | FIV1 | FIV1 | Same |
| Forced<br>inspiratory<br>volume | FIVC | FIVC | Same |
| Peak inspiratory<br>flow | PIF | PIF | Same |
| Feature | Subject Device | Primary Predicate<br>K072979 | Similarities and<br>Differences |
| Forced<br>inspiratory flow<br>in the first 25%<br>and 75% of the<br>FIV | FIF25-75 (MIF25-75) | FIF25-75 (MIF25-75) | Same |
| Ratio between<br>FEV1 and FEV6 | FEV1/FEV6 | FEV1/FEV6 | Same |
| Forced<br>expiratory flow<br>at 50% of FVC<br>divided by FVC | FEF50/FVC | FEF50/FVC | Same…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.