SPIROBANK II

K061712 · Mir Medical Intl. Research Srl · BZG · Dec 1, 2006 · Anesthesiology

Device Facts

Record IDK061712
Device NameSPIROBANK II
ApplicantMir Medical Intl. Research Srl
Product CodeBZG · Anesthesiology
Decision DateDec 1, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.1840
Device ClassClass 2
AttributesPediatric

Indications for Use

The Spirobank II spirometer and pulse oximeter is intended to be used by a physician or by a patient under the instruction of a physician or paramedic. The device is intended to test lung function and can make: spirometry testing in people of all ages, excluding infants and neonates oximetry testing in people of all ages. It can be used in any setting.

Device Story

Spirobank II is a portable, compact spirometer and pulse oximeter. It measures lung function parameters, oxygen saturation (SpO2), and pulse rate. Device inputs include airflow data from a spirometry sensor and pulse/SpO2 signals. It processes these inputs to calculate over 30 spirometric parameters and statistical metrics derived from SpO2 and pulse rate. Operated by physicians, paramedics, or patients under instruction in any setting. Connectivity options include USB, Bluetooth, RS232, and internal modem for telemedicine. Healthcare providers review outputs to assess lung function and monitor patient status, supporting clinical decision-making for respiratory conditions.

Clinical Evidence

Clinical testing performed for both spirometry and pulse oximetry functions. Spirometry testing conducted per American Thoracic Society (ATS) standards; results met accuracy requirements. Oximetry performance verified via desaturation trial; results within specification. Additional statistical parameters for SpO2 and pulse rate verified in-house.

Technological Characteristics

Portable spirometer and pulse oximeter. Connectivity: USB, Bluetooth, RS232, internal modem. Standards: EN 60601-1:1990 (safety), EN 60601-1-2:1993 (EMC). Spirometry sensor and algorithm inherited from predicate. PCB-based architecture.

Indications for Use

Indicated for lung function testing in patients of all ages (excluding infants/neonates for spirometry; all ages for oximetry). Used by physicians or patients under physician/paramedic instruction in any setting.

Regulatory Classification

Identification

A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K061712 Image /page/0/Picture/1 description: The image shows the logo for Medical International Research. The logo consists of a stylized graphic above the text "MIR MEDICAL INTERNATIONAL RESEARCH". The graphic is a rounded, stylized shape that resembles three interconnected arches. The text is in a simple, sans-serif font. OEC - 1 2006 # Summary of Safety and Effectiveness ## 1. Applicant Information | Date Prepared: | November 30, 2006 | |-----------------|-----------------------------------------------| | Submitter: | MIR Medical International Research | | Address: | Via del Maggiolino, 125<br>00155 Roma - Italy | | | | | Contact Person: | Simon Fowler | | Phone Number: | +39 06.22.754.777 | ## 2. Device Information Trade Name: Spirobank II Classification Name: spirometer and oximeter # 3. Identification of legally marketed device to which the submitter claims equivalence: | Company Name: | MIR. | |----------------|-----------| | Device Name: | Spirobank | | 510(k) number: | K983909 | MIR. Company Name: Devicc Name: Spirotel K043528 510(k) number: ## 4. Description of the device: MIR Spirobank II is a spirometer and pulse oximeter, designed to facilitate the total valuation of lung function. It is designed for use by specialist who require a simple, portable and compact device, yet at the same time being capable of calculating more than 30 spirometric parameters, and monitors the oxygen saturation and pulse rate. It also calculates several additional statistical parameters derived from the SpO2 and the pulse rate. Its connectivity capability (USB, Bluetooth, internal modem for acoustic coupling to telephone, RS232) makes it suitable also for telemedicine applications. ## 5. Statement of Intended Use: The Spirobank II spirometer and pulse oximeter is intended to be used by a physician or by a patient under the instruction of a physician or paramedic. The device is intended to test lung function and can make: spirometry testing in people of all ages, excluding infants and neonates {1}------------------------------------------------ oximetry testing in people of all ages. It can be used in any setting # 6. Summary of the technological characteristics of the new device in comparison to those of the predicate device: of the MIR Spirobank II combines in a single device the functions both of a spirometer and of a pulse oximeter. It uses the same spirometry sensor and algorithm and calculates the same parameters as Spirobank, but uses a completely new spirometry PCB. Spirodains, but asses a compress jimetry PCB as the Spirotel. But in addition to the spot SpO2 ron the pulse rate, it also calculates several additional statistical parameters derived from the SpO2 and the pulse rate. # 7. Brief discussion of the clinical and nonclinical tests relied on for a determination of SE. Testing was done to ensure that the MIR Spirotel would perform safely and accurately within the environments for which it is to be marketed. Safety and environmental testing was conducted in accordance with EN 60601-1:1990 and EN 60601-1-2:1993. Electromagnetic compatibility (EMC), electrical, mechanical durability, safety (operator and patient), and temperature/humidity testing have been completed. The results demonstrates that the MIR Spirotel is in compliance with the guideline and standards referenced and that it performs within its specifications. Testing of device performance included clinical testing of both spirometry and pulse oximetry functions. Spirometry testing was performed according with American Thoracic Society (ATS) Standards. The results obtained were within the range of accuracy required by ATS. For oximetry testing a desaturation trial was conducted. The results obtained were within specification. Specifically of SpO2 and pulse rate and the statistical parameters which the device calculates, have been verified in-house. ## 8. Conclusions Based on these results, it is our determination that the device is safe, effective, and performs as well as the legally marketed devices. This summary on 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three overlapping wings, representing the department's mission to protect the health of all Americans. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA", which is arranged in a circular fashion around the logo. #### Public Health Service ood and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 DEC - 1 2006 Mr. Simon Fowler M.I.R Medical International Research SRL Via del Maggiolino, 125 00155 Roma. ITALY Re: K061712 Trade/Device Name: Spirobank II Regulation Number: 21 CFR 868.1840 Regulation Name: Diagnostic Spirometer Regulatory Class: II Product Code: BZG, DQA Dated: November 21, 2006 Received: November 27, 2006 Dear Mr. Fowler: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Mr. Fowler Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Chiu Lin, Ph.D. Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (K061712): Device Name: Spirobank II Indications for Use: The Spirobank II spirometer and pulse oximeter is intended to be used by a physician or by a patient under the instruction of a physician or paramedic. The device is intended to test lung function and can make: - spirometry testing in people of all ages, excluding infants and neonates . - oximetry testing in people of all ages. . It can be used in any setting. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device(Eva f Anesthesiology, General Hospital, ...on Control, Dental Devices K061712 Page 1 of 1 ﮩ
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...