K231416 · Nuvoair AB · BZG · Jan 18, 2024 · Anesthesiology
Device Facts
Record ID
K231416
Device Name
Air Next (NVD-02)
Applicant
Nuvoair AB
Product Code
BZG · Anesthesiology
Decision Date
Jan 18, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1840
Device Class
Class 2
Attributes
Pediatric
Indications for Use
NuvoAir Air Next is intended to test lung function and spirometry in adults and children 5 years of age and older. It can be used in hospitals, in the clinical setting, and at home. The Air Next is not intended for use in an operating room. The user is not intended to interpret or take clinical action based on the device output without consultation of a qualified healthcare professional.
Device Story
Hand-held spirometer (75g) for lung function testing; measures FEV1, FVC, and other respiratory parameters. Input: patient exhalation into a disposable turbine; turbine propeller rotation detected via digital infrared interruption sensor. Firmware converts rotation counts to airflow (L/s). Device connects via Bluetooth to a mobile application (iOS/Android) for data display and analysis. Used in hospitals, clinics, and home settings by patients or clinicians. Provides real-time feedback on test effort and quality per ATS 2019 guidelines. Output assists clinicians in detecting/monitoring diseases like Asthma, COPD, and Cystic Fibrosis. Benefits include portable, calibration-free lung function assessment.
Clinical Evidence
Bench testing only. Device performance validated against ATS 2019 guidelines and ISO 23747/ISO 26782 standards. Biocompatibility testing performed per ISO 10993-5 and ISO 10993-10 (cytotoxicity, irritation, sensitization).
Technological Characteristics
Hand-held spirometer; ABS-PC plastic construction. Sensing: Bidirectional turbine with infrared interruption. Power: 2x AAA 1.5V alkaline batteries. Connectivity: Bluetooth. Dimensions: 75g. Software: Mobile app-based display. Standards: IEC 60601-1, IEC 60601-1-2, ISO 10993-1, ISO 23747, ISO 26782, IEC 62304.
Indications for Use
Indicated for lung function and spirometry testing in adults and children 5 years of age and older. Not for use in operating rooms. Users must consult a healthcare professional for clinical interpretation.
Regulatory Classification
Identification
A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
January 18, 2024
NuvoAir AB Daniel Keenan Senior Product Director, Medical Device 17D Riddargatan Stockholm. SE 11457 Stockholm, 11457 Sweden
Re: K231416
Trade/Device Name: Air Next Regulation Number: 21 CFR 868.1840 Regulation Name: Diagnostic Spirometer Regulatory Class: Class II Product Code: BZG Dated: December 11, 2023 Received: December 11, 2023
Dear Daniel Keenan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Rachana Visaria -S
Rachana Visaria, Ph.D. Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia,
{2}------------------------------------------------
Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231416
Device Name Air Next
Indications for Use (Describe)
NuvoAir Air Next is intended to test lung function and spirometry in adults and children 5 years of age and older.
It can be used in hospitals, in the clinical setting, and at home.
The Air Next is not intended for use in an operating room.
The user is not intended to interpret or take clinical action based on the device output without consultation of a qualified healthcare professional.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
# 510(k) Summary
| Date of Summary<br>Preparation: | January 18, 2024 |
|---------------------------------|------------------------------------------------------------------------------------------------------------------------------|
| 510(k) submitter &<br>owner | NuvoAir AB<br>17D Riddargatan<br>Stockholm, SE<br>11457 |
| Contact person | Owner & contact for 510(k) submission:<br>Furat Shawki- General Manager, CT<br>furat.shawki@nuvoair.com<br>(+44)755 188 3598 |
| Common name of the<br>device | Spirometer |
| Trade name | Air Next |
| Model number | NVD-02 |
| Classification | Class II Device |
| Classification name | Diagnostic Spirometer |
| Product code | BZG |
| Review panel | Anesthesiology |
| Regulation number | 21 CFR 868.1840 |
| Predicate device | |
| 510(k) number | K072979 |
| Trade name | Spirobank G |
| Product code | BZG |
| Manufacturer | MIR Medical International Research |
{5}------------------------------------------------
## Device Description
Air Next is intended to perform basic lung function and spirometry testing. It measures parameters such as the forced expiratory volume in 1 sec (FEV1) and the forced vital capacity (FVC) in a forced expiratory maneuver. These measures can be used for detection, assessment and monitoring of diseases affecting the lung function, such as bronchial Asthma, COPD and Cystic Fibrosis.
Air Next is a hand-held spirometer, weighing 75g and it is powered by 2 AAA alkaline 1.5V batteries. It consists of 3 main components: the Air Next device, the NuvoAir disposable turbine (delivered in one package) and the Air Next mobile application downloadable from Apple's and Google's Play Stores.
### Indications For Use
NuvoAir Air Next is intended to test lung function and spirometry in adults and children 5 years of age and older.
It can be used in hospitals, in the clinical settings, and at home.
The Air Next is not intended for use in an operating room.
The user is not intended to interpret or take clinical action based on the device output without consultation of a qualified healthcare professional.
{6}------------------------------------------------
| Predicate | K Number | Brand Name | CE Certificate | Company |
|--------------|----------|---------------|--------------------------------------|---------------------------------------|
| Primary | K072979 | Spirobank G | 93/42/EEC Certified<br>No: MED 9826 | MIR Medical International<br>Research |
| Reference | K061712 | Spirobank II | NoBo:0476<br>Valid until: 2023-02-04 | MIR Medical International<br>Research |
| Product Code | BZG | Regulation No | 21 CFR 868.1840 | |
# SUMMARY OF TECHNOLOGICAL CHARACTERISTICS AND COMPARISON TO PREDICATE DEVICE
| Characteristics | NuvoAir<br>Air Next | Spirobank G<br>K072979 | notes |
|---------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | NuvoAir | MIR Medical<br>International Research | |
| Regulation<br>Number | 21 CFR 868.1840 | 21 CFR 868.1840 | Same |
| Product Code | BZG | BZG | Same |
| Indications for Use | NuvoAir Air Next is<br>intended to test lung<br>function and<br>spirometry in adults<br>and children 5 years of<br>age and older.<br>It can be used in<br>hospitals, in the clinical<br>setting, and at home.<br>The Air Next is not<br>intended for use in an<br>operating room.<br>The user is not<br>intended to interpret or<br>take clinical action<br>based on the device<br>output without<br>consultation of a<br>qualified healthcare<br>professional. | The Spirobank G<br>spirometer is intended to<br>be used by a physician or<br>by a patient under the<br>instruction of a physician<br>or paramedic.<br>The device is intended to test<br>lung function and can<br>make spirometry testing<br>in people of all ages,<br>excluding infants and<br>neonates.<br>It can be used in any<br>setting. | Similar |
| Type of Use | Prescription Use | Prescription Use | Same |
| User Environment | Hospitals, in the<br>clinical setting, and at<br>home. | In any setting | The use environment<br>for the subject device is<br>a subset of the<br>predicate device use<br>environment. |
| Principle of<br>Operation | The Air Next is designed<br>to work with NuvoAir<br>disposable turbine. When<br>performing a spirometry<br>test, the user exhales into<br>the turbine. The airflow<br>generated is forcing a<br>propeller to rotate inside<br>the turbine. The Air Next<br>registers the speed of the<br>spinning propeller by<br>counting the rotations<br>with a digital infrared<br>interruption sensor. The<br>algorithm in the firmware<br>inside the Air Next<br>device then converts the<br>rotations into airflow<br>measured in liters per<br>second.<br>This widely spread<br>technology circumvents<br>the need for individual<br>calibration by exploiting<br>mechanical properties of<br>the turbine, which are<br>verified and calibrated at<br>the time of | Spirobank G is a powerful<br>and compact measurement<br>device, intended for use by<br>a respiratory specialist or<br>by a suitably trained<br>generalist.<br>Two different types of<br>turbine sensors can be<br>used with the device, one<br>is reusable and one is<br>single-patient disposable. A<br>disposable mouthpiece is<br>required in order to<br>connect a subject<br>to the spirometer.<br>The flow and volume<br>measurement sensor is a<br>digital turbine, based on the<br>infrared interruption principal.<br>Cleaning of the disposable<br>turbine is not required, as it<br>is supplied clean in a sealed<br>plastic bag. It must be<br>disposed of after use. | Substantially<br>equivalent (SE)<br>Differences do not<br>raise new or<br>different questions<br>of safety or<br>effectiveness |
| Flow Sensor | Bidirectional turbine<br>with infrared<br>interruption | Bidirectional turbine<br>with infrared<br>interruption | Same |
| Energy Type | Two AAA 1.5V<br>Alkaline | 9V Alkaline battery | SE |
| | batteries | | |
| Display | Touchscreen on<br>smartphone or<br>tablet | Touchscreen / LCD and<br>membrane keys<br>128 x 48 pixel, graphic<br>LCD - FSTN | Similar |
| Operating<br>Environme<br>nt | T: min $+10$ °C/max<br>$+40$ ° C RH: min 10<br>%/max 95 % ALT:<br>max 2000 m | Temperature: MIN $+ 10$<br>°C, MAX $+ 40$ °C;<br>Humidity: MIN 10% RH;<br>MAX 95%RH | Same |
| Physical Size | Light hand-held device | Light hand-held device | Same |
| Weight | 75g | 160g | Differences do not raise<br>new or different<br>questions of safety<br>and effectiveness |
| Volume accuracy | ±3% of reading or<br>$±0.050$ L,<br>whichever is<br>greater | ±3% of reading or<br>$±0.050$ L, whichever is<br>greater | Same |
| Volume range | Up to 10<br>L | Up to 10 L | Same |
| Flow range | 0-15L/s | 0-16 L/s | Similar |
| Flow accuracy | ±5% or 200 mL/s | ±5% or 200 mL/s | Same |
| Flow resistance | <0.5 cmH2O/L/s | <0.5 cmH2O/L/s | Same |
| Characteristics | NuvoAir<br>Air Next | Spirobank G<br>K072979 | notes |
| Connectivity | Bluetooth | Bluetooth or USB | Same-Bluetooth |
| Forced vital<br>capacity | FVC | FVC | Same |
| Volume expired<br>in the first<br>0.75sec, 1sec,<br>3sec, 6sec | FEV0.75, FEV1, FEV3,<br>FEV6 | FEV0.75,FEV1,FEV3,<br>FEV6 | Same |
| between<br>Ratio<br>volume expired<br>in a certain time<br>period and FVC | FEV/FVC (FER) for 0.75<br>/1 /3 / 6 | FEV/FVC (FER) for 0.75<br>/1 /3 / 6 | Same |
| Peak expiratory<br>flow | PEF | PEF | Same |
| Forced expiratory<br>flow between the<br>first 25% and 75%<br>of the FVC | FEF25-75 (MEF) | FEF25-75 (MEF) | Same |
| Volume inspired<br>in the first second<br>of the test | FIV 1 | FIV 1 | Same |
| Forced<br>inspiratory<br>volume | FIVC | FIVC | Same |
| Peak inspiratory<br>flow | PIF | PIF | Same |
| Forced inspiratory<br>flow in the first<br>25% and 75% of<br>the FIV | FIF25-75 (MIF25-75) | FIF25-75 (MIF25-75) | Same |
| Ratio between<br>FEV1 and FEV6 | FEV1/FEV6 | FEV1/FEV6 | Same |
| Forced expiratory<br>flow at 50% of<br>FVC divided by<br>FVC | FEF50/FVC | FEF50/FVC | Same |
| Characteristics | NuvoAir<br>Air Next | Spirobank G<br>K072979 | notes |
| Forced inspiratory<br>flow in the first<br>25% and 75% of<br>the FIV | FIF25-75 (MIF25-75) | FIF25-75 (MIF25-75) | Same |
| Ratio between<br>FEV1 and FEV6 | FEV1/FEV6 | FEV1/FEV6 | Same…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.