WELCH ALLYN CARDIOPERFECT WORKSTATION SOFTWARE (CPWS) V 1.62

K082478 · Welch Allyn, Inc. · BZG · Mar 9, 2009 · Anesthesiology

Device Facts

Record IDK082478
Device NameWELCH ALLYN CARDIOPERFECT WORKSTATION SOFTWARE (CPWS) V 1.62
ApplicantWelch Allyn, Inc.
Product CodeBZG · Anesthesiology
Decision DateMar 9, 2009
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 868.1840
Device ClassClass 2
AttributesSoftware as a Medical Device, Pediatric

Indications for Use

The CardioPerfect Workstation software and associated accessories are indicated for the acquisition, analysis, formatting, display, printing and storage of certain physiologic signals, as identified below, for the purpose of assisting the clinician in the diagnosis and monitoring of various diseases and/or treatment regimens. The CardioPerfect Workstation software also provides non-diagnostic functions such as patient management, data security, search tools for patient and/or test records and support for exporting data to Electronic Medical Record systems. The CardioPerfect Workstation and associated accessories are intended for use by or on the order of a physician in a hospital or clinic setting. The product is designed for use on both adult and pediatric patients, subject to any specific contraindications identified below. Stress Electrocardiograph - Intended Use Using the optional ECG module and associated accessories the user can acquire, view, store and print ECG waveforms. Resting Electrocardiograph - Intended Use The same as defined for stress ECG plus the ability to use optional algorithms (MEANS) to generate measurements, data presentations, graphical presentations and interpretive statements on an advisory basis. These are presented for review and interpretation by the clinician. Spirometry - Intended Use Using the optional spirometry module and associated accessories to acquire, view, store and print measures and waveforms of pulmonary function. The spirometer should only be used with patients able to understand the instructions for performing the test. Ambulatory Blood Pressure - Intended Use Using the optional ABP module and associated accessories the user can acquire, retrieve, view, store and print patient ambulatory blood pressure history.

Device Story

CardioPerfect Workstation Software (CPWS) operates on standard Windows-based PCs to manage ECG, ambulatory blood pressure (ABP), and spirometry data. Input signals are acquired via optional hardware modules; software processes these signals to generate waveforms, measurements, and interpretive statements. Clinicians in hospitals or clinics use the software to review advisory interpretations, manage patient records, and export data to EMRs. The system assists in diagnosing and monitoring cardiac and pulmonary conditions. Version 1.6.2 introduces the PEDMEANS algorithm for pediatric ECG interpretation. Output is displayed on-screen or printed for physician review to support clinical decision-making.

Clinical Evidence

Bench testing only. Verification of MEANS and PEDMEANS algorithms against IEC 60601-2-51:2003 requirements. Risk management conducted per ISO 14971. No clinical trials performed; equivalence established via technical comparison and software validation.

Technological Characteristics

Software-based diagnostic workstation for ECG, ABP, and spirometry. Operates on off-the-shelf Windows PCs. Includes MEANS and PEDMEANS interpretive algorithms. Complies with IEC 60601-1, IEC 60601-1-1, IEC 60601-1-2, IEC 60601-1-4, IEC 60601-2-25, and AAMI/ANSI EC11 standards. Software Level of Concern: Moderate.

Indications for Use

Indicated for adult and pediatric patients requiring cardiac screening or diagnostic differentiation (ECG), pulmonary function assessment (spirometry) for symptoms like shortness of breath, cough, or wheezing, and ambulatory blood pressure monitoring for conditions including white coat hypertension, nocturnal hypertension, resistant hypertension, diabetes, or pregnancy-related hypertension. Spirometry contraindication: patients unable to understand test instructions.

Regulatory Classification

Identification

A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(k) Summary Kof 2478 [As described in 21 CFR 807.92] 9 2009 MAR Submitted by: Welch Allyn Inc. 4341 State Street Road Skaneateles Falls, NY 13153-0220 Contact Person: John Sawyer Vice President, QA/RA Phone: (315) 685-4571 Fax: (315) 685-2532 E-mail: sawyerj@welchallyn.com August 22, 2008 Date Prepared: Trade Name: CardioPerfect Workstation Software (CPWS) Version 1.6.2 Common Name: Medical Device Software for interpretive and stress testing electrocardiographs and diagnostic spirometry. Classification Reference: Class II, Diagnostic Spirometer (21 CFR 868.1840, Product Code BZG) and Cardiovascular (Product Code LOS). Predicate Device: CardioPerfect Workstation Software Version 1.5.0 (K052158) {1}------------------------------------------------ ### Description of the Device: CardioPerfect Workstation software is used to create a computer platform on which ECG, ABP, & Spirometry applications can operate within inherent capabilities of an off-theshelf desktop or laptop personal computer utilizing a Windows operating system. ### Intended Use: The CardioPerfect Workstation software and associated accessories are indicated for the acquisition, analysis, formatting, display, printing and storage of certain physiologic signals, as identified below, for the purpose of assisting the clinician in the diagnosis and monitoring of various diseases and/or treatment regimens. The CardioPerfect Workstation software also provides non-diagnostic functions such as patient management, data security, search tools for patient and/or test records and support for exporting data to Electronic Medical Record systems. The CardioPerfect Workstation and associated accessories are intended for use by or on the order of a physician in a hospital or clinic setting. The product is designed for use on both adult and pediatric patients, subject to any specific contraindications identified below. # Stress Electrocardiograph - Intended Use Using the optional ECG module and associated accessories the user can acquire, view, store and print ECG waveforms. # Resting Electrocardiograph - Intended Use The same as defined for stress ECG plus the ability to use optional algorithms (MEANS) to generate measurements, data presentations, graphical presentations and interpretive statements on an advisory basis. These are presented for review and interpretation by the clinician. # Spirometry - Intended Use Using the optional spirometry module and associated accessories to acquire, view, store and print measures and waveforms of pulmonary function. The spirometer should only be used with patients able to understand the instructions for performing the test. # Ambulatory Blood Pressure - Intended Use Using the optional ABP module and associated accessories the user can acquire, retrieve, view, store and print patient ambulatory blood pressure history. {2}------------------------------------------------ ### Indications for Use: # Electrocardiograph - Indications for Use Indications for electrocardiography range from routine screening of cardiac health in the physician office environment to directed diagnostic differentiation in a hospital cardiology department. # Spirometry - Indications for Use Indications for spirometry include, but are not limited to, the following: - Shortness of breath . - . Chronic cough - Occupational exposure to dust or chemicals . - Assist in the diagnosis of Bronchitis . - Assist in the diagnosis of Asthma . - . Wheezing - Assist in the monitoring of bronchodilator ● # Ambulatory Blood Pressure - Indications for Use Indications for ambulatory blood pressure measurement (as listed in Journal of Hypertension 2003, 21 :821-848, E. O'Brien et. at.) include, but are not limited to, the following: - · suspected "white coat" hypertension - · Suspected nocturnal hypertension - To establish dipper status - · Resistant hypertension - · Elderly patient - · To monitor antihypertensive drug treatment - · Type 1 diabetes - · Hypertension of pregnancy - · Evaluations of hypotension - · Autonomic failure - · Masked hypertension ### Technological Characteristics: The Welch Allyn Electrocardiography and Spirometry Products are intended for use by trained operators in healthcare facilities. The 12-lead ECG interpretive algorithm provides a computer-generated analysis of potential patient cardiac abnormalities, which must be confirmed by a physician with other relevant clinical information. {3}------------------------------------------------ The optional spirometry module is indicated for use in clinical situation to assess a patient's pulmonary health status and evaluate symptoms, signs, or abnormal laboratory test results. The CPWS version 1.6.2 is substantially equivalent to the previously released CPWS version 1.5.0 (K.052158). The major difference being the change to intended use. This submission intends to change the indications for use to include interpretive ECG software for the pediation. The previously marketed CPWS version 1.5.0 (K052158) did not provide interpretive software support for patients less than 16 years of age. The Pediatric Modular ECG PEDMEANS Interpretive ECG algorithm developed by P. Rijnbeek, 2007, in Analysis System thesis entitled 'Automatic Interpretation of Pediatric Electrocardiograms', Erasmus University Rotterdam. This PEDMEANS algorithm, is also used in our CP 100™ and CP 200™ ECG device cleared under K072449. ## Summary of Effectiveness: The Welch Allyn CPWS team has determined that the software "Level of Concern" is Moderate. (See section 10 for CPWS software Level of Concern) The implementation of the PEDMEANS interpretive software in the CPWS 1.6.2 electrocardiograph device, and its performance, is equivalent in every respect to the implementation of this software in the CP 100™ and CP 200™ submitted to the agency by Welch Allyn and cleared per K072449. The MEANS algorithms were verified and found to be consistent with the requirements of IEC 60601-2-51:2003, particular requirements for safety, including essential performance, of recording and analyzing single channel and multi-channel electrocardiographs. All requirements of the recognized and applicable standards are in compliance: EC11 (AAMI/ANSI), UL 60601-1. IEC 60601-1. IEC 60601-1-1. IEC 60601-1-2. IEC 60601-1-4 and IEC 60601-2-25. Additionally, risk management (risk, SFMEA and safety analysis) activities have been conducted in accordance with ISO 14971 Medical Devices - Application of risk management to medical devices and comply with IEC 60601-1-4 Medical Electrical Equipment Part 1: General Requirements for Safety, Part 4: Programmable Electrical Medical Systems. {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" are arranged in a circular pattern around the eagle. The logo is black and white. #### Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 8 2005 Welch Allyn, Inc. c/o Mr. John Sawyer Vice President QA/RA 4341 State Street Road Skaneateles Falls, NY 13153-0220 K082478 Re: > Trade/Device Name: Welch Allyn CardioPerfect Workstation Software (CPWS), Version 1.6.2 Regulation Number: 21 CFR 868.1840 Regulation Name: Diagnostic Spirometer Regulatory Class: Class II (Two) Product Code: BZG Dated: February 4, 2009 Received: March 5, 2009 ### Dear Mr. Sawyer: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ Page 2 - Mr. John Sawyer Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market, If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, J. Atkin Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ # Indications for Use Statement KO82478 510(k) Number (if known): Device Name: Welch Allyn CardioPerfect Workstation Software (CPWS) V 1.6.2 ### Intended Use: The CardioPerfect Workstation software and associated accessories are indicated for the acquisition, analysis, formatting, display, printing and storage of certain physiologic signals, as identified below, for the purpose of assisting the clinician in the diagnosis and monitoring of various diseases and/or treatment regimens. The CardioPerfect Workstation software also provides non-diagnostic functions such as patient management, data security, search tools for patient and/or test records and support for exporting data to Electronic Medical Record systems. The CardioPerfect Workstation and associated accessories are intended for use by or on the order of a physician in a hospital or clinic setting. The product is designed for use on both adult and pediatric patients, subject to any specific contraindications identified below. # Stress Electrocardiograph - Intended Use Using the optional ECG module and associated accessories the user can acquire, view, store and print ECG waveforms. ## Resting Electrocardiograph - Intended Use The same as defined for stress ECG plus the ability to use optional algorithms (MEANS) to generate measurements, data presentations, graphical presentations and interpretive statements on an advisory basis. These are presented for review and interpretation by the clinician. # Spirometry - Intended Use Using the optional spirometry module and associated accessories to acquire, view, store and print measures and waveforms of pulmonary function. The spirometer should only be used with patients able to understand the instructions for performing the test. # Ambulatory Blood Pressure - Intended Use Using the optional ABP module and associated accessories the user can acquire, retrieve, view, store and print patient ambulatory blood pressure history. Atill Stein for B. Zuckerman (Division Sign-Off) Division of Cardiovascular Device 510(k) Number KOS2478 {7}------------------------------------------------ ### Indications for Use: ### Electrocardiograph - Indications for Use Indications for electrocardiography range from routine screening of cardiac health in the physician office environment to directed diagnostic differentiation in a hospital cardiology department. ### Spirometry - Indications for Use Indications for spirometry include, but are not limited to, the following: - Shortness of breath . - Chronic cough . - Occupational exposure to dust or chemicals ● - Assist in the diagnosis of Bronchitis . - Assist in the diagnosis of Asthma . - Wheezing ♥ - Assist in the monitoring of bronchodilator . ## Ambulatory Blood Pressure - Indications for Use Indications for ambulatory blood pressure measurement (as listed in Journal of Hypertension 2003, 21 :821-848, E. O'Brien et. at.) include, but are not limited to, the following: - · suspected "white coat" hypertension - · Suspected nocturnal hypertension - To establish dipper status - · Resistant hypertension - · Elderly patient - · To monitor antihypertensive drug treatment - · Type 1 diabetes - · Hypertension of pregnancy - · Evaluations of hypotension - · Autonomic failure - · Masked hypertension X AND/OR Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Stittser for B Zuckerman Mision Sign-Off) Division of Cardlovascular Device 510(k) Number
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...