WELCH ALLYN CARDIOPERFECT WORKSTATION VERSION 1.5.0 SOFTWARE
Applicant
Welch Allyn, Inc.
Product Code
BZG · Anesthesiology
Decision Date
Sep 8, 2005
Decision
SESE
Submission Type
Special
Regulation
21 CFR 868.1840
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The CardioPerfect Workstation software and associated accessories are indicated for the acquisition, analysis, formatting, display, printing and storage of certain physiologic signals, as identified below, for the purpose of assisting the clinician in the diagnosis and monitoring of various diseases and/or treatment regimens. The CardioPerfect Workstation software also provides non-diagnostic functions such as patient management, data security, search tools for patient and/or test records and support for exporting data to Electronic Medical Record systems. The CardioPerfect Workstation and associated accessories are intended for use by or on the order of a physician in a hospital or clinic setting. The product is designed for use on both adult and pediatric patients, subject to any specific contraindications identified below.
Device Story
CardioPerfect Workstation software operates on off-the-shelf Windows/DOS PCs; integrates ECG, ABP, and spirometry modules. Inputs: physiologic signals from connected hardware modules. Processing: software acquires, formats, displays, stores, and prints data; optional MEANS algorithm provides advisory interpretive statements for resting ECG. Output: waveforms, measurements, and interpretive reports for clinician review. Used in hospitals/clinics by physicians or under physician order. Facilitates diagnosis and monitoring of cardiac and pulmonary conditions; supports EMR data export and patient management. Benefits: centralized diagnostic data access and standardized reporting for clinical decision-making.
Clinical Evidence
No clinical data provided. Device relies on established performance of predicate modules and algorithms (MEANS). Bench testing only.
Technological Characteristics
Software-based workstation for ECG, ABP, and spirometry. Operates on standard Windows/DOS desktop or laptop PCs. Includes optional MEANS algorithm for resting ECG interpretation. Connectivity includes EMR data export. No specific hardware materials or sterilization methods applicable to the software component.
Indications for Use
Indicated for adult and pediatric patients in hospital or clinic settings for ECG (stress/resting), ambulatory blood pressure (ABP) monitoring, and spirometry. Resting ECG with optional MEANS algorithm indicated for patients 16+ years. Spirometry normal values not calculated for children under 6. Specific indications include cardiac screening, hypertension management (white coat, nocturnal, resistant, pregnancy-related, masked), hypotension evaluation, autonomic failure, diabetes, and pulmonary conditions (shortness of breath, cough, bronchitis, asthma, wheezing).
Regulatory Classification
Identification
A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.
{0}------------------------------------------------
K052158
p/13
Image /page/0/Picture/1 description: The image shows the logo for Welch Allyn. The text is in a bold, sans-serif font. The logo is black and the background is white.
SEP = 8 2005
Special 510(k) Premarket Notification CardioPerfect Workstation Software Version 1.5.0 and Accessory Product
### 11. Premarket Notification [510(k)] Summary
- Welch Allyn, Inc. Submitted By: 4341 State Street Road P.O. Box 220 Skaneateles Falls, NY 13153-0220
315 685 3694 Phone: 315 685 2532 Fax:
Contact: Christopher Klaczyk, Regulatory Affairs Manager
- Medical Device Software for interpretive and stress testing Common Name: electrocardiographs, and diagnostic spirometry
Welch Allyn CardioPerfect Workstation software (with) Trade Name:
> CardioPerfect Workstation Resting ECG Module CardioPerfect Workstation ST ECG Module CardioPerfect Workstation SpiroPerfect Module CardioPerfect Workstation ABP Module
| Classification: | |
|-----------------|-----------------|
| LOS: | None |
| DSI: | 21 CFR 870.1025 |
| DXN: | 21 CFR 870.1130 |
| BZG: | 21 CFR 868.1840 |
- ECG-resting, CardioPerfect with Means, K962854 Predicate Device: ECG-stress, CardioPerfect ST, K935732 Caird Technology Spirometer, K971336
- CardioPerfect Workstation software is used to create a computer Description: platform on which ECG, ABP, & Spirometry applications can operate within inherent capabilities of an off-the-shelf desktop or laptop personal computer utilizing a Windows or DOS based operating system.
Indications for use:
### CardioPerfect Workstation Software Version 1.5.0 and Accessories Product
{1}------------------------------------------------
K052159
p2/3
# WelchAllyn
Special 510(k) Premarket Notification CardioPerfect Workstation Software Version 1.5.0 and Accessory Product
The CardioPerfect Workstation software and associated accessories are indicated for the The Cardion, analysis, formatting, display, printing and storage of certain physiologic signals, as identified below, for the purpose of assisting the clinician in the diagnosis and monitoring of various diseases and/or treatment regimens. The CardioPerfect Workstation software also various diseases and of weatherns such as patient management, data security, search tools for provides non thaghebels rand support for exporting data to Electronic Medical Record systems.
The CardioPerfect Workstation and associated accessories are intended for use by or on the order I ho Cardio' in in a hospital or clinic setting. The product is designed for use on both adult and pediatric patients, subject to any specific contraindications identified below.
### Electrocardiograph - Stress
Using the optional ECG module and associated accessories the user can acquire, view, store and obing are specificans for electrocardiography range from routine screening of cardiac health in the physician office environment to directed diagnostic differentiation in a hospital cardiology department.
### Electrocardiograph - Resting
The same as defined for stress ECG plus the ability to use optional algorithms (MEANS) to generate measurements, data presentations, graphical presentations and interpretive statements on an advisory basis for patients sixteen (16) years of age and above. These are presented for review and interpretation by the clinician.
### Ambulatory Blood Pressure
Using the optional ABP module and associated accessories the user can acquire, retrieve, view, store and print patient ambulatory blood pressure history. Indications for ambulatory blood pressure measurement (as listed in Journal of Hypertension 2003, 21:821-848, E. O'Brien et. al.) include, but are not limited to, the following:
- suspected "white coat" hypertension .
- Suspected nocturnal hypertension .
- To establish dipper status .
- Resistant hypertension .
- . Elderly patient
- To monitor antihypertensive drug treatment .
- Type 1 diabetes .
- Hypertension of pregnancy .
- Evaluations of hypotension .
- Autonomic failure .
- Masked hypertension .
### Spirometry
Using the optional spirometry module and associated accessories to acquire, view, store and print measures and waveforms of pulmonary function. Normal values and comparative results are not calculated for children under the age of six. Indications for spirometry include, but are not limited to, the following:
- Shortness of breath .
{2}------------------------------------------------
# WelchAllyn
Special 510(k) Premarket Notification CardioPerfect Workstation Software Version 1.5.0 and Accessory Product
- Chronic cough
- . Occupational exposure to dust or chemicals
- Assist in the diagnosis of Bronchitis
- Assist in the diagnosis of Asthma .
- Wheezing .
-
- Assist in the monitoring of bronchodilator .
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines representing its body and wings. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" written around the perimeter of the circle.
SEP - 8 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Welch Allyn, Inc c/ o Mr. Christopher Klaczyk Regulatory Affairs Manager 4341 State Street Road P.O. Box 220 Skaneateles Falls, NY 13153-0220
Re: K052158
Trade Name: CardioPerfect Workstation Software Regulation Number: 21 CFR 868.1840 Regulation Name: Diagnostic Spirometer Regulatory Class: Class II (two) Product Code: BZG Dated: August 5, 2005 Received: August 9, 2005
Dear Mr. Klaczyk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Christopher Klaczyk
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. product radially w you to begin marketing your device as described in your Section 510(k) I mo letter wrification. The FDA finding of substantial equivalence of your device to a legally premated predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please it you attent office of Compliance at (240) 276-0295. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Other general mironmational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Brimmer for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### WelchAllyn
Special 510(k) Premarket Notification CardioPerfect Workstation Software Version 1.5.0 and Accessory Product
10. Statement of Indications For Use
K052158 510(k) Number:
Device Name:
CardioPerfect Workstation, Software Version 1.5.0 and Accessories Product.
### CardioPerfect Workstation Software Version 1.5.0 and Accessories Product
The CardioPerfect Workstation software and associated accessories are indicated for the acquisition, analysis, formatting, display, printing and storage of certain physiologic signals, as identified below, for the purpose of assisting the clinician in the diagnosis and monitoring of various diseases and/or treatment regimens. The CardioPerfect Workstation software also provides non-diagnostic functions such as patient management, data security, search tools for patient and/or test records and support for exporting data to Electronic Medical Record systems.
The CardioPerfect Workstation and associated accessories are intended for use by or on the order of a physician in a hospital or clinic setting. The product is designed for use on both adult and pediatric patients, subject to any specific contraindications identified below.
### Electrocardiograph - Stress
Using the optional ECG module and associated accessories the user can acquire, view, store and print ECG waveforms. Indications for electrocardiography range from routine screening of cardiac health in the physician office environment to directed diagnostic differentiation in a hospital cardiology department.
### Electrocardiograph - Resting
The same as defined for stress ECG plus the ability to use optional algorithms (MEANS) to generate measurements, data presentations, graphical presentations and interpretive statements on an advisory basis for patients sixteen (16) years of age and above. These are presented for review and interpretation by the clinician.
### Ambulatory Blood Pressure
Using the optional ABP module and associated accessories the user can acquire, retrieve, view, store and print patient ambulatory blood pressure history. Indications for ambulatory blood pressurement (as listed in Journal of Hypertension 2003, 21:821-848, E. O'Brien et. al.) include, but are not limited to, the following:
- Suspected "white coat" hypertension .
- . Suspected nocturnal hypertension
- To establish dipper status .
- Resistant hypertension �
- . Elderly patient
- To monitor antihypertensive drug treatment .
- Type 1 diabetes �
- Hypertension of pregnancy �
- Evaluations of hypotension .
- Autonomic failure .
Masked hypertension *
Page 48 of 52 This document contains information considered to be CONFIDENTIAL to Welch Allyn, Inc.
MPD FCD-0026, Rev. 2
page 1 of 2
{6}------------------------------------------------
## Welch Allyn
Special 510(k) Premarket Notification CardioPerfect Workstation Software Version 1.5.0 and Accessory Product
#### Spirometry
Using the optional spirometry module and associated accessories to acquire, view, store and print measures and waveforms of pulmonary function. Normal values and comparative results are not calculated for children under the age of six. Indications for spirometty include, but are not limited to, the following:
- Shortness of breath ◆
- Chronic cough ◆
- Occupational exposure to dust or chemicals
- Assist in the diagnosis of Bronchitis .
- . Assist in the diagnosis of Asthma
- . Wheezing
- Assist in the monitoring of bronchodilator ◆
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
AND/OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
B. Rymmmer
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number K052158
MPD FCD-0026, Rev. 2
Page 49 of 52 This document contains information considered to be CONFIDENTIAL to Welch Allyn, Inc.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.