Welch Allyn Diagnostic Cardiology Suite 2.X.X with Spirometry option

K191013 · Welch Allyn, Inc. · BZG · Sep 10, 2019 · Anesthesiology

Device Facts

Record IDK191013
Device NameWelch Allyn Diagnostic Cardiology Suite 2.X.X with Spirometry option
ApplicantWelch Allyn, Inc.
Product CodeBZG · Anesthesiology
Decision DateSep 10, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.1840
Device ClassClass 2
AttributesPediatric

Indications for Use

The Welch Allyn® Diagnostic Cardiology Suite™ is a prescription device intended for use by physicians, other licensed health care practitioners, and trained personnel who are acting on the orders of a physician. Welch Allyn Diagnostic Cardiology Suite is intended for use in medical clinics, physician offices and hospital settings to acquire, analyze, display, transmit and print certain physiological signals identified below and provide the data for consideration by a physician. Welch Allyn Diagnostic Cardiology Suite utilizes a software platform to support 12-lead diagnostic resting ECG and diagnostic Spirometry applications and is designed to operate within the inherent capabilities of an off-the-shelf laptop or PC Windows operating system. Welch Allyn Diagnostic Cardiology Suite also provides functions related to patient data management including communication with electronic medical records systems. Welch Allyn Diagnostic Cardiology Suite Resting ECG's are intended to be taken with the patient in the supine position and offer VERITAS™ resting ECG algorithm to generate measurements and advisory statements for review and interpretation by the physician. Welch Allyn Diagnostic Cardiology Suite is not intended to be used as a physiological vital signs monitor, not intended to be used in a mobile medical environment (e.g. ambulance, helicopter), in magnetic resonance (MR) environments, in operating theaters, nor in conjunction with high frequency surgical equipment.

Device Story

PC-based software platform for 12-lead resting ECG and spirometry; operates on off-the-shelf Windows PC/laptop. ECG input: surface electrodes; Spirometry input: pneumotach airflow sensor. System acquires, analyzes, displays, and prints physiological signals; integrates with EMR systems. Used in clinics, physician offices, and hospitals by physicians or trained personnel. ECG module uses VERITAS algorithm for measurements/advisory statements; Spirometry module analyzes forced expiratory efforts against norms. Output assists clinicians in diagnosing cardiac and pulmonary conditions. Benefits: standardized diagnostic data for clinical decision-making.

Clinical Evidence

No clinical studies were utilized. Substantial equivalence is supported by non-clinical performance data, including software verification/validation, usability engineering, and bench testing against standards (IEC 60601-2-25, ISO 26782, ATS 2005).

Technological Characteristics

PC-based software; Windows OS. ECG: surface electrodes, IEC 60601-2-25 compliant. Spirometry: pneumotach airflow sensor, ISO 26782 and ATS 2005 compliant. Connectivity: EMR integration. Software lifecycle: IEC 62304. Patient-contacting materials: equivalent to K133428 (spirometry) and K120865 (ECG).

Indications for Use

Indicated for cardiac health screening and diagnostic differentiation in patients of all ages (ECG). Indicated for pulmonary function testing (forced expiratory flow-volume and time measurements) in adult and pediatric patients age 6 years and older presenting with symptoms like chronic cough, shortness of breath, wheezing, or disorders such as asthma, COPD, bronchitis, or interstitial lung disease. Contraindications: hemoptysis of unknown origin, pneumothorax, unstable cardiovascular status, recent MI or pulmonary embolus, thoracic/abdominal/cerebral aneurysms, recent eye/thoracic/abdominal surgery, or acute disease interfering with test performance.

Regulatory Classification

Identification

A diagnostic spirometer is a device used in pulmonary function testing to measure the volume of gas moving in or out of a patient's lungs.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. September 10, 2019 Welch Allyn, Inc. Megan Pellenz Regulatory Engineer 7865 North 86th St. Milwaukee, Wisconsin 53224 Re: K191013 Trade/Device Name: Welch Allyn® Diagnostic Cardiology Suite™ 2.X.X with Spirometry option 901128 CARDIOPULMONARY ECG SYSTEM Regulation Number: 21 CFR 868.1840 Regulation Name: Diagnostic spirometer Regulatory Class: Class II Product Code: BZG, DPS Dated: August 22, 2019 Received: August 23, 2019 Dear Megan Pellenz: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {2}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Rachana Visaria, Ph.D. Assistant Director (Acting) DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K191013 #### Device Name Welch Allyn® Diagnostic Cardiology Suite™ 2.X.X with Spirometry option 901128 CARDIOPULMONARY ECG SYSTEM #### Indications for Use (Describe) The Welch Allyn® Diagnostic Cardiology Suite™ is a prescription device intended for use by physicians, other licensed health care practitioners, and trained personnel who are acting on the orders of a physician. Welch Allyn Diagnostic Cardiology Suite is intended for use in medical clinics, physician offices and hospital settings to acquire, analyze, display, transmit and print certain physiological signals identified below and provide the data for consideration by a physician. Welch Allyn Diagnostic Cardiology Suite utilizes a software platform to support 12-lead diagnostic Spirometry applications and is designed to operate within the inherent capabilities of an off-the-shelf laptop or PC Windows operating system. Welch Allyn Diagnostic Cardiology Suite also provides functions related to patient data management including communication with electronic medical records systems. Welch Allyn Diagnostic Cardiology Suite Resting ECG's are intended to be taken with the patient in the supine position and offer VERTAS™ resting ECG algorithm to generate measurements and advisory statements for review and interpretation by the physician. Welch Allyn Diagnostic Cardiology Suite is not intended to be used as a physiological vital signs monitor, not intended to be used in a mobile medical environment (e.g. ambulance, helicopter), in magnetic resonance (MR) environments, in operating theaters, nor in conjunction with high frequency surgical equipment. Indications for Welch Allyn Diagnostic Cardiology Suite ECG range from routine screening of cardiac health in the physician office environment to directed diagnostic differentiation in a hospital cardiology department. Welch Allyn Diagnostic Cardiology Suite ECG is indicated for patients of all ages. Welch Allyn Diagnostic Cardiology Suite Spirometry is inction testing, providing forced expiratory flowvolume and time measurements. The device is intended to be a general assessment tool assisting the diagnosis of pulmonary function. These devices are commonly used on patients with occupational exposure to dust or chemicals; presenting with symptoms such as chronic cough, shorthess of breath and wheezing; disorders such as interstitial lung disease, bronchitis, ashma, and COPD; or other patients where the clinician determines these pulmonary function measurements are needed. Welch Allyn Diagnostic Cardiology Suite with Spirometry is indicated for adult and pediatric patients age 6 years and older and should only be used with patients able to understand the instructions for performing the test. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | <div> <span> <span style="font-size: 10pt;"> <b> <span style="font-family: arial, helvetica, sans-serif;">☑</span> </b> </span> </span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span> <span style="font-size: 10pt;"> <b> <span style="font-family: arial, helvetica, sans-serif;">☐</span> </b> </span> </span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Traditional 510(k) Welch Allyn® Diagnostic Cardiology Suite™ ## 510(k) SUMMARY [As described in 21 CFR 807.92] Submitted by Manufacturer: Welch Allyn, Inc. 4341 State Street Road Skaneateles Falls, NY 13153 USA Establishment Reqistration: 2183461 Primary Contact Person: Megan Pellenz, Regulatory Affairs Engineer Secondary Contact Person: Mark Elliott, Quality Assurance Director, QMR Date Updated: September 9th, 2019 Device Trade Name: Welch Allyn® Diagnostic Cardiology Suite™ 2.X.X with Spirometry option 901128 CARDIOPULMONARY ECG SYSTEM Common Name: Diagnostic Spirometer Regulation Number: 868.1840 Device Class: Il Product Code: BZG Secondary Product code, and Regulation: DPS, 870.2340 > Predicate Device: Welch Allyn CardioPerfect®Workstation Software™ (CPWS) 901047 CARDIOPULMONAY ECG SYSTEM (K082478) Reference Device: Welch Allyn® RScribe™ (K120865) 901127 ELECTROCARDIOGRAPH ## Background The Welch Allyn® Diagnostic Cardiology Suite™ is a modification to Welch Allyn's current legally marketed RScribe cleared by FDA under K120865 and since branded "Connex Cardio". Welch Allyn's interpretation of when to submit a 510(k) for a change to an existing device and our planned device modification, the addition of indications for diagnostic spirometry, led us to the determination that the {5}------------------------------------------------ RScribe/Connex Cardio is being significantly modified and therefore this 510(k) submission is appropriate. ### Indications for Use The Welch Allyn® Diagnostic Cardiology Suite™ is a prescription device intended for use by physicians, other licensed health care practitioners, and trained personnel who are acting on the orders of a physician. Welch Allyn Diagnostic Cardiology Suite is intended for use in medical clinics, physician offices and hospital settings to acquire, analyze, display, transmit and print certain physiological signals identified below and provide the data for consideration by a physician. Welch Allyn Diagnostic Cardiology Suite utilizes a software platform to support 12-lead diagnostic resting ECG and diagnostic Spirometry applications and is designed to operate within the inherent capabilities of an off-the-shelf laptop or PC Windows operating system. Welch Allyn Diagnostic Cardiology Suite also provides functions related to patient data management including communication with electronic medical records systems. Welch Allyn Diagnostic Cardiology Suite Resting ECG's are intended to be taken with the patient in the supine position and offer VERITAS™ resting ECG algorithm to generate measurements and advisory statements for review and interpretation by the physician. Welch Allyn Diagnostic Cardiology Suite is not intended to be used as a physiological vital signs monitor, not intended to be used in a mobile medical environment (e.g. ambulance, helicopter), in magnetic resonance (MR) environments, in operating theaters, nor in conjunction with high frequency surgical equipment. Indications for Welch Allyn Diagnostic Cardiology Suite ECG range from routine screening of cardiac health in the physician office environment to directed diagnostic differentiation in a hospital cardiology department. Welch Allyn Diagnostic Cardiology Suite ECG is indicated for patients of all ages. There are no known contraindications for resting ECG. Welch Allyn Diagnostic Cardiology Suite Spirometry is indicated for pulmonary function testing, providing forced expiratory flow-volume and time measurements. The device is intended to be a general assessment tool assisting the clinician in the diagnosis of pulmonary function. These devices are commonly used on patients with occupational exposure to dust or chemicals; presenting with symptoms such as chronic cough, shortness of breath and wheezing; disorders such as interstitial lung disease, bronchitis, asthma, and {6}------------------------------------------------ COPD; or other patients where the clinician determines these pulmonary function measurements are needed. Welch Allyn Diagnostic Cardiology Suite with Spirometry is indicated for adult and pediatric patients age 6 years and older and should only be used with patients able to understand the instructions for performing the test. Contraindications for diagnostic spirometry: - hemoptysis of unknown origin (forced expiratory maneuver may ● aggravate the underlying condition) - pneumothorax ● - unstable cardiovascular status (forced expiratory maneuver may worsen ● angina or cause changes in blood pressure) or recent myocardial infarction or pulmonary embolus - . thoracic, abdominal, or cerebral aneurysms (danger of rupture due to increased thoracic pressure) - Recent eye surgery (e.g., cataract); - presence of an acute disease process that might interfere with test ● performance (e.g., nausea, vomiting) Recent surgery of thorax or abdomen. # Technoloqical Characteristics The Welch Allyn Diagnostic Cardiology Suite is a PC-based software device. Section 16 details the software technology including its requirements, architecture, detailed design, development environment, verification and validation and lifecycle management plan. Medical device software development and management is characterized in IEC 62304 Medical device software – Software life cycle processes. The technology of the hardware accessories are: cited herein as a reference device. - . For diagnostic ECG, the technology involves applying surface electrodes to the patient's skin to obtain signals of the heart's electrical activity. The electrical signals are then amplified, analyzed (VERITAS algorithms) and displayed on Welch Allyn Diagnostic Cardiology Suite ECG module user interface. Clinical Users can evaluate these ECGs, store them, save them or send them via the technological characteristics provided by the Welch Allyn Diagnostic Cardiology Suite. The diagnostic ECG technology is characterized in IEC 60601-2-25 Medical electrical equipment - Part 2-25: Particular requirements for the basic safety and essential performance of electrocardiographs. The Welch Allyn RScribe K120865 utilizes the same VERITAS algorithms as Welch Allyn Diagnostic Cardiology Suite ECG and therefore is being - For diagnostic Spirometry, a pneumotach air flow sensor obtains signals ● of a patient's forced expiratory breathing maneuvers, called "efforts", that are then analyzed against norms contained within Welch Allyn Diagnostic Cardiology Suite Spirometry module. Patient breathing efforts are {7}------------------------------------------------ displayed on the Welch Allyn Diagnostic Cardiology Suite Spirometry user interface. Clinical Users can evaluate these Spirometry efforts, store them, save them or send them via the technological characteristics provided by the Welch Allyn Diagnostic Cardiology Suite. The diagnostic Spirometry technology is characterized in ISO 26782 Anaesthetic and respiratory equipment - Spirometers intended for the measurement of time forced expired volumes in humans and follows ATS 2005 Standardization of Spirometry quidelines. ### Device Comparison It is of the opinion of Welch Allyn, Inc. that Welch Allyn Diagnostic Cardiology Suite is substantially equivalent to the predicate device. Welch Allyn CardioPerfect Workstation Software (K082478) as they share similar fundamental technical characteristics, intended use including indications for use, target population and use environment, detailed in the device comparison table below. The Welch Allyn® RScribe™ K120865 utilizes the same VERITAS algorithms as Welch Allyn Diagnostic Cardiology Suite ECG and therefore is presented as a Reference device. {8}------------------------------------------------ | Characteristic | Reference Device | Subject Device | Predicate Device | |------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Trade Name | Welch Allyn® RScribe™<br>K120865 | Welch Allyn® Diagnostic<br>Cardiology Suite™ | Welch Allyn® CardioPerfect<br>Workstation Software™ (CPWS)<br>V1.6.2 K082478 | | Model Identifier | 901127<br>ELECTROCARDIOGRAPH | 901128<br>CARDIOPULMONARY ECG SYSTEM | 901047<br>CARDIOPULMONARY ECG SYSTEM | | Manufacturer | Welch Allyn, Inc. | Welch Allyn, Inc. | Welch Allyn, Inc. | | 510(k) Number | K120865 | K082478 | K082478 | | Product Code | DPS | BZG | BZG | | Classification<br>Name | Electrocardiograph, Class II | Diagnostic Spirometer, Class II | Diagnostic Spirometer, Class II | | Regulation<br>Number | 21 CFR 870.2340 | 21 CFR 868.1840 | 21 CFR 868.1840 | | Indications for Use | The RScribe Electrocardiograph is a<br>multi-channel electrocardiograph<br>product used for acquiring, analyzing,<br>displaying and printing resting ECG's.<br>The RScribe is a 12-channel<br>diagnostic electrocardiograph intended<br>for recording and printing ECG's of<br>adult and pediatric patients. The<br>acquired ECG will be displayed for<br>quality check purpose, analyzed using<br>the Mortara VERITAS resting<br>interpretation, optionally printed, stored<br>and/or transmitted to a ECG<br>Management System or Hospital<br>Information System. The device is not<br>intended to be used as a vital signs<br>physiological monitor.<br>It is a system comprised of a Mortara<br>ECG amplifier (Wireless Acquisition<br>Module [WAM] or AM12 Patient<br>Cable) and an off-the-shelf personal<br>computer with Mortara software<br>application that allows clinicians to<br>collect ECGs on patients during | The Welch Allyn® Diagnostic Cardiology<br>Suite™ is a prescription device intended<br>for use by physicians, other licensed<br>health care practitioners, and trained<br>personnel who are acting on the orders<br>of a physician. Welch Allyn Diagnostic<br>Cardiology Suite is intended for use in<br>medical clinics, physician offices and<br>hospital settings to acquire, analyze,<br>display, transmit and print certain<br>physiological signals identified below<br>and provide the data for consideration<br>by a physician. Welch Allyn Diagnostic<br>Cardiology Suite utilizes a software<br>platform to support 12-lead diagnostic<br>resting ECG and diagnostic Spirometry<br>applications and is designed to operate<br>within the inherent capabilities of an off-<br>the-shelf laptop or PC Windows<br>operating system. Welch Allyn<br>Diagnostic Cardiology Suite also<br>provides functions related to patient<br>data management including<br>communication with electronic medical | The Welch Allyn® CardioPerfect<br>Workstation Software™ (CPWS) and<br>associated accessories are indicated<br>for the acquisition, analysis, formatting,<br>display, printing and storage of certain<br>physiologic signals, as identified<br>below, for the purpose of assisting the<br>clinician in the diagnosis and<br>monitoring of various diseases and/or<br>treatment regimens. The<br>CardioPerfect Workstation software<br>also provides non-diagnostic functions<br>such as patient management, data<br>security, search tools for patient and/or<br>test records and support for exporting<br>data to Electronic Medical Record<br>systems.<br>The CardioPerfect Workstation<br>Software and associated accessories<br>are intended for use by or on the order<br>of a physician in a hospital or clinic<br>setting. The product is designed for<br>use on both adult and pediatric<br>patients as included in the cleared | | Characteristic | Reference Device | Subject Device | Predicate Device | | | routine visits. The patient populations<br>for which the device will be used may<br>be healthy or diseased of any age.<br>ECG's are taken with the patient in the<br>supine position. The RScribe is intended<br>to be used by a licensed health care<br>practitioner in a hospital, medical clinic<br>and offices of any size, including Clinical<br>Research Organizations. | records systems. Welch Allyn<br>Diagnostic Cardiology Suite Resting<br>ECG's are intended to be taken with the<br>patient in the supine position and offer<br>VERITAS™ resting ECG algorithm to<br>generate measurements and advisory<br>statements for review and interpretation<br>by the physician. | indications for use listed below, subject<br>to any specific contraindications<br>identified below. | | | The RScribe electrocardiograph is a<br>non-invasive prescription device.<br>The device is indicated for use<br>to acquire, analyze, display,<br>transmit and print<br>electrocardiograms. The device is indicated for use<br>to provide interpretation of the<br>data for consideration by a<br>physician. The device is indicated for use<br>in a clinical setting, by a<br>physician or by trained<br>personnel who are acting on<br>the orders of a licensed<br>physician. It is not intended as<br>a sole means of diagnosis. The interpretations of ECG<br>offered by the device are only<br>significant when used in<br>conjunction with a physician<br>over-read as well as<br>consideration of all other<br>relevant patient data. The device is indicated for use<br>on adult and pediatric<br>populations. The device is not intended to<br>be used as a vital signs<br>physiological monitor. The device is not designed for<br>out of hospital transport. | Welch Allyn Diagnostic Cardiology Suite<br>is not intended to be used as a<br>physiological vital signs monitor, not<br>intended to be used in a mobile medical<br>environment (e.g. ambulance,<br>helicopter), in magnetic resonance (MR)<br>environments, in operating theaters, nor<br>in conjunction with high frequency<br>surgical equipment.<br><br>Indications for Welch Allyn Diagnostic<br>Cardiology Suite ECG range from<br>routine screening of cardiac health in<br>the physician office environment to<br>directed diagnostic differentiation in a<br>hospital cardiology department. Welch<br>Allyn Diagnostic Cardiology Suite ECG<br>is indicated for patients of all ages.<br><br>Welch Allyn Diagnostic Cardiology Suite<br>Spirometry is indicated for pulmonary<br>function testing, providing forced<br>expiratory flow-volume and time<br>measurements. The device is intended<br>to be a general assessment tool<br>assisting the clinician in the diagnosis of<br>pulmonary function. These devices are<br>commonly used on patients with<br>occupational exposure to dust or<br>chemicals; presenting with symptoms<br>such as chronic cough, shortness of<br>breath and wheezing; disorders such as<br>interstitial lung disease, bronchitis,<br>asthma, and COPD; or other patients | Electrocardiograph - Indications for<br>Use<br>Indications for electrocardiography<br>range from routine screening of<br>cardiac health in the physician office<br>environment to directed diagnostic<br>differentiation in a hospital cardiology<br>department.<br><br>Spirometry - Indications for Use<br>Indications for Spirometry include, but<br>are not limited to, the following:<br>Shortness of breath Chronic cough Occupational exposure to dust<br>or chemicals Assist in the diagnosis of<br>Bronchitis Assist in the diagnosis of<br>Asthma Wheezing Assist in the monitoring of<br>bronchodilator<br>Ambulatory Blood Pressure -<br>Indications for Use<br>Indications for ambulatory blood<br>pressure measurement (as listed in<br>Journal of Hypertension 2003, 21:821-<br>848, E. O'Brien et. at.) include, but are<br>not limited to, the following:<br>Suspected "white coat"<br>hypertension Suspected nocturnal<br>hypertension To establish dipper status Resistant hypertension Elderly patient | | Characteristic | Reference Device | Subject Device | Predicate Device | | | use in highly invasive environments, such as an operating theatre. | where the clinician determines these pulmonary function measurements are needed. Welch Allyn Diagnostic Cardiology Suite with Spirometry is indicated for adult and pediatric patients age 6 years and older and should only be used with patients able to understand the instructions for performing the test. | To monitor antihypertensive drug treatment Type 1 diabetes Hypertension of pregnancy Evaluations of hypotension Autonomic failure Masked hypertension | {9}------------------------------------------------ {10}------------------------------------------------ {11}------------------------------------------------ ## Non-clinical Performance Data Non-clinical data was utilized in this submission. The non-clinical standards and tests that were submitted. referenced or relied upon to support a determination of substantial equivalence are: ISO 14971 Medical devices - Application of risk management to medical devices (2007) ANSI/AAMI ES 60601-1 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (2005+A1:2012) IEC 60601-1:2005+A1:2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance ANSI/AAMI IEC 62304 Medical device software - Software life cycle processes (2006) IEC 62366-1 Medical devices - Part 1: Application of usability engineering to medical devices (2015) AAMI/ANSI HE75 Human Factors Engineering – Design of Medical Devices (2009 R2013) IEC 60601-1-6 General requirements for basic safety and essential performance -Collateral standard: Usability (2010+A1:2013) IEC 60601-2-25 Medical electrical equipment - Part 2-25: Particular requirements for the basic safety and essential performance of electrocardiographs (2011) ISO 26782 Anaesthetic and respiratory equipment -- Spirometers intended for the measurement of time forced expired volumes in humans (2009) ATS 2005 Standardization of Spirometry ISO 10993-1 (2009+Cor.1:2010) Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process. The spirometry patientcontacting materials are identical or equivalent to the materials cleared in K133428. The ECG patient-contacting materials are identical to the materials cleared in K120865. # Clinical Performance Data No clinical studies were utilized for the purpose of showing substantial equivalence. # Conclusion The Welch Allyn Diagnostic Cardiology Suite subject device and its predicate, the Welch Allyn CardioPerfect Workstation Software (CPWS) K082478 share the same intended use, the same modular software design, the same installation environment, the same general operating principles, the same basic technology and share similar performance claims and labeling. The Welch Allyn RScribe K120865 utilizes the same VERITAS algorithms as the subject device and has been in clinical use since 2012. Minor differences between the devices do not raise concerns of differences in safety. efficacy or intended use. Non-clinical performance tests were conducted and submitted to demonstrate that the Welch Allyn Diagnostic Cardiology Suite performs similarly to the predicate. It is Welch Allyn's opinion that the Welch Allyn Diagnostic Cardiology {12}------------------------------------------------ Suite subject device is substantially equivalent to its predicate, the Welch Allyn CardioPerfect Workstation Software (CPWS) cleared through K082478.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...