K202776 · Shenzhen Lepu Intelligent Medical Equipment Co., Ltd. · DQA · Mar 1, 2021 · Cardiovascular
Device Facts
Record ID
K202776
Device Name
Fingertip pulse oximeter
Applicant
Shenzhen Lepu Intelligent Medical Equipment Co., Ltd.
Product Code
DQA · Cardiovascular
Decision Date
Mar 1, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Fingertip pulse oximeter is a handheld non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate of adult, adolescent and child patients in hospital-type facilities, and homecare.
Device Story
Handheld, non-invasive fingertip pulse oximeter; measures SpO2 and pulse rate via photoplethysmography. Device uses dual-wavelength LED (660nm red, 905nm infrared) and photodiode detector to measure light absorption changes in pulsating arteriolar blood. Signal acquisition module captures data; Analog front-end IC amplifies signals; MCU processes data to calculate SpO2 and pulse rate based on Lambert-Beer law. Output displayed on OLED screen. Used by clinicians or patients for spot-checking in hospitals or homecare. Provides real-time physiological data to assist in monitoring patient oxygenation status.
Clinical Evidence
Clinical study conducted per ISO 80601-2-61 Annex EE. Study included 12 healthy adult volunteers (ages 18-50, diverse pigmentation). Evaluated SpO2 accuracy during steady-state/non-motion conditions across 70-100% SaO2 range compared to arterial blood co-oximeter. Results showed an Arms of 1.52% for SpO2 accuracy.
Technological Characteristics
Materials: Silicone gel pad, PCTG/PMMA enclosure. Sensing: Dual-wavelength LED (660nm, 905nm) and photodiode. Power: 2x AAA batteries. Connectivity: Standalone. Standards: ISO 80601-2-61, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 10993. Software: Moderate level of concern.
Indications for Use
Indicated for spot-checking SpO2 and pulse rate in adult, adolescent, and child patients in hospital and homecare settings.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services-USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is a symbol of the agency's mission to protect and promote public health through the regulation and supervision of food safety, tobacco, pharmaceuticals, medical devices, and other related products.
March 1, 2021
Shenzhen LEPU Intelligent Medical Equipment Co., Ltd. % Arthur Goddard President FDA Regulatory and Quality Systems Consultant 31853 Cedar Road Mayfield Heights, Ohio 44124-4445
Re: K202776
Trade/Device Name: Fingertip pulse oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: January 8, 2021 Received: January 25, 2021
Dear Arthur Goddard:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm_identifies_combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Todd Courtney Assistant Director DHT1C: Division of ENT. Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202776
Device Name Fingertip pulse oximeter
Indications for Use (Describe)
The Fingertip pulse oximeter is a handheld non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate of adult, adolescent and child patients in hospital-type facilities, and homecare.
Type of Use (Select one or both, as applicable)
| <span> <b> </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|-----------------------------------------------------------------------------|
| <span> <b> </b> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# Section 5: 510(K) Summary
This summary of 510(K) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1900 and 21 CFR 807.92.
The assigned 510(K) Number: K202776
#### 5. 510(K) Summary
### 5.1. Date of Preparation: September 11th, 2020
### 5.2. Sponsor
Shenzhen Lepu Intelligent Medical Equipment Co., Ltd. North side of floor 3, BLD 9 BaiWangxin High-Tech Industrial Park Songbai Road, Xili Street, Nanshan District 518055 Shenzhen, Guangdong, CHINA. Tel: + 86-0755-86952278 Fax: + 86-0755-86952287
Contact Person: Aaron Lin Position: General Manager Email: aaron.lin@lepu-medica1.com
### 5.3. Submission Correspondent
Mr. Arthur Goddard 31853 Cedar Road, Ohio, 44124-4445, U.S.A. Tel: (216) 233-5722 Email: asigoddard@aol.com
### 5.4. Subject Device Identification
Subject Device Name: Fingertip pulse oximeter Model: LOX100A, LOX100B, LOX100C, LOX100D Common name: Oximeter Classification Name(s): Oximeter Product Code: DOA Regulation Number: 21 CFR 870.2700 Review Panel: Anesthesiology Classification: II
### 5.5. Predicate Device
510(k) Number: K161560 Device Name: Fingertip Pulse Oximeter MD300CN310 Manufacturer: Beijing Choice Electronic Technology Co., Ltd.
{5}------------------------------------------------
#### 5.6. Indications for use
The Fingertip pulse oximeter is a handheld non-invasive device intended for spotchecking of oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate of adult, adolescent and child patients in hospitals, hospital-type facilities, and homecare.
# 5.7. Device Description
Fingertip pulse oximeter, LOX100A, LOX100B, LOX100C, LOX100D, mainly include of signal acquisition module, signal processing module, prompt module, detector, and emitter LED, display and user interface module, power supply module, the device is used to measure the patient's blood oxygen saturation (SpO2) and pulse rate (PR). The device is mainly composed of main board PCB, lamp panel PCB, sensor, OLED screen, button, silicone gel pad and enclosure.
All the models, LOX100A, LOX100B, LOX100C, LOX100D of proposed device, have difference in appearance, but follow the same design principle and technical specification. The device is a stand-alone device is intended only for spot checking, and the device is reusable and do not need sterilization and re-sterilization.
# 5.8. Predicate Devices and Subject Device Comparison
| Item | Subject Device<br>Finger pulse oximeter | Predicate Device<br>K161560<br>Fingertip Pulse Oximeter | Remark |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Product Name | Finger Pulse Oximeter | Fingertip Pulse Oximeter | SE |
| Product Code | DQA | DQA | |
| Regulation<br>Number | 21 CFR 870.2700 | 21 CFR 870.2700 | |
| Classification<br>Name(s) | Oximeter | Oximeter | |
| Classification | II | II | |
| Indications for<br>use | The Fingertip Pulse Oximeter<br>is a handheld non-invasive<br>device intended for spot-<br>checking of oxygen saturation<br>of arterial hemoglobin (SpO2)<br>and Pulse Rate of adult,<br>adolescent and child patients<br>in hospitals, hospital-type<br>facilities, and homecare. | The Fingertip Pulse<br>Oximeter MD300CN310 is a<br>handheld non-invasive<br>device intended for spot-<br>checking of oxygen<br>saturation of arterial<br>hemoglobin (SpO2) and<br>Pulse Rate of adult,<br>adolescent and child patients<br>in hospitals, hospital-type<br>facilities, and homecare | SE |
Table 5-1 Feature Comparison with Predicate Devices
{6}------------------------------------------------
| Shenzhen LEPU Intelligent Medical Equipment Co., Ltd. | | Fingertip pulse oximeter | |
|-------------------------------------------------------|-----------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|--------------|
| Item | Subject Device<br>Finger pulse oximeter | Predicate Device<br>K161560<br>Fingertip Pulse Oximeter | Remark |
| Intended patient<br>population | Adult, adolescent and child<br>patients | Adult, adolescent and child<br>patients | SE |
| Intended<br>application site | Fingertip | Fingertip | SE |
| Basic functions | Spot-checking of oxygen<br>saturation of arterial<br>hemoglobin (SpO2) and Pulse<br>Rate | Spot-checking of oxygen<br>saturation of arterial<br>hemoglobin (SpO2) and<br>Pulse Rate | SE |
| Components | Detector and emitter LED,<br>Analog front-end IC, MCU,<br>data display unit and power<br>unit | Detector and emitter LED,<br>signal amplify unit, CPU,<br>data display unit and power<br>unit | Discussion 1 |
## Discussion 1:
The device composition of the subject device and the predicate device is different in the signal processing unit, the predicate device is the signal amplify unit, and the target device is the Analog front-end IC. The function of the signal amplify unit is to convert the current signal of the receiving tube into a voltage signal, and then amplify the voltage signal. The voltage signal received by the CPU is the waveform and amplitude recognized after ADC conversion; The function of the Analog front-end IC is to directly amplify the current signal and then convert it into pulse signals of different frequencies. The MCU captures the pulse signal and converts the frequency signal into a waveform and amplitude signal. The essence of the two is the same, they both amplify the signal. In addition, we verify the performance of the subject device through clinical study (See Section 20 for details) and ISO80601-2-61 performance test (See Section 18 for details), and the verification results meet the requirements for device performance. So, the difference does not raise any new issues of safety or efficacy. The subject device can be considered to be substantially equivalent to that of the predicate device.
| Contacting Material | |
|---------------------|--|
| Fingertip pad | Silicone Gel | Silicone Gel | SE |
|---------------|--------------|--------------|--------------|
| Enclosure | PCTG, PMMA | ABS | Discussion 2 |
### Discussion 2:
The Enclosure of the two devices is made of different materials. The materials used in subject device have excellent performance and safety, and can meet the requirements of the device for the material. The biological safety of all materials has been verified, including cytotoxicity, sensitization, and irritation tests. The test results meet the requirements of the ISO10993 series of standards (See Section 15 for details). So, the difference does not raise any new issues of safety or efficacy. The subject device can be considered to be substantially equivalent to that of the predicate device.
{7}------------------------------------------------
Fingertip pulse oximeter
| Item | Subject Device<br>Finger pulse oximeter | Predicate Device<br>K161560<br>Fingertip Pulse Oximeter | Remark |
|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Software Level<br>of concern | Moderate | Moderate | SE |
| Working<br>principle | The measuring principles of<br>pulse oximeter is based on<br>Lambert-Beer law. The<br>emitter LED of oximeter<br>contains a dual wavelength<br>light source. One wavelength<br>of light source is 660nm,<br>which is red light source. The<br>other is 905nm, which is near<br>infrared light source. The dual<br>wavelength light source<br>alternately irradiates the<br>surface of the finger, and the<br>photo diode detector at the<br>other end of the finger will<br>detect the red light and near-<br>infrared signal. In the process<br>of pulse beating, with the<br>increase and decrease of blood<br>volume, different amplitude of<br>light will be absorbed at the<br>end of the detector. According<br>to the absorption ratio of two<br>kinds of wavelength light, DC<br>signal and AC signal are<br>extracted, and the oxygen<br>saturation value and pulse rate<br>value are obtained by the<br>above signals. | The pulse oximeter works by<br>applying a sensor to a<br>pulsating arteriolar vascular<br>bed. The sensor contains a<br>dual light source and photo<br>detector. The one<br>wavelength of light source is<br>660nm, which is red light;<br>the other is 905nm, which is<br>infrared-red light. Skin,<br>bone, tissue, and venous<br>vessels normally absorb a<br>constant amount of light over<br>time. The photo detector in<br>finger sensor collects and<br>converts the light into<br>electronic signal which is<br>proportional to the light<br>intensity. The arteriolar bed<br>normally pulsates and<br>absorbs variable amounts of<br>light during systole and<br>diastole, as blood volume<br>increases and decreases. The<br>ratio of light absorbed at<br>systole and diastole is<br>translated into an oxygen<br>saturation measurement.<br>This measurement is referred<br>to as SpO2. | Discussion 3 |
| Measurement<br>wavelength-Red<br>light | 660±3nm | 660±3nm | |
| Measurement<br>wavelength-<br>Infrared | 905±10nm | 905±10nm | |
{8}------------------------------------------------
| Item | Subject Device | Predicate Device | Remark |
|------|-----------------------|--------------------------|--------|
| | Finger pulse oximeter | K161560 | |
| | | Fingertip Pulse Oximeter | |
Fingertip pulse oximeter
# Discussion 3:
Though the working principle expressions of the subject device and predicate device are a little different, the working principle is basically the same. Both the ratio of red light and infrared light of different wavelengths are converted into electronic signals through the sensor, thereby calculating the oxygen saturation measurement value. The wavelength of red light is 660mm and the wavelength of infrared light is 905nm. So, the difference does not raise any new issues of safety or efficacy. The subject device can be considered to be substantially equivalent to that of the predicate device.
| Table 5-2 Specification Comparison | | | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------|---------------------------------------------------------|--------------|
| Item | Subject Device<br>Finger pulse oximeter | Predicate Device<br>K161560<br>Fingertip Pulse Oximeter | Remark |
| Display Type | OLED | OLED | SE |
| Power supply | 2*AAA batteries | 2*AAA batteries | SE |
| Working Time | Approximately 25 hours of<br>continuous operation | Approximately 24 hours of<br>continuous operation | Discussion 4 |
| Discussion 4: | | | |
| The working time of the subject device and the predicate device are slightly different. The<br>working time of the predicate device is covered by the working time of the subject device. In<br>addition, the working hours of the two devices are sufficient to meet the needs of use. So, the<br>difference does not raise any new issues of safety or efficacy. The subject device can be<br>considered to be substantially equivalent to that of the predicate device. | | | |
| User Interface | 3 display modes | 6 display modes | Discussion 5 |
| Discussion 5: | | | |
| There are 3 display modes for the subject device and 6 display modes for the predicate device.<br>The difference between them is the display direction. There are 2 display directions for the<br>subject device, which form 3 display modes when combined with bar-graph and<br>plethysmography. There are 4 display directions for the predicate device, which form 6 display<br>modes when combined with bar-graph and plethysmography. The difference is only in display<br>directions, and the display content on the screen is the same. So, the difference does not raise<br>any new issues of safety or efficacy. The subject device can be considered to be substantially<br>equivalent to that of the predicate device. | | | |
| Oxygen saturation | | | |
| SpO2<br>Measurement<br>Range | 70%~100% | 70%~100% | SE |
| Accuracy | 70%~100%· +2% | 70%~100%· +2%· | Discussion 6 |
{9}------------------------------------------------
| Shenzhen LEPU Intelligent Medical Equipment Co., Ltd. | | Fingertip pulse oximeter | |
|-------------------------------------------------------|-----------------------------------------|---------------------------------------------------------|--------|
| Item | Subject Device<br>Finger pulse oximeter | Predicate Device<br>K161560<br>Fingertip Pulse Oximeter | Remark |
| | 70%~80%: $\pm$ 2% | 0~69% no definition | |
| | 80%~90%: $\pm$ 2% | | |
| | 90%~100%: $\pm$ 2% | | |
| | No requirement for 70%<br>below | | |
# Discussion 6:
The SpO2 accuracy of the subject device and the predicate device are described in different ways. The accuracy of SpO2 of the subject device is ±2% at different levels (70%~80%, 80%~90%, 90%~100%, 70%~100%), which meets the requirements of ISO 80601-2-61. The accuracy of the predicate device at different levels is not mentioned. The accuracy of SpO2 of the two devices is 2% within the 70%~100% level. So, the difference does not raise any new issues of safety or efficacy. The subject device can be considered to be substantially equivalent to that of the predicate device.
| Resolution | 1% | 1% | SE |
|---------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------|--------------|
| Low Perfusion | 70%~100%: $ \u00b12% $<br>70%~80%: $ \u00b12% $<br>80%~90%: $ \u00b12% $<br>90%~100%: $ \u00b12% $<br>No requirement for 70% below | Not mentioned | Discussion 7 |
# Discussion 7 :
The accuracy of SpO2 at different levels (70%~80%, 80%~90%, 90%~100%, 70%~100%) of the subject device under low perfusion conditions is ± 2%, and the accuracy of SpO2 of the predicate device under low perfusion conditions is not mentioned. LEPU Intelligent Medical has verified the accuracy of SpO2 of the subject device under low perfusion conditions, and the results met the requirements of ISO 80601-2-61. So, the difference does not raise any new issues of safety or efficacy. The subject device can be considered to be substantially equivalent to that of the predicate device.
| | C<br>C | 466<br>O |
|--------|--------|----------|
| I<br>1 | | ) |
| Measuring range | 30bpm~250bpm | 30bpm~250bpm | SE |
|-----------------|---------------------------------------------------|-------------------------------------------|--------------|
| Accuracy | 30bpm~250bpm<br>±2bpm or ±2% (which is<br>larger) | 30bpm~99bpm, ±2bpm;<br>100bpm~250bpm, ±2% | Discussion 8 |
# Discussion 8:
The pulse rate accuracy of the subject device and the predicate device are described in different ways. The detail comparison is shown in the table below:
| Pulse rate range | Subject device | Predicate device | Conclusion |
|------------------|----------------|------------------|------------|
|------------------|----------------|------------------|------------|
{10}------------------------------------------------
| Shenzhen LEPU Intelligent Medical Equipment Co., Ltd. | | Fingertip pulse oximeter | |
|---------------------------------------------------------------------------------------------|-----------------------------------------------------|---------------------------------------------------------|--------------|
| Item | Subject Device<br>Finger pulse oximeter | Predicate Device<br>K161560<br>Fingertip Pulse Oximeter | Remark |
| 30bpm~99bpm | $\pm$ 2bpm ( $\pm$ 2bpm is<br>larger than $\pm$ 2%) | $\pm$ 2bpm | SE |
| 100bpm~250bpm | $\pm$ 2% ( $\pm$ 2% is larger<br>than $\pm$ 2bpm) | $\pm$ 2% | SE |
| So, the accuracy of pulse rate of the subject device and the predicate device are the same. | | | |
| Resolution | 1bpm | 1% | Discussion 9 |
## Discussion 9:
There is a difference between the resolution of the subject device and the predicate device. The display value of the pulse rate can only be an integer. The minimum resolution setting is 1bpm. According to the requirement of the pulse rate unit in ISO 80601-2-61, it is the number of pulses per minute. The pulse rate resolution of the subject device is set to 1bpm to meet the requirements of ISO 80601-2-61 and actual use. So, the difference does not raise any new issues of safety or efficacy. The subject device can be considered to be substantially equivalent to that of the predicate device.
| Low perfusion | 30bpm~250bpm<br>±2bpm or ±2% (which is larger) | Not mentioned | Discussion<br>10 |
|---------------|------------------------------------------------|---------------|------------------|
|---------------|------------------------------------------------|---------------|------------------|
# Discussion 10:
The accuracy of Pulse rate of the subject device under low perfusion conditions is ±2% (which is larger within 30bpm), and the accuracy of Pulse rate of the predicate device under low perfusion conditions is not mentioned. LEPU Intelligent Medical has verified the accuracy of Pulse rate of the subject device under low perfusion conditions, and the results met the requirements of ISO 80601-2-61. So, the difference does not raise any new issues of safety or efficacy. The subject device can be considered to be substantially equivalent to that of the predicate device.
| Environment requirement | | | |
|----------------------------------|------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|------------------|
| Operating<br>Temperature | 5°C~40°C | 5°C~40°C | Discussion<br>11 |
| Storage/Transport<br>temperature | -20°C~55°C | -25°C~70°C | |
| Relative<br>Humidity | ≤80%; No condensation in<br>working status<br>≤93%; No condensation in<br>storage status | 15%~93% no condensation<br>in operation;<br>≤93% no condensation in<br>storage/transport | |
| Atmospheric<br>pressure | 70kPa-106 kPa | 70kPa-106 kPa | |
| Discussion 11: | | | |
{11}------------------------------------------------
| Shenzhen LEPU Intelligent Medical Equipment Co., Ltd. | | Fingertip pulse oximeter | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------|--------------------------|--------|
| Item | Subject Device | Predicate Device | Remark |
| | Finger pulse oximeter | K161560 | |
| | | Fingertip Pulse Oximeter | |
| The subject device and the predicate device are different in storage/transportation temperature and<br>relative humidity. LEPU Intelligent Medical has verified the environmental requirements of the<br>device in accordance with the requirements in ISO 80601-2-61, and the results meet the | | | |
requirements. So, the difference does not raise any new issues of safety or efficacy. The subject
Item Subject Device Predicate Device Remark Finger pulse oximeter K161560 Fingertip Pulse Oximeter Particular Meeting the requirements of Meeting the requirements SE ISO 80601-2-61 of ISO 80601-2-61 requirements for basic safety and essential performance Electrical Safety Meeting the requirements of Meeting the requirements SE IEC 60601-1 and IEC 60601-1of IEC 60601-1 and IEC 11 60601-1-11 Electromagnetic Meeting the requirements of Meeting the requirements SE Compatibility IEC 60601-1-2 of IEC 60601-1-2 Biocompatibility Meeting the requirements of Not mentioned Discussion ISO 10993-1, ISO 10993-5, 12 ISO 10993-10 Clinical study Meeting the requirements of Meeting the requirements SE Annex EE of ISO 80601-2-61 of Annex EE of ISO 80601-2-61
# Table 5-3 Performance and Safety Comparison
device can be considered to be substantially equivalent to that of the predicate device.
Discussion 12: The fingertip pad and enclose of the subject device are in direct contact with the patient's tissue. According to the requirements of ISO10993 series standard and FDA Guidance: Pulse Oximeters - Premarket Notification Submissions [510(k)s], LEPU Intelligent Medical conducted the cytotoxicity test, irritation test and sensitization test on the device, and the test results met the requirements. So, the difference does not raise any new issues of safety or efficacy. The subject device can be considered to be substantially equivalent to that of the predicate device.
# 5.9. Performance Tests Summary
Bench test were conducted to verify that the subject device met all design specifications, as was Substantially Equivalent (SE) to the predicate device.
#### A Biocompatibility Testing
The Fingertip pulse oximeter was assessed against the International Standard ISO
{12}------------------------------------------------
Fingertip pulse oximeter
10993-1, "Biological evaluation of medical devices. Part 1. Guidance on selection of tests." The subject device would be classified as a Surface Medical Device in contact with the intact skin for a Limited Duration (<24 hours). The following test were performed for any user contacting material:
| Test | Standard | Results |
|-------------------------------------------------------------|--------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Cytotoxicity Study using<br>MTT Method | ISO 10993-5 | Under the conditions of this study, the MEM<br>extracts of test article would be considered no<br>cytotoxicity potential. The negative controls,<br>blank controls, and the positive controls<br>performed as anticipated. |
| Skin Sensitization Study<br>Guinea Pig Maximization<br>Test | ISO 10993-10 | Under the condition of this study, the test<br>article extracts showed no evidence of<br>causing delayed dermal contact sensitization<br>in the guinea pig. The test article was not<br>considered a sensitizer in the guinea pig<br>maximization test. |
| Skin Irritation Study | ISO 10993-10 | Under the conditions of this study, the<br>irritation response category of the test article<br>is classified as Negligible for polar extract and<br>Negligible for non-polar extract. |
#### Non-clinical Tests A
The Fingertip pulse oximeter is tested per the following standard, to evaluate its performance. The test results demonstrated that the proposed device comply with the standard requirements.
IEC 60601-1: 2005+AMD1:2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.
IEC 60601-1-2:2014 Medical Electrical Equipment Part 1-2: General Requirements for Basic Safety and Essential Performance Collateral Standard: Electromagnetic disturbances - Requirements and tests.
IEC 60601-1-11 Edition 2.0:2015-01 Medical Electrical Equipment Part 1-11: General requirements for basic safety and essential performance-Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
ISO 80601-2-61 Second Edition 2017-12 Medical Electrical Equipment Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment.
#### A Clinical data
A clinical study was conducted per the requirement of Annex EE of ISO 80601-2-61 to validate the SpO2 accuracy of Fingertip pulse oximeter. The purpose of the clinical
{13}------------------------------------------------
study was to evaluate the SpO2 accuracy performance of the Fingertip pulse oximeter during stationary (non-motion) conditions over a wide range of arterial blood oxygen saturation levels as compared to arterial blood co-oximeter. 12 healthy adult volunteer subjects (ages 18-50yr, with light to dark pigmentation, include male and female) were included in the study conducted to evaluate the SpO2 accuracy performance of proposed devices. Each system was evaluated during steady state/non-motion conditions with various levels of induced hypoxia resulting in stable oxygen saturation levels between 100% and 70% SaO2. Arterial blood samples were drawn during simultaneous data collection from the test devices. The blood was immediately analyzed on reference co-oximeter providing functional SaO2 for the basis of the SpO2 accuracy comparison. The SpO2 accuracy performance results showed the Fingertip pulse oximeter to have an Arms of 1.52% during steady state conditions over the range of 70-100% .
#### Software A
The software embedded in Fingertip pulse oximeter has been developed, documented, and validated in accordance with industry standards (IEC 62304 - Medical device software - Software life cycle processes) and FDA guidance (GUIDANCE FOR THE CONTENT OF PRE-MARKET SUBMISSIONS FOR SOFTWARE CONTAINED IN DEVICES). The software is of Moderate Level of Concern
#### A Cleaning and disinfection validation
The Fingertip pulse oximeter is reusable, and Shenzhen LEPU Intelligent Medical Equipment Co., Ltd. has verified the cleaning and disinfection process of the device. The Cleaning and disinfection Validation is compliance with FDA Guidance to Compliance with FDA Guidance for the Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling.
### 5.10. Substantially Equivalent Conclusion
The subject device and the predicate device have intended use, and the technological differences do not raise different questions of safety and effectiveness. The subject device, Fingertip pulse oximeter, is determined to be Substantially Equivalent (SE) to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.