The Fingertip Pulse Oximeter MD300CN310 is a handheld non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate of adult, adolescent and child patients in hospitals, hospital-type facilities and homecare.
Device Story
Battery-powered, handheld fingertip pulse oximeter; measures SpO2 and pulse rate non-invasively. Uses dual-wavelength LED (660nm red, 905nm infrared) and photo detector applied to pulsating arteriolar vascular bed. Device calculates SpO2 based on ratio of light absorption during systole and diastole. Features OLED display for SpO2, pulse rate, pulse bar, and waveform; 6 display modes. Operated by clinicians or patients for spot-checking in hospitals or homecare. Provides real-time physiological data to assist healthcare providers in monitoring patient oxygenation status.
Clinical Evidence
Bench testing only. Performance validated via SpO2 and pulse rate accuracy tests, weak perfusion testing, environmental (temperature/humidity) testing, and performance after cleaning. Compliance with ISO 80601-2-61, IEC 60601-1, IEC 60601-1-2, and IEC 60601-1-11 standards confirmed.
Technological Characteristics
Materials: ABS enclosure, medical silicone gel fingertip cushion. Sensing: Dual-wavelength LED (660nm/905nm) and photo detector. Power: 2x AAA alkaline batteries. Connectivity: Standalone. Software: Moderate level of concern. Sterilization: Non-sterile, reusable transducers.
Indications for Use
Indicated for spot-checking SpO2 and pulse rate in adult, adolescent, and child patients in hospital, hospital-type, and homecare settings.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three profiles facing to the right, representing the department's focus on people and health.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 5, 2016
Beijing Choice Electronic Technology Co., Ltd. Lei Chen Quality Director North Building 3F No.9 Shuangyuan Road, Badachu Hi-tech Zone Shijingshan District Beijing, 100041 CHINA
Re: K161560
Trade/Device Name: Fingertip Pulse Oximeter MD300CN310 Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: August 25, 2016 Received: August 29, 2016
Dear Lei Chen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{2}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
### Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Section II Indications for Use Statement
## Indications for Use
510(k) Number (if known): K161560
Device Name: Fingertip Pulse Oximeter MD300CN310
Indications for Use:
The Fingertip Pulse Oximeter MD300CN310 is a handheld non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate of adult, adolescent and child patients in hospitals, hospital-type facilities and homecare.
| Prescription Use | √ |
|-----------------------------|--------|
| (Part 21 CFR 801 Subpart D) | |
| | AND/OR |
| Over-The-Counter Use | |
| (21 CFR 801 Subpart C) | |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE
OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{4}------------------------------------------------
# Section III 510(k) Summary
This summary of 510(k) is being submitted in accordance with the requirements of 21 CFR 807.92.
The assigned 510(k) number is K161560 .
### 3.1 Submitter Information
#### ● Manufacturer Name:
Establishment Registration Number: 3005569927 Beijing Choice Electronic Technology Co., Ltd. Room 4104.No.A12 Yuquan Road Haidian District 100143 Beijing, P.R.China
#### ● Contact Person:
Mr.Lei Chen Beijing Choice Electronic Technology Co., Ltd. North Building 3F, No. 9 Shuangyuan Road, Badachu Hi-tech Zone, Shijingshan District Beijing China 100041 Phone: +86-10-88798300 Ext 6020 Fax:215-4052545 Email: cc@choicemmed.com
- Date prepared: September 19, 2016
### 3.2 Proposed Device Information
Device Common Name: Pulse Oximeter Device Trade/Proprietary Name: Fingertip Pulse Oximeter Model: MD300CN310 Classification Name: Oximeter Regulation Number: 870.2700 Product Code: DQA Class: II Panel: Anesthesiology
### 3.3 Predicate Device
510(k) Number: K140682
{5}------------------------------------------------
Premarket Notification 510(k) Submission-Section III 510(k) Summary
Common Name: Pulse Oximeter Device Trade/Proprietary Name: Fingertip Pulse Oximeter Model: MD300C29-H Classification Name: Oximeter Product Code: DQA Regulation Number: 870.2700
Device Class: II
Panel: Anesthesiology
Manufacturer: Beijing Choice Electronic Technology Co., Ltd.
Intended Use: The Fingertip Pulse Oximeter MD300C29-H is a portable, non-invasive device intended for spot checking of oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult, adolescent and child patient in hospital.
### 3.4 Device Description
The proposed device Fingertip Pulse Oximeter MD300CN310 is a battery powered device, which can detect and display the measured %SpO2 and pulse rate value. The device is adopted dual color OLED screen to display SpO2 and pulse rate value, pulse bar and waveform. The device has 6 display modes. The device is normally applied to adult, adolescent and child patients in hospital facilities and homecare environment.
The proposed device consists of power supply module, detector and emitter LED, signal collection and process module, display module, user interface and button control circuit.
Principle of the oximeter is as follows:
The pulse oximeter works by applying a sensor to a pulsating arteriolar vascular bed. The sensor contains a dual light source and photo detector. The one wavelength of light source is 660nm, which is red light; the other is 905nm, which is infrared-red light. Skin, bone, tissue and venous vessels normally absorb a constant amount of light over time. The photo detector in finger sensor collects and converts the light into electronic signal which is proportional to the light intensity. The arteriolar bed normally pulsates and absorbs variable amounts of light during systole and diastole, as blood volume increases and decreases. The ratio of light absorbed at systole and diastole is translated into an oxygen saturation measurement. This measurement is referred to as SpO2.
The proposed device is not for life-supporting or life-sustaining, not for implant.
The device is not sterile, and the transducers are reusable and do not need sterilization and re-sterilization.
The device is for prescription.
The device does not contain drug or biological products.
{6}------------------------------------------------
#### Premarket Notification 510(k) Submission—Section III 510(k) Summary
The device is software-driven and software validation is provided in Section XI Software.
{7}------------------------------------------------
# 3.5 Comparison list of the technological characteristics
| Comparison Elements | Proposed Device | Predicate Device | |
|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| Product Name | Fingertip Pulse Oximeter | Fingertip Pulse Oximeter | |
| Model | MD300CN310 | MD300C29-H | |
| Regulation No. | 21 CFR 870.2700 | 21 CFR 870.2700 | |
| Classification | II | II | |
| Classification Name | Oximeter | Oximeter | |
| Product Code | DQA | DQA | |
| Indications for Use | The Fingertip Pulse Oximeter MD300CN310 is a handheld non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate of adult, adolescent and child patients in hospitals, hospital-type facilities and homecare. | The Fingertip Pulse Oximeter MD300C29-H is a portable, non-invasive device intended for spot checking of oxygen saturation of arterial hemoglobin (SpO2) and pulse rate of adult, adolescent and child patient in hospital. | |
| Comparison Statement | The proposed device and the predicated device have the same classification and similar intended use. The difference exists that the proposed device can be used in the homecare environment while the predicate device can not. | | |
| Components | detector and emitter LED, signal amplify unit, CPU, data display unit and power unit | detector and emitter LED, signal amplify unit, CPU, data display unit and power unit | |
| Design Principle | The pulse oximeter works by applying a | The pulse oximeter works by applying a | |
| | | | |
| | | sensor to a pulsating arteriolar vascular bed.<br>The sensor contains a dual light source and | sensor to a pulsating arteriolar vascular<br>bed. The sensor contains a dual light |
| | | photo detector. The one wavelength of light | source and photo detector.<br>The one |
| | | source is 660nm, which is red light; the other | wavelength of light source is 660nm, |
| | | is 905nm, which is infrared-red light. Skin, | which is red light; the other is 905nm, |
| | | bone, tissue and venous vessels normally | which is infrared-red light. Skin, bone, |
| | | absorb a constant amount of light over time. | tissue and venous vessels normally absorb |
| | | The photo detector in finger sensor collects | a constant amount of light over time. The |
| | | and converts the light into electronic signal | photo detector in finger sensor collects and |
| | | which is proportional to the light intensity. | converts the light into electronic signal |
| | | The arteriolar bed normally pulsates and | which is proportional to the light intensity. |
| | | absorbs variable amounts of light during | The arteriolar bed normally pulsates and |
| | | diastole, as blood<br>systole<br>and<br>volume | absorbs variable amounts of light during |
| | | increases and decreases. The ratio of light | systole and diastole, as blood volume |
| | | absorbed at systole and diastole is translated | increases and decreases. The ratio of light |
| | | into an oxygen saturation measurement. This | absorbed<br>at systole and<br>diastole<br>is |
| | | measurement is referred to as SpO2. | into an<br>oxygen saturation<br>translated |
| | | | is<br>This<br>measurement.<br>measurement |
| | | | referred to as SpO2. |
| Measurement | Red | 660±3nm | 660±3nm |
| Wavelength | Infrared | 905 ± 10nm | 905 ± 8nm |
| Comparison Statement<br>measurement wavelength. | | The proposed device and the predicate device have the same design principle and | |
| | Display Type | OLED | OLED |
| Performance Specification | Working Time | Approximately 24 hours of continuous operation | Approximately 25 hours of continuous operation |
| | User Interface | 6 display modes | 6 directions for display |
| | Power supply | 2*AAA alkaline batteries | 2*AAA alkaline batteries |
| | Display Data | SpO2, PR | SpO2 ,PR |
| | SpO2 Display Range | 0%~100% | 35%~100% |
| | SpO2 Measurement Range | 70%~100% | 70%~100% |
| | SpO2 Accuracy | 70%~100%, ±2%;<br>0~69% no definition | 70%~100%, ±2%;<br>0~69% no definition |
| | SpO2 Resolution | 1% | 1% |
| | PR Display Range | 30bpm~250bpm | 30bpm~250bpm |
| | PR Measurement Range | 30bpm~250bpm | 30bpm~250bpm |
| | PR Accuracy | 30bpm~99bpm, ±2bpm;<br>100bpm~250bpm, ±2% | 30bpm~99bpm, ±2bpm;<br>100bpm~250bpm, ±2% |
| | PR Resolution | 1% | 1% |
Table 3-1 Performance Specification Comparison Table between the Proposed Device (MD300CN310) and Predicate Device
{8}------------------------------------------------
# Premarket Notification 510(k) Submission—Section III 510(k) Summary
{9}------------------------------------------------
Premarket Notification 510(k) Submission—Section III 510(k) Summary
{10}------------------------------------------------
| Premarket Notification 510(k) Submission-Section III 510(k) Summary | | |
|---------------------------------------------------------------------|--|--|
|---------------------------------------------------------------------|--|--|
| | | Operating<br>Temperature | 5°C~40°C | 5°C~40°C |
|-------------------------------|----------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|
| | | Storage/Transport<br>temperature | -25°C~70°C | -40°C~55°C |
| | | Relative<br>Humidity | 15%~93% no condensation in operation;<br>≤93% no condensation in storage/transport | ≤80% no condensation in operation;<br>≤93% no condensation in storage |
| | | Atmosphere<br>Pressure | 70kPa~106kpa | 86kPa~106kpa |
| | The proposed device has similar product specification as predicate device.<br>Comparison Statement | | | |
| Contacting<br>Material | Battery Cover | ABS | ABS | |
| | Enclosure | | | |
| | Fingertip Cushion<br>Power Button | Laser Etching Medical Silicone Gel | Laser Etching Medical Silicone Gel | |
| Comparison Statement | | | The contacting materials of the proposed device are same to those of the predicate<br>device. | |
| Performance Testing | | Laboratory Testing | The laboratory tests include SpO2 and PR<br>accuracy Test, Weak Perfusion Test, High and<br>Low Temperature and Humidity Test,<br>Performance Test After Cleaning and<br>ISO80601-2-61 | Meet the requirements of FDA Guidance |
| | | | | |
| | Electrical Safety | Conformed to IEC60601-1, IEC 60601-1-11 | Conformed to IEC60601-1 | |
| EMC<br>and<br>Electrical Safe | Electromagnetic<br>Compatibility | Conformed to IEC60601-1-2:2014 | Conformed to IEC60601-1-2:2007 | |
| | Software | Moderate level of concern | Moderate level of concern | |
| | | Compliance with FDA Guidance for the<br>content of Premarket Submissions for<br>Software Contained in Medical Devices | Compliance with FDA Guidance for the<br>content of Premarket Submissions for<br>Software Contained in Medical Devices | |
| | | Risk Management in Compliance with<br>ISO14971:2007 | Risk Management in Compliance with<br>ISO14971:2007 | |
| | Label and Labeling | Compliance with the Guidance of pulse<br>oximeter-premarket notification issued on<br>March 4.2013 | Compliance with FDA guidance | |
{11}------------------------------------------------
# Premarket Notification 510(k) Submission—Section III 510(k) Summary
{12}------------------------------------------------
# 3.6 Intended use
The Fingertip Pulse Oximeter MD300CN310 is a handheld non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate of adult, adolescent and child patients in hospitals, hospital-type facilities and homecare.
# 3.7 Testing
The Fingertip Pulse Oximeter MD300CN310 was supported by laboratory testing in order to ensure that they were appropriate performance and functional features to fully comply with recognized standards and is substantially equivalent to the predicate device. Non-clinical Test
The Fingertip Pulse Oximeter MD300CN310 is designed and tested and will be manufactured in accordance with both mandatory and voluntary standards, including: IEC 60601-1:2012 Medical Electrical Equipment-Part 1: General requirements for basic safety and essential performance.
IEC 60601-1-2:2014 Medical Electrical Equipment Part 1-2: General Requirements for Basic Safety and Essential Performance Collateral Standard: Electromagnetic disturbances - Requirements and tests.
IEC 60601-1-11:2010 Medical Electrical Equipment Part 1-11: General requirements for basic safety and essential performance-Collateral Standard: Requirements for mrdical electrical equipment and medical electrical systems used in the home healthcare environment.
ISO 80601-2-61:2011 Medical Electrical Equipment- Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment.
We have also conducted other performance test including SpO2 and PR Accuracy Test, Device Output Time and Finger Out Time Test, Device Response Time Test, Weak Perfusion Test, High and Low Temperature & Humidity Test Per Guidance for Industry and FDA Staff: Pulse Oximeter-Premarket Notification submission [510(k)s].
The Software Validation is in compliance with FDA Guidance to Compliance with FDA
Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.
| No. | Test Name |
|-----|----------------------------------------------|
| 1 | System Performance Test |
| 2 | Performance Test according to ISO 80601-2-61 |
The list of non-clinical test performed on the proposed devices.
{13}------------------------------------------------
| 3 | Electromagnetic Compatibility Test According to IEC 60601-1-2 |
|---|--------------------------------------------------------------------------|
| 4 | Electrical Safety Test According to IEC 60601-1 |
| 5 | Used in the home healthcare environment test according to IEC 60601-1-11 |
Premarket Notification 510(k) Submission-Section III 510(k) Summary
The test results indicate that the safety and effectiveness of the proposed device is identical to that of the predicate device.
#### 3.8 Determination of substantial equivalence
The proposed device of Fingertip Pulse Oximeter MD300CN310 has the same classification information, similar intended use, same design principle, similar product design and specifications as the predicated device. The main difference is that the proposed device can be used in the home healthcare environment while the predicate device can not. And we have conducted test according IEC 60601-1-11:2010 Medical Electrical Equipment Part 1-11: General requirements for basic safety and essential performance-Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment. According to the test results, the proposed device is as safe and as effective as the predicate device. So the proposed device is Substantially Equivalent (SE) to the predicate device which is US legally market device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.