The VS Newborn Heart Rate Monitor is indicated for the continuous monitoring of heart rate in neonatal patients. The device uses reflectance photoplethysmography to determine heart rate. The sensor is integrated into a cap placed on the head of the patient. The VS Newborn Heart Rate Monitor is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility.
Device Story
Non-invasive heart rate monitor for neonates; used in delivery rooms and NICUs. Device utilizes reflectance photoplethysmography (PPG) via a sensor integrated into a single-use soft cap worn on the newborn's head. PPG signals processed to determine heart rate; output displayed on monitor. Operated by healthcare professionals; provides real-time heart rate data to assist clinical decision-making during neonatal care. Benefits include continuous, non-invasive monitoring without requiring traditional adhesive electrodes.
Clinical Evidence
Clinical study at Nottingham University Hospitals NHS Trust (n=74 newborns, 13 mothers, 40 healthcare professionals). Assessed thermal properties, device function in NICU and C-section newborns. Primary endpoint: reliability and accuracy (Bland-Altman limits of agreement) compared to ECG electrodes (K024264) and LNCS pulse oximetry sensors (K110723). Secondary endpoints: acquisition time, user feedback. Safety endpoints: skin damage, adverse events. Results: device maintained temperature, demonstrated reliability/accuracy comparable to comparators, and showed no skin irritation or adverse events.
Technological Characteristics
Reflectance photoplethysmography (PPG) sensor integrated into single-use soft cap. Uses green visible light for HR determination. Biocompatibility per ISO 10993-1. Electrical safety per ISO 60601-1, IEC 60601-1-2, IEC 62133, ISO 80601-2-61. LED safety per IEC 62471. Usability per IEC 62366-1 and IEC 60601-1-6. Algorithm testing per JIS T 1303.
Indications for Use
Indicated for continuous heart rate monitoring in neonatal patients. Intended for use in professional healthcare facilities under supervision of licensed practitioners or trained personnel.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Predicate Devices
Masimo Rainbow SET® Rad 87 CO-Oximeter and Accessories (K091241)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
July 23, 2021
SurePulse Medical Limited % Allison Komiyama Prinicipal Consultant AcKnowledge Regulatory Strategies, LLC 2251 San Diego Ave, Suite B-257 San Diego, California 92110
Re: K201887
Trade/Device Name: VS Newborn Heart Rate Monitor Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DOA Dated: July 6, 2020 Received: July 8, 2020
Dear Allison Komiyama:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K201887
#### Device Name VS Newborn Heart Rate Monitor
#### Indications for Use (Describe)
The VS Newborn Heart Rate Monitor is indicated for the continuous monitoring of heart rate in neonatal patients. The device uses reflectance photoplethysmography to determine heart rate. The sensor is integrated into a cap placed on the head of the patient. The VS Newborn Heart Rate Monitor is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| <span style="text-decoration: overline;">\u2612</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: overline;">\u2610</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Surepulse. The logo consists of a stylized heartbeat symbol in blue, followed by the word "surepulse" in a gradient from blue to green. Below the word "surepulse" is the tagline "Every second counts" in a smaller, light gray font.
510(k) Summary
# 510(k) Summary
#### DATE PREPARED
July 6, 2020
### MANUFACTURER AND 510(k) OWNER
SurePulse Medical Limited Medicity, D6 Thane Road, Nottingham, NG90 6BH, United Kingdom Telephone: Official Contact: James Carpenter, CEO
#### REPRESENTATIVE/CONSULTANT
Allison C. Komiyama, Ph.D., R.A.C. Michelle Rubin-Onur, Ph.D. Lucie Dalet, Ph.D. AcKnowledge Regulatory Strategies, LLC Telephone: Email: akomiyama@acknowledge-rs.com Website: www.AcKnowledge-RS.com
### DEVICE INFORMATION
| Proprietary Name/Trade Name: | VS Newborn Heart Rate Monitor |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Oximeter |
| Regulation Number: | 21 CFR 870.2700 |
| Class: | II |
| Product Code: | DQA |
| Premarket Review: | Anesthesiology |
| Review Panel: | Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices (OHT1)<br>ENT, Sleep Disordered Breathing, Respiratory and<br>Anesthesia Devices (DHT1C) |
#### PREDICATE DEVICE IDENTIFICATION
The VS Newborn Heart Rate Monitor is substantially equivalent to the following predicate:
| 510(k) Number | Predicate Device Name / Manufacturer | Primary Predicate |
|---------------|--------------------------------------------------------------------------------|-------------------|
| K091241 | Masimo Rainbow SET® Rad 87 CO-Oximeter and<br>Accessories / Masimo Corporation | ✓ |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Surepulse. The logo consists of a stylized heartbeat symbol in blue, followed by the word "surepulse" in a gradient from blue to green. Below the word "surepulse" is the tagline "Every second counts" in a smaller, light gray font.
# 510(k) Summary DEVICE DESCRIPTION
The VS Newborn Heart Rate Monitor is a non-invasive device intended to continuously monitor heart rate (HR) in neonates immediately after delivery and in newborn intensive care units (NICU). The VS Newborn Heart Rate Monitor measures HR using reflectance photoplethysmography (PPG). The sensor is integrated into a single-use soft cap placed on the head of the newborn such that the sensor contacts the forehead of the newborn. The PPG signal and HR are displayed on the monitor. The VS Newborn Heart Rate Monitor is designed for use in hospitals.
# INDICATIONS FOR USE
The VS Newborn Heart Rate Monitor is indicated for the continuous monitoring of heart rate in neonatal patients. The device uses reflectance photoplethysmography to determine heart rate. The sensor is integrated into a cap placed on the head of the patient. The VS Newborn Heart Rate Monitor is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility.
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
SurePulse believes that the VS Newborn Heart Rate Monitor is substantially equivalent to the predicate device based on the information summarized here:
The subject device has a similar design and uses similar materials as the device cleared in K091241. The subject device has an equivalent intended use and similar technological characteristics as the device cleared in K091241. Unlike the predicate device that uses both green visible light and infrared light to measures heart rate and pulse oximetry, the VS Newborn Heart Rate Monitor only uses green visible light to determine heart rate. Unlike the predicate which is indicated for neonates, pediatric, infant, and adult populations, the VS Newborn Heart Rate Monitor is indicated for neonates only. To address the differences in technological characteristics, the subject device underwent biocompatibility testing, performance bench testing, and a clinical study.
# SUMMARY OF NON-CLINICAL TESTING
No FDA performance standards have been established for the VS Newborn Heart Rate Monitor.
Patient-contacting material was subjected to biocompatibility testing in compliance with ISO 10993-1 Biological evaluation of medical devices – Part 1: Evaluation and testing with a risk management process.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Surepulse. The logo features a stylized heartbeat symbol in blue, followed by the word "surepulse" in a gradient from blue to green. Below the word "surepulse" is the tagline "Every second counts" in a smaller, light gray font.
510(k) Summary
The following tests were performed to demonstrate safety based on current industry standards:
- . Basic safety and essential performance (per ISO 60601-1)
- Electromagnetic disturbances (per IEC 60601-1-2) .
- . Usability (per IEC 62366-1 and IEC 60601-1-6)
- Safety requirements for batteries (per IEC 62133)
- Basic safety and essential performance of pulse oximeter equipment (per ISO 80601-2-61)
- LED safety (per IEC 62471)
- o Algorithm (per JIS T 1303)
The results of these tests indicate that the VS Newborn Heart Rate Monitor is substantially equivalent to the predicate devices.
# SUMMARY OF CLINICAL TESTING
The VS Newborn Heart Rate Monitor underwent clinical testing at the Nottingham University Hospitals NHS Trust, in Nottingham, England. Three phases of the study were 1) assessing the thermal properties of the cap component 2) testing device function with NICU newborns, and 3) testing device function with C-section newborns. The study assessed multiple endpoints including primary, secondary, and safety endpoints. The primary endpoint of the study was to demonstrate the reliability and accuracy (i.e., Bland-Altman limits of agreement) of the subject device. The secondary endpoints included acquisition time and to obtain feedback about the subject device from mothers and healthcare professionals. Finally, there were two safety endpoints that were also considered throughout the study: any sign of skin damage due to the device, and any other unanticipated adverse events attributed to the device. The subject device was compared to ECG electrodes (K024264) and the LNCS pulse oximetry sensors (K110723). Those two devices were connected to the CARESCAPE Monitor B450 (132533) which was connected to a laptop computer. The subject device was connected to the same laptop computer and all data recorded synchronously. In total, 74 newborns, 13 mothers, and 40 healthcare professionals participated in the study. The subject device was found to maintain temperature as well as the standard of care and was as reliable and accurate as the devices it was compared to. Furthermore, there were no signs of skin irritation or adverse events therefore demonstrating that the VS Newborn Heart Rate Monitor is safe.
#### CONCLUSION
Based on the testing performed, including biocompatibility, electrical safety, and clinical testing, it can be concluded that the subject device does not raise new issues of safety or effectiveness compared to the predicate devices. The similar indications for use, technological characteristics, and performance characteristics for the proposed VS Newborn Heart Rate Monitor are assessed to be substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.