The Masimo MightySat Rx Fingertip Pulse Oximeter is indicated for the noninvasive spot checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). The Masimo MightySat Rx Fingertip Pulse Oximeter is indicated for use with adult and pediatric patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, and mobile environments.
Device Story
Fingertip pulse oximeter; utilizes Masimo SET technology for noninvasive measurement of SpO2, pulse rate (PR), perfusion index (PI), and optional pleth variability index (PVI). Device clips onto patient finger; uses spectrophotometry (red/infrared light absorption) and photoplethysmography to detect arterial blood flow. Operates via internal AAA batteries; features OLED display; includes optional Bluetooth for data transfer to mobile devices (secondary display). Used by clinicians or patients in hospitals, clinical settings, or mobile environments for spot checks; no alarms. Output provides real-time physiological data to assist healthcare providers in assessing patient oxygenation and perfusion status.
Clinical Evidence
Clinical validation performed per ISO 80601-2-61 on 14 healthy adult volunteers (male/female, diverse skin pigmentations). Total of 356 blood samples compared against laboratory CO-Oximeter. Results: SpO2 accuracy (Arms) of 1.66% (no motion) and 2.30% (motion), meeting requirements of ≤2% and ≤3% respectively.
Technological Characteristics
Fingertip pulse oximeter; plastic housing; OLED color display; powered by two AAA batteries. Sensing principle: spectrophotometry and photoplethysmography. Connectivity: optional Bluetooth. Standards: ISO 80601-2-61 (pulse oximetry), IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), ISO 10993-1/5/10 (biocompatibility).
Indications for Use
Indicated for noninvasive spot checking of SpO2 and pulse rate in adult and pediatric patients. Suitable for patients who are well or poorly perfused, under both motion and no-motion conditions, in hospital, hospital-type, and mobile environments.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
Masimo Root Monitoring System (Radius-7) (K142394)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, arranged in a way that suggests depth and connection. The profiles are simple and abstract, with smooth lines. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 22, 2015
Masimo Corporation Marguerite Thomlinson Sr. Director, Regulatory Affairs 52 Discovery Irvine, CA 92618
Re: K150314
> Trade/Device Name: Masimo MightySat Rx Fingertip Pulse Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: September 16, 2015 Received: September 18, 2015
Dear Ms. Thomlinson,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Teiashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Section 4. Indications for Use Statement
#### Indications for Use
510(k) Number: K150314
Device Name: Masimo MightySat Rx Fingertip Pulse Oximeter
#### Indications for Use:
The Masimo MightySat Rx Fingertip Pulse Oximeter is indicated for the noninvasive spot checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). The Masimo MightySat Rx Fingertip Pulse Oximeter is indicated for use with adult and pediatric patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, and mobile environments.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over The Counter Use (Part 21 CFR 801 Subpart D)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Masimo Corp. The logo consists of a red circle with a white plus sign inside, followed by the word "Masimo" in black, stylized font. To the right of the logo is the text "MASIMO CORP" followed by the address "Forty Parker, Irvine, CA 92618".
| Submitter and Address of<br>Manufacturing Facility: | Masimo Corporation<br>52 Discovery<br>Irvine, CA 92618<br>Phone: (949) 297-7000<br>FAX: (949) 297-7592 |
|-----------------------------------------------------|--------------------------------------------------------------------------------------------------------|
| Date: | October 21, 2015 |
| Contact: | Marguerite Thomlinson<br>Senior Director, Regulatory Affairs |
| Trade Name: | Masimo MightySat Rx Fingertip Pulse Oximeter |
| Common Name: | Oximeter |
| Classification Regulation/<br>Product Code: | 21 CFR 870.2700, Class II/DQA |
| Establishment Registration<br>Number: | 2031172 |
| Reason for Premarket<br>Notification: | New Device |
| Predicate Device: | K081285 – Nonin Medical, Inc. Onyx II Model 9560<br>K142394 - Masimo Root Monitoring System (Radius-7) |
| Performance Standards | No performance standards for the above device have been<br>promulgated pursuant to Section 514. |
#### Device Description
The subject device, MightySat Rx, is a fingertip pulse oximeter that includes Masimo SET technology for the measurement of functional oxygen arterial hemoglobin (SpO2) and pulse rate in adults and pediatrics. The device is a spot check pulse oximeter and does not include alarms. The device has the combined function of a pulse oximeter monitor and a reusable sensor. It includes an OLED color display, enclosed by plastic housing and powered by two alkaline AAA batteries. The MightySat Rx also includes optional Bluetooth wireless technology for the wireless transfer of patient data to mobiles devices, such as a smartphone.
The mobile device functions as a secondary display and it is not required for the intended use of pulse oximetry measurements. After measurements are made by the subject device, the data can be transferred to the mobile device via Bluetooth for the display and/or storage of the data.
{4}------------------------------------------------
# Section 5. 510(k) Summary
The MightySat Rx noninvasively measures and displays the following Masimo SET parameters:
- Functional Oxygen Saturation of Arterial Hemoglobin (SpO2): The amount of oxyhemoglobin expressed a percentage of the hemoglobin
- Pulse Rate (PR): Measured in beats per minute (BPM) is based on the optical detection of . peripheral flow pulse
- Perfusion Index (PI): The ratio of the pulsatile blood flow to the non-pulsatile in peripheral tissue. PI thus represents a non-invasive measurement of peripheral perfusion that can be obtained from the pulse oximeter.
- . Optional Pleth Variablility Index (PVI): Measure of dynamic changes in perfusion index that occur during the respiratory cycle. The calculation is accomplished by measuring changes in PI over a time interval where one or more complete respiratory cycles have occurred. PVI is displayed as a percentage (0-100%).
The MightySat Rx base device includes SpO2, PR and PI parameters. It can also include the optional PVI and/or Bluetooth wireless features.
| FEATURE | SPECIFICATION |
|---------------------------|-----------------------------------------------------------|
| Display | |
| | Display Type |
| | Display Range |
| | Display Waveform |
| | Display Resolution |
| | Measurement Accuracy in<br>Accordance with ISO 80601-2-61 |
| SpO2, No Motion | 70 – 100%, 2%, Arms, adults/pediatrics |
| SpO2, Motion | 70-100%, 3% Arms, adults/pediatrics |
| SpO2, Low Perfusion | 70–100%, 2%, Arms, adults/pediatrics |
| Pulse Rate, No Motion | 25 – 240 bpm, 3 bpm Arms, adults/pediatrics |
| Pulse Rate, Motion | 25 – 240 bpm, 5 bpm Arms, adults/pediatrics |
| Pulse Rate, Low Perfusion | 25 – 240 bpm, 3 bpm Arms, adults/pediatrics |
| Power | |
| | Internal battery |
| Interface | |
| | Wireless |
| Mechanical | |
| | Enclosure Material |
| | Dimensions/Weight |
| Environmental | |
| | Operating Temperature |
See the table below for the MightySat Rx specifications.
{5}------------------------------------------------
## Section 5. 510(k) Summary
| FEATURE | SPECIFICATION |
|-----------------------------|----------------------------------|
| Storage Temperature | -40°C to +70°C, ambient humidity |
| Operating/ Storage Humidity | 10% to 95%, non-condensing |
| Altitude | Up to 5,486 meters (18,000 feet) |
| Mode of Operation | |
| Mode of Operation | Spot check |
#### Intended Use/Indications for Use
The Masimo MightySat Rx Fingertip Pulse Oximeter is indicated for the noninvasive spot checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). The Masimo MightySat Rx Fingertip Pulse Oximeter is indicated for use with adult and pediatric patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, and mobile environments.
### Technological Characteristics
#### Principle of Operation
Pulse oximetry is governed by the following principles:
- Oxyhemoglobin (oxygentated blood) and deoxyhemoglobin (non-oxygenated blood) ● differ in their absorption of red and infrared light (spectrophotometry).
- The amount of arterial blood in tissue changes with your pulse (photoplethysymography). Therefore, the amount of light absorbed by the varying quantities of arterial blood changes as well.
#### Mechanism of Action for Achieving the Intended Effect
MightySat Rx automatically turns on when the device is opened . The device is then positioned on the patient's finger. Once the device is applied on the finger, it collects, processes physiological signals, and then displays Masimo SET measurements on the device's display screen. The MightySat Rx automatically turns off after removing the device from the finger.
#### Summary of Technological Characteristics of Subject Device Compared to Predicate Device
Similarities and Differences between Primary Predicate Device, Nonin Onyx II 9560 Pulse Oximeter (K081285) and Subject Device, MightySat Rx
The subject device, MightySat Rx, and the primary predicate device, Nonin Onyx II 9560 Pulse Oximeter (K081285), have the following key similarities:
{6}------------------------------------------------
# Section 5. 510(k) Summary
- · Both devices have the same intended use as a spot checking fingertip pulse oximeter with SpO2 and PR measurements;
- · Both devices have the same principle of operations of pulse oximetry;
- · Both devices do not have alarms;
- · Both devices have the same measurement site;
- · Both devices include wireless Bluetooth feature for transferring patient information from the device to a mobile device, such as a smartphone; and
- · Both devices are internally powered by two AAA batteries
The subject device, MightySat Rx, and the primary predicate device, Nonin Onyx II 9560 Pulse Oximeter (K081285), have the following key differences:
- · The subject device is indicated for use under motion conditions, whereas the predicate is not:
- · The subject device includes a color OLED display and the predicate includes an LED display; and
- · The subject device includes PVI calculation and the predicate does not have this feature.
#### Similarities and Differences between Predicate Device, Masimo Radius-7 (K142394) and Subject Device, MightySat Rx
The subject device, MightySat Rx, and the predicate device, Masimo Radius-7 (K142394), have the following key similarities:
- · Both devices include the Masimo SET technology;
- · Both devices display SpO2, PR, PL, PVI, and pleth and Signal IQ waveforms;
- · Both devices have the same principle of operation for pulse oximetry;
- · Both devices include wireless Bluetooth feature; and
- · Both devices have an OLED color display and touchpad interface.
The subject device, MightySat Rx, and the predicate device, Masimo Radius-7 (K142394), have the following key differences:
- The subject device is a spot check device without alarms and the predicate is a continuous monitoring device with alarms;
- The subject device combines the functionality of an instrument and a sensor into a single device, whereas the predicate is an instrument that needs to connect with a sensor; and
- · The subject device does not have acoustic respiratory rate measurement (RRa), whereas the predicate includes the RRa measurement.
{7}------------------------------------------------
# Section 5. 510(k) Summary
#### Non-clinical Testing
The subject device was subjected bench testing. The following non-clinical testing, as applicable, was performed in accordance with Masimo design control requirements and quality system to demonstrate substantial equivalence of the subject device with its predicates:
- Electrical safety testing per IEC60601-1 ●
- EMC testing per IEC60601-1-2 ●
- Pulse oximetry testing per ISO 80601-2-61
- Biocompatibility testing per ISO-10993-1, ISO-10993-5 and ISO-10993-10 ●
- . Usability testing per FDA Human Factors and Usability Draft Guidance
- Software verification per FDA Software Guidance
- Mechanical and environmental testing including: ●
- shock and vibration, O
- shipping drop test, o
- mechanical durability and O
- storage/transport and operational temperatures о
### Clinical Testing
The MightySat Rx was subjected to clinical testing. The functional oxygen saturation (SpO2) measurement has been validated in accordance with ISO 80601-2-61. The clinical testing was completed on a total of 14 healthy adult male and female volunteers (with at least a total of 356 blood samples) with light to dark skin pigmentations in the range of 70% to 100% against a laboratory CO-Oximeter. The SpO2 accuracy results include:
- 1.66% which meets the accuracy requirement of less than or equal to 2% under no ● motion condition and
- 2.30% which meets the accuracy requirement of less than or equal to 3% under motion conditions.
The accuracy specification is reported as accuracy root mean square (Arms).
#### Conclusion
The non-clinical and clinical testing provided in this 510(k) submission demonstrates that the subject device, MightySat Rx Fingertip Pulse Oximeter, as safe and as effective, and substantially equivalent to its predicates.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.