The Pulse Oximeter is a handheld non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate of adult, adolescent, child and infant patients, respiratory rate (RR) for adult in hospitals, hospital-type facilities and homecare environment.
Device Story
Handheld, battery-powered pulse oximeter; uses dual-wavelength (660nm red, 905nm infrared) LED sensor on fingertip to measure light absorption in arteriolar vascular bed; calculates SpO2, pulse rate, respiratory rate, and perfusion index. Used in hospitals, hospital-type facilities, and homecare by patients or clinicians for spot-checking. Device displays data on OLED screen; transmits data via Bluetooth 4.0 to mobile app for health tracking. Provides non-invasive monitoring; aids clinicians in assessing patient oxygenation and respiratory status. Not for life-support or continuous monitoring.
Clinical Evidence
Clinical study conducted on 33 subjects (11 healthy volunteers, 22 patients) to validate respiratory rate (RR) performance. Results showed mean error of ±1 rpm and root mean square error ≤2 rpm. SpO2 clinical performance relied on previously cleared data from predicate K181503 as no changes were made to the SpO2 measurement function. No adverse events reported.
Technological Characteristics
Battery-powered (2x AAA) handheld device; OLED display; Bluetooth 4.0 connectivity. Dual-wavelength sensor (660nm/905nm). Materials: ABS (enclosure/button), silicone gel (fingertip cushion). Standards: ISO 10993-1/5/10 (biocompatibility), IEC 60601-1/1-11 (safety), IEC 60601-1-2 (EMC), ISO 80601-2-61 (pulse oximeter performance).
Indications for Use
Indicated for non-invasive spot-checking of SpO2 and pulse rate in adult, adolescent, child, and infant patients; respiratory rate (RR) indicated for adults only. Intended for hospital, hospital-type, and homecare environments. Contraindicated for patients with cardiopulmonary disease, cardiac dysrhythmias, unstable patients, ICU patients, respiratory insufficiency, post-surgery/trauma recovery, morbid obesity, severe arrhythmia, injury/deformity, pregnancy/lactation, or use of blood-staining drugs. Not for apnea detection.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 11, 2022
Beijing Choice Electronic Technology Co., Ltd. Haiying Zhao Quality Director No.9 Shuangyuan Road, Badachu Hi-tech Zone, Shijingshan District Beijing, 100041 China
### Re: K211400
Trade/Device Name: Pulse Oximeter Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DOA Dated: February 11, 2022 Received: August 6, 2021
Dear Haiving Zhao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Todd Courtnev Assistant Director DHT1C: Division of Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory. ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K211400
Device Name Pulse Oximeter
Indications for Use (Describe)
The Pulse Oximeter is a handheld non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate of adult, adolescent, child and infant patients.
Respiratory rate (RR) is intended for adults only.
The Pulse Oximeter is intended for use in hospitals, hospital-type facilities and homecare environment.
The Pulse Oximeter is not intended for patients with:
- · cardiopulmonary disease
- · cardiac dysrhythmias, unstable patients, or patients in the ICU
- · respiratory insufficiency or recovering from surgery or trauma
- · morbid obesity
- · severe arrhythmia
The Pulse Oximeter is not intended for patients who are:
- · injured, disabled or physically deformed
- · pregnant or lactating
- · taking drugs that stain the blood.
The Pulse Oximeter is not designed to detect apneas.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <div> <span> <svg class="bi bi-check-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path> <path d="M10.97 4.97a.75.75 0 0 1 1.071 1.05l-3.992 4.99a.75.75 0 0 1-1.08.02L4.324 8.384a.75.75 0 1 1 1.06-1.06l2.094 2.093 3.473-4.425a.236.236 0 0 1 .02-.022z" fill-rule="evenodd"></path> </svg> Research Use (Part 21 CFR 361.2, Subpart B) </span> </div> |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg class="bi bi-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path> </svg> Same As Control Use (21 CFR 361.2, Subpart C) </span> </div> |
> Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section III 510(k) Summary
This summary of 510(k) is being submitted in accordance with the requirements of 21 CFR 807.92.
There is no prior submission for the device.
# 3.1 Submitter Information
#### ● Manufacturer Name:
Establishment Registration Number: 3005569927 Beijing Choice Electronic Technology Co., Ltd. 2nd Floor, 3rd Floor and Room 410-412 4th Floor, No. 2 Building, No. 9 Shuangyuan Road , Shijingshan District BEIJING, 100041, P.R.China
#### . Contact Person:
Haiying Zhao Beijing Choice Electronic Technology Co., Ltd. North Building 3F, No. 9 Shuangyuan Road, Badachu Hi-tech Zone, Shijingshan District Beijing China 100041 Phone: +86-10-88204631 Fax:861088204632 Email: cc(@choicemmed.com
- Date prepared : January 6, 2022
# 3.2 Proposed Device Information
Device Common Name: Pulse Oximeter Device Trade/Proprietary Name: Pulse Oximeter Model: MD300CI218R Classification Name: Oximeter Regulation Number: 870.2700 Product Code: DQA
Regulatory Class: Class II
Panel: Anesthesiology
{4}------------------------------------------------
# 3.3 Predicate Device
| Elements | Primary Predicate Device | Secondary Predicate Device |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K181956 | K181503 |
| Device Name | Masimo MightySat Rx Fingertip<br>Pulse Oximeter | Fingertip Pulse Oximeter |
| Model | / | MD300CI218 |
| Classification Name | Oximeter | Oximeter |
| Product Code | DQA,BZQ | DQA |
| Regulation Number | 870.2700 | 870.2700 |
| Regulatory Class | Class II | Class II |
| Panel | Anesthesiology | Anesthesiology |
| Applicant | Masimo Corporation | Beijing Choice Electronic<br>Technology Co., Ltd. |
| Indications for Use | The Masimo MightySat Rx<br>Fingertip Pulse Oximeter is<br>intended for hospitals,<br>hospital-type facilities,home<br>environments, and transport.<br>The Masimo MightySat Rx<br>Fingertip Pulse Oximeter is<br>indicated for the noninvasive<br>spot checking of functional<br>oxygen saturation of arterial<br>hemoglobin (SpO2) and pulse<br>rate (PR) for adult and pediatric<br>patients during both no motion<br>and motion conditions, and for<br>patients who are well or poorly<br>perfused.<br>The Masimo MightySat Rx<br>Fingertip Pulse Oximeter is<br>indicated for the noninvasive<br>spot checking of respiration rate<br>(RRn) for adult patients | The MD300CI218 is a handheld<br>non-invasive device intended for<br>spot-checking of oxygen<br>saturation of arterial hemoglobin<br>(SpO2) and Pulse Rate of adult,<br>adolescent, child and infant<br>patients in hospitals,<br>hospital-type facilities and<br>homecare environment. |
{5}------------------------------------------------
# 3.4 Device Description
The proposed device Fingertip Pulse Oximeter MD300CI218R is a battery powered device. It can detect and display the measured %SpO2 ,pulse rate,respiratory rate and perfusion index and will automatically power off when there is no signal for longer than 8 seconds. The proposed device is adopted colorful color OLED screen to display SpO2, PR , respiratory rate, perfusion index, and waveform which can be displayed in 2 directions. And it is designed with the battery indicator function to warn the user that the battery power mav be low.
The proposed device is a handheld non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate of adult, adolescent, child and infant patients, respiratory rate (RR) for adult in hospital-type facilities and homecare environment. And it can transmit the measurements to Smart Device installed the mobile APP via Bluetooth 4.0 to help the users to organize and track their health information.
The proposed device consists of power supply module, detector and emitter LED, signal collection and processor module, display module, Bluetooth module, user interface and button control.
The fingertip pulse oximeter works by applying a sensor to a pulsating arteriolar vascular bed. The sensor contains a dual light source and photo detector. The one wavelength of light source is 660nm, which is red light; the other is 905nm, which is infrared-red light. Skin, bone, tissue and venous vessels normally absorb a constant amount of light over time. The photo detector in finger sensor collects and converts the light into electronic signal which is proportional to the light intensity. The arteriolar bed normally pulsates and absorbs variable amounts of light during systole and diastole, as blood volume increases and decreases. The ratio of light absorbed at systole and diastole is translated into an oxygen saturation measurement. This measurement is referred to as SpO2.
The proposed device is not for life-supporting or life-sustaining, not for implant.
The device is not sterile and the transducers are reusable and do not need sterilization and re-sterilization.
The device is for prescription.
The device does not contain drug or biological products.
The device is software-driven and the software validation is provided in software.
{6}------------------------------------------------
# 3.5 Comparison list of the technological characteristics
| Comparison Elements | Proposed Device | Secondary Predicate Device | Primary Predicate Device |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Name | Pulse Oximeter | Fingertip Pulse Oximeter | Masimo MightySat Rx Fingertip<br>Pulse Oximeter |
| Model | MD300CI218R | MD300CI218 | Not Specified |
| Regulation No. | 21 CFR 870.2700 | 21 CFR 870.2700 | 21 CFR 870.2700 |
| Classification | Ⅱ | Ⅱ | Ⅱ |
| Classification Name | Oximeter | Oximeter | Oximeter |
| Product Code | DQA | DQA | DQA,BZQ |
| Indications for Use | The Pulse Oximeter is a handheld non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate of adult, adolescent, child and infant patients.<br>Respiratory rate (RR) is intended for adults only. | The MD300CI218 is a handheld non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate of adult, adolescent, child and infant patients in hospitals, hospital-type facilities and homecare environment. | The Masimo MightySat Rx Fingertip Pulse Oximeter is intended for hospitals hospital-type facilities, home environments, and transport.<br>The Masimo MightySat Rx Fingertip Pulse Oximeter is indicated for the noninvasive spot checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR) for |
Table 3-1 Performance Specification Comparison Table between the Proposed Device and Predicate Device
{7}------------------------------------------------
| Premarket Notification 510(k) Submission-Section III 510(k) Summary | |
|---------------------------------------------------------------------|--|
|---------------------------------------------------------------------|--|
| | intended for use in<br>hospitals, hospital-type<br>facilities and homecare<br>environment.<br>The Pulse Oximeter is not<br>intended for patients with:<br>● cardiopulmonary disease<br>● cardiac dysrhythmias,<br>unstable patients, or<br>patients in the ICU<br>● respiratory insufficiency<br>or recovering from surgery<br>or trauma<br>● morbid obesity<br>● severe arrhythmia<br>The Pulse Oximeter is not<br>intended for patients who<br>are:<br>● injured, disabled or<br>physically deformed<br>● pregnant or lactating<br>● taking drugs that stain the<br>blood.<br>The Pulse Oximeter is not<br>designed to detect apneas. | | both no motion and motion<br>conditions, and for patients who<br>are well or poorly perfused.<br>The Masimo Mighty Sat Rx<br>Fingertip Pulse Oximeter is<br>indicated for the noninvasive spot<br>checking of respiration rate (RRp)<br>for adult patients. |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Comparison Statement | The proposed device and the primary predicate device have the same intended use expect that the<br>proposed device can not be used in transport and poor perfused. | | |
{8}------------------------------------------------
| Components | Power supply module, detector and emitter LED, signal collection and processor module, display module, Bluetooth module, user interface and button control. | Power supply module, detector and emitter LED, signal collection and process module, display module, Bluetooth module, indicator module, user interface. | It includes an OLED color display, enclosed by plastic housing and powered by two alkaline AAA batteries. |
|------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Design Principle | The fingertip pulse oximeter works by applying a sensor to a pulsating arteriolar vascular bed. The sensor contains a dual light source and photo detector. The one wavelength of light source is 660nm, which is red light; the other is 905nm, which is infrared-red light. Skin, bone, tissue and venous vessels normally absorb a constant amount of light over time. The photo detector in finger sensor collects and converts the light into electronic signal which is proportional to the light | The fingertip pulse oximeter works by applying a sensor to a pulsating arteriolar vascular bed. The sensor contains a dual light source and photo detector. The one wavelength of light source is 660nm, which is red light intensity. The arteriolar bed normally pulsates and absorbs variable amounts of light during systole and diastole, as blood volume increases and decreases. The ratio of light absorbed at systole and diastole is translated into an oxygen saturation measurement. This measurement is referred to as SpO2. | The MightySat Rx adopts Masimo SET technology which uses a two-wavelength sensor to measure the indicated parameters based on light absorption principles of oxygenated blood and deoxygenated blood which generates a photoplethysmogram. Respiration rate (RRp) measures the respiration rate by analyzing cyclic variations in the photoplethysmogram due to respiration. |
{9}------------------------------------------------
| | | intensity. The arteriolar bed | | | |
|--|------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|----------------------|
| | | normally pulsates and | | | |
| | | absorbs variable amounts | | | |
| | | of light during systole and | | | |
| | | diastole, as blood volume | | | |
| | | increases and decreases. | | | |
| | | The ratio of light absorbed | | | |
| | | at systole and diastole is | | | |
| | | translated into an oxygen | | | |
| | | saturation measurement.<br>This measurement is<br>referred to as SpO2. | | | |
| | Respiration rate (RR) | | | | |
| | measures the respiration<br>rate by analyzing cyclic<br>variations in the<br>photoplethysmogram due<br>to respiration. | | | | |
| | Measurement<br>Wavelength | Red | 660±3nm | 660±3nm | Not Specified |
| | | Infrared | 905 ± 10nm | 905 ± 10nm | Not Specified |
| | Comparison Statement | | The proposed device and the primary predicate device have the same design principle.It has identical measurement wavalength as the second predicate device. | | |
| | | Display Type | OLED | OLED | OLED |
| | | User Interface | 2 display directions | 2 display directions | 2 display directions |
{10}------------------------------------------------
| | Power supply | 2* AAA alkaline batteries | 2* AAA alkaline batteries | 2*AAA alkaline batteries | RR<br>Measurement<br>Range | 4-45rpm | NA | 4-70 |
|--|------------------------------|-----------------------------------------------------------------|-------------------------------------------|--------------------------------------------------------------------------|-----------------------------------------|----------------------------------------------------------------------------------|-------------------------------------------------------------------------------|------------------------------|
| | Display Data | SpO2, PR,Respiratory<br>Rate, Perfusion Index,pulse<br>waveform | SpO2, PR | SpO2, PR,Respiratory<br>Rate,Perfusion Index, Pleth<br>Variability Index | RR Accuracy | Mean Error: ±1rpm | NA | 3 RPM ARMS, 1 RPM Mean Error |
| | SpO2 Display<br>Range | 0~100% | 0%~100% | 0%~100% | Perfusion Index<br>Measurement<br>Range | 0.1%-20% | NA | 0.02%-20% |
| | SpO2<br>Measurement<br>Range | 70%~100% | 70%~100% | 70%~100% | Perfusion Index<br>Resolution | 0.1% | NA | Not specified |
| | SpO2 Accuracy | 70%~100%, ±2%;<br>0~69% no definition | 70%~100%, ±2%;<br>0~69% no definition | 70%~100%, ±2% | Wireless<br>Transmission<br>Range | 0~10m | 0~10m | Not specified |
| | SpO2<br>Resolution | 1% | 1% | 1% | Antenna Type | Internal | Internal | Not specified |
| | PR Display<br>Range | 30bpm~250bpm | 30bpm~250bpm | 25bpm~240bpm | Transmitter | Bluetooth<br>Version 4.0 | Compliance: Bluetooth Compliance: Version 4.0 | Not specified |
| | PR<br>Measurement<br>Range | 30bpm~250bpm | 30bpm~250bpm | 25bpm~240bpm | Operating<br>Temperature | 5°C~40°C | 5°C~40°C | 5°C~40°C |
| | PR Accuracy | 30bpm~99bpm, ±2bpm;<br>100bpm~250bpm, ±2% | 30bpm~99bpm, ±2bpm;<br>100bpm~250bpm, ±2% | 3 BPM ARMS | Relative<br>Humidity | 15% ~93%, no<br>condensation in operation;<br>≤93% no condensation in<br>storage | 15% ~93%, no condensation in<br>operation;<br>≤93% no condensation in storage | 10% to 95%, non-condensing |
| | PR Resolution | 1 BPM | 1 BPM | 1 BPM | Atmosphere<br>Pressure | 70kPa~106kpa | 70kPa~106kpa | 54kPa~106kpa |
Premarket Notification 510(k) Submission—Section III 510(k) Summary
{11}------------------------------------------------
NA
4-70
# Premarket Notification 510(k) Submission-Section III 510(k) Summary
{12}------------------------------------------------
Premarket Notification 510(k) Submission—Section III 510(k) Summary
| Comparison Statement | The Proposed device has similar product specification as the primary predicate device . The proposed device and the secondary predicate device have identical product specification except that the proposed device can measure PI and RR. | | | |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contacting<br>Material | Battery Cover | ABS | ABS | Not specified |
| | Enclosure | | | |
| | Power Button | | | |
| | Fingertip<br>Cushion | Silicone Gel | Silicone Gel | Not specified |
| Comparison Statement | The contacting materials of the proposed device are same to those of the secondary predicate device. | | | |
| Performance Testing | Laboratory<br>Testing | The laboratory tests include SpO2 and PR accuracy Test, Weak Perfusion Test, High and Low Temperature and Humidity Test, Performance Test After Cleaning and ISO80601-2-61 | The laboratory tests include SpO2 and PR accuracy Test, Weak Perfusion Test, High and Low Temperature and Humidity Test, Performance Test After Cleaning and ISO80601-2-61 | Mechanical and environmental testing, Performance Test After Cleaning and ISO80601-2-61 |
| | Electrical<br>Safety | Conformed to IEC60601-1, IEC 60601-1-11 | Conformed to IEC60601-1, IEC 60601-1-11 | Conformed to IEC60601-1, IEC 60601-1-11 |
| EMC<br>Electrical Safe<br>and | Electromagnetic<br>Compatibility | Conformed to<br>IEC60601-1-2<br>Conformed to FCC<br>certification | Conformed to IEC60601-1-2<br>Conformed to FCC certification | Conformed to IEC60601-1-2<br>Conformed to FCC certification |
| Software and Cybersecurity | | Moderate level of concern | Moderate level of concern | Moderate level of concern |
| | | Compliance with:<br>FDA Guidance for the<br>content of Premarket<br>Submissions for Software<br>Contained in Medical<br>Devices and,<br>FDA Guidance for Content<br>of Premarket Submission<br>for Management of<br>Cybersecurity in Medical<br>Device | Compliance with:<br>FDA Guidance for the content of<br>Premarket Submissions for<br>Software Contained in Medical<br>Devices and,<br>FDA Guidance for Content of<br>Premarket Submission for<br>Management of Cybersecurity in<br>Medical Device | Compliance with:<br>FDA Guidance for the content of<br>Premarket Submissions for<br>Software Contained in Medical<br>Devices and,<br>FDA Guidance for Content of<br>Premarket Submission for<br>Management of Cybersecurity in<br>Medical Device |
| | | Risk Management in<br>Compliance with<br>ISO14971 | Risk Management in Compliance<br>with ISO14971 | Risk Management in Compliance<br>with ISO14971 |
| | Label and Labeling | Compliance with the<br>Guidance of pulse<br>oximeter-premarket<br>notification issued on<br>March 4 2013 | Compliance with the Guidance of<br>pulse oximeter-premarket<br>notification issued on<br>March 4,2013 | Compliance with the Guidance of<br>pulse oximeter-premarket<br>notification issued on<br>March 4,2013 |
{13}------------------------------------------------
Premarket Notification 510(k) Submission—Section III 510(k) Summary
{14}------------------------------------------------
# 3.6 Intended use
The Pulse Oximeter is a handheld non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin (SpO2) and Pulse Rate of adult, adolescent, child and infant patients, respiratory rate (RR) for adult in hospitals, hospital-type facilities and homecare environment.
# 3.7 Performance data
The following performance data were provided in support of the substantial equivalence determination.
### Biocompatibility Testing
The biocompatibility testing of the patient contacted materials used in the proposed device have been conducted according to the requirement of ISO 10993-1, ISO 10993-5, ISO 10993-10.
### Electrical safety and electromagnetic compatibility (EMC)
We have conducted IEC 60601-1,IEC 60601-1-11,IEC 60601-1-2,ISO 80601-2-61 test for the proposed device.
We have also conducted other performance test including SpO2 Accuracy Test, PR Accuracy Test, Reperatory Rate Test , Perfusion Index Test Device, Output Time and Finger Out Time Test, Device Response Time Test, , High and Low Temperature & Humidity Test Per Guidance for Industry and FDA Staff: Pulse Oximeter-Premarket Notification submission [510(k)s].
#### Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "Moderate" level of concern, since a failure or latent flaw could indirectly result in minor injury to the patient or operator through incorrect or delayed information or through the action of a care provider.
#### Wireless Testing
The hardare of the proposed device is same to those of the secondary predicate device
{15}------------------------------------------------
which was cleared in October 11.2018 by FDA. So testing reports performed and submitted with K181503 with repect to FCC and wireless coexistence also apply to the proposed device.
#### Clinical Test
In K181503, Choice include the clinical test of the secondary predicate device for the SpO2 measurement. Because there was no change in SpO2 measurement function of the device as part of this submission, the clinical study for SpO2 was not repeated.
For this submission to add the Respiratory Rate.Choice performed clinical test to validate the performance of Respiratory Rate. Choice conduted the clinial test on 11 heathy adult volunteer subjects and 22 patient. There were no adverse events during the study. The results of clinical study of respiratory rate is that the Mean Error is ± 1 and the Root mean square error is ≤ 2. The clinical study report was presented in Performance Testing-Clinical Study Report.
# 3.8 Conclusion
The proposed device has the same classification information, same intended use, similar design principle as the primary predicated device. The hardware of the proposed device is same as the secondary predicate device . The results of the testing demonstrate that all requirements and performance specifications were satisfied and the subject device is substantially equivalent to its predicate.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.