K052544 · American I.V. Products, Inc. · DQA · Sep 29, 2006 · Cardiovascular
Device Facts
Record ID
K052544
Device Name
PULSE OXIMETER ADAPTER CABLES
Applicant
American I.V. Products, Inc.
Product Code
DQA · Cardiovascular
Decision Date
Sep 29, 2006
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 870.2700
Device Class
Class 2
Indications for Use
These devices are intended to be used as replacement accessories to Original Equipment Manufacturer monitors used in the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. These systems are used to monitor patients who are either well or poorly perfused, in hospitals, hospital type facilities, intra-hospital transport and home environments.
Device Story
Pulse oximeter adapter cables; passive electrical interface between OEM pulse oximeter sensors and OEM patient monitors. Cables transmit raw sensor signals to monitor for processing; no active signal processing or data analysis performed by cable. Used in clinical and home environments; operated by healthcare professionals or patients. Facilitates continuous SpO2 and pulse rate monitoring; enables clinical decision-making based on monitor-generated data. Reusable; non-sterile.
Clinical Evidence
Bench testing only. No clinical data provided. Performance verified through electrical, mechanical, and safety testing, including conformity to IEC 60601-1 and IEC 60601-1-2 standards to demonstrate equivalence to OEM cables.
Technological Characteristics
Flexible, shielded, multi-conductor electrical cable. Keyed connectors for specific OEM monitor compatibility. Non-sterile. Passive electrical transmission; no active sensing or energy source. Complies with IEC 60601-1 (Safety) and IEC 60601-1-2 (EMC).
Indications for Use
Indicated for patients, well or poorly perfused, requiring continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate in hospital, intra-hospital transport, or home settings.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
K052544
Image /page/0/Picture/1 description: The image shows a logo with the letters "aiw" in a stylized, bold font. A five-pointed star is positioned above and slightly to the left of the letters, adding a visual element to the logo. The entire logo is black, creating a strong contrast against what is presumably a white or light-colored background.
Supporting Clinical Engineering Worldwide
SEP 2 9 2006
Appendix C Page 1 of 2
510(k) Summarv
## Submitter Information:
American IV Products, Inc. 7485 Shipley Avenue Hanover, MD 21076
### Contact:
Gregory Falk Director of Regulatory Affairs and Quality Assurance Telephone: 410-787-1300 ext. 129 Fax: 410-787-1337 e-mail: gfalk@aiv-inc.com
### Date Prepared:
September 15, 2005
#### Product Name:
Classification Name: Pulse Oximeter Adapter Cables Common Name: Pulse Oximeter Adapter Cables Proprietary Name: Pulse Oximeter Adapter Cables
## Predicate Device:
These AIV devices are equivalent to the following legally marketed devices:
Criticare Model: 518DD - K001020 Corometrics Model: 4033CAX - K934959 Datascope Model: 0012-00-0516-02 - K97006 & K993531 GE Medical Systems Model: E9004GE & 2006644-001 - K012467 & K033304 Nellcor: MC10 - K991823, K993637 & K973147 Nellcor: DEC8 - K971946 Nellcor: SCP10 - K991823, K993637 & K973147 Phillips Model: M1940A - K014159 Phillips Model: M1941A - K993383 Phillips Model: M1943A - K021453 Spacelabs Model: 700-0002-00 - K901209 Spacelabs Model: 700-0030-00 - K972502 Spacelabs Model: 175-0646-00 - K901209 Spacelabs Model: 700-0029-00 - K972502
#### Description:
AIV's Pulse Oximeter Adapter Cables are direct replacements for similar cables manufactured by the Original Equipment Manufacturers (OEM) for their respective monitors. They use the same type of construction and have the same technological characteristics as the predicate devices (OEM).
{1}------------------------------------------------
The OEM monitors are used for continuous, non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate.
The adapter cable connects the OEM pulse oximeter sensor with the OEM monitor.
## Intended Use:
These devices are intended to be used as replacement accessories to Original Equipment Manufacturer monitors used in the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. These systems are used to monitor patients who are either well or poorly perfused, in hospital type facilities, intra-hospital transport and home environments.
# Comparison to Predicate Device:
| | AIV | OEM |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| Intended Use | The adapter cable connects the OEM pulse oximeter sensor with the OEM monitor for continuous, non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. | Same |
| Type of Construction | Flexible, shielded, multi conductor electrical cable. | Same |
| Connector Design | Adapter cable connectors are keyed to fit the appropriate monitors. | Same |
| Installation into system | Installed in system between monitor and sensor using keyed connectors. | Same |
| Target Patient Population | Patients that are well or poorly perfused. | Same |
| Patient Use/Reuse | Reusable. | Same |
| Sterility | Non-sterile. | Same |
| IEC 60601-1 Testing | Passed Safety Testing. | Same |
| IEC 60601-1-2 Testing | Passed or when OEM system not passing with AIV cable, then OEM system tested with OEM adapter cable to show AIV equivalent to OEM adapter cable. | Same |
## Performance Data and Conclusions:
- . AIV design is equivalent to predicate device design.
- Bench Testing demonstrates that the AIV devices perform as intended. .
- . The company has declared conformity to consensus standards relating to Electrical/EMC/Mechanical/Safety
- . These devices do not raise new issues of safety and effectiveness, nor do they alter the fundamental technology of the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows a black and white abstract drawing. The drawing features three thick, curved lines that appear to be flowing or waving. The lines start from the upper right corner and descend towards the lower left, with each line having a distinct bend or curve. The overall design is simple yet dynamic, creating a sense of movement and fluidity.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 2 9 2006
Mr. Gregory Falk Director of Regulatory Affairs and Quality Assurance American I.V. Products. Incorporated 7485 Shipley Avenue Hanover, Maryland 21076
Re: K052544
Trade/Device Name: Pulse Oximeter Adapter Cables Regulation Number: 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: September 5, 2006 Received: September 6, 2006
Dear Mr. Falk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may. therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice. labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Falk
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that vour device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act s requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable. the electronic product radiation control provisions (Sections 531-542 of the Acti: 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): Page 1 of
Device Name: Pulse Oximeter Adapter Cables
Indications For Use:
These devices are intended to be used as replacement accessories to Original Equipment Manufacturer monitors used in the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate. These systems are used to monitor patients who are either well or poorly perfused, in hospitals, hospital type facilities, intra-hospital transport and home environments.
| AIV Part # | OEM Part # | OEM Monitor use |
|------------|------------------------|------------------------------------------------------------------------------------------------------------------------|
| CBCSI | 518DD | Criticare 504DX, 506DX & Alaris 8100 |
| CBCORO | 4033CAX | Corometrics 118 and Fetal Center Monitor |
| CBDAT | 0012-00-0516-02 | Datascope Passport, Accutor Plus, Sat 3 & Sat 4 |
| CB10640 | E9004GE<br>2006644-001 | GE Medical Systems Solar 8000M & 9500M, Solar Tram<br>Modules T451N & T851N, Dash 2000, 3000 & 4000 |
| CB10587 | MC10 | Nellcor NPB-290, NPB-295 & N-395 |
| CB10588 | DEC8 | Nellcor N-20, NPB-40, NPB-75, NPB-190<br>Welch Allyn Protocol Propaq<br>Most BCI Monitors |
| CB10641 | SCP10 | Nellcor NPB-290, NPB-295, N395 & N-3000 |
| CBHPHP1 | M1940A | Philips M1175A, M1176A, 78352A, 78352C, 78354A,<br>78354C, 7833A, M78833B, M1722A, M1722B, M1723A,<br>M1723B, & M2475B |
| CB10586 | M1941A | Philips M3000A Viridia, M3500B Heartstream XLT<br>Defib/Monitor & M4735A XL Defib/Monitor |
| CB10590 | M1943A | Philips M2601A Viridia, M3000A Viridia, M3500B<br>Heartstream XLT Defib/Monitor & M4735A XL<br>Defib/Monitor |
| CBSP1 | 700-0002-00 | Spacelabs Monitors with modules: 90351-D/G,<br>90465/66/67, 90489, 90651-A & 90651-08 |
| CBSP2 | 700-0030-00 | Spacelabs Ultraview 1050, 1600, 1700 & all monitors<br>using Ultraview Command Module 90496 |
{5}------------------------------------------------
# Appendix B Page 2 of 2
Page 2 of _ 2__
| CBSPNV1 | 175-0646-00 | Spacelabs Monitors with modules: 90351-D/G,<br>90465/66/67, 90489 90651-A & 90651-08 |
|---------|-------------|---------------------------------------------------------------------------------------------|
| CBSPNV2 | 700-0029-00 | Spacelabs Ultraview 1050, 1600, 1700 & all monitors<br>using Ultraview Command Module 90496 |
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Oles
mesthesiology, General Hosp Is, Junian Dental Devices
ter K052564
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.