Stethophone Pro

K252595 · Sparrow Acoustics, Inc. · DQD · Sep 12, 2025 · Cardiovascular

Device Facts

Record IDK252595
Device NameStethophone Pro
ApplicantSparrow Acoustics, Inc.
Product CodeDQD · Cardiovascular
Decision DateSep 12, 2025
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1875
Device ClassClass 2
AttributesAI/ML, Software as a Medical Device

AI Performance

OutputAlgorithmAcceptanceObservedDev DSDev ReadersTest DSTest Readers
Heart murmur detection———————
Heart sound timing and characterization———————

Indications for Use

Stethophone Pro is an electronic stethoscope that enables detection, amplification, filtering, and transmission of auscultation sound data (heart and lungs), whereby a clinician at one location can listen to the auscultation sounds of a patient acquired on site or at a different location. Stethophone Pro is intended for use on adult patients. Stethophone Pro is not intended for self-diagnosis and not intended to be used as a sole means of diagnosis. Stethophone Pro can be used in clinical and nonclinical environments. For Rx-only: Stethophone Pro is intended to provide decision support to clinicians in their evaluation of patients' heart sounds. The software analyzes heart sounds and phonocardiograms and can automatically detect murmurs that may be present, sound timing and character, including S1, S2, and the absence of a heart murmur. The interpretations of heart sounds offered by the software are not diagnoses and are meant only to provide decision support to the clinician, who may use the result in conjunction with their own evaluation and clinical judgment. For OTC: When used without access to the automatic analysis results or under supervision of healthcare professional, Stethophone Pro is also intended to be used by lay users.

Device Story

Stethophone Pro is a smartphone-based electronic stethoscope application; captures chest sounds via smartphone microphone; provides real-time listening, recording, and cloud storage. Features include three audio filters (Bell, Diaphragm, Starling) and sound visualization (oscillogram, spectrogram). Used by clinicians in clinical/non-clinical settings and by lay users at home. Rx-only features include automated murmur detection, S1/S2 identification, and heart rate calculation. Clinicians use output as decision support alongside clinical judgment; lay users can share recordings with physicians for remote assessment. Benefits include improved access to auscultation in telemedicine and remote monitoring contexts.

Clinical Evidence

No clinical data. Bench testing only; performance characteristics and algorithms are identical to previously cleared predicates K240901 and K231551.

Technological Characteristics

Standalone smartphone software (iOS). Uses smartphone microphone for sound capture. Features digital signal processing for three audio filters (Bell, Diaphragm, Starling) and visualization (oscillogram, spectrogram). Automated analysis for murmur detection and S1/S2 timing. Connectivity via cloud storage for data transmission.

Indications for Use

Indicated for adult patients for detection, amplification, filtering, and transmission of heart and lung sounds. Rx-only: provides decision support for heart sound evaluation, including murmur detection and S1/S2 timing. OTC: intended for lay users to capture and transmit sounds.

Regulatory Classification

Identification

Manual stethoscope: A mechanical device used to project the sounds associated with the heart, arteries, and veins and other internal organs. Electronic stethoscope: An electrically amplified device used to project the sounds associated with the heart, arteries, and veins and other internal organs.

Special Controls

*Classification.* Class II (special controls). The device, when it is a lung sound monitor, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION September 12, 2025 Sparrow Acoustics Inc. Nadia Ivanova Chief Product Officer 2416 Natura Dr. Lucasville, NS B4B 0X3 Canada Re: K252595 Trade/Device Name: Stethophone Pro Regulation Number: 21 CFR 870.1875 Regulation Name: Stethoscope Regulatory Class: Class II Product Code: DQD Dated: August 11, 2025 Received: August 15, 2025 Dear Nadia Ivanova: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" {1} K252595 - Nadia Ivanova Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. {2} K252595 - Nadia Ivanova Page 3 For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Stephen C. Browning -S LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. 510(k) Number (if known) K252595 Device Name Stethophone Pro Indications for Use (Describe) Stethophone Pro is an electronic stethoscope that enables detection, amplification, filtering, and transmission of auscultation sound data (heart and lungs), whereby a clinician at one location can listen to the auscultation sounds of a patient acquired on site or at a different location. Stethophone Pro is intended for use on adult patients. Stethophone Pro is not intended for self-diagnosis and not intended to be used as a sole means of diagnosis. Stethophone Pro can be used in clinical and nonclinical environments. For Rx-only: Stethophone Pro is intended to provide decision support to clinicians in their evaluation of patients' heart sounds. The software analyzes heart sounds and phonocardiograms and can automatically detect murmurs that may be present, sound timing and character, including S1, S2, and the absence of a heart murmur. The interpretations of heart sounds offered by the software are not diagnoses and are meant only to provide decision support to the clinician, who may use the result in conjunction with their own evaluation and clinical judgment. For OTC: When used without access to the automatic analysis results or under supervision of healthcare professional, Stethophone Pro is also intended to be used by lay users. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (6/20) Page 1 of 1 PSC Publishing Services (301) 443-6740 {4} SPARROW ACOUSTICS Stethophone Pro – Special 510(k): Device Modification – K252595 # 510(k) SUMMARY | 1 | Summary Date | September 10, 2025 | | --- | --- | --- | | 2 | Submitter | Sparrow Acoustics Inc. 2416 Natura Dr. Lucasville, NS Canada, B4B 0X3 Tel: +1 (902) 989-3908 | | 3 | Correspondent | Nadia Ivanova | | 4 | Device Trade Name | Stethophone Pro | | 5 | Device Common Name | Smartphone stethoscope | | 6 | Classification | Electronic Stethoscope 21 CFR 870.1875(b) Class II Product Code: DQD Panel: Cardiovascular | | 7 | Intended Use / Indications for Use | | Stethophone Pro is an electronic stethoscope that enables detection, amplification, filtering, and transmission of auscultation sound data (heart and lungs), whereby a clinician at one location can listen to the auscultation sounds of a patient acquired on site or at a different location. Stethophone Pro is intended for use on adult patients. Stethophone Pro is not intended for self-diagnosis and not intended to be used as a sole means of diagnosis. Stethophone Pro can be used in clinical and nonclinical environments. For Rx-only: Stethophone Pro is intended to provide decision support to clinicians in their evaluation of patients' heart sounds. The software analyzes heart sounds and phonocardiograms and can automatically detect murmurs that may be present, sound timing and character, including S1, S2, and the absence of a heart murmur. The interpretations of heart sounds offered by the software are not diagnoses and are meant only to provide decision support to the clinician, who may use the result in conjunction with their own evaluation and clinical judgment. For OTC: When used without access to the automatic analysis feature or under supervision of healthcare professional, Stethophone Pro is also intended to be used by lay users. 510(k) Summary {5} SPARROW ACOUSTICS Stethophone Pro – Special 510(k): Device Modification – K252595 # 8 Device Description Stethophone Pro is an electronic stethoscope software application that operates on smartphones. Stethophone enables the capture and amplification of chest sounds for real-time or recorded listening. Cloud storage with sound record sending capabilities, filtering for selected frequency ranges, and visualization all assist with sound analysis. Stethophone is designed to assist healthcare professionals in both hearing and visualizing heart and lung sounds during a physical examination of a patient and in storing recorded sounds in the cloud for later analysis. It also enables home users to record and send chest sounds to their physicians. Stethophone can be used for the assessment of chest sounds of adult patients in clinical and non-clinical environments. Assessment should be performed by healthcare professionals, while sound capturing can be performed by both healthcare professionals and home users. While all functions described above are available to both prescription and OTC users, the following analysis features and the associated report are accessible exclusively with a prescription: Stethophone Pro performs basic analysis of heart sounds, including detection of murmurs, identification of S1 and S2 heartbeats on the timeline, and calculation of timing intervals between them. ## Key Product Features Available for both prescription and OTC users - Capturing chest sounds using the smartphone microphone: - Real-time listening to chest sounds - Recording of chest sounds - Sending examinations to specialists for assessment - Two modes of sound visualization: oscillogram and spectrogram - Selecting from three audio filters for listening: - Bell — Traditional filter used in stethoscopes for low frequency sounds - Diaphragm — Traditional filter used for higher frequency sounds of heart and lungs - Starling — Filter for listening to the full frequency of chest sounds All of these features are present in both the primary predicate, Stethophone Pro (K240901), and the reference device, Stethophone (K231551), and are now available for OTC use. ## Product Features For prescription users only - Detecting murmurs, timing for S1 and S2 sounds, and calculating heart rate These features are derived from the Stethophone Pro (K240901, Rx-only). Collectively, these features enable home users to acquire their own sounds, share them with healthcare professionals, and control their health, as well as allow healthcare professionals to examine and monitor patients on site and remotely, seek out second opinions from specialists, and use the device in a telemedicine context. # 9 Predicate Devices ## Stethophone Pro (primary) 510(k) Number: K2409011 Manufacturer: Sparrow Acoustics Inc. Product Code: DQD Classification: 870.1875 ## Stethophone (reference) 510(k) Number: K2315512 Manufacturer: Sparrow Acoustics Inc. Product Code: DQD Classification: 870.1875 510(k) Summary {6} SPARROW ACOUSTICS Stethophone Pro – Special 510(k): Device Modification – K252595 # Comparison to Predicates The comparison chart below provides evidence to facilitate the substantial equivalence determination between Stethophone to the predicate devices with respect to intended use, technological characteristics, and principles of operation: | | Stethophone Pro version 3.6.0 | Stethophone Pro versions 3.0.0 - 3.5.0 | Stethophone version 2.x.x | | --- | --- | --- | --- | | Classification | Class II device | Class II device | Class II device | | US FDA Regulation | 21 CFR 870.1875(b) | 21 CFR 870.1875(b) | 21 CFR 870.1875(b) | | Product code | DQD | DQD | DQD | | Device | Stethoscope, Electronic | Stethoscope, Electronic | Stethoscope, Electronic | | US FDA Clearance | Pending | K240901 | K231551 | | Intended Use / Indications for Use | Stethophone Pro is an electronic stethoscope that enables detection, amplification, filtering, and transmission of auscultation sound data (heart and lungs), whereby a clinician at one location can listen to the auscultation sounds of a patient acquired on site or at a different location. Stethophone Pro is intended for use on adult patients. Stethophone Pro is not intended for self-diagnosis and not intended to be used as a sole means of diagnosis. Stethophone Pro can be used in clinical and nonclinical environments. For Rx-only: Stethophone Pro is intended to provide decision support to clinicians in their evaluation of patients' heart sounds. The software analyzes heart sounds and phonocardiograms and can automatically detect murmurs that may be present, sound timing and character, including S1, S2, and the absence of a heart murmur. The interpretations of heart sounds offered by the software are not diagnoses and are meant only to provide decision support to the clinician, who may use the result in conjunction with their own evaluation and clinical judgment. For OTC: When used without access to the automatic analysis feature or under supervision of healthcare professional, Stethophone Pro is also intended to be used by lay users. | Stethophone Pro is an electronic stethoscope that enables detection, amplification, filtering, and transmission of auscultation sound data (heart and lungs). Stethophone Pro is intended to provide decision support to clinicians in their evaluation of patients' heart sounds. The software analyzes heart sounds and phonocardiograms and can automatically detect murmurs that may be present, sound timing and character, including S1, S2, and the absence of a heart murmur. Stethophone Pro is not intended to be used as a sole means of diagnosis and is for use in environments where health care is provided by clinicians. The interpretations of heart sounds offered by the software are meant only to provide decision support to the clinician, who may use the result in conjunction with their own evaluation and clinical judgment. The interpretations are not diagnoses. Stethophone Pro is intended for use on adult patients. | Stethophone is an electronic stethoscope that enables detection, amplification, filtering, and transmission of auscultation sound data (heart and lungs), whereby a clinician at one location can listen to the auscultation sounds of a patient acquired on site or at a different location. Stethophone is intended for use on adult patients. Stethophone is intended to be used by professional or lay users in a clinical or nonclinical environment. Stethophone is not intended for self-diagnosis. | | Prescription/ OTC | Prescription and OTC | Prescription | Prescription and OTC | | Population Type | Adults | Adults | Adults | | Device Type | Standalone software for smartphone | Standalone software for smartphone | Standalone software for smartphone | | Dedicated Device vs. Smartphone | Operates on smartphone using its hardware and operating system | Operates on smartphone using its hardware and operating system | Operates on smartphone using its hardware and operating system | 510(k) Summary {7} SPARROW ACOUSTICS Stethophone Pro – Special 510(k): Device Modification – K252595 | iPhone model Compatibility | XS, XS Max, XR, SE 2^{nd} gen, 11, 11 Pro, 11 Pro Max, 12, 12 Pro, 12 Mini, 12 Pro Max, 13, 13 mini, 13 Pro, 13 Pro Max, SE 3^{rd} gen, 14, 14 Plus, 14 Pro, 14 Pro Max, 15, 15 Plus, 15 Pro, 15 Pro Max, 16, 16 Plus, 16 Pro, 16 Pro Max, 16e | XS, XS Max, XR, SE 2^{nd} gen, 11, 11 Pro, 11 Pro Max, 12, 12 Pro, 12 Mini, 12 Pro Max, 13, 13 mini, 13 Pro, 13 Pro Max, SE 3^{rd} gen, 14, 14 Plus, 14 Pro, 14 Pro Max, 15, 15 Plus, 15 Pro, 15 Pro Max, 16, 16 Plus, 16 Pro, 16 Pro Max, 16e | XS, XS Max, XR, SE 2^{nd} gen, 11, 11 Pro, 11 Pro Max, 12, 12 Pro, 12 Mini, 12 Pro Max, 13, 13 mini, 13 Pro, 13 Pro Max, SE 3^{rd} gen, 14, 14 Plus, 14 Pro, 14 Pro Max, 15, 15 Plus, 15 Pro, 15 Pro Max, 16, 16 Plus, 16 Pro, 16 Pro Max, 16e | | --- | --- | --- | --- | | Software Platform | iOS | iOS | iOS | | Sound Type | Heart, lungs | Heart, lungs | Heart, lungs | | Sound Processing | Yes | Yes | Yes | | Ability to Visualize Sound | Spectrogram, oscillogram | Spectrogram, oscillogram | Spectrogram, oscillogram | | Murmur detection | Yes | Yes | No | | Heart sound timing | Yes | Yes | No | | Signal quality assessment | Yes | Yes | No | 11 Performance Data The acoustics algorithms for sound filtering (that are used for both OTC and Rx parts of the software) capabilities remain entirely unchanged from those validated and cleared in both K240901 and K231551. Automated sound analysis capabilities (that are available for Rx use only) remain entirely unchanged from those validated and cleared in K240901. Therefore, no new performance data is required. 12 Biocompatibility Testing Not Applicable (Standalone Software) 13 Sterilization & Shelf-life Testing Not Applicable (Standalone Software) 14 Electrical safety and electromagnetic compatibility (EMC) Not Applicable (Standalone Software) 15 Statement of Substantial Equivalence Stethophone Pro is a modified version of the previously cleared primary predicate Stethophone Pro K240901 and reference Stethophone K231551. All technological characteristics are the same, including the entire acoustics features and automated analysis. This submission was solely intended to address a modification to the indications for use. Stethophone Pro and the predicates have the same intended use and similar indications, technological characteristics and principles of operation, and had no modifications to affect performance. The minor differences in indications for use do not present different questions of safety or effectiveness than the predicate devices. This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92. 510(k) Summary Page 4 of 4
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...