K251983 · Aug 26, 2025
Brainomix 360 Triage Stroke
Brainomix Limited
Retrospective clinical NCCT head image datasets
Retrospective clinical data was used to validate the sensitivity and specificity of the device's ICH and LVO detection algorithms across various clinical subgroups.
Retrospective study; NCCT; ICH detection; LVO detection; Performance validation
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
ICH Standalone Performance Study; Retrospective study
Adult patients in the acute setting; Sample Size: 341 cases (167 ICH positive; 174 ICH negative)
Not applicable for this study
Sensitivity and specificity of ICH detection
LVO and ICH Standalone Performance Study; Retrospective study
Adult patients in the acute setting; Sample Size: 267 cases (112 LVO positive; 40 ICH positive; 115 negative)
Reader study comparing device sensitivity to radiologists.
>1 (expert readers) + >1 (general radiologists)
Indications for Use
Brainomix 360 Triage Stroke is a radiological computer aided triage and notification software indicated for use in the analysis of non-contrast head CT (NCCT) images to assist hospital networks and trained clinicians in workflow triage by flagging and communicating suspected positive findings of head NCCT images for large vessel occlusion (LVO) of the intracranial ICA and M1 or intracranial hemorrhage (ICH). Specifically, the device is intended to be used for the triage of images acquired from adult patients in the acute setting, within 24 hours of the onset of the acute symptoms, or where this is unclear, since last known well (LKW) time. It is not intended to detect symmetrical bilateral MCA occlusions. Brainomix 360 Triage Stroke uses an artificial intelligence algorithm to analyze images and highlight cases with detected NCCT LVO or ICH on the Brainomix server on premise or in the cloud in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected LVO or ICH findings via a web user interface or mobile application. Notifications include compressed preview images that are meant for informational purposes only and are not intended for diagnostic use beyond notification. The device does not alter the original medical image, and it is not intended to be used as a primary diagnostic device. The results of Brainomix 360 Triage Stroke are intended to be used in conjunction with other patient information and based on professional judgment to assist with triage/prioritization of medical images. Notified clinicians are ultimately responsible for reviewing full images per the standard of care.
Device Story
Software-only device; processes non-contrast head CT (NCCT) images; operates on-premise or cloud server. Uses AI/ML algorithms to detect suspected ICH (hyperdense parenchyma) and LVO (ASPECTS and hyperdense vessel sign). Outputs priority notifications to clinicians via web UI or mobile app; provides compressed preview images for informational triage only. Used in hospital/clinical settings by trained clinicians; operates in parallel to standard-of-care image interpretation. Does not remove cases from reading queues; does not replace primary diagnostic review. Benefits include workflow prioritization; faster identification of suspected stroke/hemorrhage cases; improved clinical decision-making speed.
Clinical Evidence
Three retrospective studies conducted. ICH standalone study (n=341): 96.41% sensitivity, 96.55% specificity. LVO/ICH study (n=267): LVO sensitivity 69.64%, specificity 89.57%; ICH sensitivity 95.00%, specificity 88.11%. Reader study (n=unknown): Device sensitivity 47.94% vs. radiologist sensitivity 47.18% (non-expert) and expert readers. Time-to-notification met 3.5-minute acceptability criterion (range 58.3–150.7 seconds).
Technological Characteristics
Software-only; DICOM-compliant; runs on off-the-shelf physical/virtual servers. Uses AI/ML (CNN architecture) for image analysis. Connectivity: Web UI and mobile app. Workflow: Automated DICOM routing, image processing, and notification. Validated for Siemens, GE, and Philips scanners. Compliant with IEC 62304 (software development) and ISO 14971 (risk management).
Indications for Use
Indicated for adult patients in acute setting (within 24 hours of symptom onset or LKW) to triage NCCT images for suspected intracranial hemorrhage (ICH) or large vessel occlusion (LVO) of intracranial ICA and M1. Contraindicated for patients with recent (within 6 weeks) neurosurgery/endovascular intervention, recent (within 4 weeks) stroke, or scan data containing tumors, abscesses, coils, shunts, embolization, movement artifacts, or vascular pathologies (aneurysms, AVMs, venous thrombosis). Not for detecting symmetrical bilateral MCA occlusions.
Regulatory Classification
Identification
Radiological computer aided triage and notification software is an image processing prescription device intended to aid in prioritization and triage of radiological medical images. The device notifies a designated list of clinicians of the availability of time sensitive radiological medical images for review based on computer aided image analysis of those images performed by the device. The device does not mark, highlight, or direct users' attention to a specific location in the original image. The device does not remove cases from a reading queue. The device operates in parallel with the standard of care, which remains the default option for all cases.
Special Controls
Radiological computer aided triage and notification software must comply with the following special controls: 1. Design verification and validation must include: i. A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations. ii. A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (e.g., improved time to review of prioritized images for pre-specified clinicians). iii. Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment. iv. Standalone performance testing protocols and results of the device. v. Appropriate software documentation (e.g., device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results). 2. Labeling must include the following: i. A detailed description of the patient population for which the device is indicated for use. ii. A detailed description of the intended user and user training that addresses appropriate use protocols for the device. iii. Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality for certain subpopulations), as applicable. iv. A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images. v. Device operating instructions. vi. A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (e.g., improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (e.g., confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians).(iii) Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (
*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment.(iv) Stand-alone performance testing protocols and results of the device.
(v) Appropriate software documentation (
*e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use;
(ii) A detailed description of the intended user and user training that addresses appropriate use protocols for the device;
(iii) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality for certain subpopulations), as applicable;(iv) A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images;
(v) Device operating instructions; and
(vi) A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (*e.g.,* confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
August 26, 2025
Brainomix Limited
Delanyo Mensah
Regulatory Affairs Specialist
First Floor, Seacourt Tower
West Way
Oxford, OX2 0JJ
United Kingdom
Re: K251983
Trade/Device Name: Brainomix 360 Triage Stroke
Regulation Number: 21 CFR 892.2080
Regulation Name: Radiological Computer Aided Triage And Notification Software
Regulatory Class: Class II
Product Code: QAS
Dated: June 25, 2025
Received: June 27, 2025
Dear Delanyo Mensah:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251983 - Delanyo Mensah
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K251983 - Delanyo Mensah
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Jessica Lamb, PhD
Assistant Director
Imaging Software Team
DHT8B: Division of Radiological Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
Brainomix 360 Triage Stroke
Page 10 of 52
| Indications for Use | |
| --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | ? |
| Please provide the device trade name(s). | ? |
| Brainomix 360 Triage Stroke | |
| Please provide your Indications for Use below. | ? |
{4}
Brainomix 360 Triage Stroke is a radiological computer aided triage and notification software indicated for use in the analysis of non-contrast head CT (NCCT) images to assist hospital networks and trained clinicians in workflow triage by flagging and communicating suspected positive findings of head NCCT images for large vessel occlusion (LVO) of the intracranial ICA and M1 or intracranial hemorrhage (ICH). Specifically, the device is intended to be used for the triage of images acquired from adult patients in the acute setting, within 24 hours of the onset of the acute symptoms, or where this is unclear, since last known well (LKW) time. It is not intended to detect symmetrical bilateral MCA occlusions.
Brainomix 360 Triage Stroke uses an artificial intelligence algorithm to analyze images and highlight cases with detected NCCT LVO or ICH on the Brainomix server on premise or in the cloud in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected LVO or ICH findings via a web user interface or mobile application. Notifications include compressed preview images that are meant for informational purposes only and are not intended for diagnostic use beyond notification.
The device does not alter the original medical image, and it is not intended to be used as a primary diagnostic device. The results of Brainomix 360 Triage Stroke are intended to be used in conjunction with other patient information and based on professional judgment to assist with triage/prioritization of medical images. Notified clinicians are ultimately responsible for reviewing full images per the standard of care.
## Cautions:
- All patients should get adequate care for their symptoms, including angiography and/or other appropriate care per the standard clinical practice, irrespective of the output of Brainomix 360 Triage Stroke.
- Brainomix 360 Triage Stroke is not intended to be a rule-out device and for cases that have been processed by the device without notification for "Suspected LVO" should not be viewed as indicating that LVO is excluded. All cases should undergo angiography, per the standard stroke workup.
## Limitations:
1. Brainomix 360 Triage Stroke is not intended for mobile diagnostic use. Images viewed on a mobile platform are compressed preview images and not for diagnostic interpretation.
2. Brainomix 360 Triage Stroke does not replace the need for angiography in ischemic stroke workup - it provides workflow prioritization and notification only.
3. Brainomix 360 Triage Stroke has been validated and is intended to be used on Siemens, GE and Philips scanners.
4. Brainomix 360 Triage Stroke is not intended to be used on patients with recent (within 6 weeks) neurosurgery or endovascular neurointervention or recent (within 4 weeks) previous diagnosis of stroke.
5. Brainomix 360 Triage Stroke is not intended to detect symmetrical bilateral MCA occlusions.
## Contraindications:
Brainomix 360 Triage Stroke is not suitable for use with scan data containing image features associated with:
- tumours or abscesses
Brainomix 360 Triage Stroke
Page 11 of 52
{5}
Brainomix 360 Triage Stroke
Page 12 of 52
Please select the types of uses (select one or both, as applicable).
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
{6}
BRAINOMIX
Brainomix Limited
First Floor, Seacourt Tower, West Way
Oxford OX2 0JJ, United Kingdom
# K251983
# 510(K) Summary Brainomix 360 Triage Stroke
Date Prepared: 27Jun2025
Applicant's name: Brainomix Limited
Applicant's address: First Floor, Seacourt Tower, West Way
Oxford, OX2 0JJ
United Kingdom
Official contact: Delanyo Mensah
+44 (0)1865 582730
regulatory@brainomix.com
Device Trade Name: Brainomix 360 Triage Stroke
Device Common Name: Radiological Computer-Assisted Triage and Notification Software
Regulatory Class: Class II
Product Code: QAS
Regulation No: 21 C.F.R. §892.2080
Classification Panel: Radiology Devices
# 1 Predicate Device
The Brainomix 360 Triage Stroke device is claimed to be substantially equivalent to the following legally marketed device:
Trade Name: Brainomix 360 Triage Stroke
Manufacturer: Brainomix Limited
Regulation Number: 21 C.F.R. §892.2080
Regulatory Class: Class II
Regulation Name: Radiological computer-assisted triage and notification software
Product Code: QAS
Submission Number: K232496
# 2 Device Description
Brainomix 360 Triage Stroke (also referred to as Triage Stroke in this submission) is a radiological computer aided triage and notification software package compliant with the DICOM standard and running on an off-the-shelf physical or virtual server. Triage Stroke is a non-contrast CT processing software-only medical device which operates within the integrated Brainomix 360 Platform to provide triage and notification prioritization of suspected large vessel occlusion (LVO) or intracranial hemorrhage (ICH). The device uses machine learning algorithms such as advanced non adaptive imaging algorithms, artificial intelligence, and large data analytics.
Brainomix 360 Triage Stroke is available to users in three configurations, featuring three individual processing modules:
- Triage ICH Module, which can only flag positive findings of suspected ICH;
{7}
BRAINOMIX
Brainomix Limited
First Floor, Seacourt Tower, West Way
Oxford OX2 0JJ, United Kingdom
- Triage Stroke Module, which can flag positive findings of suspected ICH or LVO; And
- NCCT LVO Module, which can flag positive finding of suspected LVO
The Triage ICH Module automatically identifies suspected ICH, the NCCT LVO module automatically identifies suspected LVO, while the Triage Stroke Module automatically identifies suspected ICH or LVO on non-contrast CT (NCCT) imaging acquired from adult patients in the acute setting, within 24 hours of the onset of acute symptoms, or where this is unclear, since last known well (LKW) time. The output of the device is a priority notification to clinicians indicating the suspicion of just ICH for Triage ICH Module, the suspicion of just LVO for the NCCT LVO Module, and suspicion of ICH or LVO for Triage Stroke Module based on positive findings. Specifically, the ICH analysis algorithm is optimized to identify findings of hyperdense volume in the parenchyma typically associated with acute intracranial hemorrhage; and the NCCT LVO suspicion uses the combined analysis of the ASPECTS and hyperdense vessel sign (HDVS) algorithms to identify hyper attenuation in vessels and hypodense regions typically associated with a large vessel occlusion in a non-contrast CT scan.
Brainomix 360 Triage Stroke is not intended to detect symmetrical bilateral MCA occlusions. The device uses the basic services supplied by the Brainomix 360 Platform including DICOM processing, job management, imaging module execution and imaging output including notification and compressed image.
Brainomix 360 Triage Stroke notification capabilities enable clinicians to review and preview images via mobile app notification. Alternatively, intended users can also access the notification (a "Suspected LVO" or "Suspected hemorrhage" flag) and straightened images via the Brainomix 360 web user interface. Images that are previewed via mobile app are compressed, are for preview informational purposes only, and not intended for diagnostic use beyond notification.
The device is intended for use as an additional tool for assisting study triage within existing patient pathways. It does not replace any part of the current standard of care. It is designed to assist in prioritization of studies for reading within a worklist, in addition to any other pre-existing formal or informal methods of study prioritization in place. Specifically, it does not remove cases from a reading queue and operates in parallel to the standard of care. This device is not intended to replace the usual methods of communication and transfer of information in the current standard of care.
The Brainomix 360 Triage Stroke device is made available to the user through the Brainomix 360 Platform. The Brainomix 360 Platform is a central control unit which coordinates the execution image processing modules which support various analysis methods used in clinical practice today:
- Brainomix 360 e-ASPECTS (K243294)
- Brainomix 360 e-CTA (K242123)
- Brainomix 360 e-CTP (K223555)
- Brainomix 360 e-MRI (K231656)
- Brainomix 360 Triage ICH (K231195)
- Brainomix 360 Triage LVO (K231837)
- Brainomix 360 Triage Stroke (K232496) (predicate device)
{8}
BRAINOMIX
Brainomix Limited
First Floor, Seacourt Tower, West Way
Oxford OX2 0JJ, United Kingdom
# 3 Clinical Characteristics
The primary users of the Brainomix 360 platform are medical imaging professionals. The outputs generated by Brainomix 360 Triage Stroke may be used by clinicians to inform patient triaging decisions as part of a wider set of other imaging to identify signs of ischemic damage.
# 4 Indications for Use
Brainomix 360 Triage Stroke is a radiological computer aided triage and notification software indicated for use in the analysis of non-contrast head CT (NCCT) images to assist hospital networks and trained clinicians in workflow triage by flagging and communicating suspected positive findings of head NCCT images for large vessel occlusion (LVO) of the intracranial ICA and M1 or intracranial hemorrhage (ICH). Specifically, the device is intended to be used for the triage of images acquired from adult patients in the acute setting, within 24 hours of the onset of the acute symptoms, or where this is unclear, since last known well (LKW) time. It is not intended to detect symmetrical bilateral MCA occlusions.
Brainomix 360 Triage Stroke uses an artificial intelligence algorithm to analyze images and highlight cases with detected NCCT LVO or ICH on the Brainomix server on premise or in the cloud in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected LVO or ICH findings via a web user interface or mobile application. Notifications include compressed preview images that are meant for informational purposes only and are not intended for diagnostic use beyond notification.
The device does not alter the original medical image, and it is not intended to be used as a primary diagnostic device. The results of Brainomix 360 Triage Stroke are intended to be used in conjunction with other patient information and based on professional judgment to assist with triage/prioritization of medical images. Notified clinicians are ultimately responsible for reviewing full images per the standard of care.
## Cautions:
- All patients should get adequate care for their symptoms, including angiography and/or other appropriate care per the standard clinical practice, irrespective of the output of Brainomix 360 Triage Stroke.
- Brainomix 360 Triage Stroke is not intended to be a rule-out device and for cases that have been processed by the device without notification for "Suspected LVO" should not be viewed as indicating that LVO is excluded. All cases should undergo angiography, per the standard stroke workup.
## Limitations:
1. Brainomix 360 Triage Stroke is not intended for mobile diagnostic use. Images viewed on a mobile platform are compressed preview images and not for diagnostic interpretation.
2. Brainomix 360 Triage Stroke does not replace the need for angiography in ischemic stroke workup - it provides workflow prioritization and notification only.
3. Brainomix 360 Triage Stroke has been validated and is intended to be used on Siemens, GE and Philips scanners.
4. Brainomix 360 Triage Stroke is not intended to be used on patients with recent (within 6 weeks) neurosurgery or endovascular neurointervention or recent (within 4 weeks) previous diagnosis of stroke.
5. Brainomix 360 Triage Stroke is not intended to detect symmetrical bilateral MCA occlusions.
## Contraindications:
Brainomix 360 Triage Stroke is not suitable for use with scan data containing image features associated with:
- tumours or abscesses
- coils, shunts, embolization or movement artifacts
{9}
BRAINOMIX
Brainomix Limited
First Floor, Seacourt Tower, West Way
Oxford OX2 0JJ, United Kingdom
- intracranial vascular pathologies such as arterial aneurysms, arteriovenous malformations or venous thrombosis.
# 5 Technological Characteristics
Brainomix 360 Triage Stroke principal workflow for NCCT includes the following key steps:
1. NCCT image loading. Brainomix 360 Triage Stroke provides an automated workflow which will automatically process image data received by the system in accordance with pre-configured user DICOM routing preferences.
2. Image Processing function. The workflow in this step will depend on the module configured for the user's healthcare facility (Triage ICH Module, Triage Stroke Module or NCCT LVO Module).
3. Notification of 'Suspected hemorrhage' or 'Suspected LVO' sent to the user and image viewing. This is a non-diagnostic DICOM viewing application through mobile app, allowing a trained clinician to preview unprocessed CT studies with basic viewing functions (scroll through a cine, adjust window level and window width, pan, and zoom) prior to definitive review of images on a diagnostic workstation. Alternatively, the straightened images can be visualized in the Brainomix 360 web user interface on a radiologist workstation.
All functions listed here are performed automatically by the software. Notifications and alerts for users (when processing has completed) is the step that includes user interaction and present information to users. This is considered to be the output of the processing functions.
# 6 Performance Data
The performance of Brainomix 360 Triage Stroke has been validated in three clinical studies:
- A standalone study of ICH detection performance
- A standalone study of LVO and ICH detection performance
- A reader study to compare the sensitivity of LVO detection to that of radiologists.
# 6.1 Summary of ICH Standalone Performance
A retrospective study was conducted with the primary endpoint of assessing the performance of Brainomix 360 Triage Stroke in identifying ICH findings in NCCT head images. The dataset comprised 341 cases (167 ICH positive; 174 ICH negative). Truthing was conducted by consensus of three experienced US board certified neuroradiologists.
Sensitivity and specificity of ICH detection exceeded the pre-specified performance goal of 80%: sensitivity was 96.41% (95% CI: 92.65-98.65%) and specificity was 96.55% (CI: 92.94-98.70%).
As a secondary outcome measure, performance for detection of subarachnoid hemorrhages (SAHs) was assessed against a pre-defined performance goal: sensitivity was 85.71% (CI: 60.99-97.67%) and specificity was 96.55% (CI: 80.60-98.87%).
# 6.2 ICH Performance Over Subgroups
ICH detection performance was stratified over a range of relevant clinical variables, including age, gender, slice thickness and scanner manufacturer subgroups. Additional subgroup analysis is reported in the labelling.
| Metrics | 21 < Age < 50 | 50 ≤ Age < 70 | Age ≥ 70 |
| --- | --- | --- | --- |
| Total Positives | 36 | 69 | 62 |
| Sensitivity | 94.44% (82.88-99.20) | 98.55% (93.01-99.91) | 95.16% (87.37-98.97) |
| Specificity | 95.77% (88.90-99.10) | 96.25% (90.10-99.21) | 100% (88.74-100.00) |
Metrics
Female
Male
{10}
BRAINOMIX
Brainomix Limited
First Floor, Seacourt Tower, West Way
Oxford OX2 0JJ, United Kingdom
| Total Positives | 79 | 88 |
| --- | --- | --- |
| Sensitivity | 96.20% (89.98-99.20) | 96.59% (90.97-99-28) |
| Specificity | 98.78% (94.08-99.92) | 94.57% (88.34-98.18) |
| Metrics | Slice Thickness < 1.5 mm | 1.5mm ≤ Slice Thickness < 3 mm | Slice Thickness ≥ 3 mm |
| --- | --- | --- | --- |
| Total Positives | 112 | 23 | 32 |
| Sensitivity | 97.32% (92.84-99.44) | 95.65% (80.63-99.73) | 93.75% (80.94-99.09) |
| Specificity | 99.17% (95.94-99.95) | 100.00% (86.70-100.00%) | 85.29% (70.47-94.80) |
| Metrics | GE MEDICAL SYSTEM | Philips | SIEMENS |
| --- | --- | --- | --- |
| Total Positives | 87 | 72 | 7 |
| Sensitivity | 98.85% (94.41-99.92) | 94.44% (87.12-98.47) | 85.71% (49.44-98.90) |
| Specificity | 95.70% (89.92-98.83) | 96.88% (90.02-99.58) | 100.00% (84.61-100.00) |
## 6.3 Summary of LVO and ICH Standalone Performance
A retrospective study was conducted with the primary endpoint of assessing the performance of Brainomix 360 Triage Stroke in identifying NCCT head images containing large vessel occlusion (LVO) or intracranial hemorrhage (ICH) in 267 cases (LVO positive: 112 cases; ICH positive: 40 cases; Negative for ICH or LVO: 115; excluded: 3 cases due to technical inadequacy). Truthing was conducted by consensus of three experienced US board certified neuroradiologists.
Sensitivity and specificity exceeded the pre-specified performance goals for NCCT LVO and ICH. NCCT LVO performance was observed at 69.64% sensitivity (CI: 60.65-77.70%) and 89.57% specificity (CI: 82.92-94.36%). NCCT ICH performance in this smaller subset of cases was consistent with that of the ICH standalone study, as reported above (sensitivity: 95.00% [84.47-99.29%]; specificity: 88.11% [83.39-91.92%]).
The Triage Stroke time-to-notification analysis includes three steps: (1) the time taken to transfer the image for DICOM transfer, (2) the device processing time, and (3) the time taken to deliver a notification after processing. The combined time-to-notification was compared to an acceptability criterion of 3.5 minutes. The minimum and maximum estimates for each of the three steps were assessed and combined to create a range of expected values for the total time-to-notification. The minimum time-to-notification was 58.3 seconds, and the maximum was 150.7 seconds. This met the acceptability criterion of time-to-notification of under 3.5 minutes.
## 6.4 Summary of Reader LVO Performance
In addition, a reader study was conducted to compare NCCT LVO sensitivity of the device to that of radiologists. Secondary endpoints of expert non-inferiority and non-expert superiority were used. Triage Stroke passed both conditions, with a sensitivity for all readers (experts and non-experts) of 47.94% (CI: 37.91-57.97%). The difference between the device's sensitivity and that of all readers was 20.52% (CI: 8.26-32.78%). The general radiologists (non-experts) performed with a sensitivity of 47.18% (CI: 33.62-60.75%). The difference between the device and non-expert sensitivity was 21.28% (CI: 5.84-36.72%).
## 6.5 LVO Performance Across Subgroups
Subgroup analyses were conducted for the LVO indication. The analyses assessed performance across age, gender, slice thickness and scanner manufacturer subgroups. Additional subgroup analysis is reported in the labelling.
| Metrics | 21 < Age < 50 | 50 ≤ Age < 70 | Age ≥ 70 |
| --- | --- | --- | --- |
{11}
BRAINOMIX
Brainomix Limited
First Floor, Seacourt Tower, West Way
Oxford OX2 0JJ, United Kingdom
| Total Positives | 13 | 39 | 60 |
| --- | --- | --- | --- |
| Sensitivity | 84.62% (58.59-97.46) | 74.36% (59.09-86.41) | 63.33% (50.62-74.88) |
| Specificity | 94.74% (77.09-99.67) | 89.09% (78.7-95.74) | 87.80% (75.07-95.75) |
| Metrics | Female | Male |
| --- | --- | --- |
| Total Positives | 48 | 64 |
| Sensitivity | 75.0% (61.4-85.89) | 65.62% (53.39-76.57) |
| Specificity | 88.14% (77.94-94.92) | 91.07% (81.32-96.95) |
| Metrics | Slice Thickness < 1 mm | 1 mm ≤ Slice Thickness < 3 mm | Slice Thickness ≥ 3 mm |
| --- | --- | --- | --- |
| Total Positives | 44 | 25 | 42 |
| Sensitivity | 61.36% (46.46-74.93) | 80.00% (61.30-92.70) | 71.43% (56.52-83.7) |
| Specificity | 87.76% (76.29-95.19) | 96.67% (84.74-99.79) | 86.11% (71.95-95.11) |
| Metrics | SIEMENS | GE MEDICAL SYSTEMS | Philips |
| --- | --- | --- | --- |
| Total Positives | 48 | 33 | 27 |
| Sensitivity | 62.5% (48.25-75.4) | 72.73% (55.90-86.04) | 77.78% (59.56-90.82) |
| Specificity | 88.24% (77.15-95.39) | 83.87% (67.93-94.24) | 96.97% (86.02-99.81) |
In summary, the performance validation data demonstrated that the proposed device provides accurate detection of acute ICH and LVO under a range of clinically relevant variables associated with the intended use of the software. The notification functionality of the device also met the acceptability criterion.
## 7 Prescriptive Statement
Caution: Federal law restricts this device to sale by or on the order of a physician.
## 8 Safety and Effectiveness
Brainomix 360 Triage Stroke has been designed, verified and validated in compliance with 21 CFR, Part 820.30 requirements. The device has been designed to meet the requirements associated with ISO 14971:2019 for risk management and the software development process conforms to IEC 62304:2015. The performance of Brainomix 360 Triage Stroke has been validated through retrospective case data based on expert reader truthing of the data and reader testing.
## 9 Cybersecurity
Brainomix 360 Triage Stroke has been designed to follow the FDA Cybersecurity Guidance and IEC 81001-5-1.
## 10 Substantial Equivalence
The subject device includes similar features as compared to the predicate device. Both devices have substantially equivalent indications for use, technological characteristics and roles within a clinical workflow relative to triage and notification. Both devices are intended to automatically process and analyze non-contrast CT scans to provide a notification to users in case of a suspected LVO or ICH being identified. The subject device does raise different
{12}
BRAINOMIX
Brainomix Limited
First Floor, Seacourt Tower, West Way
Oxford OX2 0JJ, United Kingdom
questions of safety and effectiveness when compared to the predicate and both include similar mitigation strategies for reducing the risk of off label use of preview images which are shared via mobile.
Where the proposed device differs from the predicate device are:
- Improved ICH algorithm: The proposed device features an updated ICH algorithm using a different deep learning framework, CNN architecture, training data and post-processing capabilities of the algorithm. The improved algorithm resulted in improved sensitivity in detecting SAH and SDH subtypes. Furthermore, the algorithm improvements have also resulted in increased sensitivity and specificity of the software in detecting ICH. The performance of the updated algorithm has been validated in clinical studies, which provides evidence of effective triaging of all ICH subtypes with an increased sensitivity and specificity.
- Indications for use extended to SAH patients: Following the improvements made to the ICH algorithm the proposed device removes a previous SAH limitation and extends the indications of use for Triage Stroke to include subarachnoid hemorrhages (SAH) sub types. The change in the algorithm and the subsequent removal of the limitation has been validated through a clinical study and showed improvements on the performance of Triage Stroke in relation to different ICH subtypes.
- Configurable workflows: The proposed device offers three configurations with three workflows. The Triage ICH Module features a workflow to detect ICH in NCCT scans, the NCCT LVO Module features a workflow to detect LVO in NCCT scans, while the Triage Stroke Module's workflow allows the detection of either ICH or LVO in NCCT scans. The workflow of the Triage ICH Module remains the same as cleared under Brainomix 360 Triage ICH (K231195) and the workflow of the Triage Stroke Module remains the same as cleared under Brainomix 360 Triage ICH (K232496). The NCCT LVO module utilizes the workflow like that of the Triage Stroke Module but with the device output limited to LVO notification only. The proposed device differs from the predicate in that it offers a choice of these three workflows within a single device and as such, this technological characteristic does not raise different questions of safety or effectiveness compared to the predicate device.
# 11 Substantial Equivalence Comparison
A table comparing the key features of the subject and predicate devices is provided below.
| Characteristic/ Parameter | Brainomix 360 Triage Stroke Predicate Device (K232496) | Brainomix 360 Triage Stroke Subject Device |
| --- | --- | --- |
| Product Code | QAS | QAS |
| Regulation | 21 CFR. §892.2080 | 21 CFR. §892.2080 |
| Indications for Use | Brainomix 360 Triage Stroke is a radiological computer aided triage and notification software indicated for use in the analysis of non-contrast head CT (NCCT) images to assist hospital networks and trained clinicians in workflow triage by flagging and communicating suspected positive findings of head NCCT images for large vessel occlusion (LVO) of the intracranial ICA and M1 and intracranial hemorrhage (ICH). Specifically, the device is intended to be used for the triage of images acquired from adult patients in the acute setting, within 24 hours of the onset of the acute symptoms, or where this is unclear, since last known well (LKW) time. It is not intended to detect isolated | Brainomix 360 Triage Stroke is a radiological computer aided triage and notification software indicated for use in the analysis of non-contrast head CT (NCCT) images to assist hospital networks and trained clinicians in workflow triage by flagging and communicating suspected positive findings of head NCCT images for large vessel occlusion (LVO) of the intracranial ICA and M1 or intracranial hemorrhage (ICH). Specifically, the device is intended to be used for the triage of images acquired from adult patients in the acute setting, within 24 hours of the onset of the acute symptoms, or where this is unclear, since last known well (LKW) time. It is not intended to detect isolated |
{13}
B
BRAINOMIX
Brainomix Limited
First Floor, Seacourt Tower, West Way
Oxford OX2 0JJ, United Kingdom
| | subarachnoid hemorrhage and symmetrical bilateral MCA occlusions. | symmetrical bilateral MCA occlusions. |
| --- | --- | --- |
| | Brainomix 360 Triage Stroke uses an artificial intelligence algorithm to analyze images and highlight cases with detected NCCT LVO or ICH on the Brainomix server on premise or in the cloud in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected LVO or ICH findings via a web user interface or mobile application. Notifications include compressed preview images that are meant for informational purposes only and are not intended for diagnostic use beyond notification. | Brainomix 360 Triage Stroke uses an artificial intelligence algorithm to analyze images and highlight cases with detected NCCT LVO or ICH on the Brainomix server on premise or in the cloud in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected LVO or ICH findings via a web user interface or mobile application. Notifications include compressed preview images that are meant for informational purposes only and are not intended for diagnostic use beyond notification. |
| | The device does not alter the original medical image, and it is not intended to be used as a primary diagnostic device. The results of Brainomix 360 Triage Stroke are intended to be used in conjunction with other patient information and based on professional judgment to assist with triage/prioritization of medical images. Notified clinicians are ultimately responsible for reviewing full images per the standard of care. | The device does not alter the original medical image, and it is not intended to be used as a primary diagnostic device. The results of Brainomix 360 Triage Stroke are intended to be used in conjunction with other patient information and based on professional judgment to assist with triage/prioritization of medical images. Notified clinicians are ultimately responsible for reviewing full images per the standard of care. |
| | Cautions: | Cautions: |
| | • All patients should get adequate care for their symptoms, including angiography and/or other appropriate care per the standard clinical practice, irrespective of the output of Brainomix 360 Triage Stroke. | • All patients should get adequate care for their symptoms, including angiography and/or other appropriate care per the standard clinical practice, irrespective of the output of Brainomix 360 Triage Stroke. |
| | • Brainomix 360 Triage Stroke is not intended to be a rule-out device and for cases that have been processed by the device without notification for "Suspected LVO" should not be viewed as indicating that LVO is excluded. All cases should undergo angiography, per the standard stroke workup. | • Brainomix 360 Triage Stroke is not intended to be a rule-out device and for cases that have been processed by the device without notification for "Suspected LVO" should not be viewed as indicating that LVO is excluded. All cases should undergo angiography, per the standard stroke workup. |
| | Limitations: | Limitations: |
| | • Brainomix 360 Triage Stroke is not intended for mobile diagnostic use. Images viewed on a mobile platform are compressed preview images and not for diagnostic interpretation. | 1. Brainomix 360 Triage Stroke is not intended for mobile diagnostic use. Images viewed on a mobile platform are compressed preview images and not for diagnostic interpretation. |
| | • Brainomix 360 Triage Stroke does not replace the need for CTA in ischemic stroke workup - it provides workflow prioritization and notification only. | 2. Brainomix 360 Triage Stroke does not replace the need for angiography in ischemic stroke |
{14}
B
BRAINOMIX
Brainomix Limited
First Floor, Seacourt Tower, West Way
Oxford OX2 0JJ, United Kingdom
| | • Brainomix 360 Triage Stroke has been validated and is intended to be used on Siemens, GE and Philips scanners.
• Brainomix 360 Triage Stroke is not intended to be used on patients with recent (within 6 weeks) neurosurgery or endovascular neurointervention or recent (within 4 weeks) previous diagnosis of stroke.
• Brainomix 360 Triage Stroke is not intended to detect isolated subarachnoid hemorrhage and symmetrical bilateral MCA occlusions.
**Contraindications:**
Brainomix 360 Triage Stroke is not suitable for use with scan data containing image features associated with:
• tumors or abscesses
• coils, shunts, embolization or movement artifacts
• Brainomix 360 Triage Stroke is not intended to be used for analyzing CT images in intracranial vascular pathologies such as arterial aneurysms, arteriovenous malformations or venous thrombosis. | workup - it provides workflow prioritization and notification only.
3. Brainomix 360 Triage Stroke has been validated and is intended to be used on Siemens, GE and Philips scanners.
4. Brainomix 360 Triage Stroke is not intended to be used on patients with recent (within 6 weeks) neurosurgery or endovascular neurointervention or recent (within 4 weeks) previous diagnosis of stroke.
5. Brainomix 360 Triage Stroke is not intended to detect symmetrical bilateral MCA occlusions.
**Contraindications:**
Brainomix 360 Triage Stroke is not suitable for use with scan data containing image features associated with:
• tumours or abscesses
• coils, shunts, embolization or movement artifacts
• intracranial vascular pathologies such as arterial aneurysms, arteriovenous malformations or venous thrombosis. |
| --- | --- | --- |
| Environment of Use | Clinical/Hospital environment | Clinical/Hospital environment |
| User | Clinician | Clinician |
| Anatomical Region | Head | Head |
| Input Data | NCCT (ICH and LVO) | NCCT (ICH and LVO) |
| Technical Implementation | AI/ML/Neural Network | AI/ML/Neural Network |
| Diagnostic application | Notification-only | Notification-only |
| Results of image analysis | Internal no image marking | Internal no image marking |
| Segmentation of ROI | The device does not highlight or direct user’s attention to a specific location in the image file | The device does not highlight or direct user’s attention to a specific location in the image file |
| Notification Display | Web user interface and mobile device | Web user interface and mobile device |
| Preview Images | Presentation of a preview of the study for initial assessment not meant for diagnostic purposes.
The device operates in parallel with the standard of care. | Presentation of a preview of the study for initial assessment not meant for diagnostic purposes.
The device operates in parallel with the standard of care. |
| Annotation / Localization | Device does not mark, highlight, or direct users’ attention to a specific location in the original image | Device does not mark, highlight, or direct users’ attention to a specific location in the original image |
| Prioritization Notification | Yes | Yes |
| Clinical SoC Workflow | In parallel to | In parallel to |
{15}
B
BRAINOMIX
Brainomix Limited
First Floor, Seacourt Tower, West Way
Oxford OX2 0JJ, United Kingdom
| Technical Pipeline | Two cascaded functions (ICH then LVO) using three integrated algorithms | Triage Stroke Configuration: two cascaded functions (ICH then LVO) using three integrated algorithms
Triage ICH Configuration: ICH analysis function using one algorithm
NCCT LVO Configuration: LVO analysis function using three integrated algorithms |
| --- | --- | --- |
| Removal of Cases from SoC review | No | No |
# 12 Conclusion
Brainomix 360 Triage Stroke includes similar features as compared to the predicate device. Both devices have similar indications for use, technological approaches and roles within a clinical workflow. The differences in indications for use and technological characteristics for the proposed device do not raise different questions of safety or effectiveness.
The proposed device therefore does not raise different questions with regards to safety and efficacy and demonstrates substantial equivalence to the predicate, Brainomix 360 Triage Stroke (K232496).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.