AI/ML, Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K232496 · Nov 21, 2023
Brainomix 360 Triage Stroke
Brainomix Limited
Retrospective clinical NCCT head images; Patient clinical information
A retrospective study was conducted to assess the performance of the device in identifying intracranial hemorrhage (ICH) and large vessel occlusion (LVO) in 267 clinical cases. The data was used to validate sensitivity and specificity against pre-specified performance goals.
Large Vessel Occlusion (LVO) detection (reader improvement)
AI/ML/Neural Network
—
Sensitivity for all readers: 47.94% (CI: 37.91-57.97%); General radiologists sensitivity: 47.18% (CI: 33.62-60.75%)
—
—
Reader study comparing NCCT LVO sensitivity of the device to that of radiologists
>1 (expert readers) + >1 (general radiologists)
Indications for Use
Brainomix 360 Triage Stroke is a radiological computer aided triage and notification software indicated for use in the analysis of non-contrast head CT (NCCT) images to assist hospital networks and trained clinicians in workflow triage by flagging and communicating suspected positive findings of head NCCT images for large vessel occlusion (LVO) of the intracranial ICA and M1 and intracranial hemorrhage (ICH). Specifically, the device is intended to be used for the trage of images acquired from adult patients in the acute setting, within 24 hours of the acute symptoms, or where this is unclear, since last known well (LKW) time. It is not intended to detect isolated subarachnoid hemorrhage and symmetrical bilateral MCA occlusions. Brainomix 360 Triage Stroke uses an artificial intelligence algorithm to analyze images and highlight cases with detected NCCT LVO or ICH on the Brainomix server on premise or in the cloud in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected LVO or ICH findings via a web user interface or mobile applications include compressed preview images that are meant for informational purposes only and are not intended for diagnostic use beyond notification. The device does not alter the original mage, and it is not intended to be used as a primary diagnostic device. The results of Brainomix 360 Triage Stroke are intended to be used in conjunction with other patient information and based on professional judgment to assist with triage/prioritization of medical images. Notified clinicians are ultimately responsible for reviewing full images per the standard of care.
Device Story
Brainomix 360 Triage Stroke is a radiological computer-aided triage and notification software. It processes non-contrast head CT (NCCT) images from adult patients in acute settings. The device operates on an off-the-shelf physical or virtual server, either on-premise or in the cloud. It uses AI/ML algorithms to automatically identify suspected large vessel occlusion (LVO) and intracranial hemorrhage (ICH) in parallel to standard-of-care interpretation. The system outputs priority notifications to clinicians via a web user interface or mobile application, including compressed preview images. The device does not alter original images, mark findings on images, or serve as a primary diagnostic tool. It assists clinicians in prioritizing worklists; notified clinicians remain responsible for reviewing full images per standard of care. The device benefits patients by potentially reducing time-to-notification for critical stroke findings, facilitating faster clinical decision-making and intervention.
Clinical Evidence
Retrospective study of 267 cases (40 ICH, 112 LVO, 115 negative). ICH sensitivity 92.5% (95% CI: 80.97-98.36%), specificity 87.22% (CI: 82.39-91.18%). NCCT LVO sensitivity 68.75% (CI: 59.71-76.90%), specificity 89.57% (CI: 82.92-94.36%). Reader study showed device sensitivity (47.94%) was superior to general radiologists (47.18%) with a difference of 21.28% (CI: 5.84-36.72%). Time-to-notification ranged from 62 to 134 seconds, meeting the <3.5-minute criterion.
Technological Characteristics
Software-based radiological triage system compliant with DICOM. Operates on off-the-shelf physical/virtual servers. Uses AI/ML algorithms (non-adaptive) for image analysis. Supports Siemens, GE, and Philips scanners. Connectivity via local hospital network to backend server (on-premise or cloud). Development conforms to IEC 62304:2015; risk management per EN ISO 14971:2019.
Indications for Use
Indicated for adult patients in acute setting (within 24 hours of symptoms or LKW) to triage non-contrast head CT (NCCT) images for suspected large vessel occlusion (LVO) of intracranial ICA/M1 and intracranial hemorrhage (ICH). Contraindicated for patients with recent (within 6 weeks) neurosurgery/endovascular intervention, recent (within 4 weeks) stroke, tumors, abscesses, coils, shunts, embolization, movement artifacts, or intracranial vascular pathologies (aneurysms, AVMs, venous thrombosis). Not for isolated subarachnoid hemorrhage or symmetrical bilateral MCA occlusions.
Regulatory Classification
Identification
Radiological computer aided triage and notification software is an image processing prescription device intended to aid in prioritization and triage of radiological medical images. The device notifies a designated list of clinicians of the availability of time sensitive radiological medical images for review based on computer aided image analysis of those images performed by the device. The device does not mark, highlight, or direct users' attention to a specific location in the original image. The device does not remove cases from a reading queue. The device operates in parallel with the standard of care, which remains the default option for all cases.
Special Controls
Radiological computer aided triage and notification software must comply with the following special controls: 1. Design verification and validation must include: i. A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations. ii. A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (e.g., improved time to review of prioritized images for pre-specified clinicians). iii. Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment. iv. Standalone performance testing protocols and results of the device. v. Appropriate software documentation (e.g., device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results). 2. Labeling must include the following: i. A detailed description of the patient population for which the device is indicated for use. ii. A detailed description of the intended user and user training that addresses appropriate use protocols for the device. iii. Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality for certain subpopulations), as applicable. iv. A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images. v. Device operating instructions. vi. A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (e.g., improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (e.g., confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations.
(ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians).(iii) Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (
*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment.(iv) Stand-alone performance testing protocols and results of the device.
(v) Appropriate software documentation (
*e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use;
(ii) A detailed description of the intended user and user training that addresses appropriate use protocols for the device;
(iii) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality for certain subpopulations), as applicable;(iv) A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images;
(v) Device operating instructions; and
(vi) A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (
*e.g.,* improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (*e.g.,* confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Brainomix Limited % Zsolt Szrnka Regulatory Affairs Manager First Floor Seacourt Tower West Way Oxford. OX2 0JJ United Kingdom
November 21, 2023
Re: K232496 Trade/Device Name: Brainomix 360 Triage Stroke Regulation Number: 21 CFR 892.2080 Regulation Name: Radiological computer aided triage and notification software Regulatory Class: Class II Product Code: QAS Dated: November 6, 2023 Received: November 6, 2023
Dear Zsolt Szrnka:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb
Jessica Lamb. Ph.D. Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K232496
Device Name Brainomix 360 Triage Stroke
#### Indications for Use (Describe)
Brainomix 360 Triage Stroke is a radiological computer aided triage and notification software indicated for use in the analysis of non-contrast head CT (NCCT) images to assist hospital networks and trained clinicians in workflow triage by flagging and communicating suspected positive findings of head NCCT images for large vessel occlusion (LVO) of the intracranial ICA and M1 and intracranial hemorrhage (ICH). Specifically, the device is intended to be used for the trage of images acquired from adult patients in the acute setting, within 24 hours of the acute symptoms, or where this is unclear, since last known well (LKW) time. It is not intended to detect isolated subarachnoid hemorrhage and symmetrical bilateral MCA occlusions.
Brainomix 360 Triage Stroke uses an artificial intelligence algorithm to analyze images and highlight cases with detected NCCT LVO or ICH on the Brainomix server on premise or in the cloud in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected LVO or ICH findings via a web user interface or mobile applications include compressed preview images that are meant for informational purposes only and are not intended for diagnostic use beyond notification.
The device does not alter the original mage, and it is not intended to be used as a primary diagnostic device. The results of Brainomix 360 Triage Stroke are intended to be used in conjunction with other patient information and based on professional judgment to assist with triage/prioritization of medical images. Notified clinicians are ultimately responsible for reviewing full images per the standard of care.
#### Cautions:
· All patients should get adequate care for their symptoms, including angiography and/or other appropriate care per the standard clinical practice, irrespective of the output of Brainomix 360 Triage Stroke.
· Brainomix 360 Triage Stroke is not intended to be a rule-out device and for cases that have been processed by the device without notification for "Suspected L VO" should not be viewed as indicating that LVO is excluded. All cases should undergo angiography, per the standard stroke workup.
#### Limitations:
• Brainomix 360 Triage Stroke is not intended for mobile diagnostic use. Images viewed on a mobile platform are compressed preview images and not for diagnostic interpretation.
· Brainomix 360 Triage Stroke does not replace the need for CT A in ischemic stroke workflow prioritization and notification only.
- Brainomix 360 Triage Stroke has been validated and is intended to be used on Siemens, GE and Philips scanners.
· Brainomix 360 Triage Stroke is not intended to be used on patients with recent (within 6 weeks) neurosurgery or endovascular neurointervention or recent (within 4 weeks) previous diagnosis of stroke.
· Brainomix 360 Triage Stroke is not intended to detect isolated subarachnoid hemorrhage and symmetrical bilateral MCA occlusions.
Contraindications: Brainomix 360 Triage Stroke is not suitable for use with scan data containing image features associated with:
· tumors or abscesses
· coils, shunts, embolization or movement artifacts
{3}------------------------------------------------
Brainomix 360 Triage Stroke is not intended to be used for analyzing CT images in intracranial vascular pathologies such as arterial aneurysms, arteriovenous malformations or venous thrombosis.
Type of Use (Select one or both, as applicable)
区 Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for BRAINOMIX. The logo consists of a stylized brain icon on the left, followed by the word "BRAINOMIX" in a sans-serif font. The brain icon is teal, while the word "BRAINOMIX" is mostly gray, with the "AI" portion in teal.
# K232496
## 510(K) Summary Brainomix 360 Triage Stroke
| Date Prepared: | 16Aug2023 |
|-----------------------|----------------------------------------------------------------------------|
| Applicant's name: | Brainomix Limited |
| Applicant's address: | First Floor, Seacourt Tower, West Way<br>Oxford, OX2 0JJ<br>United Kingdom |
| Official contact: | Zsolt Szrnka<br>+44 (0)1865 582730<br>regulatory@brainomix.com |
| Device Trade Name: | Brainomix 360 Triage Stroke |
| Device Common Name: | Radiological Computer-Assisted Triage and Notification Software |
| Regulatory Class: | Class II |
| Product Code: | QAS |
| Regulation No: | 21 C.F.R. §892.2080 |
| Classification Panel: | Radiology Devices |
## Predicate Device
The Brainomix 360 Triage Stroke device is claimed to be substantially equivalent to the following legally marketed predicate device: iSchemaView's Rapid NCCT Stroke (K222884).
## Indications for Use
Brainomix 360 Triage Stroke is a radiological computer aided triage and notification software indicated for use in the analysis of non-contrast head CT (NCCT) images to assist hospital networks and trained clinicians in workflow triage by flagging and communicating suspected positive findings of head NCCT images for large vessel occlusion (LVO) of the intracranial ICA and M1 and intracranial hemorrhage (ICH). Specifically, the device is intended to be used for the triage of images acquired from adult patients in the acute setting, within 24 hours of the acute symptoms, or where this is unclear, since last known well (LKW) time. It is not intended to detect isolated subarachnoid hemorrhage and symmetrical bilateral MCA occlusions.
Brainomix 360 Triage Stroke uses an artificial intelligence algorithm to analyze images and highlight cases with detected NCCT LVO or ICH on the Brainomix server on premise or in the cloud in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected LVO or ICH findings via a web user interface or mobile applications include compressed preview images that are meant for informational purposes only and are not intended for diagnostic use beyond notification.
The device does not alter the original medical image, and it is not intended to be used as a primary diagnostic device. The results of Brainomix 360 Triage Stroke are intended to be used in conjunction with other patient information and based on professional judgment to assist with triage/prioritization of medical images. Notified clinicians are ultimately responsible for reviewing full images per the standard of care.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for BRAINOMIX. The logo consists of a stylized brain icon on the left, followed by the word "BRAINOMIX" in a sans-serif font. The brain icon is a gradient of teal and blue, while the letters "AI" in "BRAINOMIX" are also in the same gradient, providing a visual connection between the brain icon and the artificial intelligence aspect implied by the letters.
## Cautions:
- All patients should get adequate care for their symptoms, including angiography and/or other appropriate care per the standard clinical practice, irrespective of the output of Brainomix 360 Triage Stroke.
- . Brainomix 360 Triage Stroke is not intended to be a rule-out device and for cases that have been processed by the device without notification for "Suspected LVO" should not be viewed as indicating that LVO is excluded. All cases should undergo angiography, per the standard stroke workup.
#### Limitations:
- Brainomix 360 Triage Stroke is not intended for mobile diagnostic use. Images viewed on a mobile platform are compressed preview images and not for diagnostic interpretation.
- . Brainomix 360 Triage Stroke does not replace the need for CTA in ischemic stroke workup - it provides workflow prioritization and notification only.
- . Brainomix 360 Triage Stroke has been validated and is intended to be used on Siemens, GE and Philips scanners.
- Brainomix 360 Triage Stroke is not intended to be used on patients with recent (within 6 weeks) neurosurgery or endovascular neurointervention or recent (within 4 weeks) previous diagnosis of stroke.
- Brainomix 360 Triage Stroke is not intended to detect isolated subarachnoid hemorrhage and symmetrical bilateral MCA occlusions.
### Contraindications:
Brainomix 360 Triage Stroke is not suitable for use with scan data containing image features associated with:
- tumors or abscesses
- coils, shunts, embolization or movement artifacts
Brainomix 360 Triage Stroke is not intended to be used for analyzing CT images in intracranial vascular pathologies such as arterial aneurysms, arteriovenous malformations or venous thrombosis.
## Device Description
Brainomix 360 Triage Stroke is a radiological computer aided triage and notification software package compliant with the DICOM standard and running on an off-the-shelf physical or virtual server.
The Triage Stroke module is a non-contrast CT processing module which operates within the integrated Brainomix 360 Platform to provide triage and notification of suspected large vessel occlusion (LVO) and intracranial hemorrhage (ICH). Brainomix 360 Triage Stroke is a stand-alone software device which uses machine learning algorithms that uses advanced non adaptive imaging artificial intelligence, and large data analytics to automatically identify suspected LVO and ICH on non-contrast CT (NCCT) imaging acquired from adult patients in the acute setting, within 24 hours of the acute symptoms, or where this is unclear, since last known well (LKW) time. The output of the module is a priority notification to clinicians indicating the suspicion of LVO or ICH based on positive findings. Specifically, Brainomix 360 Triage Stroke's ICH analysis is optimized to identify findings of hyperdense volume in the parenchyma typically associated with acute intracranial hemorrhage; and the NCCT LVO suspicion uses the combined analysis of the ASPECTS and hyperdense vessel sign (HDVS) algorithms. It is not intended to detect isolated subarachnoid hemorrhage and symmetrical bilateral MCA occlusions. The Triage Stroke module uses the basic services supplied by the Brainomix 360 Platform including DICOM processing, job management, imaging module execution and imaging output including the notification and compressed image.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for BRAINOMIX. The logo consists of a stylized brain icon on the left, followed by the word "BRAINOMIX" in a sans-serif font. The letters "AI" in the word "BRAINOMIX" are colored in a gradient of blue and green, while the rest of the letters are black.
Brainomix 360 Triage Stroke notification capabilities enable clinicians to review images via mobile app notification. Alternatively, intended users can also access the notification (a "Suspected LVO" or "Suspected hemorrhage" flag) and straightened images via the Brainomix 360 web user interface. Images that are previewed via mobile app are compressed, are for preview informational purposes only, and not intended for diagnostic use beyond notification.
The device is intended for use as an additional tool for assisting study triage within existing pathways. It does not replace any part of the current standard of care. It is designed to assist in prioritization of studies for reading within a worklist, in addition to any other pre-existing formal methods of study prioritization in place. Specifically, it does not remove cases from a reading queue and operates in parallel to the standard of care. This device is not intended to replace the usual methods of communication and transfer of information in the current standard of care.
## Comparison of Technological Characteristics
Both subject and predicate device use automatic data identification and transfer to send images from the local hospital network to a remote server for image processing and analysis, i.e., via a DICOM router that automatically identifies relevant images on a local IT Network and transfers them to a Backend Server using a DICOM compliant communication protocol. The Backend Server software for each device has the same additional software functionality that interacts with the image management architecture, including a notifier module.
Both subject and predicate device provide triage and notification of suspected ICH and LVO. In case of both devices, the ICH analysis is optimized to identify findings of hyperdense volume in the parenchyma typically associated with acute ICH; and the NCCT LVO suspicion uses the combined analysis of the ASPECTS and hyperdense vessel sign (HDVS) algorithms. The algorithms do not externalize any internal segmentation, analysis, or intermediate outputs used in determining if an ICH or LVO is present in the NCCT, nor does either algorithm mark the analyzed NCCT image. Both devices support a mobile application that allows a user to receive push notifications, preview related images, and view patient details associated with a series. The mobile application of both devices is subject to nondiagnostic viewing limitations and have a non-diagnostic warning on the image viewing screen.
Where subject device differs from the predicate is that Brainomix 360 Triage Stroke may also notify the user of a suspected LVO or ICH via a web user interface, as an additional output. The unprocessed images are subjected to rotation and resampling (pre-processing and registration) which do not alter the original imaging in relation to aid for diagnostic. Identically to the predicate, the notified clinicians are responsible for viewing non-compressed images on a diagnostic viewer and engaging in appropriate patient evaluation and relevant discussion with a treating physician before making care-related decisions or requests. Therefore, the technical differences in how the notification may be sent to the user between the proposed and predicate devices do not raise different questions of safety and effectiveness. The predicate device offers two additional channels of notification via email and PACS, which the subject device does not, and therefore, the risks associated with these are not applicable to the subject device.
The below table summarizes and compares data on the Brainomix 360 Triage Stroke device, the subject of this filing, with the Rapid NCCT Stroke device (K222884).
| Characteristic/Parameter | Brainomix 360 Triage Stroke<br>Subject Device (K232496) | iSchemaView Rapid NCCT Stroke<br>Predicate Device (K222884) |
|--------------------------|---------------------------------------------------------|-------------------------------------------------------------|
| Product Code | QAS | QAS |
| Regulation | 21 C.F.R. §892.2080 | 21 C.F.R. §892.2080 |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Brainomix. The logo consists of a stylized brain icon on the left, followed by the word "BRAINOMIX" in a sans-serif font. The brain icon is split into two halves, with a gradient of blue and teal. The letters "AI" in "BRAINOMIX" are also in the same blue and teal gradient, while the rest of the letters are in a dark gray color.
#### Indications for Use
Brainomix 360 Triage Stroke is a radiological computer aided triage and notification software indicated for use in the analysis of non-contrast head CT (NCCT) images to assist hospital networks and trained clinicians in workflow triage by flagging and communicating suspected positive findings of head NCCT images for large vessel occlusion (LVO) of the intracranial ICA and M1 and intracranial hemorrhage (ICH). Specifically, the device is intended to be used for the triage of images acquired from adult patients in the acute setting, within 24 hours of the onset of the acute symptoms, or where this is unclear, since last known well (LKW) time. It is not intended to detect isolated subarachnoid hemorrhage and symmetrical bilateral MCA occlusions.
Brainomix 360 Triage Stroke uses an artificial intelligence algorithm to analyze images and highlight cases with detected NCCT LVO or ICH on the Brainomix server on premise or in the cloud in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected LVO or ICH findings via a web user interface or mobile application. Notifications include compressed preview images that are meant for informational purposes only and are not intended for diagnostic use beyond notification.
The device does not alter the original medical image, and it is not intended to be used as a primary diagnostic device. The results of Brainomix 360 Triage Stroke are intended to be used in conjunction with other patient information and based on professional judgment to assist with triage/prioritization of medical images. Notified clinicians are ultimately responsible for reviewing full images per the standard of care.
#### Cautions:
- All patients should get adequate care for their symptoms, including angiography and/or other appropriate care per the standard clinical practice, irrespective of the output of Brainomix 360 Triage Stroke.
- Brainomix 360 Triage Stroke is not intended to be a rule-out device and for cases that have been processed by the device without notification for "Suspected LVO" should not be viewed as indicating that LVO is excluded. All cases should undergo
Rapid NCCT Stroke is a radiological computer aided triage and notification software indicated for use in the analysis of (1) nonenhanced head CT (NCCT) images. The device is intended to assist hospital networks and trained clinicians in workflow triage by flagging and communicating suspected positive findings of (1) head CT images for Intracranial Hemorrhage (ICH) and (2) NCCT large vessel occlusion (LVO) of the ICA and MCA-M1.
Rapid NCCT Stroke uses an artificial intelligence algorithm to analyze images and highlight cases with detected (1) ICH or (2) NCCT LVO on the Rapid server on premise or in the cloud in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for cases with suspected ICH or LVO findings via PACS, email or mobile device. Notifications include compressed preview images that are meant for informational purposes only, and are not intended for diagnostic use beyond notification.
The device does not alter the original medical image, and it is not intended to be used as a primary diagnostic device. The results of Rapid NCCT Stroke are intended to be used in conjunction with other patient information and based on professional judgment to assist with triage/prioritization of medical images. Notified clinicians are ultimately responsible for reviewing full images per the standard of care. Rapid NCCT Stroke is for Adults only.
#### Cautions:
- All patients should get adequate care for their symptoms including CTA or MRA and/or other appropriate care per the standard clinical practice, irrespective of the device output.
- The device is not intended to be a ruleout device and for cases that have been processed by the device without notification for "Suspected LVO" should not be viewed as indicating that LVO is excluded. All cases should undergo CTA or MRA, per the standard stroke workup.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the word "BRAINOMIX" in a stylized font. The letters are a dark gray color, except for the "AI" which is a gradient of light blue to dark blue. To the left of the word is a graphic of a brain in the same gradient of light blue to dark blue.
initial assessment not meant for diagnostic
purposes.
| Environment of use | Clinical/Hospital environment | Clinical/Hospital environment |
|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| User | Clinician | Clinician |
| Anatomical Region | Head | Head |
| Input Data | NCCT (ICH and LVO) | NCCT (ICH and LVO) |
| Technical Implementation | AI/ML/Neural Network | AI/ML/Neural Network |
| Diagnostic application | Notification-only | Notification-only |
| Results of image analysis | Internal, no image marking | Internal, no image marking |
| Segmentation of ROI | The device does not highlight or direct user's attention to a specific location in the image file | The device does not highlight or direct user's attention to a specific location in the image file |
| Notification Display | PACS, email and mobile device | Web user interface and mobile device |
| Limitations: | Brainomix 360 Triage Stroke is not intended for mobile diagnostic use. Images viewed on a mobile platform are compressed preview images and not for diagnostic interpretation. Brainomix 360 Triage Stroke does not replace the need for CTA in ischemic stroke workup - it provides workflow prioritization and notification only. Brainomix 360 Triage Stroke has been validated and is intended to be used on Siemens, GE and Philips scanners. Brainomix 360 Triage Stroke is not intended to be used on patients with recent (within 6 weeks) neurosurgery or endovascular neurointervention or recent (within 4 weeks) previous diagnosis of stroke. Brainomix 360 Triage Stroke is not intended to detect isolated subarachnoid hemorrhage and symmetrical bilateral MCA occlusions. Contraindications: Brainomix 360 Triage Stroke is not suitable for use with scan data containing image features associated with: tumors or abscesses coils, shunts, embolization or movement artifacts Brainomix 360 Triage Stroke is not intended to be used for analyzing CT images in intracranial vascular pathologies such as arterial aneurysms, arteriovenous malformations or venous thrombosis. | Rapid NCCT Stroke does not replace the need for CTA in ischemic stroke workup, it provides workflow prioritization and notification only. Rapid ICH has been shown to reliably identify hemorrhages of ≥ 0.4ml. Contraindications/Exclusions/Cautions: Patient Motion: excessive motion leading to artifacts that make the scan technically inadequate. Hemorrhagic Transformation, Hematoma Very thin or no Ventricles |
| | | |
| | The device operates in parallel with the<br>standard of care. | The device operates in parallel with the<br>standard of care. |
| Annotation / Localization | Device does not mark, highlight, or direct<br>users' attention to a specific location in the<br>original image | Device does not mark, highlight, or direct<br>users' attention to a specific location in the<br>original image |
| Results of Image Analysis | Internal, no image marking | Internal, no image marking |
| Prioritization Notification | Yes | Yes |
| Clinical SoC Workflow | In parallel to | In parallel to |
| Technical Pipeline | Two cascaded functions (ICH then LVO) using<br>three integrated algorithms | Two cascaded functions (ICH then LVO) using<br>three integrated algorithms |
| Removal of Cases from<br>SoC review | No | No |
initial assessment not meant for diagnostic
purposes.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for BRAINOMIX. The logo consists of a stylized brain icon on the left, followed by the word "BRAINOMIX" in a sans-serif font. The brain icon is teal, and the letters of the word "BRAINOMIX" are a gradient of teal to gray.
## Performance Data
A retrospective study was conducted with the primary endpoint of assessing the performance of Brainomix 360 Triage Stroke in identifying NCCT head images containing intracranial hemorrhage (ICH) or large vessel occlusion (LVO) findings in 267 cases (ICH positive: 40 cases; LVO positive: 112 cases; Negative for ICH or LVO: 115; excluded: 3 cases due to technical inadequacy). ICH positive cases were previously truthed using NCCT imaging with additional clinical information as described in the standalone study for our previously cleared device, Brainomix 360 Triage ICH (K231195). The LVO positive and negative cases were truthed using acute CTA imaging and additional clinical information by consensus of three experienced US board certified neuroradiologists.
Sensitivity and specificity exceeded the pre-specified performance goals for ICH and NCCT LVO. Specifically, ICH performance was observed at 92.5% sensitivity (95% Cl: 80.97-98.36%) and 87.22% specificity (Cl: 82.39-91.18%). These results were consistent with the Triage ICH standalone module performance (K231195). NCCT LVO performance was observed at 68.75% sensitivity (C: 59.71-76.90%) and 89.57% specificity (Cl: 82.92-94.36%).
In addition, a reader study was conducted to compare NCCT LVO sensitivity of the device to that of radiologists. Secondary endpoints of expert non-inferiority and non-expert superiority were used. Triage Stroke passed both conditions, with a sensitivity for all readers (experts and non-experts) of 47.94% (Cl: 37.91-57.97%). The difference between the device's sensitivity and that of all readers was 20.52% (Cl: 8.26-32.78%). The general radiologists (nonexperts) performed with a sensitivity of 47.18% (C: 33.62-60.75%). The difference between the device and nonexpert sensitivity was 21.28% (CI: 5.84-36.72%).
The Triage Stroke time-to-notification analysis includes three steps: (1) the time taken to transfer the image for DICOM transfer, (2) the device processing time, and (3) the time taken to deliver a notification after processing. The combined time-to-notification was compared to an acceptability criterion of 3.5 minutes. The minimum and maximum estimates for each of the three steps was assessed and combined to create a range of expected values for the total time-to-notification. The results were a minimum expected time-to-notification of 62 seconds, and a maximum of 134 seconds. This met the acceptability criterion of time-to-notification of under 3.5 minutes.
#### Performance across subgroups
Subgroup analyses were conducted for the LVO indication (the ICH algorithm is the same as the device cleared under K231195). The analyses assessed performance across age, gender, slice thickness and scanner manufacturer subgroups.
| Metrics | 21 < Age < 50 | 50 ≤ Age < 70 | Age ≥ 70 |
|-----------------|---------------|---------------|----------|
| Total Positives | 13 | 39 | 60 |
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for BRAINOMIX. The logo consists of a stylized brain icon on the left, followed by the word "BRAINOMIX" in a sans-serif font. The brain icon is teal, and the letters in "BRAINOMIX" transition from teal to dark gray.
| Sensitivity | 84.62% (58.59-97.46) | 74.36% (59.09-86.41) | 61.67% (48.93-73.39) |
|-------------|----------------------|----------------------|----------------------|
| Specificity | 94.74% (77.09-99.67) | 89.09% (78.7-95.74) | 87.8% (75.07-95.75) |
| Metrics | Female | Male |
|-----------------|----------------------|----------------------|
| Total Positives | 48 | 64 |
| Sensitivity | 75.0% (61.4-85.89) | 64.06% (51.78-75.19) |
| Specificity | 88.14% (77.94-94.92) | 91.07% (81.32-96.95) |
| Metrics | Slice Thickness<br>< 1 mm | 1 mm ≤ Slice Thickness<br>< 3 mm | Slice Thickness<br>≥ 3 mm |
|-----------------|---------------------------|----------------------------------|---------------------------|
| Total Positives | 44 | 25 | 42 |
| Sensitivity | 61.36% (46.46-74.93) | 76.0% (56.8-90.0) | 71.43% (56.52-83.7) |
| Specificity | 87.76% (76.29-95.19) | 96.67% (84.74-99.79) | 86.11% (71.95-95.11) |
| Metrics | SIEMENS | GE MEDICAL SYSTEMS | Philips |
|-----------------|----------------------|----------------------|----------------------|
| Total Positives | 48 | 33 | 27 |
| Sensitivity | 62.5% (48.25-75.4) | 69.7% (52.68-83.68) | 77.78% (59.56-90.82) |
| Specificity | 88.24% (77.15-95.39) | 83.87% (67.93-94.24) | 96.97% (86.02-99.81) |
Additional subgroup analysis is reported in the labeling. In summary, the performance validation data demonstrated that the proposed device provides accurate detection of acute ICH and LVO under a range of clinically relevant variables associated with the intended use of the software. The notification functionality of the device also met the acceptability criterion.
## Prescriptive Statement
Caution: Federal law restricts this device to sale by or on the order of a physician.
## Safety and Effectiveness
Brainomix 360 Triage Stroke has been designed, verified and validated in compliance with 21 CFR Part 820.30 requirements. The device has been designed to meet the requirements associated with EN ISO 14971:2019 for risk management and the software development process conforms to IEC 62304:2015. The performance of Brainomix 360 Triage Stroke has been validated through retrospective case data based on expert reader truthing of the data and reader testing.
## Substantial Equivalence Discussion
The subject device includes similar features as compared to the predicate devices have substantially equivalent indications for use, technical approaches and roles within a clinical workflow relative to triage and notification. Both devices are intended to automatically process and analyze non-contrast CT scans to provide a notification to users in case of a suspected LVO or ICH being identified. The subject device does not introduce any new risks when compared to the predicate and both include similar mitigation strategies for reducing the risk of off label use of preview images which are shared via mobile. Compared to the predicate device, the subject device
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for BRAINOMIX. The logo consists of a stylized brain icon on the left, followed by the word "BRAINOMIX" in a sans-serif font. The brain icon is a gradient of light blue to dark blue, and the letters "AI" in "BRAINOMIX" are also in the same gradient of blue, while the rest of the letters are in dark gray.
offers one additional channel of notification (web user interface), while the predicate offers notification via email and PACS, which the subject device does not.
## Conclusion
In conclusion, Brainomix 360 Triage Stroke has the same intended use and is substantially equivalent in technological characteristics, safety, and performance characteristics to the legally marketed predicate device, Rapid NCCT Stroke (K222884).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.