BriefCase-Triage

K251195 · Aidoc Medical , Ltd. · QAS · Jan 27, 2026 · Radiology

Device Facts

Record IDK251195
Device NameBriefCase-Triage
ApplicantAidoc Medical , Ltd.
Product CodeQAS · Radiology
Decision DateJan 27, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2080
Device ClassClass 2
AttributesAI/ML, Software as a Medical Device, Real-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K251195 · Jan 27, 2026BriefCase-TriageAidoc Medical , Ltd.Retrospective clinical CT images from 6 US-based clinical sitesA retrospective, multicenter study was conducted to evaluate the performance (sensitivity, specificity, PPV, NPV) of the BriefCase-Triage software in identifying brain aneurysms in contrast-enhanced CT images.Retrospective study; Clinical performance; Brain Aneurysm; Multicenter

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Pivotal Study; Retrospective, blinded, multicenter studyAdults/transitional adults (≥ 18 years) undergoing contrast-enhanced CT brain scans; Sample Size: 544 cases; Number of Sites: 6 US-based clinical sitesNot applicable for this studySensitivity, specificity, time-to-notification, PPV, NPV, PLR, NLR

AI Performance

OutputAlgorithmAcceptanceObservedDev DSDev ReadersTest DSTest Readers
Brain Aneurysm DetectionDeep learning algorithmSensitivity > 80%, Specificity > 80%Sensitivity 87.8% (95% CI: 83.1%, 91.6%), Specificity 91.6% (95% CI: 87.9%, 94.5%)Training data set of manually labeled (tagged) images.—Retrospective, blinded, multicenter study of 544 cases from 6 US-based clinical sites.3 (senior board-certified radiologists)

Indications for Use

BriefCase-Triage is a radiological computer aided triage and notification software indicated for use in the analysis of contrast-enhanced CT images that include the brain, in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communication of suspected positive cases of Brain Aneurysm (BA) findings that are 3.0 mm or larger. BriefCase-Triage uses an artificial intelligence algorithm to analyze images and flag suspect cases in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for suspect cases. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device. The results of BriefCase-Triage are intended to be used in conjunction with other patient information and based on professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care.

Device Story

BriefCase-Triage is a cloud-based, AI-powered radiological triage software. It receives filtered DICOM images from CT scanners; processes them chronologically using a deep-learning algorithm to detect suspected Brain Aneurysms (≥3.0 mm). Upon detection, the system sends a pop-up notification to a desktop application, providing a compressed, low-quality, unannotated grayscale preview image. The device operates in parallel to standard-of-care (SOC) workflows; it does not alter original images, remove cases from the reading queue, or de-prioritize studies. Clinicians use the notification to prioritize their review of the full images within their PACS. By flagging critical cases, the device aims to reduce turnaround time for urgent findings, potentially benefiting patients through faster clinical intervention. It is intended for use by hospital networks and trained medical specialists.

Clinical Evidence

Retrospective, blinded, multicenter study (n=544) across 6 US sites. Primary endpoints: sensitivity 87.8% (95% CI: 83.1%-91.6%) and specificity 91.6% (95% CI: 87.9%-94.5%). Secondary endpoint: mean time-to-notification 44.8s (95% CI: 41.4-48.2s). Additional metrics: NPV 98.9%, PPV 47.6%, PLR 10.5, NLR 0.13. Performance validated across default and three additional operating points.

Technological Characteristics

Software-only device; deep-learning AI algorithm. Runs on Linux-based server in cloud environment. Inputs: DICOM-compliant contrast-enhanced CT images (slice thickness 0.5-2.5 mm). Outputs: Notification and compressed, low-quality grayscale preview images. Integrates with PACS and radiology workstations via image communication platform. Cybersecurity: risk-based approach, SBOM, penetration testing, aligned with Section 524B FD&C Act.

Indications for Use

Indicated for adults and transitional adolescents (≥18 years) undergoing contrast-enhanced CT imaging of the brain. Assists clinicians in triage by flagging suspected Brain Aneurysm (BA) findings ≥3.0 mm.

Regulatory Classification

Identification

Radiological computer aided triage and notification software is an image processing prescription device intended to aid in prioritization and triage of radiological medical images. The device notifies a designated list of clinicians of the availability of time sensitive radiological medical images for review based on computer aided image analysis of those images performed by the device. The device does not mark, highlight, or direct users' attention to a specific location in the original image. The device does not remove cases from a reading queue. The device operates in parallel with the standard of care, which remains the default option for all cases.

Special Controls

Radiological computer aided triage and notification software must comply with the following special controls: 1. Design verification and validation must include: i. A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations. ii. A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage (e.g., improved time to review of prioritized images for pre-specified clinicians). iii. Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts (e.g., subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment. iv. Standalone performance testing protocols and results of the device. v. Appropriate software documentation (e.g., device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results). 2. Labeling must include the following: i. A detailed description of the patient population for which the device is indicated for use. ii. A detailed description of the intended user and user training that addresses appropriate use protocols for the device. iii. Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (e.g., poor image quality for certain subpopulations), as applicable. iv. A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images. v. Device operating instructions. vi. A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness (e.g., improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (e.g., confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.

*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include: (i) A detailed description of the notification and triage algorithms and all underlying image analysis algorithms including, but not limited to, a detailed description of the algorithm inputs and outputs, each major component or block, how the algorithm affects or relates to clinical practice or patient care, and any algorithm limitations. (ii) A detailed description of pre-specified performance testing protocols and dataset(s) used to assess whether the device will provide effective triage ( *e.g.,* improved time to review of prioritized images for pre-specified clinicians).(iii) Results from performance testing that demonstrate that the device will provide effective triage. The performance assessment must be based on an appropriate measure to estimate the clinical effectiveness. The test dataset must contain sufficient numbers of cases from important cohorts ( *e.g.,* subsets defined by clinically relevant confounders, effect modifiers, associated diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals for these individual subsets can be characterized with the device for the intended use population and imaging equipment.(iv) Stand-alone performance testing protocols and results of the device. (v) Appropriate software documentation ( *e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results).(2) Labeling must include the following: (i) A detailed description of the patient population for which the device is indicated for use; (ii) A detailed description of the intended user and user training that addresses appropriate use protocols for the device; (iii) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level ( *e.g.,* poor image quality for certain subpopulations), as applicable;(iv) A detailed description of compatible imaging hardware, imaging protocols, and requirements for input images; (v) Device operating instructions; and (vi) A detailed summary of the performance testing, including: test methods, dataset characteristics, triage effectiveness ( *e.g.,* improved time to review of prioritized images for pre-specified clinicians), diagnostic accuracy of algorithms informing triage decision, and results with associated statistical uncertainty (*e.g.,* confidence intervals), including a summary of subanalyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION Aidoc Medical, Ltd. % John Smith Partner Hogan Lovells US LLP 555 Thirteenth Street NW Washington, District of Columbia 20004 January 27, 2026 Re: K251195 Trade/Device Name: BriefCase-Triage Regulation Number: 21 CFR 892.2080 Regulation Name: Radiological Computer Aided Triage And Notification Software Regulatory Class: Class II Product Code: QAS Dated: December 22, 2025 Received: December 22, 2025 Dear John Smith: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K251195 - John Smith Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K251195 - John Smith Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, ![img-0.jpeg](img-0.jpeg) Jessica Lamb, PhD Assistant Director Imaging Software Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. 510(k) Number (if known) K251195 Device Name BriefCase-Triage Indications for Use (Describe) BriefCase-Triage is a radiological computer aided triage and notification software indicated for use in the analysis of contrast-enhanced CT images that include the brain, in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communication of suspected positive cases of Brain Aneurysm (BA) findings that are 3.0 mm or larger. BriefCase-Triage uses an artificial intelligence algorithm to analyze images and flag suspect cases in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for suspect cases. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device. The results of BriefCase-Triage are intended to be used in conjunction with other patient information and based on professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (6/20) Page 1 of 1 PSC Publishing Services (301) 443-6740 {4} K251195 aidoc 510(k) Summary Aidoc Medical, Ltd.'s BriefCase-Triage K251195 Submitter: Aidoc Medical, Ltd. 34 Hamasger St. Tel-Aviv, Israel, 6721119 Phone: +972-73-7946870 Contact Person: Amalia Schreier, LL.M Date Prepared: January 24, 2025 Name of Device: BriefCase-Triage Classification Name: Radiological computer-assisted triage and notification software device Regulatory Class: Class II Product Code: QAS (21 C.F.R. 892.2080) Primary Predicate Device: BriefCase-Triage for BA (K213721) Reference Device: BriefCase-Triage for iPE (K250248) Rapid Aneurysm Triage and Notification (K230074) Viz ANEURYSM, Viz ANX (K213319) Device Description Briefcase-Triage is a radiological computer-assisted triage and notification software device. The software is based on an algorithm programmed component and is intended to run on a linux-based server in a cloud environment. The BriefCase-Triage receives filtered DICOM Images, and processes them chronologically by running the algorithms on each series to detect suspected cases. Following the AI processing, the output of the algorithm analysis is transferred to an image review software (desktop application). When a suspected case is detected, the user receives a pop-up notification and is presented with a compressed, low-quality, grayscale image that is captioned "not for diagnostic use, for prioritization only" which is displayed as a preview function. This preview is meant for informational purposes only, does not contain any marking of the findings, and is not intended for primary diagnosis beyond notification. {5} Presenting the users with worklist prioritization facilitates efficient triage by prompting the user to assess the relevant original images in the PACS. Thus, the suspect case receives attention earlier than would have been the case in the standard of care practice alone. The algorithm was trained during software development on images of the pathology. As is customary in the field of machine learning, deep learning algorithm development consisted of training on manually labeled ("tagged") images. In that process, critical findings were tagged in all CTs in the training data set. ## Intended Use / Indications for Use BriefCase-Triage is a radiological computer aided triage and notification software indicated for use in the analysis of contrast-enhanced CT images that include the brain, in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communication of suspected positive cases of Brain Aneurysm (BA) findings that are 3.0 mm or larger. BriefCase-Triage uses an artificial intelligence algorithm to analyze images and flag suspect cases in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for suspect cases. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device. The results of BriefCase-Triage are intended to be used in conjunction with other patient information and based on professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care. ## Summary of Technological Characteristics The subject BriefCase-Triage for BA and the algorithm analysis module for the primary predicate BriefCase-Triage for BA (K213721) are identical in most aspects and differ mostly with respect to their algorithm performance. In addition, the SW architecture was changed to separate the image communication platform from the BriefCase-Triage SW. The subject device consists of only the algorithm analysis module which can be integrated with image communication platforms that meet the BriefCase-Triage input and output requirements. Both the primary predicate and subject devices are radiological computer-aided triage and notification software programs. Both devices are artificial intelligence, deep-learning algorithms incorporated in software packages for use with DICOM compliant CT scanners, PACS, and radiology workstations. Both devices are intended to aid in triage and prioritization of radiological images and utilize the same design of deep learning algorithm trained on medical images. Both devices are intended to provide the specialists with notifications and unannotated, compressed, low-quality, and grayscale preview images of suspect studies for the purpose of preemptive triage. {6} The subject and predicate BriefCase-Triage devices raise the same types of safety and effectiveness questions, namely, accurate triage of findings within the processed study. It is important to note that, like the predicate, the subject device neither removes cases from the standard of care reading queue nor de-prioritized cases. Both devices operate in parallel with the standard of care, which remains the default option for all cases. A table comparing the key features of the subject and the primary predicate devices is provided below. Table 1. Key Feature Comparison | | Subject Device Aidoc BriefCase-Triage for BA | Predicate Device Aidoc BriefCase-Triage for BA (K213721) | | --- | --- | --- | | Intended Use / Indications for Use | BriefCase-Triage is a radiological computer aided triage and notification software indicated for use in the analysis of contrast-enhanced CT images that include the brain, in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communication of suspected positive cases of Brain Aneurysm (BA) findings that are 3.0 mm or larger. BriefCase-Triage uses an artificial intelligence algorithm to analyze images and flag suspect cases in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for suspect cases. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device. The results of BriefCase-Triage are intended to be used in conjunction with other patient information and based on professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are | BriefCase is a radiological computer aided triage and notification software indicated for use in the analysis of head CT Angio (CTA) images. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communication of suspected positive cases of Brain Aneurysm (BA) findings above 5 mm size. BriefCase uses an artificial intelligence algorithm to analyze images and flag suspect cases on a standalone desktop application in parallel to the ongoing standard of care image interpretation. The user is presented with notifications for suspect cases. Notifications include compressed preview images that are meant for informational purposes only and not intended for diagnostic use beyond notification. The device does not alter the original medical image and is not intended to be used as a diagnostic device. The results of BriefCase are intended to be used in conjunction with other patient information and based on professional judgment, to assist with triage/prioritization of medical images. Notified clinicians are responsible for viewing full images per the standard of care. | {7} | | Subject Device Aidoc BriefCase-Triage for BA | Predicate Device Aidoc BriefCase-Triage for BA (K213721) | | --- | --- | --- | | | responsible for viewing full images per the standard of care. | | | User population | Hospital networks and appropriately trained medical specialists | Hospital networks and appropriately trained medical specialists | | Threshold BA Findings | Brain Aneurysm (BA) findings that are 3.0 mm size or larger | Brain Aneurysm (BA) findings above 5 mm size | | Anatomical region of interest / Organ | Brain | Head | | Data acquisition protocol | Contrast-enhanced CT images that include the brain | Computed tomography angiography (CTA) of the head | | Notification-only (/notification alerts), parallel workflow tool | Yes | Yes | | Images format | DICOM | DICOM | | Interference with standard workflow | No. No cases are removed from desktop app or deprioritized | No. No cases are removed from desktop app or deprioritized | 4 {8} | | Subject Device Aidoc BriefCase-Triage for BA | Predicate Device Aidoc BriefCase-Triage for BA (K213721) | | --- | --- | --- | | Inclusion/Exclusion criteria for clinical performance testing | Inclusion criteria • Contrast-enhanced CT images that include the brain • Scans performed on adults/transitional adults ≥ 18 years of age. • Slice thickness; 0.5 mm – 2.5 mm axial slices. Exclusion Criteria • All studies that have an inadequate field of view. | Inclusion criteria • computed tomography angiography (CTA) of the head. • Scans performed on adults/transitional adults ≥ 18 years of age. • Scans performed on CT scanners with 64 or greater number of detectors. • Slice thickness; 0.5 – 2.5 mm axial slices. Exclusion Criteria • All scans that are technically inadequate, including motion artifacts, severe metal artifacts, or an inadequate field of view. | | Additional Operating Points | 3 Additional Operating Points | N/A | | Algorithm | Artificial intelligence algorithm with database of images. | Artificial intelligence algorithm with database of images. | | Structure | - Integrated with image routing module via image communication platform (ICP) (image acquisition). - Algorithm module (image processing) - Integrated with desktop application for workflow integration (feed and non-diagnostic Image Viewer). | - AHS module (image acquisition); - ACS module (image processing); - Aidoc Desktop Application for workflow integration (Feed/Worklist (alternate names) and non-diagnostic Image Viewer). | # Performance Data Pivotal Study Summary Aidoc conducted a retrospective, blinded, multicenter study with the BriefCase-Triage software to evaluate the software's performance in identifying contrast-enhanced CT images that include the brain containing Brain {9} Aneurysm in 544 cases from 6 US-based clinical sites, compared to the ground truth as determined by three senior board-certified radiologists. The cases collected for the pivotal dataset were all distinct in time or center from the cases used to train the algorithm. Test pivotal study data was sequestered from algorithm development activities, and use of the data is managed by appropriate Quality Management System procedures. Primary endpoints were sensitivity and specificity with an $80\%$ performance goal. Secondary endpoints were BriefCase-Triage time-to-notification compared to the predicate device. Positive Predictive Value (PPV), Negative Predictive Value (NPV), Positive Likelihood Ratio (PLR), and Negative Likelihood Ratio (NLR) were also assessed. # Primary Endpoint For the primary endpoint, sensitivity was $87.8\%$ (95% CI: 83.1%, 91.6%) and specificity was $91.6\%$ (95% CI: 87.9%, 94.5%), meeting the study's primary endpoint. # Secondary Endpoint In addition, the time-to-notification metric observed for the BriefCase-Triage software, when integrated with a compatible image communication platform, was compared to the equivalent metric of the predicate devices. The BriefCase-Triage time-to-notification includes the time to get the DICOM exam, de-identify it, upload it to the cloud, analyze and send a notification on a positive suspect case back to the desktop application. The BriefCase-Triage time-to-notification was measured for all True Positive cases (i.e., identified as positive both by the reviewers as well as the BriefCase-Triage device) and is given in Table 2 below. The Table also displays the same metric reported for the predicate BriefCase-Triage for BA. The time-to-notification results obtained for the subject BriefCase-Triage device show comparability with the primary predicate with regard to time savings to the standard of care review. The BriefCase-Triage mean time-to-notification for the subject BA triage was 44.8 seconds (95% CI: 41.4-48.2). The time-to-notification for the predicate BA was 252 seconds (95% CI: 234-270). Table 2. Time-to- notification comparison for BriefCase-Triage devices (Seconds) | Time -to-notification | Mean Estimate (seconds) | N | 95% Lower CL | 95% Upper CL | Median | | --- | --- | --- | --- | --- | --- | | Predicate K213721 Processing Time | 252 | 103 | 234 | 270 | 252 | {10} | Time -to-notification | Mean Estimate (seconds) | N | 95% Lower CL | 95% Upper CL | Median | | --- | --- | --- | --- | --- | --- | | BriefCase-Triage and Image Communication Platform Time-To-Notification | 44.8 | 241 | 41.4 | 48.2 | 38.2 | Thus, the reported similar time-to-notification data demonstrates that when using the subject BriefCase-Triage for BA the clinician may have the same benefit in time saving as with the predicate BriefCase-Triage for BA. NPV was $98.9\%$ (95% CI: 98.4%-99.2%) and PPV was $47.6\%$ (95% CI: 38.4%-57.1%). PLR was 10.5 (95% CI: 7.2-15.3) and NLR was 0.13 (95% CI: 0.1-0.19). As can be seen in Table 3 the mean age of patients whose scans were reviewed for BA was 61.9 years, with a standard deviation of 17.7 years. Gender distribution was $38.1\%$ male, and $61.9\%$ female (Table 4). Scanner distribution can also be found in Table 5 below. Table 3. Descriptive Statistics for Age | | Mean | Std | Min | Median | Max | N | | --- | --- | --- | --- | --- | --- | --- | | Age (Years) | 61.9 | 17.7 | 18 | 65 | 98 | 544 | Table 4. Frequency Distribution of Gender * | Ground Truth Results | Gender | | | | All | | | --- | --- | --- | --- | --- | --- | --- | | | Female | | Male | | | | | | N | % | N | % | N | % | | Positive | 165 | 30.3% | 81 | 14.9% | 246 | 45.2% | | Negative | 172 | 31.6% | 126 | 23.2% | 298 | 54.8% | | All | 337 | 61.9% | 207 | 38.1% | 544 | 100.0% | Table 5. Frequency Distribution of Manufacturer {11} | Manufacturer | N | % | | --- | --- | --- | | GE MEDICAL SYSTEMS | 243 | 44.7% | | Philips | 69 | 12.7% | | SIEMENS | 166 | 30.5% | | TOSHIBA | 66 | 12.1% | | Total | 544 | 100% | Table 6. Frequency Distribution of slice thickness | Thickness (mm) | N | % | | --- | --- | --- | | 0-1 | 355 | 65.3% | | 1-2.5 | 189 | 34.70% | | Total | 544 | 100% | Table 7. Frequency Distribution of protocol | Protocol | N | % | | --- | --- | --- | | CT angiogram | 270 | 49.6% | | CT w/ contrast | 274 | 50.4% | | Total | 544 | 100% | Clinical Subgroups and Confounders: Pathologies present in negative cases: Inflammatory; Oncology; Trauma; Heart & vascular; Chronic Diseases, Fully Negative and None of the above. Additional Operating Points: In addition to the default operating point that was selected to maximize both sensitivity and specificity, three additional operating points were selected to enhance specificity gradually (with AOP3 chosen to maximize specificity) while meeting the $80\%$ threshold for both sensitivity and specificity. The performance results for the algorithm at the AOP1-AOP3 is shown below in Table 8: Table 8. Additional Operating Points | Operating Points | Sensitivity % (95% CI) | Specificity % (95% CI) | | --- | --- | --- | {12} | AOP1 | 87.4% (95% CI: 82.6%-91.3%) | 91.9% (95% CI: 88.3%-94.8%) | | --- | --- | --- | | AOP2 | 87.0% (95% CI: 82.1%-90.9%) | 92.6% (95% CI: 89.0%-95.3%) | | AOP3 | 86.2% (95% CI: 81.2%-90.2%) | 93.6% (95% CI: 90.2%-96.1%) | In summary, performance goals were achieved for the default and four additional operating points. Combined with the comparison results of time-to-notification metric with the predicate device, these data establish the achievement by the subject BriefCase-Triage of preemptive triage in the range of several minutes. # Cybersecurity Cybersecurity has been incorporated into the software development lifecycle in alignment with Section 524B of the FD&C Act and FDA cybersecurity guidance. Aidoc utilizes a risk-based approach to cybersecurity, including secure design practices, vulnerability assessments, a Software Bill of Materials (SBOM), and penetration testing. These efforts support the safety, effectiveness, and resilience of the software against cybersecurity threats. # Conclusions The subject BriefCase-Triage for BA and the predicate BriefCase-Triage for BA (K213721) are intended to aid in prioritization and triage of radiological images for the indications for suspected positive findings of incidental pulmonary embolism pathologies. Both devices are software packages consisting of deep learning AI algorithms that process images and produce analysis results, which are displayed to the user by a prioritization alert and a compressed, low-quality, grayscale, unannotated preview image. In both devices, the labeling clearly states that the devices are not for diagnostic use and instructs the user to further evaluate and diagnose based only on the original images in the local PACS. Both devices operate in parallel to the standard of care workflow in the sense that they do not change the original image, do not provide any marking on the output preview, do not remove images from the standard of care FIFO queue and do not de-prioritize cases, thus not disturbing standard interpretation of the images. Both devices notify the radiologist of time-sensitive critical cases within the range of several minutes, and thus contribute similarly to the standard of care workflow turnaround time reduction through preemptive triage. The technological characteristics of both devices are identical with the exception of additional operating points and minor changes in inclusion and exclusion criteria. These differences do not raise new questions of substantial equivalence. The subject BriefCase-Triage device for BA is thus substantially equivalent to the primary predicate BriefCase-Triage for BA (K213721).
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...