VBrain-OAR

K212116 · Vysioneer, Inc. · QKB · Oct 12, 2021 · Radiology

Device Facts

Record IDK212116
Device NameVBrain-OAR
ApplicantVysioneer, Inc.
Product CodeQKB · Radiology
Decision DateOct 12, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesAI/ML, Software as a Medical Device

AI Performance

OutputAlgorithmAcceptanceObservedDev DSDev ReadersTest DSTest Readers
Organ at risk contouringDeep learning neural networks——————
Image registrationIntensity-based algorithm——————

Indications for Use

VBrain-OAR is a software device intended to assist trained radiotherapy personnel including, but not limited to, radiologists, radiation oncologists, neurosurgeons, radiation therapists, dosimetrists, and medical physicists, during their clinical workflows of brain tumor radiation therapy treatment planning, by providing initial object contours of organs at risk in the brain (i.e., the region of interest, ROI) on axial T1 contrast-enhanced brain MRI images. VBrain-OAR is intended to be used on adult patients only. VBrain-OAR uses an artificial intelligence algorithm (i.e., deep learning neural networks) to contour (segment) organs at risk (brain stem, eyes, optic nerves, optic chiasm) in the brain on MRI images for trained radiotherapy personnel's attention, which is meant for informational purposes only and not intended for replacing their current standard practice of manual contouring process. VBrain-OAR does not alter the original MRI image, nor does it intend to detect tumors for diagnosis. VBrain-OAR is intended only for contouring and generating contours of organs at risk in the brain; it is not intended to be used with images of other body parts. VBrain-OAR also contains the automatic image register volumetric medical image data. (e.g., MR, CT). It allows rigid image registration to adjust the spatial position and orientation of two images. Radiation therapy treatment personnel must finalize (confirm or modify) the contours generated by VBrain-OAR, as necessary, using an external platform available at the facility that supports DICOM-RT viewinglediting functions, such as image visualization software and treatment planning system.

Device Story

Software device for radiotherapy treatment planning; assists clinicians in contouring organs at risk (OAR) in brain and performing image registration. Inputs: axial T1 contrast-enhanced brain MRI images for segmentation; volumetric medical images (MR, CT) for registration. Operation: utilizes deep learning neural networks for OAR segmentation and intensity-based algorithms for rigid image registration. Workflow: integrated into PACS network; pulls patient scans, executes modules, and exports results as DICOM-RT objects (RTSTRUCT). Output: initial contours for clinician review/modification on external DICOM-RT compatible treatment planning systems. Used in clinical radiotherapy settings by trained personnel. Benefits: provides initial contours to assist in treatment planning workflows; does not replace manual contouring or diagnose tumors.

Clinical Evidence

Bench testing only. Performance evaluated on multi-institutional datasets. Segmentation testing demonstrated clinically acceptable contours for brain OARs. Registration testing evaluated rigid registration quality across multiple modalities. Stratified analysis confirmed consistent performance across patient sex, imaging hardware, and protocols.

Technological Characteristics

Software application; Linux OS. Modules: contouring (deep learning neural networks), registration (intensity-based), and workflow management. Connectivity: PACS network integration. Output: DICOM-RT (RTSTRUCT). Software level of concern: Major.

Indications for Use

Indicated for adult patients to assist radiotherapy personnel (radiologists, oncologists, neurosurgeons, therapists, physicists) in brain tumor radiation therapy treatment planning by providing initial contours of organs at risk (brain stem, eyes, optic nerves, optic chiasm) on axial T1 contrast-enhanced brain MRI images and performing rigid image registration of volumetric data (MR, CT). Not for tumor detection or diagnosis.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ October 12, 2021 Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of the seal is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font below. Vysioneer Inc. % Vicki Lin Regulatory Specialist 33 Rogers St., # 308 CAMBRIDGE MA 02142 # Re: K212116 Trade/Device Name: VBrain-OAR Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: QKB Dated: September 10, 2021 Received: September 13, 2021 Dear Vicki Lin: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see {2}------------------------------------------------ https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, For Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K212116 Device Name VBrain-OAR ### Indications for Use (Describe) V Brain-OAR is a software device intended to assist trained radiotherapy personnel including, but not limited to, radiologists, radiation oncologists, neurosurgeons, radiation therapists, and medical physicists, during their clinical workflows of brain tumor radiation therapy treatment planning, by providing initial object contours of organs at risk in the brain (i.e., the region of interest, ROI) on axial T1 contrast-enhanced brain MRI images. VBrain-OAR is intended to be used on adult patients only. V Brain-OAR uses an artificial intelligence algorithm (i.e., deep learning neural networks) to contour (segment) organs at risk (brain stem, eyes, optic nerves, optic chiasm) in the brain on MRI images for trained radiotherapy personnel's attention, which is meant for informational purposes only and not intended for replacing their current standard practice of manual contouring process. VBrain-OAR does not alter the original MRI image, nor does it intend to detect tumors for diagnosis. VBrain-OAR is intended only for contouring and generating contours of organs at risk in the brain; it is not intended to be used with images of other body parts. V Brain-OAR also contains the automatic image register volumetric medical image data. (e.g., MR, CT). It allows rigid image registration to adjust the spatial position and orientation of two images. Radiation therapy treatment personnel must finalize (confirm or modify) the contours generated by VBrain-OAR, as necessary, using an external platform available at the facility that supports DICOM-RT viewinglediting functions, such as image visualization software and treatment planning system. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------| | <div> <span> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size: 20px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | CONTINUE ON A SEPARATE PAGE IF NEEDED## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." ============================================================================================================================================================================== {4}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for Vysioneer. The logo consists of a stylized purple "V" shape above the word "VYSIONEER" in a sans-serif font. A horizontal purple line is located below the word "VYSIONEER". #### Section 5 510(k) Summary K212116 #### 5.1 Submitter Vysioneer Inc. 33 Rogers St. #308, Cambridge, MA 02142 | Contact Person: | Vicki Lin (Regulatory Specialist) | |------------------------|-----------------------------------| | Phone: | 609-865-8659 | | Email: | vicki.lin@vysioneer.com | | Date Summary Prepared: | September 10, 2021 | #### 5.2 Device Name | Trade Name: | VBrain-OAR | |-----------------------------------|-----------------------------------------------------------------| | Common Name: | Radiological Image Processing Software for<br>Radiation Therapy | | Regulation Number / Product Code: | 21 CFR 892.2050 / QKB | #### 5.3 Predicate Device Primary Predicate Device: AccuContour™ 510(k) Holder/Submitter: Xiamen Manteia Technology LTD. 510(k) Number: K191928 (Cleared on 02/28/2020) Reference Device: iPlan 510(k) Holder/Submitter: Brainlab AG 510(k) Number: K113732 (Cleared on 05/07/2012) {5}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for Vysioneer. The logo consists of the word "Vysioneer" in a sans-serif font, with the letters in a dark blue color. Above the word is a stylized "V" shape, also in dark blue, formed by two angled lines that do not quite meet at the bottom. The overall design is simple and modern. #### 5.4 Intended Use / Indications for Use VBrain-OAR is a software device intended to assist trained radiotherapy personnel including, but not limited to, radiologists, radiation oncologists, neurosurgeons, radiation therapists, dosimetrists, and medical physicists, during their clinical workflows of brain tumor radiation therapy treatment planning, by providing initial object contours of organs at risk in the brain (i.e., the region of interest, ROI) on axial T1 contrast-enhanced brain MRI images. VBrain-OAR is intended to be used on adult patients only. VBrain-OAR uses an artificial intelligence algorithm (i.e., deep learning neural networks) to contour (segment) organs at risk (brain stem, eyes, optic chiasm) in the brain on MRI images for trained radiotherapy personnel's attention, which is meant for informational purposes only and not intended for replacing their current standard practice of manual contouring process. VBrain-OAR does not alter the original MRI image, nor does it intend to be used to detect tumors for diagnosis. VBrain-OAR is intended only for contouring and generating contours of organs at risk in the brain; it is not intended to be used with images of other body parts. VBrain-OAR also contains the automatic image registration feature to register volumetric medical image data. (e.g., MR, CT). It allows rigid image registration to adjust the spatial position and orientation of two images. Radiation therapy treatment personnel must finalize (confirm or modify) the contours generated by VBrain-OAR, as necessary, using an external platform available at the facility that supports DICOM-RT viewing/editing functions, such as image visualization software and treatment planning system. ### Device Description ર્સ્ડ VBrain-OAR is a software application system indicated for use in the contouring (segmentation) of brain MRI images for the organs at risk (OAR) in the brain during radiation treatment planning and in the registration of multi-modality images. The device consists of 2 algorithm modules, which are contouring algorithm module and registration algorithm module, and a workflow management module. The modules can work independently, and yet can be integrated with each other. The contouring (segmentation) algorithm module consists of image preprocessing, deep learning neural networks, and postprocessing components, and is intended to contour organs at risk in the brain on the axial T1 contrast-enhanced MR images. It utilizes deep learning neural networks to generate contours for the organs at risk in the brain and export the results as DICOM-RT objects (using the RT Structure Set ROI Contour attribute, RTSTRUCT). The registration algorithm module registers volumetric medical image data (e.g., MR, CT). It allows rigid image registration to adjust the spatial position and orientation of two images. {6}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the logo for Vysioneer. The logo consists of a stylized letter 'V' in purple, with the word 'VYSIONEER' written in a sans-serif font below it. A thin purple line is located under the word 'VYSIONEER'. The workflow management module is configured to work on a PACS network. Upon user's request, it will pull patient scans or users can send corresponding DICOM images, and it will trigger a predefined workflow, in which different algorithm modules are executed to generate the DICOM output. The DICOM output of a workflow are sent back to the PACS. # 5.6 Comparison of Technological Characteristics with the Predicate Device VBrain-OAR is substantially equivalent to the primary predicate device AccuContour™ (K191928). The proposed device, VBrain-OAR, and the primary predicate, AccuContour™ (K191928) are both AI-based (deep learning) software devices intended to be used in the workflow of radiation therapy treatment planning by providing tools to automatically segment/contour organs at risk on images as well as perform image registration (image fusion). In addition, VBrain-OAR and the predicate device both use an intensity-based algorithm for image registration. Both the proposed device and AccuContour™ (K191928) are regulated under 21 CFR 892.2050, Product Code QKB (Radiological Image Processing Software for Radiation Therapy). Both the proposed device and the reference device, iPlan (K113732), are designed to contour organs at risk of brain on MR images and perform image registration (image fusion). The only difference between these two software devices is that VBrain-OAR uses an AI-based algorithm, while iPlan uses an atlas-based algorithm for segmentation of images. Please see Table 5-1 comparing the intended use and key technological characteristics of VBrain-OAR and the predicate and reference device. | | Proposed Device | Primary Predicate<br>Device | Reference Device | |--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Company | Vysioneer Inc. | Xiamen Manteia<br>Technology LTD. | Brainlab AG | | Device Name | VBrain-OAR | AccuContour™ | iPlan | | 510k Number | K212116 | K191928 | K113732 | | Regulation No. | 21CFR 892.2050 | 21CFR 892.2050 | 21CFR 892.1750 | | Classification | II | II | II | | Product Code | QKB | QKB | JAK/LLZ | | Intended<br>Use/Indication for Use | VBrain-OAR is a<br>software device intended<br>to assist trained<br>radiotherapy personnel | It is used by radiation<br>oncology department to<br>register multimodality<br>images and segment<br>(non-contrast) CT | iPlan's indications for<br>use are the viewing,<br>presentation and<br>documentation of<br>medical imaging, | | Proposed Device | Primary Predicate<br>Device | Reference Device | | | including, but not<br>limited to, radiologists,<br>radiation oncologists,<br>neurosurgeons, radiation<br>therapists, dosimetrists,<br>and medical physicists,<br>during their clinical<br>workflows of brain<br>tumor radiation therapy<br>treatment planning, by<br>providing initial object<br>contours of organs at<br>risk in the brain (i.e., the<br>region of interest, ROI)<br>on axial T1 contrast-<br>enhanced brain MRI<br>images. VBrain-OAR is<br>intended to be used on<br>adult patients only.<br><br>VBrain-OAR uses an<br>artificial intelligence<br>algorithm (i.e., deep<br>learning neural<br>networks) to contour<br>(segment) organs at risk<br>(brain stem, eyes, optic<br>nerves, optic chiasm) in<br>the brain on MRI images<br>for trained radiotherapy<br>personnel's attention,<br>which is meant for<br>informational purposes<br>only and not intended<br>for replacing their<br>current standard practice<br>of manual contouring<br>process. VBrain-OAR<br>does not alter the<br>original MRI image, nor<br>does it intend to be used<br>to detect tumors for<br>diagnosis. VBrain-OAR<br>is intended only for<br>contouring and<br>generating contours of<br>organs at risk in the | images, to generate<br>needed information for<br>treatment planning,<br>treatment evaluation and<br>treatment adaptation.<br>The product has two<br>image process functions:<br>(1) Deep learning<br>contouring: it can<br>automatically contour<br>the organ-at-risk,<br>including head and neck,<br>thorax, abdomen and<br>pelvis (for both male and<br>female),<br>(2) Automatic<br>Registration, and<br>(3) Manual Contour.<br><br>It also has the following<br>general functions:<br>(1) Receive,<br>add/edit/delete, transmit,<br>input/export, medical<br>images and DICOM<br>data;<br>(2) Patient management;<br>(3) Review of processed<br>images;<br>(4) Open and Save of<br>files. | including different<br>modules for image<br>processing, image<br>fusion, atlas assisted<br>visualization and<br>segmentation,<br>intraoperative functional<br>planning where the<br>output can be used e.g.<br>with stereotactic image<br>guided surgery or other<br>devices for further<br>processing and<br>visualization.<br>Example procedures<br>include but are not<br>limited to:<br>Planning and<br>simulation of cranial<br>surgical procedures such<br>as tumor resection, shunt<br>placement, minimal-<br>invasive stereotactic<br>interventions, biopsy,<br>planning and simulation<br>of trajectories for<br>stimulation and electrode<br>recording ENT procedures<br>such as sinus surgery,<br>tumor surgery Spine<br>procedures such as<br>tumor surgery, pedicle<br>screw planning,<br>vertebroplasty planning Plan View is an<br>application which is<br>intended to be used for<br>reviewing existing<br>treatment plans Planning and<br>simulation of cranio-<br>maxillofacial procedures<br>Typical users of iPlan<br>are medical | | | | Proposed Device | Primary Predicate<br>Device | Reference Device | | | brain; it is not intended<br>to be used with images<br>of other body parts. | | professionals, including<br>but not limited to<br>surgeons and<br>radiologists. | | | VBrain-OAR also<br>contains the automatic<br>image registration<br>feature to register<br>volumetric medical<br>image data. (e.g., MR,<br>CT). It allows rigid<br>image registration to<br>adjust the spatial<br>position and orientation<br>of two images. | | | | | Radiation therapy<br>treatment personnel<br>must finalize (confirm or<br>modify) the contours<br>generated by VBrain-<br>OAR, as necessary,<br>using an external<br>platform available at the<br>facility that supports<br>DICOM-RT<br>viewing/editing<br>functions, such as image<br>visualization software<br>and treatment planning<br>system. | | | | Device Description | VBrain-OAR is a<br>software application<br>system indicated for use<br>in the contouring<br>(segmentation) of brain<br>MRI images for the<br>organs at risk (OAR) in<br>the brain during<br>radiation treatment<br>planning and in the<br>registration of multi-<br>modality images. The<br>device consists of 2<br>algorithm modules,<br>which are contouring | The proposed device,<br>AccuContour, is a<br>standalone software<br>which is used by<br>radiation oncology<br>department to register<br>multimodality images<br>and segment (non-<br>contrast) CT images, to<br>generate needed<br>information for<br>treatment planning,<br>treatment evaluation and<br>treatment adaptation. | iPlan is a software based<br>treatment planning<br>application providing<br>functionalities like<br>viewing, processing and<br>documentation of<br>medical data including<br>different modules for<br>image preparation,<br>image fusion, image<br>segmentation where the<br>result is a treatment plan<br>that can be used e.g. for<br>stereotactic and/or image<br>guided surgery. | | Proposed Device | Primary Predicate<br>Device | Reference Device | | | algorithm module and<br>registration algorithm<br>module, and a workflow<br>management module.<br>The modules can work<br>independently, and yet<br>can be integrated with<br>each other.<br><br>The contouring<br>(segmentation)<br>algorithm module<br>consists of image<br>preprocessing, deep<br>learning neural<br>networks, and<br>postprocessing<br>components, and is<br>intended to contour<br>organs at risk in the<br>brain on the axial T1<br>contrast-enhanced MR<br>images. It utilizes deep<br>learning neural networks<br>to generate contours for<br>the organs at risk in the<br>brain and export the<br>results as DICOM-RT<br>objects (using the RT<br>Structure Set ROI<br>Contour attribute,<br>RTSTRUCT).<br><br>The registration<br>algorithm module<br>registers volumetric<br>medical image data (e.g.,<br>MR, CT). It allows rigid<br>image registration to<br>adjust the spatial<br>position and orientation<br>of two images.<br><br>The workflow<br>management module is<br>configured to work on a<br>PACS network. Upon | The product has two<br>image process functions:<br>(1) Deep learning<br>contouring: it can<br>automatically contour<br>the organ-at-risk,<br>including head and neck,<br>thorax, abdomen and<br>pelvis (for both male and<br>female),<br>(2) Automatic<br>Registration, and<br>(3) Manual Contour.<br>It also has the following<br>general functions:<br>Receive, add/edit/delete,<br>transmit, input/export,<br>medical images and<br>DICOM data;<br>Patient management;<br>Review of processed<br>images;<br>Open and Save of files. | | | | | Proposed Device | Primary Predicate<br>Device | Reference Device | | | user's request, it will<br>pull patient scans or<br>users can send<br>corresponding DICOM<br>images, and it will<br>trigger a predefined<br>workflow, in which<br>different algorithm<br>modules are executed to<br>generate the DICOM<br>output. The DICOM<br>output of a workflow are<br>sent back to the PACS. | |…
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...