WakeMed dataset (5 November 2019 - 30 June 2020) including patient stays in 7 intensive care units across 2 hospitals
>1 (physicians)
WakeMed dataset (5 November 2019 - 30 June 2020) including patient stays in 7 intensive care units across 2 hospitals
>1 (physicians)
Indications for Use
CLEWICU provides the clinician with physiological insight into a patient's likelihood of future hemodynamic instability. CLEWICU is intended for use with intensive care unit (ICU) patients 18 years and over. CLEWICU is considered to provide additional information regarding the patient's predicted future risk for clinical deterioration, as well as identifying patients at low risk for deterioration. The product predictions are for reference only and no therapeutic decisions should be made based solely on the CLEWICU predictions.
Device Story
CLEWICU is a stand-alone analytical software system (ClewICUServer and ClewICUnitor) for ICU use. It ingests patient data from EHRs and medical devices, including vital signs, nursing assessments, flowsheet data, medications, and lab results. Machine learning models process these inputs to calculate the CLEWHI index (likelihood of hemodynamic instability requiring vasopressor/inotrope support) and CLEWLR (low risk for deterioration). The ClewICUnitor web interface displays these indices and notifications to clinicians. The system serves as an adjunctive decision support tool; it provides reference-only insights to assist clinicians in monitoring patient status and identifying potential deterioration. It does not replace clinical judgment or serve as the sole basis for therapeutic decisions.
Clinical Evidence
Retrospective clinical validation using a dataset from 7 ICUs (WakeMed Health System). A validated tagging system (benchmarked against physician readers) generated ground truth labels for model training and validation. HI model performance: 60.6% sensitivity (95% CI: 56.9-63.9%), 22.3% PPV (95% CI: 20.7-24.3%), median lead-time of 3.0 hours. LR model performance: 95.7% specificity (95% CI: 94.8-96.6%), 21.4% sensitivity (95% CI: 21.2-21.6%).
Technological Characteristics
Software-only analytical system installed on user-provided hardware. Uses machine learning models to process multi-modal patient data (EHR, vitals, labs, medications). Provides near-real-time risk indices and alerts. Standalone software architecture.
Indications for Use
Indicated for ICU patients 18 years and older to provide physiological insight into the likelihood of future hemodynamic instability and clinical deterioration, or to identify patients at low risk for deterioration.
Regulatory Classification
Identification
The adjunctive predictive cardiovascular indicator is a prescription device that uses software algorithms to analyze cardiovascular vital signs and predict future cardiovascular status or events. This device is intended for adjunctive use with other physical vital sign parameters and patient information and is not intended to independently direct therapy.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) A software description and the results of verification and validation testing based on a comprehensive hazard analysis and risk assessment must be provided, including:
(i) A full characterization of the software technical parameters, including algorithms;
(ii) A description of the expected impact of all applicable sensor acquisition hardware characteristics and associated hardware specifications;
(iii) A description of sensor data quality control measures;
(iv) A description of all mitigations for user error or failure of any subsystem components (including signal detection, signal analysis, data display, and storage) on output accuracy;
(v) A description of the expected time to patient status or clinical event for all expected outputs, accounting for differences in patient condition and environment; and
(vi) The sensitivity, specificity, positive predictive value, and negative predictive value in both percentage and number form.
(2) A scientific justification for the validity of the predictive cardiovascular indicator algorithm(s) must be provided. This justification must include verification of the algorithm calculations and validation using an independent data set.
(3) A human factors and usability engineering assessment must be provided that evaluates the risk of misinterpretation of device output.
(4) A clinical data assessment must be provided. This assessment must fulfill the following:
(i) The assessment must include a summary of the clinical data used, including source, patient demographics, and any techniques used for annotating and separating the data.
(ii) The clinical data must be representative of the intended use population for the device. Any selection criteria or sample limitations must be fully described and justified.
(iii) The assessment must demonstrate output consistency using the expected range of data sources and data quality encountered in the intended use population and environment.
(iv) The assessment must evaluate how the device output correlates with the predicted event or status.
(5) Labeling must include:
(i) A description of what the device measures and outputs to the user;
(ii) Warnings identifying sensor acquisition factors that may impact measurement results;
(iii) Guidance for interpretation of the measurements, including a statement that the output is adjunctive to other physical vital sign parameters and patient information;
(iv) A specific time or a range of times before the predicted patient status or clinical event occurs, accounting for differences in patient condition and environment;
(v) Key assumptions made during calculation of the output;
(vi) The type(s) of sensor data used, including specification of compatible sensors for data acquisition;
(vii) The expected performance of the device for all intended use populations and environments; and
(viii) Relevant characteristics of the patients studied in the clinical validation (including age, gender, race or ethnicity, and patient condition) and a summary of validation results.
Predicate Devices
Acumen Hypotension Prediction Index feature by Edwards Lifesciences (DEN160044, K183646, K190205)
Submission Summary (Full Text)
{0}------------------------------------------------
January 9, 2021
CLEW Medical Ltd. % Yarmela Pavlovic Partner Manatt, Phelps & Phillips, LLP 1 Embarcadero Center, 30th Floor San Francisco, California 94111
Re: K200717
Trade/Device Name: CLEWICU System (ClewICUServer and ClewICUnitor) Regulation Number: 21 CFR 870.2210 Regulation Name: Adjunctive Predictive Cardiovascular Indicator Regulatory Class: Class II Product Code: QNL Dated: January 6, 2021 Received: January 8, 2021
Dear Yarmela Pavlovic:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for LT Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200717
Device Name CLEWICU System
Indications for Use (Describe)
CLEWICU provides the clinician with physiological insight into a patient's likelihood of future hemodynamic instability. CLEWICU is intended for use with intensive care unit (ICU) patients 18 years and over.
CLEWICU is considered to provide additional information regarding the patient's predicted future risk for clinical deterioration, as well as identifying patients at low risk for deterioration. The product predictions are for reference only and no therapeutic decisions should be made based solely on the CLEWICU predictions.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: overline;"> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## K200717
# 510(k) SUMMARY CLEW Medical's CLEWICU
#### Submitter
CLEW Medical, Ltd. 5 Hamelacha St. Poleg, Netanya lsrael 4206002
Phone: +972-9-779-5995 Contact Person: Avigdor Faians
Date Prepared:January 7, 2021
Name of Device: CLEWICU System
Common or Usual Name: CLEWICU
Classification Name: Adjunctive Predictive Cardiovascular Indicator
Requlation Number: 21 CFR 870.2210
Regulatory Class: Class II
Product Code: QNL
#### Predicate Device
Acumen Hypotension Prediction Index feature by Edwards Lifesciences (DEN160044, K183646, K190205)
#### Device Description
The CLEWICU System is a stand-alone analytical software product that includes the ClewICUServer and the ClewICUnitor. It uses models derived from machine learning to calculate the likelihood of occurrence of certain clinically significant events for patients in the intensive care unit (ICU). ClewICUServer and ClewICUnitor are software-only devices that are installed on user-provided hardware.
The ClewICUServer is a backend software platform that imports patient data from various sources including Electronic Health Record (EHR) data and medical device data. The data are then used by models operating within the ClewICUServer to compute and store the CLEWHI index (likelihood of hemodynamic instability requiring vasopressor / inotrope support), and CLEWLR (indication that the patient is at "low risk" for deterioration).
The ClewICUnitor is the web-based user interface displaying CLEWHI, and CLEWLR associated notifications and related measures, as well as presenting the overall unit status.
{4}------------------------------------------------
## Intended Use / Indications for Use
The CLEWICU's Indications for Use are as follows:
CLEWICU provides the clinician with physiological insight into a patient's likelihood of future hemodynamic instability. CLEWICU is intended for use with intensive care unit (ICU) patients 18 years and over. CLEWICU is considered to provide additional information regarding the patient's predicted future risk for clinical deterioration, as well as identifying patients at low risk for deterioration. The product predictions are for reference only and no therapeutic decisions should be made based solely on the CLEWICU predictions.
While there are differences in the indications for use statements between CLEWICU and Acumen HPI, these differences are not critical to the intended use. Both devices are generally intended to calculate patient health status indices and predict the likelihood of future cardiovascular clinical events. The differences in the indications statements do not impact the safety or effectiveness, as neither device is intended to be the sole factor upon which a patient's clinical care is based. Instead, the analyses are intended to notify healthcare providers of potential patient deterioration and should be considered in conjunction with other physiological parameters before any treatment is provided.
## Summary of Technological Characteristics
Both the subject and predicate devices are software decision support systems that receive patient data, process it, and display calculated insights. At a high level, the subject and predicate devices are based on the following same technological elements:
- . Both devices receive near-real-time patient health data;
- . Both devices analyze the available data to generate predictions of a patient's risk of experiencing a future clinically significant cardiac event;
- . Both devices alert the healthcare provider when a patient's risk has reached a predefined threshold.
Although the two products produce slightly different outputs (e.g., risk of hypotensive event versus risk of hemodynamic instability), those differences do not raise different questions of safety or effectiveness, because in both instances, the key point is whether the predictions are accurate and in a timeframe that permits appropriate clinical review and action.
Additionally, though the subject device incorporates additional data inputs relative to the predicate device (i.e., vital signs, nursing assessments, flowsheet data, medications, lab data), this does not raise different questions of safety or effectiveness. The key question in both cases is whether the indices accurately predict the associated future cardiac events.
## Performance Data
The CLEWICU was subjected to extensive software verification testing, as well as risk analysis and cybersecurity review.
{5}------------------------------------------------
## K200717
Human factors testing was conducted to confirm that the user manual provides adequate instructions and that the CLEWICU can be used by the intended users without use errors. Included in the study were 15 intended users, comprised of five nurses and ten physicians. Testing was successfully completed, demonstrating that the product has no expected user errors associated with potential for serious injury to users.
Further, CLEWICU was assessed in a retrospective clinical validation with a dataset from the WakeMed Health System. As an initial matter, a tagging system was developed and validated (against human physician readers as ground truth). Once validated, the tagging system was used to generate the clinical truth labels needed, both for training of the CLEWICU predictive models and for validation of the clinical performance of those models. The WakeMed dataset included patient stays in 7 intensive care units across 2 hospitals between 5 November 2019 and 30 June 2020. Patients were male or female and at least 18 years of age.
Following validation of the tagging system (comparing performance against human physician readers), the predictive models were evaluated by assessing each model's ability to predict the events of interest. The summary of the data from this study is presented below:
| | Point Estimate | 95% Lower CI | 95% Upper CI |
|---------------------------|----------------|--------------|--------------|
| Sensitivity | 60.6% | 56.9% | 63.9% |
| Positive Predictive Value | 22.3% | 20.7% | 24.3% |
| For the Hemodynamic Instability (HI) model | Point Estimate |
|--------------------------------------------|----------------|
|--------------------------------------------|----------------|
| | Median | 25th Percentile | 75th Percentile |
|-------------------------------------|-----------|-----------------|-----------------|
| Lead-time<br>(true positive alerts) | 3.0 hours | 1.6 hours | 4.8 hours |
## For the Low Risk (LR) model
| | Point Estimate | 95% Lower Cl | 95% Upper Cl |
|-------------|----------------|--------------|--------------|
| Specificity | 95.7% | 94.8% | 96.6% |
| Sensitivity | 21.4% | 21.2% | 21.6% |
The performance was determined acceptable considering the software is intended to provide visualonly indicators with a typical time horizon of several hours for an already closely monitored population.
## Conclusions
The CLEWICU is as safe and effective as the Acumen HPI. The CLEWICU has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device. The minor differences in indications do not alter the intended use of the device and do not raise different questions of safety and effectiveness when used as labeled. In addition, the minor technological differences between the CLEWICU and its predicate device raise no new issues of safety or effectiveness. Performance data demonstrate that the CLEWICU performs as expected and in a manner that is substantially equivalent to the Acumen HPI. Thus, the CLEWICU is substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.